K Number
K251983

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-08-26

(60 days)

Product Code
Regulation Number
892.2080
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect symmetrical bilateral MCA occlusions.

Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

Cautions:

  • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
  • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.

Limitations:

  1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
  2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke workup - it provides workflow prioritization and notification only.
  3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
  4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
  5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions.

Contraindications:
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:

  • tumours or abscesses
  • coils, shunts, embolization or movement artifacts
  • intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis.
Device Description

Brainomix 360 Triage Stroke (also referred to as Triage Stroke in this submission) is a radiological computer aided triage and notification software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server. Triage Stroke is a non-contrast CT processing software-only medical device which operates within the integrated Brainomix 360 Platform to provide triage and notification prioritization of suspected large vessel occlusion (LVO) or intracranial hemorrhage (ICH). The device uses machine learning algorithms such as advanced non adaptive imaging algorithms, artificial intelligence, and large data analytics.

Brainomix 360 Triage Stroke is available to users in three configurations, featuring three individual processing modules:

  • Triage ICH Module, which can only flag positive findings of suspected ICH;
  • Triage Stroke Module, which can flag positive findings of suspected ICH or LVO; And
  • NCCT LVO Module, which can flag positive finding of suspected LVO

The Triage ICH Module automatically identifies suspected ICH, the NCCT LVO module automatically identifies suspected LVO, while the Triage Stroke Module automatically identifies suspected ICH or LVO on non-contrast CT (NCCT) imaging acquired from adult patients in the acute setting, within 24 hours of the onset of acute symptoms, or where this is unclear, since last known well (LKW) time. The output of the device is a priority notification to clinicians indicating the suspicion of just ICH for Triage ICH Module, the suspicion of just LVO for the NCCT LVO Module, and suspicion of ICH or LVO for Triage Stroke Module based on positive findings. Specifically, the ICH analysis algorithm is optimized to identify findings of hyperdense volume in the parenchyma typically associated with acute intracranial hemorrhage; and the NCCT LVO suspicion uses the combined analysis of the ASPECTS and hyperdense vessel sign (HDVS) algorithms to identify hyper attenuation in vessels and hypodense regions typically associated with a large vessel occlusion in a non-contrast CT scan.

Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions. The device uses the basic services supplied by the Brainomix 360 Platform including DICOM processing, job management, imaging module execution and imaging output including notification and compressed image.

Brainomix 360 Triage Stroke notification capabilities enable clinicians to review and preview images via mobile app notification. Alternatively, intended users can also access the notification (a "Suspected LVO" or "Suspected hemorrhage" flag) and straightened images via the Brainomix 360 web user interface. Images that are previewed via mobile app are compressed, are for preview informational purposes only, and not intended for diagnostic use beyond notification.

The device is intended for use as an additional tool for assisting study triage within existing patient pathways. It does not replace any part of the current standard of care. It is designed to assist in prioritization of studies for reading within a worklist, in addition to any other pre-existing formal or informal methods of study prioritization in place. Specifically, it does not remove cases from a reading queue and operates in parallel to the standard of care. This device is not intended to replace the usual methods of communication and transfer of information in the current standard of care.

The Brainomix 360 Triage Stroke device is made available to the user through the Brainomix 360 Platform The Brainomix 360 Platform is a central control unit which coordinates the execution image processing modules which support various analysis methods used in clinical practice today:

  • Brainomix 360 e-ASPECTS (K243294)
  • Brainomix 360 e-CTA (K242123)
  • Brainomix 360 e-CTP (K223555)
  • Brainomix 360 e-MRI (K231656)
  • Brainomix 360 Triage ICH (K231195)
  • Brainomix 360 Triage LVO (K231837)
  • Brainomix 360 Triage Stroke (K232496) (predicate device)
AI/ML Overview

The provided document describes the acceptance criteria and the study that proves the device meets those criteria for the Brainomix 360 Triage Stroke device.

Here's a breakdown of the requested information:


1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Pre-specified Goal)Reported Device Performance
ICH Detection (Standalone Study)Sensitivity > 80%Sensitivity: 96.41% (95% CI: 92.65-98.65%)
Specificity > 80%Specificity: 96.55% (95% CI: 92.94-98.70%)
SAH Detection (Secondary Outcome)Sensitivity > 80%Sensitivity: 85.71% (CI: 60.99-97.67%)
Specificity > 80%Specificity: 96.55% (CI: 80.60-98.87%)
LVO Detection (Standalone Study)(Not explicitly stated, but "exceeded pre-specified performance goals")Sensitivity: 69.64% (CI: 60.65-77.70%)
(Not explicitly stated, but "exceeded pre-specified performance goals")Specificity: 89.57% (CI: 82.92-94.36%)
Combined Time-to-Notification< 3.5 minutesMinimum: 58.3 secondsMaximum: 150.7 seconds(Both met criterion)
LVO Detection (Reader Study)Expert Non-InferiorityDevice Sensitivity: 69.64%All Readers Sensitivity: 47.94%Difference: 20.52% (8.26-32.78%) (Device demonstrated superiority, thus non-inferiority was met)
Non-Expert SuperiorityDevice Sensitivity: 69.64%Non-Expert Sensitivity: 47.18%Difference: 21.28% (5.84-36.72%) (Device demonstrated superiority)

2. Sample Size Used for the Test Set and Data Provenance

  • ICH Standalone Study Test Set: 341 cases (167 ICH positive; 174 ICH negative)
  • LVO and ICH Standalone Study Test Set: 267 cases (112 LVO positive; 40 ICH positive; 115 Negative for ICH or LVO; 3 excluded due to technical inadequacy).
  • Reader LVO Performance Study Test Set: The document does not explicitly state the number of cases used in the reader study. It refers to the same LVO and ICH Standalone Study data for performance metrics, suggesting the reader study was conducted on a subset or the entirety of that dataset.
  • Data Provenance: Retrospective study. The document does not specify the country of origin of the data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Three (3)
  • Qualifications of Experts: Experienced US board-certified neuroradiologists.

4. Adjudication Method for the Test Set

  • Method: Consensus of the three experienced US board-certified neuroradiologists.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • Yes, a reader study was conducted.
  • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance:
    The study compared the device's standalone LVO sensitivity to that of human readers without AI assistance (the document does not describe human readers using AI assistance).
    • Device's LVO Sensitivity: 69.64%
    • All Human Readers (Experts and Non-experts) LVO Sensitivity: 47.94%
    • Difference (Effect Size): The device's sensitivity was 20.52% (CI: 8.26-32.78%) higher than that of all human readers.
    • Non-expert Radiologists LVO Sensitivity: 47.18%
    • Difference (Effect Size for Non-experts): The device's sensitivity was 21.28% (CI: 5.84-36.72%) higher than that of non-expert (general) radiologists.
    • This indicates that the AI performs better than human readers alone for LVO detection in this study (i.e., human readers would need to improve significantly to match the AI's standalone performance, if this AI assistance was their only aid). The study primarily demonstrates the device's standalone performance in comparison to human unassisted performance rather than human improvement with AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, standalone performance studies were conducted.
    • A standalone study for ICH detection performance was conducted.
    • A standalone study for LVO and ICH detection performance was conducted.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Expert Consensus (consensus of three experienced US board-certified neuroradiologists).

8. The Sample Size for the Training Set

  • The document mentions that the improved ICH algorithm uses "a different deep learning framework, CNN architecture, training data and post-processing capabilities of the algorithm." However, it does not specify the sample size of the training set used for the AI models.

9. How the Ground Truth for the Training Set Was Established

  • The document implies ground truth for the training data was established due to the mention of "training data." However, it does not explicitly detail the method for establishing ground truth for the training set, only for the test sets (expert consensus).

U.S. Food & Drug Administration - FDA Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 26, 2025

Brainomix Limited
Delanyo Mensah
Regulatory Affairs Specialist
First Floor, Seacourt Tower
West Way
Oxford, OX2 0JJ
United Kingdom

Re: K251983
Trade/Device Name: Brainomix 360 Triage Stroke
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: June 25, 2025
Received: June 27, 2025

Dear Delanyo Mensah:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251983 - Delanyo Mensah Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251983 - Delanyo Mensah Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

Please provide the device trade name(s).

Brainomix 360 Triage Stroke

Please provide your Indications for Use below.


Brainomix 360 Triage Stroke Page 10 of 52

Page 5

Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect symmetrical bilateral MCA occlusions.

Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

Cautions:

  • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
  • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.

Limitations:

  1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
  2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke workup - it provides workflow prioritization and notification only.
  3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
  4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
  5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions.

Contraindications:
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:

  • tumours or abscesses

Brainomix 360 Triage Stroke Page 11 of 52

Page 6

Please select the types of uses (select one or both, as applicable).

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)


Brainomix 360 Triage Stroke Page 12 of 52

Page 7

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

510(K) Summary

Brainomix 360 Triage Stroke

K251983

Date Prepared: 27Jun2025
Applicant's name: Brainomix Limited
Applicant's address: First Floor, Seacourt Tower, West Way
Oxford, OX2 0JJ
United Kingdom

Official contact: Delanyo Mensah
+44 (0)1865 582730
regulatory@brainomix.com

Device Trade Name: Brainomix 360 Triage Stroke
Device Common Name: Radiological Computer-Assisted Triage and Notification Software
Regulatory Class: Class II
Product Code: QAS
Regulation No: 21 C.F.R. §892.2080
Classification Panel: Radiology Devices

1 Predicate Device

The Brainomix 360 Triage Stroke device is claimed to be substantially equivalent to the following legally marketed device:

Trade Name: Brainomix 360 Triage Stroke
Manufacturer: Brainomix Limited
Regulation Number: 21 C.F.R. §892.2080
Regulatory Class: Class II
Regulation Name: Radiological computer-assisted triage and notification software
Product Code: QAS
Submission Number: K232496

2 Device Description

Brainomix 360 Triage Stroke (also referred to as Triage Stroke in this submission) is a radiological computer aided triage and notification software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server. Triage Stroke is a non-contrast CT processing software-only medical device which operates within the integrated Brainomix 360 Platform to provide triage and notification prioritization of suspected large vessel occlusion (LVO) or intracranial hemorrhage (ICH). The device uses machine learning algorithms such as advanced non adaptive imaging algorithms, artificial intelligence, and large data analytics.

Brainomix 360 Triage Stroke is available to users in three configurations, featuring three individual processing modules:

  • Triage ICH Module, which can only flag positive findings of suspected ICH;

Page 8

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

  • Triage Stroke Module, which can flag positive findings of suspected ICH or LVO; And
  • NCCT LVO Module, which can flag positive finding of suspected LVO

The Triage ICH Module automatically identifies suspected ICH, the NCCT LVO module automatically identifies suspected LVO, while the Triage Stroke Module automatically identifies suspected ICH or LVO on non-contrast CT (NCCT) imaging acquired from adult patients in the acute setting, within 24 hours of the onset of acute symptoms, or where this is unclear, since last known well (LKW) time. The output of the device is a priority notification to clinicians indicating the suspicion of just ICH for Triage ICH Module, the suspicion of just LVO for the NCCT LVO Module, and suspicion of ICH or LVO for Triage Stroke Module based on positive findings. Specifically, the ICH analysis algorithm is optimized to identify findings of hyperdense volume in the parenchyma typically associated with acute intracranial hemorrhage; and the NCCT LVO suspicion uses the combined analysis of the ASPECTS and hyperdense vessel sign (HDVS) algorithms to identify hyper attenuation in vessels and hypodense regions typically associated with a large vessel occlusion in a non-contrast CT scan.

Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions. The device uses the basic services supplied by the Brainomix 360 Platform including DICOM processing, job management, imaging module execution and imaging output including notification and compressed image.

Brainomix 360 Triage Stroke notification capabilities enable clinicians to review and preview images via mobile app notification. Alternatively, intended users can also access the notification (a "Suspected LVO" or "Suspected hemorrhage" flag) and straightened images via the Brainomix 360 web user interface. Images that are previewed via mobile app are compressed, are for preview informational purposes only, and not intended for diagnostic use beyond notification.

The device is intended for use as an additional tool for assisting study triage within existing patient pathways. It does not replace any part of the current standard of care. It is designed to assist in prioritization of studies for reading within a worklist, in addition to any other pre-existing formal or informal methods of study prioritization in place. Specifically, it does not remove cases from a reading queue and operates in parallel to the standard of care. This device is not intended to replace the usual methods of communication and transfer of information in the current standard of care.

The Brainomix 360 Triage Stroke device is made available to the user through the Brainomix 360 Platform The Brainomix 360 Platform is a central control unit which coordinates the execution image processing modules which support various analysis methods used in clinical practice today:

  • Brainomix 360 e-ASPECTS (K243294)
  • Brainomix 360 e-CTA (K242123)
  • Brainomix 360 e-CTP (K223555)
  • Brainomix 360 e-MRI (K231656)
  • Brainomix 360 Triage ICH (K231195)
  • Brainomix 360 Triage LVO (K231837)
  • Brainomix 360 Triage Stroke (K232496) (predicate device)

Page 9

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

3 Clinical Characteristics

The primary users of the Brainomix 360 platform are medical imaging professionals. The outputs generated by Brainomix 360 Triage Stroke may be used by clinicians to inform patient triaging decisions as part of a wider set of other imaging to identify signs of ischemic damage.

4 Indications for Use

Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect symmetrical bilateral MCA occlusions.

Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.

The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

Cautions:

  • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
  • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.

Limitations:

  1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
  2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke workup - it provides workflow prioritization and notification only.
  3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
  4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
  5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions.

Contraindications:
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:

  • tumours or abscesses
  • coils, shunts, embolization or movement artifacts

Page 10

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

  • intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis.

5 Technological Characteristics

Brainomix 360 Triage Stroke principal workflow for NCCT includes the following key steps:

  1. NCCT image loading. Brainomix 360 Triage Stroke provides an automated workflow which will automatically process image data received by the system in accordance with pre-configured user DICOM routing preferences.

  2. Image Processing function. The workflow in this step will depend on the module configured for the user's healthcare facility (Triage ICH Module, Triage Stroke Module or NCCT LVO Module).

  3. Notification of 'Suspected hemorrhage' or 'Suspected LVO' sent to the user and image viewing. This is a non-diagnostic DICOM viewing application through mobile app, allowing a trained clinician to preview unprocessed CT studies with basic viewing functions(scroll through a cine, adjust window level and window width, pan, and zoom) prior to definitive review of images on a diagnostic workstation. Alternatively, the straightened images can be visualized in the Brainomix 360 web user interface on a radiologist workstation.

All functions listed here are performed automatically by the software. Notifications and alerts for users (when processing has completed) is the step that includes user interaction and present information to users. This is considered to be the output of the processing functions.

6 Performance Data

The performance of Brainomix 360 Triage Stroke has been validated in three clinical studies:

  • A standalone study of ICH detection performance
  • A standalone study of LVO and ICH detection performance
  • A reader study to compare the sensitivity of LVO detection to that of radiologists.

6.1 Summary of ICH Standalone Performance

A retrospective study was conducted with the primary endpoint of assessing the performance of Brainomix 360 Triage Stroke in identifying ICH findings in NCCT head images. The dataset comprised 341 cases (167 ICH positive; 174 ICH negative). Truthing was conducted by consensus of three experienced US board certified neuroradiologists. Sensitivity and specificity of ICH detection exceeded the pre-specified performance goal of 80%: sensitivity was 96.41% (95% CI: 92.65-98.65%) and specificity was 96.55% (CI: 92.94-98.70%).

As a secondary outcome measure, performance for detection of subarachnoid hemorrhages (SAHs) was assessed against a pre-defined performance goal: sensitivity was 85.71% (CI: 60.99-97.67%) and specificity was 96.55% (CI: 80.60-98.87%).

6.2 ICH Performance Over Subgroups

ICH detection performance was stratified over a range of relevant clinical variables, including age, gender, slice thickness and scanner manufacturer subgroups. Additional subgroup analysis is reported in the labelling.

Metrics21 < Age < 5050 ≤ Age < 70Age ≥ 70
Total Positives366962
Sensitivity94.44% (82.88-99.20)98.55% (93.01-99.91)95.16% (87.37-98.97)
Specificity95.77% (88.90-99.10)96.25% (90.10-99.21)100% (88.74-100.00)
MetricsFemaleMale

Page 11

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

Total Positives7988
Sensitivity96.20% (89.98-99.20)96.59% (90.97-99-28)
Specificity98.78% (94.08-99.92)94.57% (88.34-98.18)
MetricsSlice Thickness < 1.5 mm1.5mm ≤ Slice Thickness < 3 mmSlice Thickness ≥ 3 mm
Total Positives1122332
Sensitivity97.32% (92.84-99.44)95.65% (80.63-99.73)93.75% (80.94-99.09)
Specificity99.17% (95.94-99.95)100.00% (86.70-100.00%)85.29% (70.47-94.80)
MetricsGE MEDICAL SYSTEMPhilipsSIEMENS
Total Positives87727
Sensitivity98.85% (94.41-99.92)94.44% (87.12-98.47)85.71% (49.44-98.90)
Specificity95.70% (89.92-98.83)96.88% (90.02-99.58)100.00% (84.61-100.00)

6.3 Summary of LVO and ICH Standalone Performance

A retrospective study was conducted with the primary endpoint of assessing the performance of Brainomix 360 Triage Stroke in identifying NCCT head images containing large vessel occlusion (LVO) or intracranial hemorrhage (ICH) in 267 cases (LVO positive: 112 cases; ICH positive: 40 cases; Negative for ICH or LVO: 115; excluded: 3 cases due to technical inadequacy). Truthing was conducted by consensus of three experienced US board certified neuroradiologists.

Sensitivity and specificity exceeded the pre-specified performance goals for NCCT LVO and ICH. NCCT LVO performance was observed at 69.64% sensitivity (CI: 60.65-77.70%) and 89.57% specificity (CI: 82.92-94.36%). NCCT ICH performance in this smaller subset of cases was consistent with that of the ICH standalone study, as reported above (sensitivity: 95.00% [84.47-99.29%]; specificity: 88.11% [83.39-91.92%]).

The Triage Stroke time-to-notification analysis includes three steps: (1) the time taken to transfer the image for DICOM transfer, (2) the device processing time, and (3) the time taken to deliver a notification after processing. The combined time-to-notification was compared to an acceptability criterion of 3.5 minutes. The minimum and maximum estimates for each of the three steps were assessed and combined to create a range of expected values for the total time-to-notification. The minimum time-to-notification was 58.3 seconds, and the maximum was 150.7 seconds. This met the acceptability criterion of time-to-notification of under 3.5 minutes.

6.4 Summary of Reader LVO Performance

In addition, a reader study was conducted to compare NCCT LVO sensitivity of the device to that of radiologists. Secondary endpoints of expert non-inferiority and non-expert superiority were used. Triage Stroke passed both conditions, with a sensitivity for all readers (experts and non-experts) of 47.94% (CI: 37.91-57.97%). The difference between the device's sensitivity and that of all readers was 20.52% (CI: 8.26-32.78%). The general radiologists (non-experts) performed with a sensitivity of 47.18% (CI: 33.62-60.75%). The difference between the device and non-expert sensitivity was 21.28% (CI: 5.84-36.72%).

6.5 LVO Performance Across Subgroups

Subgroup analyses were conducted for the LVO indication. The analyses assessed performance across age, gender, slice thickness and scanner manufacturer subgroups. Additional subgroup analysis is reported in the labelling.

Metrics21 < Age < 5050 ≤ Age < 70Age ≥ 70

Page 12

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

Total Positives133960
Sensitivity84.62% (58.59-97.46)74.36% (59.09-86.41)63.33% (50.62-74.88)
Specificity94.74% (77.09-99.67)89.09% (78.7-95.74)87.80% (75.07-95.75)
MetricsFemaleMale
Total Positives4864
Sensitivity75.0% (61.4-85.89)65.62% (53.39-76.57)
Specificity88.14% (77.94-94.92)91.07% (81.32-96.95)
MetricsSlice Thickness < 1 mm1 mm ≤ Slice Thickness < 3 mmSlice Thickness ≥ 3 mm
Total Positives442542
Sensitivity61.36% (46.46-74.93)80.00% (61.30-92.70)71.43% (56.52-83.7)
Specificity87.76% (76.29-95.19)96.67% (84.74-99.79)86.11% (71.95-95.11)
MetricsSIEMENSGE MEDICAL SYSTEMSPhilips
Total Positives483327
Sensitivity62.5% (48.25-75.4)72.73% (55.90-86.04)77.78% (59.56-90.82)
Specificity88.24% (77.15-95.39)83.87% (67.93-94.24)96.97% (86.02-99.81)

In summary, the performance validation data demonstrated that the proposed device provides accurate detection of acute ICH and LVO under a range of clinically relevant variables associated with the intended use of the software. The notification functionality of the device also met the acceptability criterion.

7 Prescriptive Statement

Caution: Federal law restricts this device to sale by or on the order of a physician.

8 Safety and Effectiveness

Brainomix 360 Triage Stroke has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with ISO 14971:2019 for risk management and the software development process conforms to IEC 62304:2015. The performance of Brainomix 360 Triage Stroke has been validated through retrospective case data based on expert reader truthing of the data and reader testing.

9 Cybersecurity

Brainomix 360 Triage Stroke has been designed to follow the FDA Cybersecurity Guidance and IEC 81001-5-1.

10 Substantial Equivalence

The subject device includes similar features as compared to the predicate device. Both devices have substantially equivalent indications for use, technological characteristics and roles within a clinical workflow relative to triage and notification. Both devices are intended to automatically process and analyze non-contrast CT scans to provide a notification to users in case of a suspected LVO or ICH being identified. The subject device does raise different

Page 13

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

questions of safety and effectiveness when compared to the predicate and both include similar mitigation strategies for reducing the risk of off label use of preview images which are shared via mobile.

Where the proposed device differs from the predicate device are:

  • Improved ICH algorithm: The proposed device features an updated ICH algorithm using a different deep learning framework, CNN architecture, training data and post-processing capabilities of the algorithm. The improved algorithm resulted in improved sensitivity in detecting SAH and SDH subtypes. Furthermore, the algorithm improvements have also resulted in increased sensitivity and specificity of the software in detecting ICH. The performance of the updated algorithm has been validated in clinical studies, which provides evidence of effective triaging of all ICH subtypes with an increased sensitivity and specificity.

  • Indications for use extended to SAH patients: Following the improvements made to the ICH algorithm the proposed device removes a previous SAH limitation and extends the indications of use for Triage Stroke to include subarachnoid hemorrhages (SAH) sub types. The change in the algorithm and the subsequent removal of the limitation has been validated through a clinical study and showed improvements on the performance of Triage Stroke in relation to different ICH subtypes.

  • Configurable workflows: The proposed device offers three configurations with three workflows. The Triage ICH Module features a workflow to detect ICH in NCCT scans, the NCCT LVO Module features a workflow to detect LVO in NCCT scans, while the Triage Stroke Module's workflow allows the detection of either ICH or LVO in NCCT scans. The workflow of the Triage ICH Module remains the same as cleared under Brainomix 360 Triage ICH (K231195) and the workflow of the Triage Stroke Module remains the same as cleared under Brainomix 360 Triage Stroke (K232496). The NCCT LVO module utilizes the workflow like that of the Triage Stroke Module but with the device output limited to LVO notification only. The proposed device differs from the predicate in that it offers a choice of these three workflows within a single device and as such, this technological characteristic does not raise different questions of safety or effectiveness compared to the predicate device.

11 Substantial Equivalence Comparison

A table comparing the key features of the subject and predicate devices is provided below.

Characteristic/ParameterBrainomix 360 Triage Stroke Predicate Device (K232496)Brainomix 360 Triage Stroke Subject Device
Product CodeQASQAS
Regulation21 CFR. §892.208021 CFR. §892.2080
Indications for UseBrainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 and intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolatedBrainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect

Page 14

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

| | subarachnoid hemorrhage and symmetrical bilateral MCA occlusions. Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Cautions: • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke. • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup. Limitations: • Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation. • Brainomix 360 Triage Stroke does not replace the need for CTA in ischemic stroke workup - it provides workflow prioritization and notification only. | symmetrical bilateral MCA occlusions. Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Cautions: • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke. • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup. Limitations: 1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation. 2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke |

Page 15

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

| | • Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners. • Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke. • Brainomix 360 Triage Stroke is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions. Contraindications: Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with: • tumors or abscesses • coils, shunts, embolization or movement artifacts • Brainomix 360 Triage Stroke is not intended to be used for analyzing CT images in intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis. | workup - it provides workflow prioritization and notification only. 3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners. 4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke. 5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions. Contraindications: Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with: • tumours or abscesses • coils, shunts, embolization or movement artifacts • intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis. |
| Environment of Use | Clinical/Hospital environment | Clinical/Hospital environment |
| User | Clinician | Clinician |
| Anatomical Region | Head | Head |
| Input Data | NCCT (ICH and LVO) | NCCT (ICH and LVO) |
| Technical Implementation | AI/ML/Neural Network | AI/ML/Neural Network |
| Diagnostic application | Notification-only | Notification-only |
| Results of image analysis | Internal no image marking | Internal no image marking |
| Segmentation of ROI | The device does not highlight or direct user's attention to a specific location in the image file | The device does not highlight or direct user's attention to a specific location in the image file |
| Notification Display | Web user interface and mobile device | Web user interface and mobile device |
| Preview Images | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes. The device operates in parallel with the standard of care. | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes. The device operates in parallel with the standard of care. |
| Annotation / Localization | Device does not mark, highlight, or direct users' attention to a specific location in the original image | Device does not mark, highlight, or direct users' attention to a specific location in the original image |
| Prioritization Notification | Yes | Yes |
| Clinical SoC Workflow | In parallel to | In parallel to |

Page 16

Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom

Technical PipelineTwo cascaded functions (ICH then LVO) using three integrated algorithmsTriage Stroke Configuration: two cascaded functions (ICH then LVO) using three integrated algorithms Triage ICH Configuration: ICH analysis function using one algorithm NCCT LVO Configuration: LVO analysis function using three integrated algorithms
Removal of Cases from SoC reviewNoNo

12 Conclusion

Brainomix 360 Triage Stroke includes similar features as compared to the predicate device. Both devices have similar indications for use, technological approaches and roles within a clinical workflow. The differences in indications for use and technological characteristics for the proposed device do not raise different questions of safety or effectiveness.

The proposed device therefore does not raise different questions with regards to safety and efficacy and demonstrates substantial equivalence to the predicate, Brainomix 360 Triage Stroke (K232496).

§ 892.2080 Radiological computer aided triage and notification software.

(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.