K Number
K251923

Validate with FDA (Live)

Date Cleared
2026-01-29

(220 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.

Device Description

The device includes a control unit, handpiece, footswitch, blades, waste bottle, collection bottle, and suction tube. The device is a reusable morcellation device which is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Medical Shaver System (Morcellator)

Page 1

January 29, 2026

Hangzhou Hawk Optical Electronic Instruments Co., Ltd
Yayan Ding
Person Responsible for Regulatory Compliance
No.9 Xinda Road, Suoqian, Xiaoshan
Hangzhou, Zhejiang 311201
China

Re: K251923
Trade/Device Name: Medical Shaver System (Morcellator) (YSB-IIIA)
Regulation Number: 21 CFR§ 876.1500
Regulation Name: Endoscope and Accessories
Regulatory Class: II
Product Code: GCJ
Dated: December 22, 2025
Received: December 22, 2025

Dear Yayan Ding:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251923 - Yayan Ding Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K251923 - Yayan Ding Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MARK J. ANTONINO -S

Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology and Urology Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026

Indications for Use

510(k) Number (if known): K251923

Device Name: Medical Shaver System (Morcellator) (YSB-IIIA)

Indications for Use (Describe):
It is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1

Page 5

510(k) Premarket Notification Submission

510(k) Summary

In accordance with 21 CFR 807.92 the following summary information is provided:

Date Prepared: January 8, 2026
Manufacturer: Hangzhou Hawk Optical Electronic Instruments Co., Ltd
No.9 Xinda Road, Suoqian, Xiaoshan, Hangzhou,
Zhejiang 311201 P.R. China

Contact Person: Yayan DING
Person Responsible for Regulatory Compliance
Hangzhou Hawk Optical Electronic Instruments Co., Ltd
Tel: (+86) 0571-22811660
Email: christina.ding@chinahawk.cn

Identification of the Device:

Proprietary/Trade Name: Medical Shaver System (Morcellator) YSB-IIIA
Regulation Name: Endoscope and Accessories
Regulatory Number: 21 CFR 876.1500
Classification Product Code: GCJ
Device Class: Class II
Review Panel: General & Plastic Surgery

Predicate Device:

Proprietary/Trade Name: Cyber Blade
Regulation Name: Endoscope and accessories
Regulatory Number: 21 CFR 876.1500
Product Code: GCJ
Device Class: Class II
Review Panel: General & Plastic Surgery
Submitter/510(k) Holder: SIDAM s.r.l.
Clearance: K192499 (cleared July 31, 2020)

Device Description:

The device includes a control unit, handpiece, footswitch, blades, waste bottle, collection bottle, and suction tube. The device is a reusable morcellation device which is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures.

Page 6

Principle of operation: Morcellation through the cutting effect of the cutting window, produced by the relative movement of the internal tube and the external tube.

Indications for Use:

It is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.

Comparison with Predicate Device:

The Medical Shaver System (Morcellator) YSB-IIIA and its predicate device, the SIDAM s.r.l. Cyber Blade (K192499), have the same intended use, and similar technological characteristics.

Subject DeviceMedical Shaver System (Morcellator) YSB-IIIAPredicate DeviceCyber Blade (K192499)
ManufacturerHangzhou Hawk Optical Electronic Instruments Co., Ltd.Sidam S.r.l
Trade NameMedical Shaver System (Morcellator) YSB-IIIACyber Blade
Product CodeGCJGCJ
Regulation Number21 CFR 876.150021 CFR 876.1500
Regulation NameEndoscope and accessories.Endoscope and accessories.
Indications for useIt is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.Cyber BLADE is intended for use under endoscopic visualization for the morcellation and removal of dissected benign prostatic hyperplasia (BPH) tissue during endoscopic surgical procedures in urology.
UsersHealth care professionalsHealth care professionals
Principle of operationMorcellation through the cutting effect of the cutting window, produced by the relative movement of the internal tube and the external tube.Morcellation through the cutting effect of the cutting window, produced by the relative movement of the internal tube and the external tube.

Page 7

Subject DeviceMedical Shaver System (Morcellator) YSB-IIIAPredicate DeviceCyber Blade (K192499)
Vacuum intensity regulationPossible from the vacuum generatorPossible through the leverage on the morcellator
Speed regulationPossible from the vacuum generatorPossible through the button on the morcellator
Blade dimension (external tube)40 cm working length0.45 cm external diameter (N5912)0.48 cm external diameter (N5911A)38 cm working length0.47 cm external diameter
Maximum morcellation velocityAbout 10 grams per minute (blade N5912, Bovine Heart)About 20 grams per minute (blade N5911A, Bovine Heart)About 17 grams per minute
Suction flow rate (Range and adjusted)≥1.5L/minAbout 1.61 L/min
Rotational speed (tied to morcellation rate)300-1600 R/min500 - 1475RPM
Blade detachment (tensile strength)Blade and handpiece are unlikely to detach due to forceBlade and handpiece are unlikely to detach due to force
Materials in contact with the tissueStainless steelStainless steel
Power supplyAC power suppliedThe morcellator is battery powered.The battery charger is AC power supplied.
Suction pumpPart of the systemNot part of the system, must use an external suction system.
Vacuum extractor≤ 0.08MPa/

Page 8

Subject DeviceMedical Shaver System (Morcellator) YSB-IIIAPredicate DeviceCyber Blade (K192499)
Vacuum extractor flow≥ 1.5L / min/
Supplied SterileNoYes
Re-usableReusable – blade and internal motorThe internal motor is reusable, but it never comes in contact with the patient.

Technology:

The Medical Shaver System (Morcellator) YSB-IIIA employs the same fundamental scientific technology as the predicate device.

Determination of substantial equivalence:

The Proposed Medical Shaver System (Morcellator) YSB-IIIA is substantially equivalent to the predicate SIDAM s.r.l. Cyber Blade (K192499) with regards to indication for use, performance capabilities, and technological characteristics.

The following is an overview of the differences between the proposed subject device Medical Shaver System (Morcellator) YSB-IIIA and the predicate Cyber Blade (K192499):

Note 1 Vacuum intensity regulation, Speed regulation
Both proposed device and predicate device have functions of vacuum intensity regulation and speed regulation. The vacuum intensity regulation and speed regulation of proposed device is possible from the vacuum generator and on the morcellator. The predicate device is only through the button on the morcellator. These functions were verified via testing.

Note 2 Blade dimension (external tube)
The proposed device dimension is 40 cm working length, 0.45 cm (N5912), and 0.48cm (N5911A) external diameter. The predicate device's dimension is 38 cm working length and 0.47 cm external diameter. They are close and in same range. They can demonstrate substantial equivalence with predicate device.

Note 3 Morcellation velocity

Page 9

The proposed device morcellation velocity is about 10 and 20 grams per minute for N5912 and N5911A, the predicate device's morcellation velocity is about 17 grams per minute. The subject device contains a similar morcellation rate that is within range of the predicate device. The surgeons can select the appropriate blade according to morcellation velocity for patient condition. The Morcellation velocity performance was verified in bench test.

Note 4 Suction flow rate
They are close and in same range. They can demonstrate substantial equivalence with predicate device.

Note 5 Rotational speed
They are close and in same range. They can demonstrate substantial equivalence with predicate device.

Note 6 Patient Contact Materials
The patient contact materials are equivalent to the predicate device, both the subject and predicate device blade material meet ISO 10993-1 and FDA guidance requirement.

Note 7 Power supply
The proposed device's power supply is by AC power, while the predicate device morcellator is battery powered and the battery charger is AC powered. Both devices meet IEC60601-1 and IEC60601-1-2 safety and EMC requirement.

Note 8 Suction pump
Suction pump is part of the proposed device. FDA cleared suction systems are intended to be used with the predicate device. The predicate device refers to the Morce Scope Set 8970 (K041610), Richard Wolf Medical Instruments Corp. Suction pump performance was compared and verified through testing.

Subject DeviceMedical Shaver System (Morcellator) YSB-IIIADeviceMorce Scope Set 8970 (K041610)Comparison to Predicate
ManufacturerHangzhou Hawk Optical Electronic Instruments Co., Ltd.Richard Wolf Medical Instruments Corp.NA

Page 10

Subject DeviceMedical Shaver System (Morcellator) YSB-IIIADeviceMorce Scope Set 8970 (K041610)Comparison to Predicate
Trade NameMedical Shaver System (Morcellator) YSB-IIIAMorce Scope Set 8970NA
Intended useIt is used for aspirating irrigation fluidThe SUCTION PUMP is used for aspirating irrigation fluid in conjunction with a resectoscope or a morcellator following laser TURP.Same.
Physical descriptionAC power, Microprocessor, pressure controlAC power, Microprocessor, pressure controlSame
Vacuum extractor≤ -0.08Mpa-0.06MPa~-0.01MPaEquivalentClose and Better than predicate device.
Vacuum extractor flow≥ 1.5L / min(liquid)Aspiration: 1.4L/minEquivalentClose and Better than predicate device.

Note 9 Supplied Sterile/Re-usable
The proposed device is supplied non-sterile, and the device is intended to be sterilized prior to usage. Sterilization and cleaning validation were conducted and verified. The predicate morcellator is single use only. Its internal motor is reusable, but it never comes in contact with the patient. Normally a single use device raises less concerns than a re-usable one. However, the proposed device's reprocessing methods and instructions were verified. The reprocessing validation meet the sterilized assurance level.

Summary of Non-clinical test:

The Medical Shaver System (Morcellator) YSB-IIIA conducted performance bench testing.

The device was evaluated for biocompatibility, cleaning and sterilization effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and have been found to comply with applicable medical device safety standard, The Medical Shaver System (Morcellator) YSB-IIIA complies with voluntary standards:

  • IEC60601-1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 &

Page 11

A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]

  • IEC 60601-2-18-2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
  • IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC /TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
  • IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • IEC 62366-1 Edition 1.1 2020-06 Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1
  • ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
  • IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes
  • ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10:2021Biological evaluation of medical devices Part 10: Tests for skin sensitization
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
  • ISO 10993-23:2021Biological evaluation of medical devices — Part 23: Tests for irritation
  • ISO 15223-1 Fourth edition 2021-07 Medical devices - Symbols to be used with medical device labels, labeling, and information to be supplied - Part 1: General requirements
  • ISO 20417: 2021 Medical devices - Information to be supplied by the manufacturer
  • ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices
  • ISO 8600-1 Fourth Edition 2015-10-15 Endoscopes - Medical endoscopes and endotherapy devices -- Part 1: General requirements
  • ISO 17664-1:2021 Processing of health care products — Information to be

Page 12

provided by the medical device manufacturer for the processing of medical devices — Part 1: Critical and semi-critical medical devices

  • AAMI TIR12:2020/(R)2023 Designing testing and labeling medical devices intended for processing by health care facilities: A guide for device manufacturers
  • ISO 17665:2024 Sterilization of health care products — Moist heat — Requirements for the development, validation and routine control of a sterilization process for medical devices

Summary of Clinical Tests:

The subject of this premarket submission did not require clinical studies to support substantial equivalence.

Conclusion:

Hangzhou Hawk Optical Electronic Instruments Co., Ltd considers the Medical Shaver System (Morcellator) YSB-IIIA to be substantially equivalent, with respect to safety and effectiveness to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.