K Number
K253142

Validate with FDA (Live)

Date Cleared
2026-01-09

(106 days)

Product Code
Regulation Number
870.1130
Age Range
22 - 120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arm blood pressure monitor is for home use for measuring blood pressure and pulse rate. It is suitable for adult, not for neonate or pregnancy.

Device Description

The subject device, Arm Blood Pressure Monitor, is an automatic non-invasive blood pressure monitor which can be driven by dry batteries. It uses an inflatable cuff which is wrapped around the patient's upper arm to measure the systolic and diastolic blood pressure as well as the pulse rate of adult, not for neonate or pregnancy.

The proposed device consists of the main body and the arm belt, suitable for home use for measuring blood pressure and pulse rate.

This blood pressure monitor has the memory function of 60 groups of measuring data of two people, which can save the data separately. It can display the average reading of the latest 3 groups of measurement results.

This blood pressure monitor has voice broadcast function (optional). During measurement and recall the memory, there will be voice operation tips.

No operation for 1 minute the device will shut down automatically.

The subject device includes model MJ1DS/MJ1D/MJ3D/MJ5D/MJ6D/MJ8D/RN3D/MJ4D/RN1D/RN2D. These models are identical in terms of software design, cuff type, measurement range and function, principle, the core algorithm of the software, the key components. The schematic circuit diagrams are identical in all models. The PCB layout are identical in all models because of different appearance such as structure, buttons layout.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - K253142

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

January 9, 2026

Ningbo Ranor Medical Technology Co., Ltd.
℅ Boyle Wang
General Manager
Shanghai Truthful Information Technology Co., Ltd.
Room 1801, No. 161 East Lu Jiazui Rd., Pudong
Shanghai, 200120
China

Re: K253142
Trade/Device Name: Arm Blood Pressure Monitor (MJ1D,MJ1DS,MJ3D,MJ5D,MJ6D,MJ8D,RN3D,MJ4D,RN1D,RN2D)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive blood pressure measurement system
Regulatory Class: Class II
Product Code: DXN
Dated: September 25, 2025
Received: September 25, 2025

Dear Boyle Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K253142 - Boyle Wang
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

Page 3

K253142 - Boyle Wang
Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

HETAL B. ODOBASIC -S

for

Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.K253142
Please provide the device trade name(s).

Arm Blood Pressure Monitor (MJ1D,MJ1DS,MJ3D,MJ5D,MJ6D,MJ8D,RN3D,MJ4D,RN1D,RN2D)

Please provide your Indications for Use below.

The Arm blood pressure monitor is for home use for measuring blood pressure and pulse rate. It is suitable for adult, not for neonate or pregnancy.

Please select the types of uses (select one or both, as applicable).☐ Prescription Use (Part 21 CFR 801 Subpart D)☒ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

510(k) Summary

Page 1 of 5

510(k) Summary

K253142

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name: Ningbo Ranor Medical Science & Technology Co., Ltd.
Address: No. 127 Fenghui Road, Wangchun Industrial Park, Haishu District, Ningbo, China
Tel: 86-574-89258788
Contact: Xu Jia'nan

Designated Submission Correspondent

Contact: Mr. Boyle Wang
Name: Shanghai Truthful Information Technology Co., Ltd.
Address: Room 1801, No. 161 East Lu Jiazui Rd., Pudong,Shanghai, 200120 China
Tel: +86-21-50313932
Email: Info@truthful.com.cn

Date of Preparation: Dec.10,2025

2.0 Device Information

Trade name: Arm Blood Pressure Monitor
Common name: Noninvasive Blood Pressure Measurement System
Classification name: Noninvasive Blood Pressure Measurement System
Model(s): MJ1D MJ1DS MJ3D MJ5D MJ6D MJ8D RN3D MJ4D RN1D RN2D
Production code: DXN
Regulation number: 21 CFR 870.1130
Classification: Class II
Panel: Cardiovascular

3.0 Predicate Device Information

Manufacturer: Ningbo Ranor Medical Science & Technology Co., Ltd.
Trade/Device name: Arm Blood Pressure Monitor, RN-032A,RN-032C

Page 6

510(k) Summary
Page 2 of 5

510(k) number: K193456

4.0 Indication for Use Statement

The Arm blood pressure monitor is for home use for measuring blood pressure and pulse rate. It is suitable for adult, not for neonate or pregnancy.

5.0 Device Description

The subject device, Arm Blood Pressure Monitor, is an automatic non-invasive blood pressure monitor which can be driven by dry batteries. It uses an inflatable cuff which is wrapped around the patient's upper arm to measure the systolic and diastolic blood pressure as well as the pulse rate of adult, not for neonate or pregnancy.

The proposed device consists of the main body and the arm belt, suitable for home use for measuring blood pressure and pulse rate.

This blood pressure monitor has the memory function of 60 groups of measuring data of two people, which can save the data separately. It can display the average reading of the latest 3 groups of measurement results.

This blood pressure monitor has voice broadcast function (optional). During measurement and recall the memory, there will be voice operation tips.

No operation for 1 minute the device will shut down automatically.

The subject device includes model MJ1DS/MJ1D/MJ3D/MJ5D/MJ6D/MJ8D/RN3D/MJ4D/RN1D/RN2D. These models are identical in terms of software design, cuff type, measurement range and function, principle, the core algorithm of the software, the key components. The schematic circuit diagrams are identical in all models. The PCB layout are identical in all models because of different appearance such as structure, buttons layout.

6.0 Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
  • IEC 60601-1-2:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.

Page 7

510(k) Summary
Page 3 of 5

  • IEC 60601-1-11:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • IEC 80601-2-30:2018, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.

Biocompatibility Testing

The proposed Arm Blood Pressure Monitor does not introduce any new direct or indirect patient contacting materials including new color additive, as compared to the previous clearance predicate Arm Blood Pressure Monitor. So the biocompatibility testing reports provided and reviewed in K193456 remain valid and can support the subject Arm Blood Pressure Monitor.

Software Verification and Validation Testing

The subject devices utilize the same software version, software algorithm, measurement logic and functionality as the cleared predicate devices (RN-032A and RN-032C, K193456). No code changes or new functions were introduced. The software safety classification remains unchanged. The verification and validation testing previously reviewed in K193456 continues to support the subject devices, as the software operates identically to the predicate. Therefore, no additional software verification and validation testing is required to demonstrate substantial equivalence.

7.0 Clinical Test Conclusion

As the subject device utilizes the same monitoring technology as the predicate device, there is no change on the sensors, cuffs, algorithms, measurement accuracy of the subject device, additional testing was not considered necessary to support the substantial equivalence.

8.0 Technological Characteristic Comparison Table

Page 8

510(k) Summary
Page 4 of 5

Table1-General Comparison

ItemSubject Device K253142Predicate Device K193456Remark
ManufacturerNingbo Ranor Medical Science & Technology Co., Ltd.Ningbo Ranor Medical Science & Technology Co., Ltd./
Product NameArm Blood Pressure Monitor MJ1D MJ1DS MJ3D MJ5D MJ6D MJ8D RN3D MJ4D RN1D RN2DArm Blood Pressure Monitor RN-032A,RN-032C/
Product CodeDXNDXNSame
Regulation No.21 CFR 870.113021 CFR 870.1130Same
ClassIIIISame
Intended Use/Indication for UseThe Arm blood pressure monitor is for home use for measuring blood pressure and pulse rate. It is suitable for adult, not for neonate or pregnancy.The Arm blood pressure monitor is for home use for measuring blood pressure and pulse rate. It is suitable for adult, not for neonate or pregnancy.Same
Application SiteUpper ArmUpper ArmSame
Arm Circumference22cm~32cm220mm~320mmSame
Patients Contacting MaterialsPatient contact materials of the cuff: Nylon TPU (Cuff) According to ISO-10993Patient contact materials of the cuff: Nylon TPU (Cuff) According to ISO-10993Same
Patient PopulationAdultAdultSame
Measurements ItemSYS,DYS,PulseSYS,DYS,PulseSame
DisplaySegment LCD Display , Segment LED Display(MJ1DS)Segment LCD DisplaySame
Design MethodOscillometric MethodOscillometric MethodSame

Table 2 Performance Comparison

ItemSubject Device K253142Predicate Device K193456Remark
Max Cuff pressure294 mmHg300mmHgDifferent
BP RangeSystolic: 60~249 mmHg Diastolic: 30~170 mmHg0 ~ 299 mmHg
BP Accuracy±3 mmHg±3 mmHgSame
PR Range40-190 beats/min40 ~ 180 beats/minDifferent
Pulse Accuracy±5% of reading value±5% of reading valueSame
Inflation MethodAutomatic by electronic pumpAutomatic by electronic pumpSame
Deflation MethodAutomatic Pressure Release ValveAutomatic Pressure Release ValveSame

Page 9

510(k) Summary
Page 5 of 5

Memory Size2x60 sets of data2x60 sets of dataSame
Operation Condition5-40℃ 15%-80% RH 80kPa~106kPa5~40℃, Humidity: 15%-80% Atmospheric: 80kPa~106kPaSame
Storage Condition-20℃~55℃ Humidity: ≤95% (noncondensing) 50kPa~106kPa-20 ~ 55℃, ≤95% (noncondensing ) 50kPa~106kPaSame
Power Supply4 AAA dry batteries4 AAA dry batteriesSame
Performance StandardComply with IEC 80601-2-30Comply with IEC 80601-2-30Same

Table 3 Safety Comparison

ItemSubject Device K253142Predicate Device K193456Remark
Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1Same
Home UseComply with IEC 60601-1-11Comply with IEC 60601-1-11Same
EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2Same
BiocompatibilityComply with ISO 10993-1, FDA GuidanceComply with ISO 10993-1, FDA GuidanceSame

9.0 Conclusion

The conclusions drawn from the comparison and analysis above demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicated device and raises no new questions of safety or effectiveness. The differences between both devices are insignificant in terms of safety and effectiveness.

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).