K Number
K251909

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-09-03

(75 days)

Product Code
Regulation Number
882.5890
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SX2500 / SX3800 Series devices uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposes.

Device Description

The subject devices, SX2500 & SX3800, under the NFO Product Code, uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposes. The SX series uses applied accessories like probes, gloves or pads to deliver low electrical current to stimulation of facial tissues for aesthetic purposes.

The SX2500 & SX3800 is a powered microcurrent stimulator that uses electrical energy to stimulate facial tissues for aesthetic purposes. The SX2500 & SX3800 is substantially equivalent to the predicate devices compared in its product code class and has the same IFU. The SX2500 & SX3800 delivers signals independently to specific facial area points by way of channel ports via stainless steel probes and applicators substantially equivalent to the predicate device. The probes, gloves and pads referenced are the only patient contacting parts on the device that is substantially equivalent to the predicate device.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - NeurotriS Devices

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.08.00
Silver Spring, MD 20993
www.fda.gov

September 3, 2025

A-1 Engineering
Anthony Picciano
CEO
30 Mauchly Ste A
Irvine, California 92618

Re: K251909
Trade/Device Name: NeurotriS (SX2500); NeurotriS (SX3800)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: February 6, 2024
Received: June 20, 2025

Dear Anthony Picciano:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K251909 - Anthony Picciano Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K251909 - Anthony Picciano Page 3

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K251909

Device Name
NeurotriS (SX2500); NeurotriS (SX3800)

Indications for Use (Describe)
The SX2500 / SX3800 Series devices uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposes.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510K Summary

K251909

Submitters Details:
Name: A-1 Engineering
Address: 30 Mauchly Ste A
Irvine, CA 92618
Phone Number: 951 537 7240
Email: tony@neurotris.com
Contact Person: Anthony Picciano
Date Prepared: 08/14/2025

Devices Trade Name: NeurotriS SX2500 Picowave
NeurotriS SX3800
Common Name: Transcutaneous electrical nerve stimulator for pain relief
Product Code: NFO
Classification Name: Stimulator, Transcutaneous Electrical, Aesthetic Purposes
Regulation: 21 CFR 882.5890

Predicate Devices 1:
Name: TAMA Research Corp
510k Number: K173093
Product code: NFO

Predicate Devices 2:
Name: Biosonic Technologies LLC
510k Number: K130065
Product code: NFO

Device Description: The subject devices, SX2500 & SX3800, under the NFO Product Code, uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposes.

Intended Use: The SX series uses applied accessories like probes, gloves or pads to deliver low electrical current to stimulation of facial tissues for aesthetic purposes.

The SX2500 & SX3800 is a powered microcurrent stimulator that uses electrical energy to stimulate facial tissues for aesthetic purposes. The SX2500 & SX3800 is substantially equivalent to the predicate devices compared in its product code class and has the same IFU. The SX2500 & SX3800 delivers signals independently to specific facial area points by way of channel ports via stainless steel probes and applicators substantially equivalent to the predicate device. The probes, gloves and pads referenced are the only patient contacting parts on the device that is substantially equivalent to the predicate device.

Page 6

See table below of SX2500 and SX3800 Accessories / patient contacting parts substantially equivalent to the predicate devices

NeurotriSTAMA 510K #173093
Pico ProbesProbes
ANMA ProbeProbes
Infinity ProbesProbes
Conductive Gel

Page 7

SprayTonic
PadsLip & Square Electrode Pads
GlovesConductive Gloves

Non-clinical tests performed to demonstrate substantial equivalence.

Technological characteristics between the SX2500 and the SX3800 and those of the predicate devices were evaluated against the technical data found in the 510K on File with the FDA of the predicate devices.

Testing was performed at our Engineering test facility. The SX2500 and the SX3800 was BENCH tested against the Predicate devices as we have available the predicate devices at our facility to test.

Technical Data found in the 510K of the predicate devices, and data obtained from the Bench testing was compared against the SX2500 and the SX3800 is substantially equivalent.

Summary of Technological characteristics compared showing substantial equivalence.

Technical Summary:

  • Indications for Use: Substantial Equivalent
  • Number of Output Channels: Substantial Equivalent
  • Regulated Current or Voltage: Substantial Equivalent
  • Indicator Display Status: Substantial Equivalent
  • Compliance with Voluntary Standards: Substantial Equivalent
  • Compliance with 21 CFR 989: Substantial Equivalent

Page 8

  • Waveform: Substantial Equivalent
  • Shape: Substantial Equivalent
  • Symmetrical phases: Substantial Equivalent
  • Phases Duration: Substantial Equivalent
  • Net Charge: Substantial Equivalent
  • Maximum Phase Charge: Substantial Equivalent
  • Maximum Current Density: Substantial Equivalent
  • Maximum Average Current: Substantial Equivalent

Applicators:

  • Probes: Substantial Equivalent
  • Conductive Gloves: Substantial Equivalent
  • Pads: Substantial Equivalent

Detailed Technological characteristics Tabular Comparison showing substantial equivalence

NeurotriS SX3800 SX2500 Subject DeviceTAMA BEMS Device – K173093Beautiful Image Model 900 Facial Toning Device – K130065
Indications for UseThe NeurotriS SX3800 SX2500 Devices uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposesThe TAMA BEMS Device uses microcurrent electrical energy to stimulate facial tissues for aesthetic purposesThe Beautiful Image Model 900 Facial Toning Device uses microcurrent electrical energy to stimulate facial tissues for aesthetic
Number of Output PortsOne to FourOne to FourOne to Two
Number of Output Channels Synchronous or Alternating?One to Four N/AOne N/AOne N/A
ManufacturerA1 EngineeringTAMA Research CorporationBiosonic Technologies, LLC
Regulated Current or Regulated Voltage?BOTHBothBoth
Software/Firmware/MicroprocessorYESYesYes
Automatic Overload Trip?YesYesYes
Automatic No-Load Trip?YesYesYes
Automatic Shut Off?YesYesYes
User Override Control?YesYesYes
Indicator Display Status On/Off Display Status? Low Battery?YES NAYes Yes YesYes Yes Yes

Page 9

NeurotriS SX3800 SX2500TAMA BEMS Device – Subject Device K173093Beautiful Image Model 900 Facial Toning Device – K130065
Timer Range (minutes)20 minutes0-30 minutesNone
Compliance with Voluntary Standards?IEC60601-1IEC60601-1IEC 60601-1
Compliance with 21 CFR 989?YESYesYes
Weight (lbs., oz.)12 lbs9.5 ounces10 lbs
Dimensions (in.) [WxHxD]10x11x63x5x0.75.5x15.3x11.3
Housing Materials and ConstructionAnodized aluminum 6061Anodized aluminum 6061Thermoplastic
Waveform (e.g., pulsed monophasic,BiphasicBiphasicBiphasic
Shape (e.g., rectangular, spike, rectified sinusoidal)RectangularRectangularRectangular
Maximum Output Voltage (volts) (+/-5%)±0.500 @500Ω ±1.700 @2 k Ω ±7.50 @10k Ω±0.400 @500Ω ±1.600 @2 k Ω ±8.00 @10k Ω0.347 @500Ω 1.242 @2 k Ω 5.780 @10k Ω
Maximum Output Current (mA) (+/-5%)±1.0 @500 Ω ±0.850 @2 k Ω ±0.790 @10k Ω±0.800 @500 Ω ±0.800 @2 k Ω ±0.800 @10k Ω0.647 @500 Ω 0.625 @2 k Ω 0.584 @10k Ω
Duration of primary (depolarizing) phase (msec)220 – 128026.3 – 12000.648 – 322
Frequency (Hz) [or Rate (pps)].5 - 12500.045 – 25600.621 – 308.6
For multiphasic waveforms only: Symmetrical phases?YESYesYes
Phases Duration (msec), (state range, if applicable), (both phases, if asymmetrical)220 - 125026.3 – 12000.324-161
Net Charge (micro coulombs (µC) per pulse) (If zero, state method of achieving zero net charge.)0μC @500 Ω0μC @500 Ω0μC @500 Ω
Maximum Phase Charge, (µC)510 µC positive phase, (50% duty cycle), all loads400 µC positive phase, 400 µC positive phase, (50%190 @500 Ω
Maximum Current Density (mA/cm2, r.m.s.)1.421 @500 Ω (1)1.591 @500 Ω (1)1.486 @500 Ω
Maximum Average Current (average absolute value), mA0.850 @500 Ω0.800 @500 Ω0.493 @500 Ω
Maximum Average Power Density, (W/cm2), (using smallest electrode conductive surface area)357E-6 @500Ω318E-6 @500Ω (2) 0.012 @18.75 KΩ (3)366E-6 @500 Ω
ON Time (seconds)ConstantConstant10-30
OFF Time (seconds)NoneNone1-6
Additional Features (specify, if applicable)NoneNoneNone

Basis for a determination of substantial equivalence.

The facial System has two models; SX2500 and SX3800. Both models utilize the same internal electrical components. The models differ in the control/display features and exterior casing/dimensions.

Although the SX3800 model has more channels (1 to 4) than the predicate devices, identical isolated signals come from all channels. The SX2500 and SX3800 uses from 1 to 4 channels substantially equivalent to the predicate device that uses from 1 up to 4 ports. Detailed Technological characteristics Tabular Comparison above is showing substantial equivalence in all Parameters compared.

Page 10

Nonclinical tests demonstrate substantial equivalence.

Both new Devices and the Predicate devices were bench tested at our facility and by third party testing facilities to the following safety standards for safe and effective use.

StandardResult
IEC 60601-1-2:2014 [Including AMD 1:2021]Meets
Basic safety and essential performance: IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007), IEC 60601-1-11:2015+A1Meets
Edition 4.1 EMC (IEC 60601-1-2:2020)Meets
Biocompatibility testing was performed to: ISO 10993-1:2018Meets

Test show that the SX series devices are safe, as effective, and performs as well as or better than the legally marketed device.

Testing has indicated that the output from all 4 channels simultaneously is substantially equivalent as compared to the output from the 1 channel of the predicate devices tested. Detailed Technological characteristics Tabular Comparison above shows substantial equivalence in ALL technical parameters compared. (technical characteristics and parameters were obtained from the predicate 510K's filed with the FDA)

The principles of operation and performance characteristics are substantial equivalence as the predicate device and pose no harm or risk to the user or patient.

Final Summary conclusion:

Therefore, the SX2500 and SX3800 Systems are substantially equivalent to the predicate device with technical characteristics, with Accessories / patient contacting parts, IFU, and performance characteristics. Test show that the SX series devices are safe, as effective, and performs as well as or better than the legally marketed device.

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).