K Number
K253000

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-10-16

(28 days)

Product Code
Regulation Number
878.4350
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CryoFreeze Wart and Skin Tag Remover is intended for the over-the-counter treatment of common warts and plantar warts in individuals ages 4 years and older and the over-the-counter treatment of skin tags in individuals ages 22 and older.

Device Description

The CryoFreeze Wart and Skin Tag Remover is an over-the-counter (OTC) cryotherapy device designed for the treatment of common warts, plantar warts, and skin tags. The device includes:

  • A pressurized aerosol canister filled with a liquid mixture of compressed gases
  • Applicator swabs
  • Detailed instructions, including illustrated descriptions

To use the device, the user must insert an applicator swab into the actuator, saturate the swab with cryogen, and apply the foam tip of the swab directly onto the lesion, holding it in place for 40 seconds. After removal, the frozen tissue will appear white and begin to thaw, typically taking 20-40 seconds. The device is non-sterile and intended for home use by laypersons. No special training or sterilization is required for operation.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - CryoFreeze Wart and Skin Tag Remover

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov

CryoSurgery, Inc.
Kaitlyn Clements
Regulatory Affairs Coordinator
5829 Old Harding Pike
Nashville, Tennessee 37205

Re: K253000
Trade/Device Name: CryoFreeze Wart and Skin Tag Remover
Regulation Number: 21 CFR 878.4350
Regulation Name: Cryosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: GEH
Dated: September 16, 2025
Received: September 18, 2025

Dear Kaitlyn Clements:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

October 16, 2025

Page 2

U.S. FOOD & DRUG ADMINISTRATION

CryoSurgery, Inc.
Kaitlyn Clements
Regulatory Affairs Coordinator
5829 Old Harding Pike
Nashville, Tennessee 37205

October 16, 2025

Re: K253000
Trade/Device Name: CryoFreeze Wart and Skin Tag Remover
Regulation Number: 21 CFR 878.4350
Regulation Name: Cryosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: GEH
Dated: September 16, 2025
Received: September 18, 2025

Dear Kaitlyn Clements:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 3

K253000 - Kaitlyn Clements Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 4

K253000 - Kaitlyn Clements Page 3

Sincerely,

Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2025.10.16 11:08:36 -04'00'

Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 5

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.K253000
Please provide the device trade name(s).

CryoFreeze Wart and Skin Tag Remover

Please provide your Indications for Use below.

The CryoFreeze Wart and Skin Tag Remover is intended for the over-the-counter treatment of common warts and plantar warts in individuals ages 4 years and older and the over-the-counter treatment of skin tags in individuals ages 22 and older.

Please select the types of uses (select one or both, as applicable).☐ Prescription Use (Part 21 CFR 801 Subpart D)☒ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 6

510(k) Summary

CryoFreeze Wart and Skin Tag Remover

Version: 1 Page 1 of 4

1. Submission Sponsor
Ronald McDow
CryoSurgery, Inc.
5829 Old Harding Pike
Nashville, TN 37205
(615) 354-0414
info@cryosurgeryinc.com

2. Primary Correspondent
Kaitlyn Clements
5829 Old Harding Pike
Nashville, TN 37205
kclements@cryosurgeryinc.com

3. Date Prepared
10/15/2025

4. Device Identification
Trade / Proprietary Name: CryoFreeze Wart and Skin Tag Remover
Common / Used Name: OTC Cryosurgical Device
510(k) Number: K253000
Regulation: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

5. Predicate Device
Device Name: CryoFreeze Wart and Skin Tag Remover
510(k) Number: K243487
Regulation Number: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

6. Predicate Device
Device Name: Skin Clinic Nitro Clear
510(k) Number: K242932
Regulation Number: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

7. Predicate Device
Device Name: Oystershell Pixie Skin Tag Remover

K253000

Page 7

510(k) Summary

CryoFreeze Wart and Skin Tag Remover

K253000

1. Submission Sponsor
Ronald McDow
CryoSurgery, Inc.
5829 Old Harding Pike
Nashville, TN 37205
(615) 354-0414
info@cryosurgeryinc.com

2. Primary Correspondent
Kaitlyn Clements
5829 Old Harding Pike
Nashville, TN 37205
kclements@cryosurgeryinc.com

3. Date Prepared
10/15/2025

4. Device Identification
Trade / Proprietary Name: CryoFreeze Wart and Skin Tag Remover
Common / Used Name: OTC Cryosurgical Device
510(k) Number: K253000
Regulation: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

5. Predicate Device
Device Name: CryoFreeze Wart and Skin Tag Remover
510(k) Number: K243487
Regulation Number: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

6. Predicate Device
Device Name: Skin Clinic Nitro Clear
510(k) Number: K242932
Regulation Number: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

7. Predicate Device
Device Name: Oystershell Pixie Skin Tag Remover

Version: 1 Page 1 of 4

Page 8

510(k) Summary

CryoFreeze Wart and Skin Tag Remover

Version: 1 Page 2 of 4

510(k) Number: K240226
Regulation Number: 878.4350 - Cryosurgical unit and accessories
Product Code: GEH
Class: 2

8. Description of the Subject Device

The CryoFreeze Wart and Skin Tag Remover is an over-the-counter (OTC) cryotherapy device designed for the treatment of common warts, plantar warts, and skin tags. The device includes:

  • A pressurized aerosol canister filled with a liquid mixture of compressed gases
  • Applicator swabs
  • Detailed instructions, including illustrated descriptions

To use the device, the user must insert an applicator swab into the actuator, saturate the swab with cryogen, and apply the foam tip of the swab directly onto the lesion, holding it in place for 40 seconds. After removal, the frozen tissue will appear white and begin to thaw, typically taking 20-40 seconds. The device is non-sterile and intended for home use by laypersons. No special training or sterilization is required for operation.

9. Indications for Use

The CryoFreeze Wart and Skin Tag Remover is intended for the over-the-counter treatment of common warts and plantar warts in adults and children age 4 years or older and over-the-counter treatment of skin tags in adults age 22 or older.

10. Technological Comparison

The modified CryoFreeze Wart and Skin Tag Remover cryosurgical device has the same technological characteristics as the predicate device, CryoFreeze Wart and Skin Tag Remover (K243487), with the exception of a modification to the cryogen formulation. Both devices have the same design, principle of operation, mode of use, actuator and valve system, applicator materials, packaging configuration, and labeling.

CharacteristicSubject Device: CryoFreeze Wart and Skin Tag RemoverPredicate Device: CryoFreeze Wart and Skin Tag RemoverComparison
Trade Name / Device NameCryoFreeze Wart and Skin Tag RemoverCryoFreeze Wart and Skin Tag RemoverSame
510(k) NumberK253000K243487n/a
Indications for UseThe CryoFreeze Wart and Skin Tag Remover is intended for use in the over-the-counter treatment of common warts and plantar warts in adults and children age 4 years or older andThe CryoFreeze Wart and Skin Tag Remover is intended for use in the over-the-counter treatment of common warts and plantar warts in adults and children age 4 years or older andSame

K253000

Page 9

510(k) Summary

CryoFreeze Wart and Skin Tag Remover

K253000

over-the-counter treatment of skin tags in adults age 22 years or older.over-the-counter treatment of skin tags in adults age 22 years or older.
Type of UseOver-the-counterOver-the-counterSame
Design• Cryogen canister with attached safety shield• Foam-tipped applicators• Disposable tweezers• Cryogen canister with attached safety shield• Foam-tipped applicators• Disposable tweezersSame
Mechanism of ActionExtreme cold destroys the target tissueExtreme cold destroys the target tissueSame
Mode of UseCryogen is dispensed into foam applicator which is then applied to the lesionCryogen is dispensed into foam applicator which is then applied to the lesionSame
Storage and Safety Conditions-Keep away from fire or flame-Do not smoke while using the product-Do not puncture or incinerate canister-Do not expose to heat or store at temperatures above 122°F.-Store at room temperature away from heat-Keep away from fire or flame-Do not smoke while using the product-Do not puncture or incinerate canister-Do not expose to heat or store at temperatures above 122°F.-Store at room temperature away from heatSame

8.1. Comparison Table vs Additional Predicates

CharacteristicSubject Device: CryoFreeze Wart and Skin Tag RemoverPredicate Device: Skin Clinic Nitro Clear Wart RemoverPredicate Device: Oystershell Pixie Skin Tag RemoverComparison
Trade Name / Device NameCryoFreeze Wart and Skin Tag RemoverSkin Clinic Nitro Clear Wart RemoverOystershell Pixie Skin Tag RemoverSame
510(k) NumberK253000K242932K240226n/a
Indications for UseThe CryoFreeze Wart and Skin Tag Remover is intended for use in the over-the-counter treatment of common warts and plantar warts in adults and children age 4 years or older and over-the-counter treatment of skin tags inThe Skin Clinic Nitro Clear Wart Remover is intended for the OTC treatment of common warts and plantar warts in adults and children 4 years of age and older.Pixie Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.Different

K253000

Version: 1 Page 3 of 4

Page 10

510(k) Summary

CryoFreeze Wart and Skin Tag Remover

K253000

adults age 22 years or older.over-the-counter treatment of skin tags in adults age 22 years or older.
Type of UseOver-the-counter• Cryogen canister with attached safety shieldOver-the-counter• Cryogen canister with attached safety shieldSameSame
Design• Foam-tipped applicators• Disposable tweezers• Foam-tipped applicators• Disposable tweezers
Mechanism of ActionExtreme cold destroys the target tissueExtreme cold destroys the target tissueSame
Mode of UseCryogen is dispensed into foam applicator which is then applied to the lesion-Keep away from fire or flameCryogen is dispensed into foam applicator which is then applied to the lesion-Keep away from fire or flameSameSame
Storage and Safety Conditions-Do not smoke while using the product-Do not smoke while using the product
-Do not puncture or incinerate canister-Do not puncture or incinerate canister
-Do not expose to heat or store at temperatures above 122°F.-Do not expose to heat or store at temperatures above 122°F.
-Store at room temperature away from heat-Store at room temperature away from heat

8.1. Comparison Table vs Additional Predicates

CharacteristicSubject Device: CryoFreeze Wart and Skin Tag RemoverPredicate Device: Skin Clinic Nitro Clear Wart RemoverPredicate Device: Oystershell Pixie Skin Tag RemoverComparison
Trade Name / Device NameCryoFreeze Wart and Skin Tag RemoverSkin Clinic Nitro Clear Wart RemoverOystershell Pixie Skin Tag RemoverSame
510(k) NumberK253000K242932K240226n/a Different
Indications for UseThe CryoFreeze Wart and Skin Tag Remover is intended for use in the over-the-counter treatment of common warts and plantar warts in adults and children age 4 years or older and over-the-counter treatment of skin tags in adults age 22 years or older.The Skin Clinic Nitro Clear Wart Remover is intended for the OTC treatment of common warts and plantar warts in adults and children 4 years of age and older.Pixie Skin Tag is a cryogenic device for home use intended to treat skin tags in persons over the age of 21.

Version: 1 Page 3 of 4

Page 11

510(k) Summary

CryoFreeze Wart and Skin Tag Remover

Version: 1 Page 4 of 4

Mechanism of Actionadults age 22 years or older.Extreme cold destroys the target tissueExtreme cold destroys the target tissueExtreme cold destroys the target tissueSame

9. Non-Clinical Performance Data

Performance testing has demonstrated that the CryoFreeze Wart and Skin Tag Remover cryosurgical device meets the same performance specifications as the predicate device, including output temperature, freeze charge hold time, and visual confirmation of ice ball formation. The modification to the cryogen formulation does not alter the device's intended use, operating principle, or overall safety and effectiveness.

10. Statement of Substantial Equivalence

Based on the information presented, it is concluded that the CryoFreeze Wart and Skin Tag Remover device is substantially equivalent to the predicate devices.

K253000

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.