K Number
K251899

Validate with FDA (Live)

Device Name
E Blator
Date Cleared
2025-10-01

(103 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The E Blator is intended for general arthroscopic applications, which include cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator via a standard active lead and a standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.

Device Description

The device is equipped with a patented stainless steel electrode tip encased in a thermal insulator. When used in combination with a compatible electrosurgical generator, the device provides radio frequency (RF) energy for cutting, vaporizing, and coagulating soft tissue. The device enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect with handswitch control, said current then returns back to the electrosurgical unit through the return electrode and connector. This provides the capability for the surgeon to balance tissue dissection and hemostatic effects as needed during a given surgical case. For aspirating models, when coupled with a vacuum system, the surgical field can be cleared of undesirable debris simultaneously by using the aspiration port located at the bottom of the face of the electrode. Removing this debris can facilitate the surgeon's visualization of the surgical field.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - E Blator

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

New Deantronics Taiwan Ltd.
℅ Coombs Craig
President
Coombs Medical Device Consulting, Inc.
427 14th Ave
San Francisco, California 94118

Re: K251899
Trade/Device Name: E Blator
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: September 2, 2025
Received: September 2, 2025

Dear Coombs Craig:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

October 1, 2025

Page 2

New Deantronics Taiwan Ltd.
℅ Coombs Craig
President
Coombs Medical Device Consulting, Inc.
427 14th Ave
San Francisco, California 94118

October 1, 2025

Re: K251899
Trade/Device Name: E Blator
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: September 2, 2025
Received: September 2, 2025

Dear Coombs Craig:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 3

K251899 - Coombs Craig Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 4

K251899 - Coombs Craig Page 3

Sincerely,

Colin K. Chen -S
Digitally signed by Colin K. Chen -S
Date: 2025.10.01 14:46:58 -04'00'

Colin Kejing Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 5

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K251899

Please provide the device trade name(s).

E Blator

Please provide your Indications for Use below.

The E Blator is intended for general arthroscopic applications, which include cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator via a standard active lead and a standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

E Blator Page 9 of 38

Page 6

K251899: E Blator 510(k) Summary

A. Device Information:

CategoryComments
Sponsor:New Deantronics Taiwan Ltd.12F. No.51, Sec. 4, Zhongyang Rd., Tucheng Dist.,New Taipei City, 236 Taiwan
Correspondent Contact Information:Mr. Craig CoombsPresidentCoombs Medical Device Consulting, Inc.427 14th Ave,San Francisco, CA, 94118Tel: 650-380-2474Email: craigJcoombs@gmail.com
Common NameElectrosurgical accessory
Classification Number21 CFR 878.4400
Classification & Product CodeClass II, GEI
Proprietary NameE Blator

Predicate Device Information:

Predicate DeviceEblator Device
ManufacturerE Surgical, LLC
Common NameElectrosurgical Ablator
Premarket Notification #K231126
Classification21 CFR 878.4400
Classification & Product CodeClass 2, GEI

Reference Device 1:

Reference DeviceBovie Disposable Bipolar Ablator
ManufacturerBovie Medical Corporation
Common NameElectrosurgical Generator Accessory
Premarket Notification #K161558
Classification21 CFR 878.4400
Classification & Product CodeClass 2, GEI

Page 1

Page 7

K251899: E Blator 510(k) Summary Page 2

Reference Device 2:

Reference DeviceArthrex Synergy RF System
ManufacturerArthrex, Inc.
Common NameElectrosurgical, Cutting & Coagulation device and accessories
Premarket Notification #K161581
Classification21 CFR 878.4400
Classification & Product CodeClass 2, GEI

Reference Device 3:

Reference DeviceArthrex OPES Electrodes and Accessories Ablator
ManufacturerArthrex, Inc
Common NameHigh Frequency Devices and Accessories
Premarket Notification #K023986
Classification21 CFR 878.4400
Classification & Product CodeClass 2, GEI

B. Date Summary Prepared

1 October 2025

C. Description of Device

The device is equipped with a patented stainless steel electrode tip encased in a thermal insulator. When used in combination with a compatible electrosurgical generator, the device provides radio frequency (RF) energy for cutting, vaporizing, and coagulating soft tissue. The device enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect with handswitch control, said current then returns back to the electrosurgical unit through the return electrode and connector. This provides the capability for the surgeon to balance tissue dissection and hemostatic effects as needed during a given surgical case. For aspirating models, when coupled with a vacuum system, the surgical field can be cleared of undesirable debris simultaneously by using the aspiration port located at the bottom of the face of the electrode. Removing this debris can facilitate the surgeon's visualization of the surgical field.

Page 2

Page 8

K251899: E Blator 510(k) Summary Page 3

D. Indications for Use

The E Blator is intended for general arthroscopic applications, which include cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator via a standard active lead and a standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.

Page 3

Page 9

K251899: E Blator 510(k) Summary Page 4

E. Comparison to Predicate Device

This section presents the justification for the substantial equivalence between the subject device, the New Deantronics E Blator, and the predicate and reference devices, including the E Surgical Eblator (K231126, predicate), Bovie Disposable Bipolar Ablator (K161558, reference 1), Arthrex Synergy RF System (K161581, reference 2), and Arthrex OPES Electrodes and Accessories Ablator (K023986, reference 3).

The tabular comparison of E Blator is thus split up into three tables in accordance with the different electrode tip configurations.

Page 4

Page 10

K251899: E Blator 510(k) Summary Page 5

(I) Classic Tip

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Indications for UseThe E Blator is intended for general arthroscopic applications, which include cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator via a standard active lead and a standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.The E Surgical Eblator device is intended for general arthroscopic applications, which includes cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator active lead and standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.This device is intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures. This device is intended to be used with a standard electrosurgical generator with footswitch control and a standard return electrode connection, and the electrode is to be activated only when immersed in a conductive media such as standard saline solution.Identical to predicate and reference 1, except for the name of the device.
Product CodeGEIGEIGEIIdentical

Page 5

Page 11

K251899: E Blator 510(k) Summary Page 6

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Device CategoryClassic TipRB4050A-01 (4.0mm),RB4050-01 (4.0mm),RB4090A-01 (4.0mm),RB4090-01 (4.0mm),RB4090A-02 (4.0mm),RB4090-02 (4.0mm),RB4050A-02 (4.0mm),RB4050-02 (4.0mm),RB3455-01 (3.4mm),RB3490-01 (3.4mm),RB2790-01 (2.7mm),RB2760-01 (2.7mm),RB2790-00 (2.7mm),RB2760-00 (2.7mm)Classic TipRB4050A-01 (4.0mm),RB4050-01 (4.0mm),RB4090A-01 (4.0mm),RB4090-01 (4.0mm)Classic TipBA2455NA (2.4mm),BA1860NA (1.8mm)The application device presents a subset of models (Classic Tip) that are the same as the predicate and the reference 1 devices.

Technology

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Energy UseRadiofrequency electrical energySameSameIdentical to both
Power outputFrom monopolar portSameSameIdentical to both
Operation ModeBipolarSameSameIdentical to both
Operation PrincipleUse RF power to generate arcing through bubbles formed between an active electrode and tissue with the tissue being vaporized by the arcing.SameSameIdentical to both
Equipment MatedGeneral purpose electrosurgicalGeneral purpose electrosurgicalStandard electrosurgical generatorIdentical to the predicate

Page 6

Page 12

K251899: E Blator 510(k) Summary Page 7

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Equipment Matedgenerator with standard return electrode connection. Compatible foot switch controller is optional.Aspiration:Adequate vacuum source for procedure (Applicable to these models: RB4050A-01, RB4090A-01, RB4090A-02, RB4050A-02)generator with standard return electrode connection. Compatible foot switch controller is optional.Aspiration:Adequate vacuum source for procedure (Applicable to these models: RB4050A-01, RB4090A-01)with standard return electrode connection. Compatible footswitch controller is required.Aspiration:Adequate vacuum source for procedure
Use only in Conductive MediaThe electrode is to be activated only when immersed in a conductive media such as standard saline solutionSameSameIdentical to both

Design – Mechanism

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Main ComponentInsulated handle,Active electrode,Aspirating tube,Cable with ESU plug and return connectorInsulated handle,Active electrode,Aspirating tubeCable with ESU plug and return connectorInsulated handle,Active electrode,Aspirating tube,Return connectorFootswitch connectorIdentical to the predicate.
Handle Width1'' (25.2 mm)1'' (25.2 mm)0.82'' (20.8 mm)Identical to the predicate.
Device Length (without cable and tube)9.76'' (248 mm),13.5'' (343 mm),14.49'' (368 mm)13.5'' (343 mm)11.85'' (301.0 mm)Functionally identical to both
Cable Length12 feet12 feet10 feetIdentical to the predicate
User interfaceHandswitch integrated in the handle or FootswitchHandswitch integrated in the handle or FootswitchFootswitchIdentical to the predicate

Page 7

Page 13

K251899: E Blator 510(k) Summary Page 8

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Aspirating FeatureClassic Tip:Aspirating or non-aspiratingAspirating or non-aspiratingAspirating or non-aspiratingIdentical to both
Shaft Diameter3.6 mm, 4.4 mm4.4 mm3.6 mm, 4.4mmFunctionally identical to both. Identical to the reference 1
Outer Diameter at end of the ceramic insulatorClassic Tip2.7mm, 3.4mm, 4.0mmClassic Tip4.0mmClassic Tip2.7mm, 3.4mmFunctionally identical to both
Electrode AngleClassic Tip: 50°, 60°, 90°Classic Tip: 50°, 90°Classic Tip: 55°,60°Functionally identical to both
Working Length65mm, 160mm, 185mm155 mm160 mmFunctionally identical to both as demonstrated in Thermal Testing

Design- Electrical Feature

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Maximum Allowable Voltage1800 Vp1800 Vp1800 VpIdentical to both
Electrical SafetyWithstand 2160 Vp (120% of 1800Vp) for High Frequency breakdownWithstand 2800 Vp (1000+1800Vp) at Main Frequency (60Hz) breakdownWithstand 2160 Vp (120% of 1800Vp) for High Frequency breakdownWithstand 2800 Vp (1000+1800Vp) at Main Frequency (60Hz) breakdownFollow IEC 60601-2-2 requirementIdentical to the predicate

Design- Material

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
ElectrodeSUS 304SameSameIdentical to both
InsulatorCeramicSameSameIdentical to both
Aspiration TubePVCPVCTygon™Identical to the predicate
Aspiration AdaptorABSSameSameIdentical to both

Page 8

Page 14

K251899: E Blator 510(k) Summary Page 9

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Heat Shrink TubePVDF, POFSameSameIdentical to both
Inner TubeSUS 304SameSameIdentical to both
Outer TubeSUS 304SameSameIdentical to both
AdhesiveEpoxySameSameIdentical to both

Other Attributes

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 1Bovie Disposable Bipolar Ablator (K161558)Pertinence of Feature to Consideration of Substantial Equivalence
Single Use or ReusableSingle useSameSameIdentical to both
SterilizationEthylene OxideSameSameIdentical to both
Standards MetIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIdentical to both

Page 9

Page 15

K251899: E Blator 510(k) Summary Page 10

(II) Hook

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Indications for UseThe E Blator is intended for general arthroscopic applications, which include cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator via a standard active lead and a standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.The E Surgical Eblator device is intended for general arthroscopic applications, which includes cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator active lead and standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.The RF Probe are accessories to the Synergy RF Console and are intended for use as a complete system in the resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels and tissue in arthroscopic and orthopedic procedures. Specifically, the RF probes, Synergy RF Console and their accessories are used for arthroscopic surgery of the shoulder, wrist, hand, elbow, hip, knee, foot and ankle.Identical to the predicate, except for the name of the device.The applications of the E Blator are clinically equivalent (cutting, vaporization, coagulation) to that of the reference devices.
Product CodeGEIGEIGEIIdentical to both
Device CategoryHookRB0700-21 (0.75 mm, 0°)Classic TipRB4050A-01 (4.0mm, 50°),RB4050-01 (4.0mm, 50°),RB4090A-01(4.0mm, 90°),RB4090-01 (4.0mm, 90°)HookAR-9825 (0.75 mm, 0°)The application device includes Hook-type model that is identical to Reference Device 2.

Technology

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Energy UseRadiofrequency electrical energySameSameIdentical to both
Power OutputFrom monopolar portSameSameIdentical to both
Operation ModeBipolarSameSameIdentical to both
Operation PrincipleUse RF power to generate arcing through bubbles formed between an active electrode and tissue with theSameSameIdentical to both

Page 10

Page 16

K251899: E Blator 510(k) Summary Page 11

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Operation Principletissue being vaporized by the arcing.
Equipment MatedGeneral purpose electrosurgical generator with standard return electrode connection. Compatible foot switch controller is optional.General purpose electrosurgical generator with standard return electrode connection. Compatible foot switch controller is optional.Aspiration:Adequate vacuum source for procedure (Applicable to these models: RB4050A-01, RB4090A-01)Specific electrosurgical generator (SynergyRF™ Bipolar Ablation System)Identical to the predicate
Use only in Conductive MediaThe electrode is to be activated only when immersed in a conductive media such as standard saline solutionSameSameIdentical to both

Design – Mechanism

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Main ComponentInsulated handle,Active electrode,Aspirating tube,Cable with ESU plug and return connectorInsulated handle,Active electrode,Aspirating tubeCable with ESU plug and return connectorInsulated handle,Active electrode,Cable with integrated plugIdentical to the predicate.
Handle Width1'' (25.2 mm)1'' (25.2 mm)0.95" (24 mm)Identical to the predicate.
Device Length (without cable and tube)13.5'' (343 mm)13.5'' (343 mm)12.6" (320 mm)Functionally identical to both
Cable Length12 feet12 feet10 feetIdentical to the predicate
User interfaceHandswitch integrated in the handle or FootswitchHandswitch integrated in the handle or FootswitchHandswitch integrated in the handle or FootswitchIdentical to both

Page 11

Page 17

K251899: E Blator 510(k) Summary Page 12

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Aspirating FeatureHook: Non-AspiratingAspirating or Non-AspiratingNon-AspiratingIdentical to both
Shaft Diameter4.4 mm4.4 mm4.15 mmIdentical to the predicate
Outer Diameter at end of the ceramic insulatorHook: 0.75 mmClassic Tip: 4.0 mmHook: 0.75 mmIdentical to the Ref #2. Functionally identical to both
Electrode AngleHook: 0°Classic Tip: 50°, 90°Hook: 0°Identical to the reference 2 device. Functionally identical to both
Working Length160 mmSameSameIdentical to both

Design- Electrical Feature

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Maximum Allowable Voltage1800 Vp1800 Vp550V (round)Identical to the predicate
Electrical SafetyWithstand 2160 Vp (120% of 1800Vp) for High Frequency breakdownWithstand 2800 Vp (1000+1800Vp) at Main Frequency (60Hz) breakdownWithstand 2160 Vp (120% of 1800Vp) for High Frequency breakdownWithstand 2800 Vp (1000+1800Vp) at Main Frequency (60Hz) breakdownFollow IEC 60601-2-2 requirementIdentical to the predicate

Design- Material

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
ElectrodeSUS 304SameSameIdentical to the predicate
InsulatorCeramicSameSameIdentical to the predicate
Aspiration Tube(N/A)(N/A)(N/A)Identical to both. The hook electrode model does not include an aspiration tube is identical to the non-aspirating versions of the predicate

Page 12

Page 18

K251899: E Blator 510(k) Summary Page 13

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Aspiration Adaptor(N/A)(N/A)(N/A)Identical to both. The hook electrode model does not include an adaptor is identical to the non-aspirating versions of the predicate device.
Heat Shrink TubePVDF, POFSameSameIdentical to both
Inner TubeSUS 304SameSameIdentical to both
Outer TubeSUS 304SameSameIdentical to both
AdhesiveEpoxySameSameIdentical to both

Other Attributes

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 2Arthrex Synergy RF System (K161581)Pertinence of Feature to Consideration of Substantial Equivalence
Single Use or ReusableSingle useSameSameIdentical to both
SterilizationEthylene OxideSameSameIdentical to both
Standards MetIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIdentical to both

Page 13

Page 19

K251899: E Blator 510(k) Summary Page 14

(III) Ball

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Indications for UseThe E Blator is intended for general arthroscopic applications, which include cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator via a standard active lead and a standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.The E Surgical Eblator device is intended for general arthroscopic applications, which includes cutting, vaporization, and coagulation. This device is intended to be used in conjunction with a general purpose electrosurgical generator active lead and standard return lead connection. The device is only operable when activated in an appropriate conductive media, such as a standard saline solution.The Arthrex electrosurgical cutting and coagulation devices and accessories are monopolar in structure, allowing for use in a range of surgical procedures. The Arthrex electrosurgical ablation devices and accessories are intended for use in resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels and tissue in general, arthroscopic, and orthopedic procedures. Specifically, these devices and their accessories will be used for general surgeries, and open and arthroscopic surgery of the shoulder, wrist, hand, elbow, hip, knee, and ankle.Identical to the predicate, except for the name of the device.The applications of the E Blator are clinically equivalent (cutting, vaporization, coagulation) to that of the reference devices.
Product CodeGEIGEIGEIIdentical to both

Page 14

Page 20

K251899: E Blator 510(k) Summary Page 15

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Device CategoryBallRB2000-11 (2.0 mm, 0°),RB2040-11 (2.0 mm, 40°)Classic TipRB4050A-01 (4.0mm, 50°),RB4050-01 (4.0mm, 50°),RB4090A-01(4.0mm, 90°),RB4090-01 (4.0mm, 90°)BallAR-9808SJ-45 (0.8 mm, 45°),AR-9808SJ (0.8 mm, 0°)The application device presents a subset of models (Ball) that are present in the reference 3 device.

Technology

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Energy UseRadiofrequency electrical energySameSameIdentical to both
Power outputFrom monopolar portSameSameIdentical to both
Operation ModeBipolarBipolarMonopolarIdentical to the predicate. Technically identical to the reference 3 device
Operation PrincipleUse RF power to generate arcing through bubbles formed between an active electrode and tissue with the tissue being vaporized by the arcing.SameSameIdentical to both
Equipment MatedGeneral purpose electrosurgical generator with standard return electrode connection. Compatible foot switch controller is optional.General purpose electrosurgical generator with standard return electrode connection. Compatible foot switch controller is optional.Aspiration:Adequate vacuum source for procedure (Applicable to these models: RB4050A-01, RB4090A-01)General purpose electrosurgical generator with standard return electrode connectionIdentical to both

Page 15

Page 21

K251899: E Blator 510(k) Summary Page 16

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Use only in Conductive MediaThe electrode is to be activated only when immersed in a conductive media such as standard saline solutionSameSameIdentical to both

Design – Mechanism

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Main ComponentInsulated handle,Active electrode,Aspirating tube,Cable with ESU plug and return connectorInsulated handle,Active electrode,Aspirating tubeCable with ESU plug and return connectorInsulated handle,Active electrodeCable with ESU plugIdentical to the predicate.
Handle Width1'' (25.2 mm)1'' (25.2 mm)0.95" (24 mm)Identical to the predicate.
Device Length (without cable and tube)13.5'' (343 mm)13.5'' (343 mm)8.86" (225 mm)Functionally identical to both
Cable Length12 feet12 feet10 feetIdentical to the predicate
User interfaceHandswitch integrated in the handle or FootswitchHandswitch integrated in the handle or FootswitchHandswitch integrated in the handle or FootswitchIdentical to both
Aspirating FeatureBall: Non-AspiratingAspirating or non-aspiratingNon-AspiratingIdentical to both
Shaft Diameter4.4 mm4.4 mm4.15 mmIdentical to the predicate
Outer Diameter at end of the ceramic insulatorBall: 2.0 mmClassic Tip: 4.0 mmBall: 0.8 mmIdentical geometry to the reference 3 device. Functionally identical to predicate in Thermal Testing.

Page 16

Page 22

K251899: E Blator 510(k) Summary Page 17

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Electrode AngleBall: 40°, 0°Classic Tip: 50°, 90°Ball: 45°, 0°Functionally identical to both
Working Length160 mm160 mm65 mmIdentical to the predicate

Design- Electrical Feature

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Maximum Allowable Voltage1800 Vp1800 Vp6750pIdentical to the predicate
Electrical SafetyWithstand 2160 Vp (120% of 1800Vp) for High Frequency breakdownWithstand 2800 Vp (1000+1800Vp) at Main Frequency (60Hz) breakdownWithstand 2160 Vp (120% of 1800Vp) for High Frequency breakdownWithstand 2800 Vp (1000+1800Vp) at Main Frequency (60Hz) breakdownFollow IEC 60601-2-2 requirementIdentical to the predicate

Design- Material

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
ElectrodeSUS 304SameSameIdentical to both
InsulatorHeatshrinkCeramicHeatshrinkFunctionally identical to both
Aspiration Tube(N/A)(N/A)(N/A)Identical to both. The ball electrode models do not include an aspiration tube are identical to the non-aspirating versions of the predicate device.
Aspiration Adaptor(N/A)(N/A)(N/A)Identical to both. The ball electrode models do not include an adaptor and are identical to the non-aspirating versions of

Page 17

Page 23

K251899: E Blator 510(k) Summary Page 18

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Aspiration Adaptorthe predicate device.
Heat Shrink TubePVDF, POFSameSameIdentical to both
Inner TubeSUS 304SameSameIdentical to both
Outer TubeSUS 304SameSameIdentical to both
AdhesiveEpoxySameSameIdentical to both

Other Attributes

FeatureApplication DeviceNew Deantronics E BlatorPredicate DeviceE Surgical Eblator (K231126)Reference Device 3Arthrex OPES Electrodes and Accessories Ablator(K023986)Pertinence of Feature to Consideration of Substantial Equivalence.
Single Use or ReusableSingle useSameSameIdentical to both
SterilizationEthylene OxideSameSameIdentical to both
Standards MetIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIEC 60601-1IEC 60601-1-2IEC 60601-2-2FDA RF GuidanceIdentical to both

Page 18

Page 24

K251899: E Blator 510(k) Summary Page 19

F. Summary of Supporting Data

The E Blator was tested and found to be in compliance with the pertinent portions of the following standards:

StandardsTitleVersion
IEC 60601-1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance2005 + AMD1:2012 +AMD2:2020
IEC 60601-1-2Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests2014 +AMD1:2020
IEC 60601-2-2Medical electrical equipment –Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories2017 + AMD1:2023
ISO10993-1Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process2018
ISO 11135Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices2014 + AMD1: 2018
ISO11607-1Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems2019 + AMD1:2023
ISO11607-2Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming, sealing and assembly processes2019+ AMD1:2023

In addition, the E Blator was fully tested and in compliance with the FDA guideline Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery: Guidance for Industry and Food and Drug Administration Staff (9 March 2020).

G. Conclusion

After comparing the indications for use, technology and design of the E Blator, along with all electrical safety and performance testing, in accordance with the FDA's guidelines and FDA-recognized consensus standards for electrical safety, New Deantronics concludes that the E Blator is substantially equivalent to the predicate and reference devices.

Page 19

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.