K Number
K252762

Validate with FDA (Live)

Device Name
HALO AP Dx
Manufacturer
Date Cleared
2025-11-25

(88 days)

Product Code
Regulation Number
864.3700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the interoperable components specified in the Table below.

Device Description

HALO AP Dx, version 2.4 (abbreviated as v2.4), is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management, and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner or the Leica Biosystems Imaging, Inc. Aperio GT 450 DX.

HALO AP Dx is typically operated as follows:

  1. Image acquisition is performed using the predicate device. The operator performs quality control of the digital slides per the instructions of the predicate device and lab specifications to determine if re-scans are necessary.

  2. Once image acquisition is complete, according to its Instructions for Use, the unaltered image is saved in an external image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.

  3. Depending upon a laboratory's workflow, the scanned images may first be reviewed by histotechnicians to confirm image quality and initiate any re-scans. After review, the histotechnician may modify the case status and make it available to the pathologist.

  4. The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.

  5. The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
    a. Zoom and pan the image.
    b. Measure distances and areas in the image.
    c. Annotate images.
    d. View multiple images side by side in a synchronized fashion.

  6. The above steps are repeated as necessary.

After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System.

HALO AP Dx operates and is validated for use with the FDA cleared components listed in the intended use statement table above.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - HALO AP Dx

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

November 25, 2025

Indica Labs, LLC
Michele Fredrickson
Director, Quality & Regulatory Affairs
8700 Education Pl NW, Building B
Albuquerque, NM 87114

Re: K252762
Trade/Device Name: HALO AP Dx
Regulation Number: 21 CFR 864.3700
Regulation Name: Whole Slide Imaging System
Regulatory Class: Class II
Product Code: QKQ
Dated: August 25, 2025
Received: August 29, 2025

Dear Michele Fredrickson:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252762 - Michele Fredrickson
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252762 - Michele Fredrickson
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shyam Kalavar -S

Shyam Kalavar
Deputy Branch Chief
Division of Molecular Genetics and Pathology
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K252762

Device Name
HALO AP Dx

Indications for Use (Describe)

HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the interoperable components specified in the Table below.

Table: Interoperable Components of HALO AP Dx

Scanner HardwareScanner Output File FormatInteroperable Displays
Leica Aperio GT 450 DX scannerSVSDell U3223QE
Barco MDPC-8127
DICOMDell U3223QE
Hamamatsu NanoZoomer S360MD Slide scannerNDPIBarco MDPC-8127

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

HALO AP Dx

Date Prepared: November 24, 2025

Submitter

Indica Labs, LLC
8700 Education Place NW, Building B
Albuquerque, NM, 87114 USA

Contact Person

Michele Fredrickson, Director Quality & Regulatory
Tel: 505-492-0972

Device

Proprietary/Trade Name: HALO AP Dx
Version Number: 2.4
Regulation Name: Whole Slide Imaging System
Regulation: 21 CFR §864.3700
Regulatory Class: Class II
Product Classification Code: QKQ
Review Panel: 88 – Pathology
510 (k) Submission Number: K252762

Predicate Devices

Predicate Manufacturer: Hamamatsu Photonics K.K.
Predicate Trade Name: NanoZoomer S360MD Slide scanner system
Predicate 510(k): K233027

Predicate Manufacturer: Leica Biosystems Imaging, Inc.
Predicate Trade Name: Aperio GT 450 DX
Predicate 510(k): K232202

No reference devices were used in this submission.

Intended Use

HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens.

It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy

Page 6

review when making a diagnostic decision. HALO AP Dx is intended for use with the interoperable components specified in the Table below.

Table: Interoperable Components of HALO AP Dx

Scanner HardwareScanner Output File FormatInteroperable Displays
Leica Aperio GT 450 DX scannerSVSDell U3223QE, Barco MDPC-8127
DICOMDell U3223QE
Hamamatsu NanoZoomer S360MD Slide scannerNDPIBarco MDPC-8127

Device Description

HALO AP Dx, version 2.4 (abbreviated as v2.4), is a browser-based software-only device intended to aid pathology professionals in viewing, manipulating, management, and interpretation of digital pathology whole slide images (WSI) of glass slides obtained from the Hamamatsu Photonics K.K. NanoZoomer S360MD scanner or the Leica Biosystems Imaging, Inc. Aperio GT 450 DX.

HALO AP Dx is typically operated as follows:

  1. Image acquisition is performed using the predicate device. The operator performs quality control of the digital slides per the instructions of the predicate device and lab specifications to determine if re-scans are necessary.

  2. Once image acquisition is complete, according to its Instructions for Use, the unaltered image is saved in an external image storage location. HALO AP Dx ingests the image, and a copy of image metadata is stored in the subject device's database to improve viewing response times.

  3. Depending upon a laboratory's workflow, the scanned images may first be reviewed by histotechnicians to confirm image quality and initiate any re-scans. After review, the histotechnician may modify the case status and make it available to the pathologist.

  4. The reading pathologist selects a patient case from a selected worklist within HALO AP Dx whereby the subject device fetches the associated images from external image storage.

  5. The reading pathologist uses the subject device to view the images and can perform the following actions, as needed:
    a. Zoom and pan the image.
    b. Measure distances and areas in the image.
    c. Annotate images.
    d. View multiple images side by side in a synchronized fashion.

  6. The above steps are repeated as necessary.

After viewing all images belonging to a particular case (patient), the pathologist will make a diagnosis which is documented in another system, such as a Laboratory Information System.

HALO AP Dx operates and is validated for use with the FDA cleared components listed in the intended use statement table above.

Page 7

Table 1: Computer Environment / System for Use with HALO AP Dx

ComponentRequirement
Operating SystemWindows 11
Memory16 GB or more
ProcessorIntel Core i7 CPU
Supported BrowsersGoogle Chrome version 138 or higher, Microsoft Edge version 138 or higher

Table 2: Server System Requirements

ServerOperating SystemMemoryProcessor
Component Server, Processing Node Server, File Monitor ServerWindows Server 2022 Only x64 (64 bit) operating systems are supported16 GB or more8 CPU cores or more
Traefik Reverse ProxyWindows Server 2022, Ubuntu 22.042GB, per 150 concurrent users2 CPU Cores, per 150 concurrent users
MySQL ServerWindows Server 2022, Ubuntu 22.04 Only x64 (64 bit) operating systems are supported16 GB or more4 CPU Cores or more

Table 3: Server Configurations

ComponentSpecifications
Network Connectivity1 Gbps (10 Gbps recommended) LAN connection between services
Antivirus SoftwareThe following antivirus software has been validated for use with Windows Server: - Microsoft Windows Defender - Trend Micro - Webroot

Page 8

Summary of Technological Characteristics

Comparison with Predicate – Hamamatsu NanoZoomer S360MD Slide scanner system

ItemPredicate (K233027)Subject Device (K252762)
Device Trade NameHamamatsu NanoZoomer S360MD Slide scanner systemHALO AP Dx
Indications for UseThe NanoZoomer S360MD Slide scanner system ("NanoZoomer System") is an automated digital slide creation, viewing, and management system. The NanoZoomer System is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded ("FFPE") tissue. The NanoZoomer System is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. The NanoZoomer System comprises the NanoZoomer S360MD Slide scanner, the NZViewMD Software and a compatible display that has been 510(k) cleared for use with the NanoZoomer system or a 510(k)-cleared display that has been assessed in accordance with the Predetermined Change Control Plan (PCCP) for qualifying additional compatible displays. The NanoZoomer System is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using NanoZoomer System.HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision. HALO AP Dx is intended for use with the interoperable components specified in the Table below. Table: Interoperable Components of HALO AP Dx Scanner Hardware: Leica Aperio GT 450 DX scanner, Hamamatsu NanoZoomer S360MD Slide scanner; Scanner Output File Format: SVS, DICOM, NDPI; Interoperable Displays: Dell U3223QE, Barco MDPC-8127
Specimen TypeSurgical pathology slides prepared from FFPE tissueSame

Page 9

Comparison with Predicate – Aperio GT 450 DX

ItemPredicate (K233027)Subject Device (K252762)
Diagnostic Image File FormatHamamatsu NDPI FileSVS and DICOM – Aperio GT 450 DX, NDPI – Hamamatsu NanoZoomer S360MD Slide scanner system
Image Manipulation and Review FunctionsFunctions for continuous panning and zooming, annotations, distance/area measurements, track visited areas, and image adjustments.Same
Type of Software ApplicationPC-based installed applicationInternet browser-based application
Principle of OperationDuring review, the pathologist opens the WSI images acquired with the scanner from the network storage, performs further QC, and reviews and interprets the WSI images to make a diagnosis.Same
Image StorageUser-supplied network attached storageSame
End User's InterfaceNZViewMDHALO AP Dx
ScannerHamamatsu NanoZoomer S360MDAperio GT 450 DX, Hamamatsu Nanozoomer S360MD
ItemPredicate (K232202)Subject Device (K252762)
Device Trade NameAperio GT 450 DXHALO AP Dx
Indications for UseThe Aperio GT 450 DX is an automated digital slide creation and viewing system. The Aperio GT 450 DX is intended for in vitro diagnostic use as an aid to the pathologist to review and interpret digital images of surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue. The Aperio GT 450 DX is for creation and viewing of digital images of scanned glass slides that would otherwise be appropriate for manual visualization by conventional light microscopy.HALO AP Dx is a software only device intended as an aid to the pathologist to review, interpret and manage digital images of scanned surgical pathology slides prepared from formalin-fixed paraffin embedded (FFPE) tissue for the purposes of pathology primary diagnosis. HALO AP Dx is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the quality of the images obtained and, where necessary, use conventional light microscopy review when making a diagnostic decision.

Page 10

ItemPredicate (K232202)Subject Device (K252762)
Aperio GT 450 DX is comprised of the Aperio GT 450 DX scanner, which generates images in the Digital Imaging and Communications in Medicine (DICOM) and in the ScanScope Virtual Slide (SVS) file formats, the Aperio WebViewer DX viewer, and the displays. The Aperio GT 450 DX is intended to be used with the interoperable components specified in Table 1. Table 1: Interoperable components of Aperio GT 450 DX Scanner Hardware: Leica Aperio GT 450 DX scanner; Scanner Output File Format: SVS, DICOM; Interoperable Viewing Software: Aperio WebViewer DX, Sectra Digital Pathology Module (3.3); Interoperable Displays: Barco MDPC8127, Dell UP3017, Dell U3023E, Dell U3223QE. The Aperio GT 450 DX is not intended for use with frozen section, cytology, or non-FFPE hematopathology specimens. It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using the Aperio GT 450 DX.HALO AP Dx is intended for use with the interoperable components specified in the Table below. Table: Interoperable Components of HALO AP Dx Scanner Hardware: Leica Aperio GT 450 DX scanner, Hamamatsu NanoZoomer S360MD Slide scanner; Scanner Output File Format: SVS, DICOM, NDPI; Interoperable Displays: Dell U3223QE, Barco MDPC8127
Specimen TypeSurgical pathology slides prepared from FFPE tissueSame
Diagnostic Image File FormatSVS and DICOMSVS and DICOM – Aperio GT 450 DX, NDPI – Hamamatsu NanoZoomer S360MD Slide scanner system
Image Manipulation andFunctions for continuous panning and zooming, annotations, distance/areaSame

Page 11

ItemPredicate (K232202)Subject Device (K252762)
Review Functionsmeasurements, track visited areas, and image adjustments.
Type of Software ApplicationInternet browser-based applicationSame
Principle of OperationDuring review, the pathologist opens the WSI images acquired with the scanner from the network storage, performs further QC, and reviews and interprets the WSI images to make a diagnosis.Same
Image StorageUser-supplied network attached storageSame
End User's InterfaceWebViewer DX for Leica SVS, Sectra Digital Pathology Module (3.3) for Leica SVS and DICOMHALO AP Dx
ScannerAperio GT 450 DXAperio GT 450 DX, NanoZoomer S360MD Slide scanner

Summary of Studies (Performance Data)

Performance DataDescription
Identical image reproductionPixel-level comparisons were made to compare images displayed by HALO AP Dx (v2.4) and the predicate devices' image review and manipulation software (IRMS) as described below for the same scanned image file format as captured from the same FDA cleared scanning system to validate identical image display on the same compatible display monitors.
Scanning SystemFile FormatPredicate Device IRMSSubject DeviceDisplay
Aperio GT 450 DXSVSWebViewer DX (Google Chrome)HALO AP Dx (Google Chrome)Barco MDPC8127
WebViewer DX (Microsoft Edge)HALO AP Dx (Microsoft Edge)Dell U3223QE
DICOMSectra Digital Pathology Module (IDS7)HALO AP Dx (Google Chrome)Dell U3223QE
Sectra Digital Pathology Module (IDS7)HALO AP Dx (Microsoft Edge)
Hamamatsu NanoZoomer S360MD Slide scannerNDPINZViewMDHALO AP Dx (Google Chrome)Barco MDPC8127
NZViewMDHALO AP Dx (Microsoft Edge)

Page 12

Thirty (30) formalin-fixed paraffin-embedded (FFPE) slides representing a range of anatomical sites and diagnoses were scanned on the predicate scanning systems at 40x magnification. Three ROIs were identified per slide and captured at two magnification levels (10x and 40x) across each viewer, browser (if applicable), and display combination. Each predicate IRMS and subject device ROI was registered and the difference between each pixel was computed using the 2000 International Commission on Illumination color difference CIEDE2000 metric (ΔE00). The 95th percentile of the pixelwise ΔE00 differences for each ROI in the compared image sets was reported. An acceptance criterion where each analyzed ROI image pair shall not have a 95th percentile of the pixelwise ΔE00 differences value greater than 3 was applied.

The final test results showed that 95th percentile of the pixelwise ΔE00 differences for each ROI was less than 3 for all compared image sets. Therefore, HALO AP Dx and the predicate IRMS image display has been found to be pixelwise equivalent.

Performance DataDescription
Turnaround timeTurnaround times for image loading, panning, and zooming have been determined and found to be adequate for the intended use of the subject device. The system requirements have been fulfilled: • When selecting a case, it should not take longer than 4 seconds until the image is fully loaded. • When panning the image, it should not take longer than 3 seconds until the image is fully loaded.
MeasurementsMeasurement accuracy has been verified using a test image containing objects with known sizes.
Human factors testingA human factors study was conducted per FDA's Guidance on Applying Human Factors and Usability Engineering to Medical Devices (2016). The study was designed around critical and essential user tasks and use scenarios performed by representative users was conducted for previously cleared HALO AP Dx (version 2.1) in K232833. No new human factors study was performed for HALO AP Dx (version 2.4).

Substantial Equivalence Comparison

The proposed subject device has similar Indications for Use, principles of operation, and technical characteristics as compared to each predicate device's IRMS. Performance testing results confirm equivalent image reproducibility, measurement accuracy, and similar or better turnaround times for common operations. Therefore, the subject device is confirmed to be substantially equivalent to the predicate devices.

Conclusion

Based on the information provided, when HALO AP Dx is used with the Hamamatsu Photonics K.K. NanoZoomer S360MD Slide scanner or Leica Biosystems Imaging, Inc. Aperio GT 450 DX and the interoperable monitor displays, it has similar Indications for Use, Functional, and Technological Characteristics as the applicable predicate IRMS. The performance data support substantial equivalence of HALO AP Dx to the predicate devices.

§ 864.3700 Whole slide imaging system.

(a)
Identification. The whole slide imaging system is an automated digital slide creation, viewing, and management system intended as an aid to the pathologist to review and interpret digital images of surgical pathology slides. The system generates digital images that would otherwise be appropriate for manual visualization by conventional light microscopy.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information:
(i) The indications for use must specify the tissue specimen that is intended to be used with the whole slide imaging system and the components of the system.
(ii) A detailed description of the device and bench testing results at the component level, including for the following, as appropriate:
(A) Slide feeder;
(B) Light source;
(C) Imaging optics;
(D) Mechanical scanner movement;
(E) Digital imaging sensor;
(F) Image processing software;
(G) Image composition techniques;
(H) Image file formats;
(I) Image review manipulation software;
(J) Computer environment; and
(K) Display system.
(iii) Detailed bench testing and results at the system level, including for the following, as appropriate:
(A) Color reproducibility;
(B) Spatial resolution;
(C) Focusing test;
(D) Whole slide tissue coverage;
(E) Stitching error; and
(F) Turnaround time.
(iv) Detailed information demonstrating the performance characteristics of the device, including, as appropriate:
(A) Precision to evaluate intra-system and inter-system precision using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(B) Reproducibility data to evaluate inter-site variability using a comprehensive set of clinical specimens with defined, clinically relevant histologic features from various organ systems and diseases. Multiple whole slide imaging systems, multiple sites, and multiple readers must be included.
(C) Data from a clinical study to demonstrate that viewing, reviewing, and diagnosing digital images of surgical pathology slides prepared from tissue slides using the whole slide imaging system is non-inferior to using an optical microscope. The study should evaluate the difference in major discordance rates between manual digital (MD) and manual optical (MO) modalities when compared to the reference (
e.g., main sign-out diagnosis).(D) A detailed human factor engineering process must be used to evaluate the whole slide imaging system user interface(s).
(2) Labeling compliant with 21 CFR 809.10(b) must include the following:
(i) The intended use statement must include the information described in paragraph (b)(1)(i) of this section, as applicable, and a statement that reads, “It is the responsibility of a qualified pathologist to employ appropriate procedures and safeguards to assure the validity of the interpretation of images obtained using this device.”
(ii) A description of the technical studies and the summary of results, including those that relate to paragraphs (b)(1)(ii) and (iii) of this section, as appropriate.
(iii) A description of the performance studies and the summary of results, including those that relate to paragraph (b)(1)(iv) of this section, as appropriate.
(iv) A limiting statement that specifies that pathologists should exercise professional judgment in each clinical situation and examine the glass slides by conventional microscopy if there is doubt about the ability to accurately render an interpretation using this device alone.