K Number
K252178

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-10-03

(84 days)

Product Code
Regulation Number
876.1520
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ME-APDS (Magentiq Eye's Automatic Polyp Detection System) is intended to be used by endoscopists as an adjunct to the common video colonoscopy procedure (screening and surveillance), aiming to assist the endoscopist in identifying lesions during colonoscopy procedure by highlighting regions with visual characteristics consistent with different types of mucosal abnormalities that appear in the colonoscopy video during the procedure. Highlighted regions can be independently assessed by the endoscopist and appropriate action taken according to standard clinical practice.

ME-APDS is trained to process video images which may contain regions consistent with polyps.

ME-APDS is limited for use with standard white-light endoscopy imaging only.

ME-APDS is intended to be used as an adjunct to endoscopy procedures and is not intended to replace histopathological sampling as means of diagnosis.

Device Description

ME-APDS™/MAGENTIQ-COLO is intended to be used as an adjunct to the common video colonoscopy procedure. The system application aims to assist the endoscopist in identifying lesions, such as polyps, during the colonoscopy procedures in real time. The device is not intended to be used for diagnosis or characterization of lesions, and does not replace clinical decision making.

The system acquires the digital video output signal from the local endoscopy camera and processes the video frames. It runs deep machine learning and additional supporting algorithms in real time on the video frames in order to detect and identify regions having characteristics consistent with different types of mucosal abnormalities such as polyps. The output video with the detected lesions is presented on a separate screen, highlighting the suspicious areas on the original video. The user can also take snapshots of the videos, with and without the highlighting of the suspicious areas, record videos and view in full screen mode. The ME-APDS™ includes an additional feature that removes the presentation of the bounding box once the physician has inserted an endoscopic tool (e.g. snare or forceps) into the endoscopic field of view.

AI/ML Overview

Acceptance Criteria and Study for MAGENTIQ-COLO (ME-APDS)

Based on the provided FDA 510(k) clearance letter, here's a detailed breakdown of the acceptance criteria and the study conducted for MAGENTIQ-COLO (ME-APDS):

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "All testing met pre-specified acceptance criteria," implying that the reported performance metrics achieved or exceeded predefined thresholds, though the specific numerical thresholds for acceptance are not explicitly stated within this document. The table below presents the reported performance metrics based on the standalone algorithm performance testing.

MetricContextReported Performance
Polyp Detection
Polyp-wise Recall (PRc1)With Histology99.7%
Polyp-wise Recall (PRc3)With Histology99.7%
Polyp-wise Recall (PRc5)With Histology99.7%
Polyp-wise Recall (PRc7)With Histology99.7%
Polyp Coverage (MPC)With Histology81.7% (median)
Polyp-wise Recall (PRc1)Without Histology100.0%
Polyp-wise Recall (PRc3)Without Histology100.0%
Polyp-wise Recall (PRc5)Without Histology100.0%
Polyp-wise Recall (PRc7)Without Histology100.0%
Polyp Coverage (MPC)Without Histology81.3% (median)
Frame-wise Recall (FRecall)Full Testing Dataset73.0%
Polyp Coverage (MPC)Full Testing Dataset72.0% (median)
Polyp-wise Recall (PRc1)Full Testing Dataset98.1%
Polyp-wise Recall (PRc3)Full Testing Dataset95.8%
Polyp-wise Recall (PRc5)Full Testing Dataset93.8%
Polyp-wise Recall (PRc7)Full Testing Dataset91.2%
False Positives Per Full Video (procedure) rateAlgorithm PerformanceTarget achieved (specific rate not stated)
False Positives Per Frame (TPP)Algorithm Performance0.0334 (threshold achieved)
Tool Detection
Tool Detection Frame-wise Recall (FRecall)Tool Detection85.9%
Median Tool Coverage (MTC)Tool Detection96.5%
Tool Recall (TRc1)Tool Detection99.7%
Tool Recall (TRc3)Tool Detection99.7%
Tool Recall (TRc5)Tool Detection99.7%
Tool Recall (TRc7)Tool Detection99.7%
False Positives Per Frame (TPP)Tool Detection0.0118 (calculated)
System Performance
Marker Annotation DelayAll cases (new system, with/without tool detection)Adequate performance (specific metric not stated, but criteria met)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 252 unique full colonoscopy videos. These videos contained 806 polyps (660 small, 111 medium, 35 large). 24 videos contained no polyps.
  • Data Provenance: The document does not explicitly state the country of origin. The test set appears to be retrospective, as it is referred to as "the algorithm was tested offline on... videos."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not explicitly state the number of experts or their specific qualifications (e.g., number of years of experience) used to establish the ground truth for the test set. It mentions "polyps verified by histology" and "polyps that were reported in the procedure report," implying a clinical and pathological review process, but details on the experts involved are missing.

4. Adjudication Method for the Test Set

The document does not specify an explicit adjudication method (e.g., 2+1, 3+1). The ground truth for polyps is based on "histology findings" and "procedure report," suggesting a consensus or standard clinical reporting process but not a formal adjudication process by a panel for the study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study is not mentioned in the provided document. The study described is a standalone performance test of the algorithm.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a standalone study was performed. The document explicitly states: "The algorithm was tested offline on 252 unique full colonoscopy videos..." and "Standalone algorithm performance testing was conducted..."

7. Type of Ground Truth Used

The ground truth used for the test set was primarily based on:

  • Histology findings: This is considered a high-quality ground truth for polyp characteristics.
  • Procedure report: This likely refers to the endoscopist's findings documented during the colonoscopy.
  • The presence and size of polyps were noted, and 375 polyps had histology findings (302 adenoma polyps).

8. Sample Size for the Training Set

The document does not specify the sample size used for the training set. It only states that the device "is trained to process video images."

9. How the Ground Truth for the Training Set Was Established

The document does not specify how the ground truth for the training set was established. It only mentions that "ME-APDS is trained to process video images which may contain regions consistent with polyps."

FDA 510(k) Clearance Letter - MAGENTIQ-COLO (ME-APDS)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

October 3, 2025

Magentiq Eye LTD
℅ John J. Smith
Partner
Hogan Lovells US LLP
555 13th Street, NW
Washington, District of Columbia 20004

Re: K252178
Trade/Device Name: MAGENTIQ-COLO (ME-APDS)
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal Lesion Software Detection System
Regulatory Class: Class II
Product Code: QNP
Dated: July 11, 2025
Received: September 5, 2025

Dear John J. Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K252178 - John J. Smith Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K252178 - John J. Smith Page 3

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K252178

Device Name
MAGENTIQ-COLO (ME-APDS)

Indications for Use (Describe)
ME-APDS (Magentiq Eye's Automatic Polyp Detection System) is intended to be used by endoscopists as an adjunct to the common video colonoscopy procedure (screening and surveillance), aiming to assist the endoscopist in identifying lesions during colonoscopy procedure by highlighting regions with visual characteristics consistent with different types of mucosal abnormalities that appear in the colonoscopy video during the procedure. Highlighted regions can be independently assessed by the endoscopist and appropriate action taken according to standard clinical practice.

ME-APDS is trained to process video images which may contain regions consistent with polyps.

ME-APDS is limited for use with standard white-light endoscopy imaging only.

ME-APDS is intended to be used as an adjunct to endoscopy procedures and is not intended to replace histopathological sampling as means of diagnosis.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

Magentiq Eye's MAGENTIQ-COLO

Submitter:
Magentiq Eye Ltd.
6 Ben-Gurion Blvd.,
Haifa, 3541416,
Israel
Phone: +972 (77) 2018838
Contact Person: Dr. Dror Zur

Date Prepared: August 4, 2025

Name of Device: Magentiq Eye's Automatic Polyp Detection System (ME-APDS™)

Trade Name: MAGENTIQ-COLO™

Common or Usual Name: Computer aided detection software for colorectal polyps

Classification Name: Gastrointestinal Lesion Software Detection System

Regulatory Class: II

Product Code: QNP

Predicate Device: Magentiq Eye's Automatic Polyp Detection System (K244023)

Device Description:

ME-APDS™/MAGENTIQ-COLO is intended to be used as an adjunct to the common video colonoscopy procedure. The system application aims to assist the endoscopist in identifying lesions, such as polyps, during the colonoscopy procedures in real time. The device is not intended to be used for diagnosis or characterization of lesions, and does not replace clinical decision making.

The system acquires the digital video output signal from the local endoscopy camera and processes the video frames. It runs deep machine learning and additional supporting algorithms in real time on the video frames in order to detect and identify regions having characteristics consistent with different types of mucosal abnormalities such as polyps. The output video with the detected lesions is presented on a separate screen, highlighting the suspicious areas on the original video. The user can also take snapshots of the videos, with and without the highlighting of the suspicious areas, record videos and view in full screen mode. The ME-APDS™ includes an additional feature that removes the presentation of the bounding box once the physician has inserted an endoscopic tool (e.g. snare or forceps) into the endoscopic field of view.

Intended Use / Indications for Use:

ME-APDS™ (Magentiq Eye's Automatic Polyp Detection System) is intended to be used by endoscopists as an adjunct to the common video colonoscopy procedure (screening and surveillance),

K252178 Page 1 of 4

Page 6

aiming to assist the endoscopist in identifying lesions during colonoscopy procedure by highlighting regions with visual characteristics consistent with different types of mucosal abnormalities that appear in the colonoscopy video during the procedure. Highlighted regions can be independently assessed by the endoscopist and appropriate action taken according to standard clinical practice.

ME-APDS™ is trained to process video images which may contain regions consistent with polyps.

ME-APDS™ is limited for use with standard white-light endoscopy imaging only.

ME-APDS™ intended to be used as an adjunct to endoscopy procedures and is not intended to replace histopathological sampling as means of diagnosis.

Summary of Technological Characteristics:

Computer-Aided Polyp Detection (CADe) engine is the technological principle for both the subject and predicate devices. The major roles of CADe engine during colonoscopy is to process a video frame and to indicate the presence and location of detected lesions (such as polyps) in real time during colonoscopy procedure in order to improve mucosal lesion detection rates, thus improving the performance of the endoscopist.

The proposed device is a modification to the company cleared device. The proposed device labeling has been updated to identify an additional compatible endoscopy systems, and the system includes a feature that removes the presentation of the bounding box once the physician has inserted an endoscopic tool (e.g. snare or forceps) into the endoscopic field of view. The AI Polyp detection algorithm remains identical to the predicate device.

Performance Data:

Conducted Testing:

The additional compatible endoscopy system was evaluated on the standalone performance testing database. The following testing was conducted:

  • Standalone algorithm performance testing was conducted to assess polyp detection frame-wise recall, polyp-wise recall, and median of polyp coverage. All testing met pre-specified acceptance criteria demonstrating that polyp detection performance is appropriately maintained when the tool detection is active
  • Marker Annotation delay testing. Marker annotation delay was assessed which demonstrated adequate performance. All testing met pre-specified acceptance criteria.

The Tool Detection feature was evaluated on a subset of the standalone performance testing database videos containing forceps, pumps, shafts, cold snares, and hot snares. The following testing was conducted:

  • Standalone algorithm performance testing was conducted to assess polyp detection frame-wise recall, polyp-wise recall, and median of polyp coverage. All testing met pre-specified acceptance criteria demonstrating that polyp detection performance is appropriately maintained when the tool detection is active.

K252178 Page 2 of 4

Page 7

  • Standalone algorithm performance testing was conducted to assess tool detection frame-wise recall, tool-wise recall, and median tool coverage. All testing met pre-specified acceptance criteria demonstrating tool detection reliability.

Marker annotation delay testing. Marker annotation delay was assessed including cases from new compatible endoscopy systems, with and without tool detection, which demonstrated adequate performance. All testing met pre-specified acceptance criteria.

  • Software verification and validation. All testing was as expected.

Detailed Summary of Standalone Performance Testing:

The algorithm was tested offline on 252 unique full colonoscopy videos, containing 806 polyps. 24 videos contained no polyps. Of the 806 polyps, 660 were small (s≤5mm), 111 have medium size (5mm<s<10mm), and 35 were large (s≥10mm). 375 polyps had histology findings where 302 were found to be adenoma polyps. Polyps evaluated varied by subject sex (270 Male, 170 Female, 366 Unknown), age (60 under 50 years, 133 50-60 years, 256 older than 60 years, 357unknown), race (343 Caucasian, 6 African American, 457 Unknown).

ME-APDS recall and false positive performance was evaluated. In addition, the number of False Positives Per Full Video (procedure) rate was assessed. Polyp-wise Recall was defined as the number of polyps detected, each for a set number consecutive frames, out of the total number of polyps in the testing dataset, Polyp-wise Recall was evaluated a 1, 3, 5 and 7 consecutive frames as Precall1, Precall3, Precall5, and Precall7, respectively. The system detects 99.7% to 99.7% (PRecall1 to PRecall7) of polyps verified by histology and 98.1% to 91.2% (PRecall1 to PRecall7) of the polyps when polyps without histology verification were included, showing the ability of the system to adequately aid in the detection of polyps when working with the MAGENTIQ-COLO. The median of the coverage of polyps with histology was high (81.7%). The False Positives Per Frame (FPPF) threshold of 0.0334 was achieved.

Results on polyps that were reported in the procedure report, classified according to polyps with histology and without.

Reported Results on Polyps with Histology and Without

FRecallCIMPCPRecall1CIPRecall3CIPRecall5CIPRecall7CI
With Histology76.5%[73.77%, 79.47%]81.7%99.7%[99.17%, 100.0%]99.7%[99.18%, 100.0%]99.7%[99.08%, 100.0%]99.7%[99.09%, 100.0%]
Without Histology73.2%[67.0%, 78.7%]81.3%100.0%[100.0%, 100.0%]100.0%[100.0%, 100.0%]100.0%[100.0%, 100.0%]100.0%[100.0%, 100.0%]

K252178 Page 3 of 4

Page 8

Results on the Entire Testing Dataset

FRecallCIMPCPRecall1CIPRecall3CIPRecall5CIPRecall7CI
73.0%[70.58%, 75.65%]72.0%98.1%[97.0%, 99.06%]95.8%[94.12%, 97.24%]93.8%[91.81%, 95.46%]91.2%[89.02%, 93.14%]

As specified above, the Tool Detection feature was assessed and verified on a subset of this standalone performance testing database. The following results demonstrated performance of the tool detection on videos containing the following tool types: e.g. forceps, pumps, shafts, cold snares, and hot snares. The False Positives Per Frame (FPPF) of 0.0118 was calculated with the following results:

FRecallCIMTCTRecall1CITRecall3CITRecall5CITRecall7CI
85.9%[83.21%, 88.83%]96.5%99.7%[99.15%, 100%]99.7%[99.07%, 100%]99.7%[99.11%, 100%]99.7%[99.14%, 100%]

All testing met pre-specified acceptance criteria demonstrating tool detection reliability.

In summary, the testing results discussed above were observed to be as expected and support that the device has similar performance to the predicate device.

Conclusions:

The modified ME-APDS/MAGENTIQ-COLO™ is as safe and effective as the company's prior cleared ME-APDS. The modified ME-APDS has the same intended uses and indications, similar technological characteristics, and principles of operation as its predicate device. The minor differences do not alter the intended use and raise no new issues of safety or effectiveness. Performance data demonstrate that the modified ME-APDS is as safe and effective as the cleared ME-APDS. Thus, the modified ME-APDS /MAGENTIQ-COLO™ is substantially equivalent.

K252178 Page 4 of 4

§ 876.1520 Gastrointestinal lesion software detection system.

(a)
Identification. A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.