(117 days)
a2z-Unified-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominal/pelvic CT images in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 7 specified abdominopelvic findings: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. These findings are intended to be used together as one device. The device supports both cloud-based and on-premises deployment, with integration either directly with healthcare facility systems or through third-party healthcare technology platforms.
a2z-Unified-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for primary diagnosis beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of a2z-Unified-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
a2z-Unified-Triage is a radiological computer-assisted triage and notification software device. The software consists of an algorithmic component that supports both cloud-based and on-premises deployment on standard server hardware. The device processes abdomen/pelvis CT images from clinical imaging systems, analyzing them using artificial intelligence algorithms to detect suspected cases of 7 abdominopelvic conditions: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction.
Following the AI processing, the analysis results are returned to the client system for worklist prioritization. When a suspected case is detected, the software provides analysis results that enable the client system to generate appropriate notifications. These results can include DICOM instance UIDs for key images, which are for informational purposes only, do not contain any marking of the findings, and are not intended for primary diagnosis beyond notification.
Integration with clinical imaging systems facilitates efficient triage by enabling prioritization of suspect cases for review of the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
Here's a detailed summary of the acceptance criteria and the study proving the device meets them, based on the provided FDA clearance letter:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
a2z-Unified-Triage differentiates between two types of findings for regulatory purposes: QAS (Qualitative, Automated, and Subjective) and QFM (Quantitative, Functional, and Measurable).
| Condition Type | Acceptance Criteria | Device Performance (with 95% Confidence Intervals) |
|---|---|---|
| QFM Findings | AUC > 0.95 | |
| Acute Cholecystitis | AUC > 0.95 | AUC: 0.985 [0.972-0.998] (Also provided: High Sensitivity: Se 96.1% [89.2-98.7%], Sp 89.3% [86.6-91.5%]; Sensitivity Biased: Se 92.2% [84.0-96.4%], Sp 95.8% [93.9-97.2%]; Balanced: Se 92.2% [84.0-96.4%], Sp 95.8% [93.9-97.2%]) |
| Acute Pancreatitis | AUC > 0.95 | AUC: 0.994 [0.985-1.000] (Also provided: High Sensitivity: Se 98.0% [92.9-99.4%], Sp 87.8% [84.9-90.3%]; Sensitivity Biased: Se 98.0% [92.9-99.4%], Sp 97.0% [95.3-98.1%]; Balanced: Se 98.0% [92.9-99.4%], Sp 97.0% [95.3-98.1%]; High Specificity: Se 92.9% [86.1-96.5%], Sp 99.8% [99.0-100.0%]) |
| Unruptured AAA | AUC > 0.95 | AUC: 0.995 [0.991-0.999] (Also provided: High Sensitivity: Se 100.0% [95.2-100.0%], Sp 86.3% [83.3-88.8%]; Sensitivity Biased: Se 97.4% [90.9-99.3%], Sp 95.8% [93.9-97.2%]; Balanced: Se 97.4% [90.9-99.3%], Sp 97.5% [95.9-98.5%]) |
| Acute Diverticulitis | AUC > 0.95 | AUC: 0.995 [0.990-1.000] (Also provided: High Sensitivity: Se 98.7% [92.9-99.8%], Sp 89.3% [86.6-91.5%]; Sensitivity Biased: Se 97.4% [90.9-99.3%], Sp 96.8% [95.1-98.0%]; Balanced: Se 97.4% [90.9-99.3%], Sp 96.8% [95.1-98.0%]; High Specificity: Se 94.7% [87.2-97.9%], Sp 98.7% [97.4-99.3%]) |
| Hydronephrosis | AUC > 0.95 | AUC: 0.976 [0.960-0.991] (Also provided: High Sensitivity: Se 89.7% [82.1-94.3%], Sp 92.9% [90.5-94.7%]) |
| QAS Findings | Sensitivity > 80% and Specificity > 80% | |
| Small Bowel Obstruction | Sensitivity > 80%, Specificity > 80% | High Sensitivity: Se 94.9% [88.7-97.8%], Sp 91.7% [89.1-93.7%]; Sensitivity Biased: Se 91.9% [84.9-95.8%], Sp 96.0% [94.1-97.3%]; Balanced: Se 88.9% [81.2-93.7%], Sp 98.1% [96.6-98.9%] |
| Free Air | Sensitivity > 80%, Specificity > 80% | Balanced: Se 89.3% [82.2-93.8%], Sp 88.6% [85.7-91.0%]; High Specificity: Se 88.4% [81.1-93.1%], Sp 90.8% [88.1-92.9%] |
Turnaround Time Acceptance Criteria and Performance:
| Metric | Acceptance Criteria (Implied by Predicate) | Device Performance |
|---|---|---|
| Triage Turn-around Time | Mean < 81.6 seconds (Predicate's Mean) | Mean: 58.39 seconds (95% CI: 56.11-60.68) |
| Median: 55.02 seconds | ||
| 95th percentile: 90.36 seconds |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: 675 cases from 643 unique patients (after excluding 3 cases due to quality control failures from an initial 678 cases).
- Data Provenance: The data was sourced from multiple clinical sites within the United States. Specific states mentioned are New York (45.2%), Kansas (21.2%), Missouri (18.4%), Texas (15.0%), and Nebraska (0.3%). The study evaluated against clinical standards consistent with U.S. practice patterns. The data appears to be retrospective, as it was used for development and testing after collection.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: A minimum of two U.S. board-certified radiologists, with a third U.S. board-certified expert adjudicator for discordant cases.
- Qualifications: All experts were U.S. board-certified radiologists. The third adjudicator was specifically fellowship-trained in body imaging.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: 2+1 methodology. Each case was independently reviewed by two U.S. board-certified radiologists. If the two initial readers disagreed, a third U.S. board-certified expert adjudicator (fellowship-trained in body imaging) provided the tie-breaking determination.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- The provided document does not indicate that an MRMC comparative effectiveness study was performed or submitted for this clearance. The study described is a standalone performance assessment of the algorithm itself against ground truth.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, a standalone performance assessment was done. The document explicitly states: "A standalone performance assessment was performed for a2z-Unified-Triage to validate the accuracy of detecting the 7 findings against a reference standard established by U.S. board-certified radiologists."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: Expert consensus, specifically a 2+1 consensus of U.S. board-certified radiologists, with the third adjudicator being fellowship-trained in body imaging.
8. The sample size for the training set
- The document states, "The algorithms were developed on an extensive dataset of abdomen/pelvis CT studies from multiple clinical sites." However, a specific numerical sample size for the training set is not provided. It only mentions that strict protocols ensured complete independence between development and testing datasets (mutually exclusive patients).
9. How the ground truth for the training set was established
- The document does not explicitly detail how the ground truth for the training set was established. It only describes the ground truth establishment for the test set (2+1 radiologist consensus). It states that the algorithms were developed on an "extensive dataset" and implies internal processes for data collection and annotation during development.
U.S. Food & Drug Administration FDA Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
November 24, 2025
a2z Radiology AI Inc.
Samir Rajpurkar
Chief Executive Officer
292 Newbury Street
Unit 235
Boston, MA 02115
Re: K252366
Trade/Device Name: a2z-Unified-Triage
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS, QFM,
Dated: November 1, 2025
Received: November 3, 2025
Dear Samir Rajpurkar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
Page 2
K252366 – Samir Rajpurkar Page 2
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Page 3
K252366 – Samir Rajpurkar Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K252366
Please provide the device trade name(s).
a2z-Unified-Triage
Please provide your Indications for Use below.
a2z-Unified-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominal/pelvic CT images in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 7 specified abdominopelvic findings: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. These findings are intended to be used together as one device. The device supports both cloud-based and on-premises deployment, with integration either directly with healthcare facility systems or through third-party healthcare technology platforms.
a2z-Unified-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for primary diagnosis beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of a2z-Unified-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Page 5
510(k) Summary
1. Administrative Information
| Field | Value |
|---|---|
| Submitter Name | a2z Radiology AI Inc. |
| Address | 292 Newbury Street, Unit 235, Boston, MA 02115, USA |
| Phone Number | +1 508-293-1822 |
| Fax Number | N/A |
| Company Representative | Samir Rajpurkar |
| support@a2zradiology.ai | |
| Date Summary Prepared | November 01, 2025 |
2. Subject Device Information
| Field | Value |
|---|---|
| Trade Name | a2z-Unified-Triage |
| Subject Device K Number | K252366 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS, QFM |
| Regulation Number | 892.2080 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
3. Predicate Device Information
| Field | Value |
|---|---|
| Predicate Device Name | Annalise Enterprise CTB Triage Trauma |
| Predicate Device K Number | K223240 |
| Common Name | Radiological computer aided triage and notification software |
| Product Code | QAS |
| Regulation Number | 892.2080 |
| Regulatory Class | Class II |
| Review Panel | Radiology |
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
4. Device Description
a2z-Unified-Triage is a radiological computer-assisted triage and notification software device. The software consists of an algorithmic component that supports both cloud-based and on-premises
Page 6
510(k) Summary
deployment on standard server hardware. The device processes abdomen/pelvis CT images from clinical imaging systems, analyzing them using artificial intelligence algorithms to detect suspected cases of 7 abdominopelvic conditions: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction.
Following the AI processing, the analysis results are returned to the client system for worklist prioritization. When a suspected case is detected, the software provides analysis results that enable the client system to generate appropriate notifications. These results can include DICOM instance UIDs for key images, which are for informational purposes only, do not contain any marking of the findings, and are not intended for primary diagnosis beyond notification.
Integration with clinical imaging systems facilitates efficient triage by enabling prioritization of suspect cases for review of the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
Algorithm Architecture: The artificial intelligence algorithms implement deep learning models that analyze CT studies and return binary outputs indicating the presence or absence of each of the 7 target abdominopelvic findings. When a finding is detected, the algorithm identifies a key image slice most likely to show the condition, provided for informational purposes only.
Training Database: The algorithms were developed on an extensive dataset of abdomen/pelvis CT studies from multiple clinical sites. The training data encompassed diverse imaging characteristics including multiple CT scanner manufacturers, both contrast-enhanced and non-contrast studies, variable slice thicknesses, and a wide range of patient demographics.
Data Independence: Strict protocols ensured complete independence between development and testing datasets, with mutually exclusive patients. No overlap existed between training/validation data and the independent test dataset used for performance evaluation.
5. Indications for Use
a2z-Unified-Triage is a radiological computer-aided triage and notification software indicated for use in the analysis of abdominal/pelvic CT images in adults aged 22 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communicating suspected positive cases of the 7 specified abdominopelvic findings: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. These findings are intended to be used together as one device. The device supports both cloud-based and on-premises deployment, with integration either directly with healthcare facility systems or through third-party healthcare technology platforms.
a2z-Unified-Triage uses an artificial intelligence algorithm to analyze images and flag cases with detected findings in parallel to the ongoing standard of care image interpretation. The device provides analysis results that enable client systems to generate notifications for cases with suspected findings. These results can include DICOM instance UIDs for key images, which are meant for informational purposes only and not intended for primary diagnosis beyond notification. The
Page 7
510(k) Summary
device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of a2z-Unified-Triage are intended to be used in conjunction with other patient information and based on clinicians' professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Comparison of Technological Characteristics
The subject device, a2z-Unified-Triage, is substantially equivalent to the primary predicate Annalise Enterprise CTB Triage Trauma (K223240).
The subject and predicate are both radiological computer-aided triage and notification software using artificial intelligence algorithms. Both aid radiological image triage through deep learning algorithms trained on medical images, providing specialists with notifications and preview images for preemptive triage. Neither removes cases from standard care reading queues nor de-prioritizes cases. Both operate in parallel to standard care as the default option.
While the anatomical regions of interest differ (abdomen/pelvis for the subject device vs. head for the predicate), the core technology, intended use for triage, and principles of operation are similar. Both devices raise identical safety and effectiveness questions regarding accurate triage. Performance testing, summarized below, demonstrated that the subject device maintains equivalent safety and effectiveness for its intended use. A detailed comparison of key features is provided in the table below.
6. Comparison of Technological Characteristics with the Predicate Device
Table 1. Comparison with predicate device.
| Characteristic | Subject Device: a2z-Unified-Triage | Predicate Device: Annalise Enterprise CTB Triage Trauma (K223240) |
|---|---|---|
| Manufacturer | a2z Radiology AI Inc. | Annalise-AI Pty Ltd |
| Regulation Number | 892.2080 | 892.2080 |
| Regulatory Class | Class II | Class II |
| Product Code | QAS, QFM | QAS |
| Regulation Name | Radiological computer aided triage and notification software | Radiological computer aided triage and notification software |
| Device Property | SaMD (Software as a Medical Device) | SaMD (Software as a Medical Device) |
Page 8
510(k) Summary
| Characteristic | Subject Device: a2z-Unified-Triage | Predicate Device: Annalise Enterprise CTB Triage Trauma (K223240) |
|---|---|---|
| Indications for Use | Radiological computer-aided triage and notification software for analysis of abdominal/pelvic CT images in adults aged 22+. Assists in workflow triage by flagging 7 abdominopelvic findings. | Device to aid in triage and prioritization of studies with features suggestive of 4 intracranial hemorrhage findings in patients 22+. |
| Technical Characteristics | ||
| Input | Abdomen/Pelvis CT images (with or without contrast) | Non-contrast brain CT images |
| Output | JSON object with binary classification for 7 findings and DICOM instance UIDs for positive cases | Notification to image and order management system for worklist prioritization |
| Algorithm Type | Non-adaptive machine learning (Deep Learning - Neural Networks) | Non-adaptive machine learning (Deep Learning - Neural Networks) |
| Intended Users | Appropriately trained medical specialists | Trained clinicians qualified to interpret brain CT studies |
| Target Population | Adults (22 years and older) | Adults (22 years and older) |
| Location of anatomical structures | Abdomen and Pelvis | Head (Brain) |
| Imaging Modality | Computed Tomography (CT) | Computed Tomography (CT) |
| Intended Use Environment | Hospital environment or other clinical settings with DICOM-compliant CT imaging and IT infrastructure | Medical care environment |
| Performance | 7 findings met targets: QAS findings >80% sensitivity/specificity; QFM findings >0.95 AUC | Met performance targets with sensitivity and specificity >80% across findings |
| Triage Turn-around Time | Mean: 58.39 seconds (95% CI: 56.11-60.68), Median: 55.02 seconds, 95th percentile: 90.36 seconds | Mean: 81.6 seconds (95% CI: 80.3-82.9) |
Page 9
510(k) Summary
| Characteristic | Subject Device: a2z-Unified-Triage | Predicate Device: Annalise Enterprise CTB Triage Trauma (K223240) |
|---|---|---|
| Indications for Use | Radiological computer-aided triage and notification software for analysis of abdominal/pelvic CT images in adults aged 22+. Assists in workflow triage by flagging 7 abdominopelvic findings. | Device to aid in triage and prioritization of studies with features suggestive of 4 intracranial hemorrhage findings in patients 22+. |
| Technical Characteristics | ||
| Input | Abdomen/Pelvis CT images (with or without contrast) | Non-contrast brain CT images |
| Output | JSON object with binary classification for 7 findings and DICOM instance UIDs for positive cases | Notification to image and order management system for worklist prioritization |
| Algorithm Type | Non-adaptive machine learning (Deep Learning - Neural Networks) | Non-adaptive machine learning (Deep Learning - Neural Networks) |
| Intended Users | Appropriately trained medical specialists | Trained clinicians qualified to interpret brain CT studies |
| Target Population | Adults (22 years and older) | Adults (22 years and older) |
| Location of anatomical structures | Abdomen and Pelvis | Head (Brain) |
| Imaging Modality | Computed Tomography (CT) | Computed Tomography (CT) |
| Intended Use Environment | Hospital environment or other clinical settings with DICOM-compliant CT imaging and IT infrastructure | Medical care environment |
| Performance | 7 findings met targets: QAS findings >80% sensitivity/specificity; QFM findings >0.95 AUC | Met performance targets with sensitivity and specificity >80% across findings |
| Triage Turn-around Time | Mean: 58.39 seconds (95% CI: 56.11-60.68), Median: 55.02 seconds, 95th percentile: 90.36 seconds | Mean: 81.6 seconds (95% CI: 80.3-82.9) |
| Software device that operates on off-the-shelf hardware | Yes. Software supports both cloud-based and on-premises deployment on standard server hardware | Yes. Interfaces with image and order management systems |
7. Performance Data
7.1. Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Device Software Functions".
7.2. Performance Assessment
A standalone performance assessment was performed for a2z-Unified-Triage to validate the accuracy of detecting the 7 findings against a reference standard established by U.S. board-certified radiologists.
7.2.1. Ground Truth
The ground truth for the 678-case Test Data was established using a 2+1 methodology exclusively with U.S. board-certified radiologists. Specifically, each case was independently reviewed by two U.S. board-certified radiologists. For cases where the two readers disagreed (discordant cases), a third U.S. board-certified expert adjudicator, fellowship-trained in body imaging, provided the tiebreaking determination to establish the final ground truth. This ensures that all ground truth determinations are made by qualified U.S. board-certified radiologists using clinical standards consistent with U.S. practice patterns.
7.2.2. Development and Test Dataset
a2z-Unified-Triage was trained and tested on an extensive and diverse set of abdominal/pelvic CT studies from multiple clinical sites. The final test cohort consisted of 678 cases, with 3 cases excluded due to quality control failures during model inference, resulting in an analytic cohort of 675 cases from 643 unique patients. Development and test cohorts had mutually exclusive patients. The test dataset was constructed to ensure adequate representation for each of the findings and included a wide range of patient demographics and imaging characteristics, as summarized below.
Test Cohort Demographics and Imaging Characteristics (N=675)
| Sex | N | Percentage |
|---|---|---|
| Female | 345 | 51.1 |
Page 10
510(k) Summary
| Sex | N | Percentage |
|---|---|---|
| Male | 330 | 48.9 |
Age (years): Median (IQR) = 61 (44-75); Mean (SD) = 59.9 (18.2); Range = 22-90
Site and State Distribution
| Site | State | Number of Cases |
|---|---|---|
| Site 1 | New York | 223 (33.0%) |
| Site 2 | Kansas | 120 (17.8%) |
| Site 3 | New York | 82 (12.1%) |
| Site 4 | Missouri | 66 (9.8%) |
| Site 5 | Texas | 61 (9.0%) |
| Other (7 Sites) | Various | 123 (18.2%) |
| Total | 675 |
Sites in 'Other' category were located in Missouri, Texas, Kansas, and Nebraska. Total unique sites: 12
State Distribution:
| State | N | Percentage |
|---|---|---|
| New York | 305 | 45.2 |
| Kansas | 143 | 21.2 |
| Missouri | 124 | 18.4 |
| Texas | 101 | 15.0 |
| Nebraska | 2 | 0.3 |
7.3.3. Imaging Characteristics
Manufacturer Distribution:
| Manufacturer | N | Percentage |
|---|---|---|
| GE | 277 | 41.0 |
| SIEMENS | 217 | 32.1 |
| Canon | 107 | 15.9 |
| TOSHIBA | 52 | 7.7 |
| Other | 22 | 3.3 |
Page 11
510(k) Summary
Manufacturers in 'Other' category: FUJI, Hitachi, Philips
Contrast Status:
| Contrast Status | N | Percentage |
|---|---|---|
| With Contrast | 385 | 57 |
| Without Contrast | 290 | 43 |
Slice Thickness:
| Slice Thickness | N | Percentage |
|---|---|---|
| 5mm | 464 | 69 |
| 2mm to <5mm | 211 | 31 |
Mean (SD): 4.30 (1.06) mm; Median (IQR): 5.00 (3.00-5.00) mm; Range: 2-5mm
7.2.4. Clinical Subgroups and Confounders
The test dataset included diverse disease presentations across all 7 target conditions. U.S. board-certified radiologists annotated clinically relevant anatomical variations and disease subtypes on subsets of positive cases, including: calculous and acalculous acute cholecystitis, gallbladder wall thickening, and gallbladder hydrops; acute pancreatitis with and without fluid collections; sigmoid and other colonic locations of acute diverticulitis including cases with perforation; fusiform and saccular unruptured AAA morphologies across multiple size categories and anatomical locations; varying severities of hydronephrosis; small bowel obstruction from adhesive and mechanical etiologies; and free air from various etiologies and volumes. The test cohort also included cases with concurrent findings. Comprehensive stratified analyses demonstrated consistent device performance across all demographic subgroups, imaging parameters, clinical sites, and disease presentations.
7.2.5. Primary Endpoints
| Condition | Positive Cases | Total Cases | Prevalence (%) |
|---|---|---|---|
| Acute Cholecystitis | 77 | 675 | 11.4% |
| Acute Pancreatitis | 99 | 675 | 14.7% |
| Unruptured AAA | 76 | 675 | 11.3% |
| Acute Diverticulitis | 76 | 675 | 11.3% |
| Free Air | 112 | 675 | 16.6% |
| Hydronephrosis | 97 | 675 | 14.4% |
| Small Bowel Obstruction | 99 | 675 | 14.7% |
Page 12
510(k) Summary
7.2.4. QFM Conditions
| Condition | Available Operating Points and Performance |
|---|---|
| Acute Cholecystitis (QFM) Total N=675, Positive=77 | AUC: 0.985 [0.972-0.998]High Sensitivity: Se 96.1% [89.2-98.7%], Sp 89.3% [86.6-91.5%]Sensitivity Biased: Se 92.2% [84.0-96.4%], Sp 95.8% [93.9-97.2%]Balanced: Se 92.2% [84.0-96.4%], Sp 95.8% [93.9-97.2%] |
| Acute Pancreatitis (QFM) Total N=675, Positive=99 | AUC: 0.994 [0.985-1.000]High Sensitivity: Se 98.0% [92.9-99.4%], Sp 87.8% [84.9-90.3%]Sensitivity Biased: Se 98.0% [92.9-99.4%], Sp 97.0% [95.3-98.1%]Balanced: Se 98.0% [92.9-99.4%], Sp 97.0% [95.3-98.1%]High Specificity: Se 92.9% [86.1-96.5%], Sp 99.8% [99.0-100.0%] |
| Acute Diverticulitis (QFM) Total N=675, Positive=76 | AUC: 0.995 [0.990-1.000]High Sensitivity: Se 98.7% [92.9-99.8%], Sp 89.3% [86.6-91.5%]Sensitivity Biased: Se 97.4% [90.9-99.3%], Sp 96.8% [95.1-98.0%]Balanced: Se 97.4% [90.9-99.3%], Sp 96.8% [95.1-98.0%]High Specificity: Se 94.7% [87.2-97.9%], Sp 98.7% [97.4-99.3%] |
| Unruptured AAA (QFM) Total N=675, Positive=76 | AUC: 0.995 [0.991-0.999]High Sensitivity: Se 100.0% [95.2-100.0%], Sp 86.3% [83.3-88.8%]Sensitivity Biased: Se 97.4% [90.9-99.3%], Sp 95.8% [93.9-97.2%]Balanced: Se 97.4% [90.9-99.3%], Sp 97.5% [95.9-98.5%] |
| Hydronephrosis (QFM) Total N=675, Positive=97 | AUC: 0.976 [0.960-0.991]High Sensitivity: Se 89.7% [82.1-94.3%], Sp 92.9% [90.5-94.7%] |
7.2.5. QAS Conditions
Page 13
510(k) Summary
| Condition | Available Operating Points and Performance |
|---|---|
| Small Bowel Obstruction (QAS) Total N=675, Positive=99 | High Sensitivity: Se 94.9% [88.7-97.8%], Sp 91.7% [89.1-93.7%]Sensitivity Biased: Se 91.9% [84.9-95.8%], Sp 96.0% [94.1-97.3%]Balanced: Se 88.9% [81.2-93.7%], Sp 98.1% [96.6-98.9%] |
| Free Air (QAS) Total N=675, Positive=112 | Balanced: Se 89.3% [82.2-93.8%], Sp 88.6% [85.7-91.0%]High Specificity: Se 88.4% [81.1-93.1%], Sp 90.8% [88.1-92.9%] |
7.2.6. Performance Discussion
The comprehensive performance evaluation of a2z-Unified-Triage demonstrates consistently exceptional performance across all seven abdominopelvic findings, meeting or substantially exceeding FDA-specified acceptance criteria. All QAS findings (Free Air, Small Bowel Obstruction) met the acceptance criteria of sensitivity and specificity >80%, while 5 QFM findings (Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Hydronephrosis) met the acceptance criteria of AUC >0.95.
The extensive stratified analyses across patient demographics (sex, age, site, state), imaging parameters (contrast status, slice thickness, equipment manufacturer), and clinical scenarios (concurrent findings) reveal robust performance maintenance with no clinically significant degradation across any subgroup. Additionally, comprehensive stratified analyses by anatomical variations and edge cases were performed for all seven conditions, with U.S. board-certified radiologists annotating clinically relevant attributes on subsets of positive cases. These analyses–including disease subtypes, confounders, and morphological features specific to each condition–demonstrate robust device performance across all clinically relevant variations, validating that the device independently and accurately detects each indication based on its primary imaging features rather than confounding concurrent findings. This validates the algorithm's generalizability and reliability across diverse clinical presentations. The triage effectiveness study confirms rapid processing times well within clinical workflow requirements, establishing a2z-Unified-Triage as a clinically viable solution for enhancing radiological workflow efficiency.
8. Predetermined Change Control Plan (PCCP)
This submission contains a Predetermined Change Control Plan (PCCP). The PCCP does not include provisions for implementation of adaptive algorithms that will continuously learn in the field. All algorithm modifications will be trained, validated, and locked prior to release of the software to the field. A procedure has been established for updating device labeling to inform users about changes implemented under this PCCP, including a summary of the changes, characterization of algorithm performance, and compatibility information.
The PCCP specifies possible modifications to a2z-Unified-Triage, as well as verification and validation activities to implement changes in a controlled manner such that the modified device remains as safe and effective as the originally cleared device. The PCCP applies to the seven conditions val-
Page 14
510(k) Summary
idated in this 510(k) submission: Acute Cholecystitis, Acute Pancreatitis, Unruptured Abdominal Aortic Aneurysm, Acute Diverticulitis, Free Air, Hydronephrosis, and Small Bowel Obstruction. For each condition, the PCCP covers all operating points that meet pre-specified performance criteria upon revalidation.
The planned modifications include:
- Training data volume and diversity expansion
- Training data quality and annotation refinement
- Ensemble composition optimization
- Data augmentation parameter adjustments
- Training parameter optimization
- Weight initialization method selection
- Neural network architecture component modifications
- Input data preprocessing adjustments
- Validated slice thickness range expansion
All modifications implement bounded specifications with specific parameter ranges to ensure changes remain within validated frameworks. Comprehensive validation requirements apply to all modifications, including minimum test dataset size, ground truth determination by U.S. board-certified radiologists, complete data sequestration, stratified subgroup analysis, and acceptance criteria requiring sensitivity and specificity >80% for all findings plus AUC >0.95 for QFM findings.
The modification protocol incorporates impact assessment considerations and specifies requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data re-use practices. Detailed validation activities, testing methodologies, and performance requirements have been established for each modification. All changes will undergo appropriate verification and validation testing before implementation to ensure the modified device maintains the safety and effectiveness profile of the originally cleared device.
9. Conclusion
The subject device, a2z-Unified-Triage, and the predicate, Annalise Enterprise CTB Triage Trauma, are substantially equivalent. Both are software devices intended to aid in the prioritization and triage of radiological images using AI algorithms. They share the same fundamental scientific technology and principles of operation.
While the specific anatomical regions and findings differ, the intended use for triage, the AI-based approach, the integration into clinical workflows, and the parallel nature of operation are highly similar. Both devices aim to reduce turnaround time through preemptive triage without altering the standard of care.
The performance data for a2z-Unified-Triage demonstrates that the device is safe and effective for its intended use and performs as intended. Therefore, a2z-Unified-Triage is substantially equivalent to the predicate device.
§ 892.2080 Radiological computer aided triage and notification software.
(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.