K Number
K253057

Validate with FDA (Live)

Date Cleared
2026-01-22

(122 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images.

AI-Rad Companion Brain MR provides the following functionalities:
• Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities
• Quantitative comparison of each brain structure with normative data from a healthy population
• Presentation of results for reporting that includes all numerical values as well as visualization of these results

Device Description

AI-Rad Companion Brain MR runs two distinct and independent algorithms for Brain Morphometry analysis and White Matter Hyperintensities (WMH) segmentation, respectively. In overall, comprises four main algorithmic features:

• Brain Morphometry
• Brain Morphometry follow-up
• White Matter Hyperintensities (WMH)
• White Matter Hyperintensities (WMH) follow-up

The feature for Brain Morphometry is available since the first version of the device (VA2x), while segmentation of White Matter Hyperintensities was added since VA4x and the follow-up analysis for both is available since VA5x. The brain morphometry and brain morphometry follow-up feature have not been modified and remain identical to previous VA5x mainline version.

AI-Rad Companion Brain MR VA60 is an enhancement to the predicate, AI-Rad Companion Brain MR VA50 (K232305). Just as in the predicate, the brain morphometry feature of AI-Rad Companion Brain MR addresses the automatic quantification and visual assessment of the volumetric properties of various brain structures based on T1 MPRAGE datasets. From a predefined list of brain structures (e.g. Hippocampus, Caudate, Left Frontal Gray Matter, etc.) volumetric properties are calculated as absolute and normalized volumes with respect to the total intracranial volume. The normalized values are compared against age-matched mean and standard deviations obtained from a population of healthy reference subjects. The deviation from this reference population can be visualized as 3D overlay map or out-of-range flag next to the quantitative values.

Additionally, identical to the predicate, the white matter hyperintensities feature addresses the automatic quantification and visual assessment of white matter hyperintensities on the basis of T1 MPRAGE and T2 weighted FLAIR datasets. The detected WMH can be visualized as a 3D overlay map and the quantification in count and volume as per 4 brain regions in the report.

AI/ML Overview

Here's a structured overview of the acceptance criteria and study details for the AI-Rad Companion Brain MR, based on the provided FDA 510(k) clearance letter:

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (AI-Rad Companion Brain MR WMH Feature)Reported Device Performance (AI-Rad Companion Brain MR WMH Follow-up Feature)
WMH Segmentation AccuracyPearson correlation coefficient between WMH volumes and ground truth annotation: 0.96Interclass correlation coefficient between WMH volumes and ground truth annotation: 0.94Dice score: 0.60F1-score: 0.67Detailed Dice Scores for WMH Segmentation:Mean: 0.60Median: 0.62STD: 0.1495% CI: [0.57, 0.63]Detailed ASSD Scores for WMH Segmentation:Mean: 0.05Median: 0.00STD: 0.1595% CI: [0.02, 0.08]
New or Enlarged WMH Segmentation Accuracy (Follow-up)Pearson correlation coefficient between new or enlarged WMH volumes and ground truth annotation: 0.76Average Dice score: 0.59Average F1-score: 0.71Detailed Dice Scores for New/Enlarged WMH Segmentation (by Vendor - Siemens, GE, Philips):Siemens: Mean 0.64, Med 0.67, STD 0.15, 95% CI [0.60, 0.69]GE: Mean 0.56, Med 0.60, STD 0.14, 95% CI [0.51, 0.61]Philips: Mean 0.55, Med 0.59, STD 0.16, 95% CI [0.50, 0.61]Detailed ASSD Scores for New/Enlarged WMH Segmentation (by Vendor - Siemens, GE, Philips):Siemens: Mean 0.02, Med 0.00, STD 0.06, 95% CI [0.00, 0.04]GE: Mean 0.09, Med 0.01, STD 0.23, 95% CI [0.03, 0.19]Philips: Mean 0.04, Med 0.00, STD 0.11, 95% CI [0.00, 0.08]

Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • White Matter Hyperintensities (WMH) Feature: 100 subjects (Multiple Sclerosis patients (MS), Alzheimer's patients (AD), cognitive impaired (CI), and healthy controls (HC)).
    • White Matter Hyperintensities (WMH) Follow-up Feature: 165 subjects (Multiple Sclerosis patients (MS) and Alzheimer's patients (AD)).
    • Data Provenance: Data acquired from Siemens, GE, and Philips scanners. Testing data had balanced distribution with respect to gender and age of the patient according to target patient population, and field strength (1.5T and 3T). This indicates a retrospective, multi-vendor, multi-national (implied by vendor diversity) dataset.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Number of Experts: Three radiologists.
    • Qualifications: Not explicitly stated beyond "radiologists." It is not specified if they are board-certified, or their years of experience.
  3. Adjudication Method for the Test Set:

    • For each dataset, three sets of ground truth annotations were created manually.
    • Each set was annotated by a disjoint group consisting of an annotator, a reviewer, and a clinical expert.
    • The clinical expert was randomly assigned per case to minimize annotation bias.
    • The clinical expert reviewed and corrected the initial annotation of the changed WMH areas according to a specified annotation protocol. Significant corrections led to re-communication with the annotator and re-review.
    • This suggests a 3+1 Adjudication process, where three initial annotations are reviewed by a clinical expert.
  4. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was not done. The study focuses on the standalone performance of the AI algorithm against expert ground truth.
  5. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done:

    • Yes, a standalone performance study was done. The "Accuracy was validated by comparing the results of the device to manual annotated ground truth from three radiologists." This evaluates the algorithm's performance directly.
  6. The Type of Ground Truth Used:

    • Expert Consensus / Manual Annotation: The ground truth for both WMH and WMH follow-up features was established through "manual annotated ground truth from three radiologists" and involved a "standard annotation process" with annotators, reviewers, and clinical experts.
  7. The Sample Size for the Training Set:

    • The document states that the "training data used for the fine tuning the hyper parameters of WMH follow-up algorithm is independent of the data used to test the white matter hyperintensity algorithm follow up algorithm." However, the specific sample size for the training set is not provided in the given text.
  8. How the Ground Truth for the Training Set Was Established:

    • The document implies that the WMH follow-up algorithm "does not include any machine learning/ deep learning component," suggesting a rule-based or conventional image processing algorithm. Therefore, "training" might refer to parameter tuning rather than machine learning model training.
    • For the "fine-tuning the hyper parameters of WMH follow-up algorithm," the ground truth establishment method for this training data is not explicitly detailed in the provided text. It only states that this data was "independent of the data used to test" the algorithm.

FDA 510(k) Clearance Letter - AI-Rad Companion Brain MR

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

January 22, 2026

Siemens Healthcare GmbH
Kira Morales
Regulatory Affairs Manager
Henkestrasse 127
Erlangen, 91052
Germany

Re: K253057
Trade/Device Name: AI-Rad Companion Brain MR
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: December 5, 2025
Received: December 5, 2025

Dear Kira Morales:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K253057 - Kira Morales Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K253057 - Kira Morales Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K253057

Please provide the device trade name(s).
AI-Rad Companion Brain MR

Please provide your Indications for Use below.
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images.

AI-Rad Companion Brain MR provides the following functionalities:
• Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities
• Quantitative comparison of each brain structure with normative data from a healthy population
• Presentation of results for reporting that includes all numerical values as well as visualization of these results

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

510(k) SUMMARY FOR AI-Rad Companion Brain MR

K253057

Submitted by:
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19355
Date Prepared: December 5, 2025

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92.

1. Submitter

Importer/Distributor
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 193552
Registration Number: 2240869

Manufacturing Site
Siemens Healthcare GmbH
Henkestrasse 127
Erlangen, Germany 91052
Registration Number: 3002808157

2. Contact Person

Kira Morales
Regulatory Affairs Manager
Siemens Medical Solutions USA, Inc.
40 Liberty Boulevard
Malvern, PA 19335
Phone: +1 (484) 901 - 9471
Email: kira.morales@siemens-healthineers.com

3. Device Name and Classification

Product Name: AI-Rad Companion Brain MR
Trade Name: AI-Rad Companion Brain MR
Classification Name: Medical Image Management and Processing System
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050

Page 6

Device Class: Class II
Product Code: QIH

4. Predicate Device

Product Name: AI-Rad Companion Brain MR
Propriety Trade Name: AI-Rad Companion Brain MR
510(k) Number: K232305
Clearance Date: October 23, 2023
Classification Name: Medical Image Management and Processing System
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Primary Product Code: QIH
Recall Information: N/A

5. Reference Device

Product Name: icobrain
510(k) Number: K192130
Clearance Date: December 13, 2019
Classification Name: Medical Image Management and Processing System
Classification Panel: Radiology
CFR Section: 21 CFR §892.2050
Device Class: Class II
Primary Product Code: LLZ
Recall Information: N/A

6. Indications for Use

AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images.

AI-Rad Companion Brain MR provides the following functionalities:
• Automated segmentation and quantitative analysis of individual brain structures and white matter hyperintensities
• Quantitative comparison of each brain structure with normative data from a healthy population
• Presentation of results for reporting that includes all numerical values as well as visualization of these results

7. Device Description

Page 7

AI-Rad Companion Brain MR runs two distinct and independent algorithms for Brain Morphometry analysis and White Matter Hyperintensities (WMH) segmentation, respectively. In overall, comprises four main algorithmic features:

• Brain Morphometry
• Brain Morphometry follow-up
• White Matter Hyperintensities (WMH)
• White Matter Hyperintensities (WMH) follow-up

The feature for Brain Morphometry is available since the first version of the device (VA2x), while segmentation of White Matter Hyperintensities was added since VA4x and the follow-up analysis for both is available since VA5x. The brain morphometry and brain morphometry follow-up feature have not been modified and remain identical to previous VA5x mainline version.

AI-Rad Companion Brain MR VA60 is an enhancement to the predicate, AI-Rad Companion Brain MR VA50 (K232305). Just as in the predicate, the brain morphometry feature of AI-Rad Companion Brain MR addresses the automatic quantification and visual assessment of the volumetric properties of various brain structures based on T1 MPRAGE datasets. From a predefined list of brain structures (e.g. Hippocampus, Caudate, Left Frontal Gray Matter, etc.) volumetric properties are calculated as absolute and normalized volumes with respect to the total intracranial volume. The normalized values are compared against age-matched mean and standard deviations obtained from a population of healthy reference subjects. The deviation from this reference population can be visualized as 3D overlay map or out-of-range flag next to the quantitative values.

Additionally, identical to the predicate, the white matter hyperintensities feature addresses the automatic quantification and visual assessment of white matter hyperintensities on the basis of T1 MPRAGE and T2 weighted FLAIR datasets. The detected WMH can be visualized as a 3D overlay map and the quantification in count and volume as per 4 brain regions in the report.

8. Substantially Equivalent (SE) and Technological Characteristics

The intended use of the predicate device and the subject device are equivalent. The main difference is that AI-Rad Companion Brain MR VA60 adds improvements to the White matter hyperintensities and White Matter Hyperintensities Follow-up as compared to the predicate, AI-Rad Companion Brain MR VA50.

The subject device, AI-Rad Companion Brain MR VA60 is substantially equivalent with regard to the intended use and technical characteristics compared to the predicate device, AI-Rad Companion Brain MR VA50 (K232305) with respect to the software features, functionalities, and core algorithms. The additional enhancements and improvements provided in AI-Rad Companion Brain MR VA60 increase the usability and reduce the complexity of the imaging workflow for the clinical user.

Page 8

Icobrain serves as a reference device within this submission and a dedicated comparison of technological characteristics is provided. Siemens Healthineers has determined that AI-Rad Companion Brain MR VA60 is comparable to icobrain (K192130) as it has similar technological and performance characteristics with respect to the white matter hyperintensities & follow-up feature improvements (algorithms cleared in K232305). AI-Rad Companion Brain MR VA60 used equivalent validation methodology to analyze the performance of the white matter hyperintensities follow-up feature compared to icobrain (K192130).

Page 9

The risk analysis and non-clinical data support that both devices perform equivalently and do not raise different questions of the safety and effectiveness.

FeatureSubject Device: AI-Rad Companion Brain MR VA60Predicate Device: AI-Rad Companion Brain MR VA50 (K232305)Reference Device: icobrain (K192130)Comparison Results
Brain Morphometry SegmentationPre-processing functionality for automatic segmentation and volumetry of MPRAGE data.Pre-processing functionality for automatic segmentation and volumetry of MPRAGE data.Image processing for automatic segmentation and volumetry of MPRAGE data.Same as predicate
Brain Morphometry QuantificationCalculation of label maps (display of brain segmentation) and partially combined label maps (fused with the processed MPRAGE data).Calculation of label maps (display of brain segmentation) and partially combined label maps (fused with the processed MPRAGE data).Normalized and unnormalized volume and volume changes of different brain structures.Same as predicate
Brain Morphometry: Deviation MapCalculation of deviation map (representation of brain status in relation to reference data) and partially combined deviation maps (fused with the processed MPRAGE data) User customizable color labels for the overlay map.Calculation of deviation map (representation of brain status in relation to reference data) and partially combined deviation maps (fused with the processed MPRAGE data) User customizable color labels for the overlay map.Not availableSame as predicate
Brain Morphometry Follow-UpAutomatic calculation of the atrophy range inAutomatic calculation of the atrophy range inNot availableSame as predicate

Page 10

FeatureSubject Device: AI-Rad Companion Brain MR VA60Predicate Device: AI-Rad Companion Brain MR VA50 (K232305)Reference Device: icobrain (K192130)Comparison Results
Brain Morphometry Follow-Up Time Between Studiespercentage for each segmented brain structure Configurable between 14-180 dayspercentage for each segmented brain structure Configurable time interval between the current and the prior scan should be ≥ 180 days and < the retention period.Not availableEnhanced from the predicate
Brain White Matter Hyperintensities SegmentationPre-processing functionality for automatic segmentation and volumetry of MPRAGE and FLAIR data.Pre-processing functionality for automatic segmentation and volumetry of MPRAGE and FLAIR data.Image processing for automatic segmentation and volumetry of FLAIR data.Same as predicate
Brain White Matter Hyperintensities QuantificationCalculation of white matter hyperintensities count and volume as per 4 brain regions.Calculation of white matter hyperintensities count and volume as per 4 brain regions.Unnormalized volume and volume changes of FLAIR white matter hyperintensities as per 4 brain regionsSame as predicate
Brain White Matter Hyperintensities MapCalculation of white matter hyperintensities map fused with the processed FLAIR data User customizable color labels for the overlay map.Calculation of white matter hyperintensities map fused with the processed FLAIR data User customizable color labels for the overlay map.Calculation of white matter hyperintensities map overlaid with the FLAIR dataSame as predicate
Brain White Matter Hyperintensities Input DataT2-weighted 3D FLAIR image series.T1-weighted MPRAGE and T2-weighted 3D FLAIR image seriesT1-weighted MPRAGE and T2-weighted 3D FLAIR image seriesStreamlined from the predicate

Page 11

FeatureSubject Device: AI-Rad Companion Brain MR VA60Predicate Device: AI-Rad Companion Brain MR VA50 (K232305)Reference Device: icobrain (K192130)Comparison Results
White Matter Hyperintensities Follow-UpMR images from two time points to identify new or enlarged areas. The results of follow-up scan are then refined by using morphological operations to generate WMH changed areas between baseline and follow-up scanMR images from two time points to identify new or enlarged areas. The results of follow-up scan are then refined by using morphological operations to generate WMH changed areas between baseline and follow-up scanAssessment of New/Enlarging Lesion countSame as predicate
White Matter Hyperintensities Follow Up Input DataWMH follow-up supports 1.5T and 3T input data in both prior and current studyWMH follow-up is only supported for 3T input data in both prior and current study.Both 1.5T and 3T is supportedEnhanced from the predicate
White Matter Hyperintensities Follow-Up Data CriteriaNot mandatory to have the same Slice Thickness, Pixel Spacing, and Magnetic Field Strength consistent between prior and current studies. *In case if the prior and current Magnetic Field Strength (0018, 0087) are not same, the system will process the input data with an information message.Mandatory to have the same Slice Thickness, Pixel Spacing, and Magnetic Field Strength consistent between prior and current studies.Not availableEnhanced from the predicate
White Matter Hyperintensities & Follow-Up Multi-Vendor SupportValidated with data from Siemens Healthineers, GE and PhilipsOnly supports data acquired on 3 Tesla Siemens Healthineers MR scannersMulti-vendor supportEnhanced from the predicate
Distribution & ArchivingCreation of an image series for a morphometry report. Automatic transfer ofCreation of an image series for a morphometry report. Automatic transfer ofAutomatic transfer of generated imageSame as predicate

Page 12

FeatureSubject Device: AI-Rad Companion Brain MR VA60Predicate Device: AI-Rad Companion Brain MR VA50 (K232305)Reference Device: icobrain (K192130)Comparison Results
Distribution & Archivinggenerated maps and morphometry report to a PACS system.generated maps and morphometry report to a PACS system.series and report to a PACS system.Same as predicate
User Interface ConfirmationConfirmation UI with basic visualization functionalityConfirmation UI with basic visualization functionalityNot available.Same as predicate
User Interface ConfigurationConfiguration UIConfiguration UINot availableSame as predicate
LayoutsSimplified layout dedicated for confirmation of resultsSimplified layout dedicated for confirmation of resultsNot availableSame as predicate
ArchitectureCloud solution and Edge components deployed on customer premise.Cloud solution and Edge components deployed on customer premise.Cloud only solution with no components deployed on customer premise.Same as predicate
DICOM SRDICOM structured report representation of a natural language reportDICOM structured report representation of a natural language reportDICOM structured reportSame as predicate

Table 1: Comparison table for AI-Rad Companion Brain MR VA60, predicate device AI-Rad Companion Brain MR VA50 (K232305) and reference device icobrain (K192130)

The conclusions from all verification and validation data suggest that these enhancements are equivalent with respect to safety and effectiveness of the predicate device. These modifications do not change the intended use of the product. Siemens is of the opinion that AI-Rad Companion Brain MR VA60 is substantially equivalent to the currently marketed device, AI-Rad Companion Brain MR VA50.

Page 13

9. Nonclinical Tests

Non-clinical tests were conducted to test the functionality of AI-Rad Companion Brain MR. Software validation and bench testing have been conducted to assess the performance claims as well as the claim of substantial equivalence to the predicate device.

AI-Rad Companion has been tested to meet the requirements of conformity to multiple industry standards. Non-clinical performance testing demonstrates that AI-Rad Companion Organs RT complies with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Device Software Functions" (June 2023) as well as with the following voluntary FDA recognized Consensus Standards listed in Table 2.

Recognition NumberProduct AreaTitle of StandardReference Number and DateStandards Development Organization
5-129GeneralMedical Devices – Application of usability engineering to medical devices [including Corrigendum 1 (2016)]IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSIONIEC
5-125GeneralMedical Devices – application of risk management to medical devicesISO 14971 Third Edition 2019-12ISO
13-79Software/InformaticsMedical device software – software life cycle processes [Including Amendment 1 (2016)]IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSIONAAMI ANSI IEC
12-349RadiologyDigital Imaging and Communications in Medicine (DICOM) SetPS 3.1 – 3.20 2021eNEMA
5-134GeneralMedical devices – symbols to be used with information to be supplied by the manufacturer – Part 1: General Requirements15223-1 Fourth edition 2021-07ISO IEC
13-97Software/InformaticsHealth software – Part 1: General requirements for product safety82304-1 Edition 1.0 2016-10IEC

Table 2: List of recognized standards

Verification and Validation

Software documentation level, per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on June 14, 2023, is also included as part of this submission. The performance data demonstrates continued

Page 14

conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the subject device during product development.

Software bench testing in the form of Unit, System and Integration tests were performed to evaluate the performance and functionality of the new features and software updates. All testable requirements in the Requirement Specifications and the Risk Analysis have been successfully verified and traced in accordance with the Siemens Healthineers DH product development process. Human factor usability validation is addressed in system testing and usability validation test records. Software verification and regression testing have been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.

Siemens Healthineers adheres to the cybersecurity recommendations as defined the FDA Guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" (September 2023) by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.

10. Performance Software Validation

AI-Rad Companion Brain MR VA60A brain morphometry and brain morphometry follow-up features are identical to the predicate device AI-Rad Companion Brain MR VA50A.

White Matter Hyperintensities Features

Performance testing for AI-Rad Companion Brain MR WMH was performed on test data from 100 subjects, which included Multiple Sclerosis patients (MS), Alzheimer's patients (AD), cognitive impaired (CI) and healthy controls (HC) from Siemens, GE, and Philips scanners. Testing data has balanced distribution with respect to gender and age of the patient according to target patient population and field strength (1.5T and 3T) of the MR scanner used.

Accuracy was validated by comparing the results of the device to manual annotated ground truth from three radiologists. Acceptance criteria for above experiments were defined based on a literature review. In all validation experiments, AI-Rad Companion Brain MR WMH feature passed the acceptance criteria, the Pearson correlation coefficient between the WMH volumes estimated by our software and ground truth annotation was 0.96. The interclass correlation coefficient between between the WMH volumes estimated by our software and ground truth annotation was 0.94. The segmentation accuracy of the WMH reaches a Dice score of 0.60. The F1-score of WMH detection is 0.67.

Page 15

Subject DeviceIcobrain (K192130)AllGenderField Strength
MF1.5 T3.0 T
# Data1003961386251
Dice
Mean0.600.600.590.620.580.58
Med0.620.600.630.640.60N.A.
STD0.140.120.170.150.16N.A.
95% CI[0.57, 0.63][0.56, 0.64][0.54, 0.63][0.57, 0.66][0.54, 0.62]N.A.
ASSD
Mean0.050.040.050.080.03N.A.
Med0.000.000.000.010.00N.A.
STD0.150.110.170.210.08N.A.
95% CI[0.02, 0.08][0.01, 0.08][0.02, 0.10][0.03, 0.16][0.01, 0.05]N.A.

DICE and ASSD Results for the AI-Rad Companion Brain MR White Matter Hyperintensities Algorithm

White Matter Hyperintensities Follow-up Features

Performance testing for AI-Rad Companion Brain MR WMH follow-up feature was performed on 1.5T and 3T test datasets from 165 subjects, which included Multiple Sclerosis patients (MS) and Alzheimer's patients (AD) scanned using Siemens, GE and Philips scanners. Testing data has balanced distribution with respect to gender and age of the patient according to target patient population.

Accuracy was validated by comparing the results of the subject device to manual annotated ground truth from three radiologists. Acceptance criteria for above experiments were defined based on a literature review. In all validation experiments, AI-Rad Companion Brain MR WMH follow-up feature passed the acceptance criteria, the Pearson correlation coefficient between the new or enlarged WMH volumes estimated by our software and ground truth annotation was 0.76. The segmentation accuracy of the new or enlarged WMH reaches an average Dice score of 0.59. The average F1-score of new or enlarged WMH detection is 0.71.

Subject DeviceSiemensGEPhilips
# Data403030
Dice
Mean0.640.560.55
Med0.670.600.59
STD0.150.140.16
95% CI[0.60, 0.69][0.51, 0.61][0.50, 0.61]
ASSD
Mean0.020.090.04
Med0.000.010.00
STD0.060.230.11
95% CI[0.00, 0.04][0.03, 0.19][0.00, 0.08]

Page 16

Subject Device
GenderField StrengthIcobrain (K192130)
AllMF1.5 T3.0 T
# Data1003961386251
Dice
Mean0.600.600.590.620.580.58
Med0.620.600.630.640.60N.A.
STD0.140.120.170.150.16N.A.
95% CI[0.57, 0.63][0.56, 0.64][0.54, 0.63][0.57, 0.66][0.54, 0.62]N.A.
ASSD
Mean0.050.040.050.080.03N.A.
Med0.000.000.000.010.00N.A.
STD0.150.110.170.210.08N.A.
95% CI[0.02, 0.08][0.01, 0.08][0.02, 0.10][0.03, 0.16][0.01, 0.05]N.A.
Subject Device
SiemensGEPhilips
# Data403030
Dice
Mean0.640.560.55
Med0.670.600.59
STD0.150.140.16
95% CI[0.60, 0.69][0.51, 0.61][0.50, 0.61]
ASSD
Mean0.020.090.04
Med0.000.010.00
STD0.060.230.11
95% CI[0.00, 0.04][0.03, 0.19][0.00, 0.08]

DICE and ASSD Results for the AI-Rad Companion Brain MR White Matter Hyperintensities Algorithm

White Matter Hyperintensities Follow-up Features

Performance testing for AI-Rad Companion Brain MR WMH follow-up feature was performed on 1.5T and 3T test datasets from 165 subjects, which included Multiple Sclerosis patients (MS) and Alzheimer's patients (AD) scanned using Siemens, GE and Philips scanners. Testing data has balanced distribution with respect to gender and age of the patient according to target patient population.

Accuracy was validated by comparing the results of the subject device to manual annotated ground truth from three radiologists. Acceptance criteria for above experiments were defined based on a literature review. In all validation experiments, AI-Rad Companion Brain MR WMH follow-up feature passed the acceptance criteria, the Pearson correlation coefficient between the new or enlarged WMH volumes estimated by our software and ground truth annotation was 0.76. The segmentation accuracy of the new or enlarged WMH reaches an average Dice score of 0.59. The average F1-score of new or enlarged WMH detection is 0.71.

Page 17

For each dataset, three sets of ground truth of white matter hyperintensity changes between two time points are annotated manually. Each set is annotated by a disjoin group of annotator, reviewer and clinical expert, with the expert randomly assigned per case. For each test dataset, the three initial annotations are annotated by three different in-house annotators, then each initial annotation is reviewed by the in-house reviewer. Afterwards, each initial annotation is reviewed by the referred clinical expert.

Standard Annotation Process:

For each dataset, three sets of ground truth of white matter hyperintensity changes between two time points are annotated manually. Each set is annotated by a disjoint group of annotator, reviewer, and clinical expert, with the expert randomly assigned per case to minimize annotation bias. For each test dataset, the three initial annotations are annotated by three different in-house annotators. Then, each initial annotation is reviewed by the in-house reviewer. Afterwards, each initial annotation is reviewed by the referred clinical expert. The clinical expert reviews and corrects the initial annotation of the changed WMH areas according to the annotation protocol. If the corrections are significant and time-consuming, the corrections are communicated to the annotator for correction and then re-reviewed.

Testing & Training Data Independence:

WMH follow-up algorithm does not include any machine learning/ deep learning component. The training data used for the fine tuning the hyper parameters of WMH follow-up algorithm is independent of the data used to test the white matter hyperintensity algorithm follow up algorithm.

11. Summary of Nonclinical Tests

Based on the nonclinical performance documented within the Scientific Evaluation, AI-Rad Companion Brain MR VA60 was found to have a safety and effectiveness profile that is similar to the predicate. Since the predicate device was cleared based on the results of the prior conducted scientific evaluation, the same methodology was required to support the substantial equivalence. The nonclinical data and verification and validation results supports the safety and effectiveness of the subject device in that it should performs comparable to the predicate device that is currently marketed.

12. Summary of Clinical Tests

The predicate (K232305) was not validated using clinical tests and therefore no clinical tests were conducted to test the performance and functionality of the modifications introduced within AI-Rad Companion Brain MR. Verification and validation of the enhancements and improvements have been performed and these modifications have been validated for their intended use. No animal testing has been performed on the subject device.

13. Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.

Risk management is ensured via ISO 14971:2019 compliance to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the

Page 18

development of the product. These risks are controlled via measures realized during software development, testing and product labeling.

Furthermore, the device is intended for healthcare professionals familiar with the post processing of magnetic resonance images.

14. Conclusion

Based on the discussion and validation testing and performance data above, the proposed device is determined to be as safe and effective as its predicate device, AI-Rad Companion Brain MR VA50 (K232305).

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).