K Number
K252588

Validate with FDA (Live)

Date Cleared
2025-12-04

(111 days)

Product Code
Regulation Number
890.3800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

The Mobility Scooter (T5) is a motor driven, indoor and outdoor transportation vehicle. It has an overall dimension of 1240mm610mm875mm and a steel frame. It features a foldable tiller, removable battery pack, and disassembles for transport. It has 2 rear wheels, a ground clearance of 20 mm, and a maximum loading capacity of 150kg on level ground. The minimum turn radius is 2175mm, and it is powered by a 24V 350W motor with rear-wheel drive. Braking is handled by an electromagnetic brake. It uses a Li-ion Battery 25.6Vdc 15Ah with a 29.4V/2A charger. The maximum speed is 7.92 km/h, and it can handle a maximum slope of 6°. The travel distance is 18.5 km, with a brake time of less than 1 second and a brake distance of ≤1.3m (horizontal-forward at max speed). The total mass is 54 kg. It is intended for indoor use and restricted outdoor use on pavements or paved footpaths only. It does not have a remote control. The folding mechanism includes a foldable column and a removable basket. The front and rear tire diameters are 9 inches.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Mobility Scooter (T5)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.02

December 4, 2025

Zhejiang Taotao Vehicle Co.,Ltd.
℅ Eva Li
Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave.
Shanghai, 200122
China

Re: K252588
Trade/Device Name: Mobility Scooter (T5)
Regulation Number: 21 CFR 890.3800
Regulation Name: Motorized Three-Wheeled Vehicle
Regulatory Class: Class II
Product Code: INI
Dated: November 4, 2025
Received: November 4, 2025

Dear Eva Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K252588 - Eva Li
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K252588 - Eva Li
Page 3

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K252588

Please provide the device trade name(s).
Mobility Scooter (T5)

Please provide your Indications for Use below.
It is motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Please select the types of uses (select one or both, as applicable).
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

Mobility Scooter Page 11 of 50

Page 5

510(K) Summary

Application Information: Zhejiang Taotao Vehicle Co.,LTD.
Device Name: Mobility Scooter
Model: T5
Classification: II
Product Code: INI
Regulation: 890.3800
Panel: Motorized three-wheeled vehicle

Date prepared: Aug 5,2025

Substantial Equivalence Information:
510K number: K231428
Device Name: Mobility Scooter
Model: W3431D
Zhejiang Innuovo Rehabilitation Devices Co.,Ltd.

1. Product Parameter

Table 1 General Comparison

Elements of ComparisonSubject Device (K252588)Predicate Device (K231428)Remark
ManufacturerZhejiang Taotao Vehicle Co.,LTD.Zhejiang Innuovo Rehabilitation Devices Co.,LtdNA
Device nameMobility ScooterMobility ScooterSame
Model(s)T5W3431D--
Indications for useIt is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.Same
Overall dimension1240mm610mm875mm1020 mm x 500 mm x 840 mmSimilar
Frame MaterialSteelSteel
Frame styleFoldable tiller, removable battery pack, disassemble for transportFoldable seat, removable battery pack, disassemble for transportSimilar
Rear Wheels Quantity22Same
Ground clearance20 mm45 mmSimilar

Page 6

Max Loading( on level ground)150kg120kgSimilar
Min Turn Radius2175mm1650mmSimilar
Motor output24 V 350W24 V 180WSimilar
Drive SystemRear Wheel DriveRear Wheel DriveSame
BrakesElectromagnetic brakeElectromagnetic brakeSame
BatteryLi-ion Battery 25.6Vdc 15AhLead-acid 12V12Ah*2Similar
Charger29.4V/2A24V/2ASimilar
Max Speed7.92 km/h6 km/hSame
Max SlopeSimilar
Travel Distance18.5 km15 kmSimilar
Time to brake< 1 s0.7-1sSimilar
Brake Distance Normal operation (Horizontal-Forward Max speed)≤1.3m≤1.5mSimilar
Total mass54 kg42kgSimilar
Operating surface & environmentIndoor use and restricted outdoor use on pavements or paved footpaths only.Indoor use and restricted outdoor use on pavements or paved footpaths only.Same
Remote controlNoneNoneSame
Folding mechanismfoldable column, removable basketFoldable seat, removable battery packdifferent
Stowage size(HLW)6001240560mmNo revelationN/A
Front Tire Diameter9inch190 x 55 mm(15inch)different
Rear Tire Diameter9inch190 x 55 mm(15inch)different

Analysis: Subject device control system has passed the requirements of ISO 7176 series and test results meet its design specification.

Table 2 safety comparison

ItemProposed DevicePredicate DeviceResults
BiocompatibilityAll user directly contacting materials are compliance with ISO10993-1All user directly contacting materials are compliance with ISO10993-1Same
EMCISO7176-21& IEC 60601-1-2:2014+A1:2020ISO7176-21& IEC 60601-1-2:2014+A1:2020Same

Page 7

PerformanceISO7176 seriesISO7176 seriesSame
Label and labelingConforms to FDA RegulatoryConforms to FDA RegulatorySame
ItemProposed DevicePredicate DeviceResults
ISO7176-1The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification.The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification.Same
ISO7176-2The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification.The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification.Same
ISO7176-3The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification.The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification.Same
ISO7176-4The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification.The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification.Same
ISO7176-5The dimensions, mass has been determined after the testing according to the ISO 7176-5,The dimensions, mass has been determined after the testing according to the ISO 7176-5,Same
ISO7176-6The dimensions, mass has been determined after the testing according to the ISO 7176-5,The dimensions, mass has been determined after the testing according to the ISO 7176-5,Same
ISO7176-7The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7,The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7,Same
ISO7176-8All test results meet the requirements in Clause 4 of ISO 7176-8All test results meet the requirements in Clause 4 of ISO 7176-8Same
ISO7176-9The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9Same

Page 8

ISO7176-10The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10,The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10,Same
ISO7176-11The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11Same
ISO7176-13The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involvedThe coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involvedSame
ISO7176-14All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14Same
ISO7176-15The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15Same
ISO7176-16The performance of resistance to ignition meet the requirements of ISO 7176-16The performance of resistance to ignition meet the requirements of ISO 7176-16Same
ISO 7176-21The EMC performance results meet the requirements of ISO 7176-21, IEC 60601-1-2:2014+A1:2020The EMC performance results meet the requirements of ISO 7176-21, IEC 60601-1-2:2014+A1:2020Same
ISO 7176-22The performance of resistance to ignition meet the requirements of ISO 7176-22The performance of resistance to ignition meet the requirements of ISO 7176-22Same
ISO7176-25The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25Same

Page 9

Substantial Equivalence Discussion

The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 7176-16, ISO 7176-21, ISO 7176-22, ISO 7176-25, and FDA guidance Submission for Scooter.

The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/ control criteria following above mentioned standards. And the test results show that the subject product is substantially equivalent to the predicate device in performance.

The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability (Scooter tipping angle), The Dynamic stability (Safe Gradient Maximum Gradient), Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to ignition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers.

The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.

3. Substantially Equivalency Conclusion

Based on the comparison and analysis above, the subject device is determined to be Substantially Equivalent (SE) to the predicate devices, W3431D Scooter from Zhejiang Innuovo Rehabilitation Devices Co.,Ltd. K231428.

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).