K Number
K251801

Validate with FDA (Live)

Date Cleared
2025-08-08

(57 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Diode laser device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI),including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Device Description

Diode laser device consists of a main unit, a handheld, a foot switch and a power cord.

Device composition and handle characteristics: The device includes a handle with a single-band 808nm semiconductor laser, and the semiconductor lasers of the handle are powered by the laser power supply in the host to emit the laser of the corresponding wavelength.

Working principle of semiconductor laser: Semiconductor laser uses semiconductor materials of different doping types as laser working substance, uses natural cleavage surface to form resonant cavity for laser oscillation and amplification, and adds forward voltage to PN junction area of semiconductor laser to form particle number inversion of non-equilibrium carrier between conduction band and valence band of semiconductor substance. When a large number of electrons and holes in particle number inversion state are recombined, excess energy will be released, and these energies will be expressed in the form of photons, that is, laser is formed. Due to resonance amplification of cleavage surface resonant cavity, stimulated feedback is realized, so that laser can be directionally emitted and output from semiconductor laser.

AI/ML Overview

This 510(k) clearance letter and summary primarily focus on demonstrating substantial equivalence to a predicate device, which is typical for Class II medical devices. The information provided does not contain a detailed study report with specific acceptance criteria related to a performance study for the device's efficacy (hair removal/reduction). Instead, it relies on comparison to a predicate device and adherence to recognized standards for safety and performance.

Here's an analysis based on the provided text:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in the context of a clinical or performance study with numerical targets for efficacy (e.g., a certain percentage of hair reduction). Instead, it relies on demonstrating substantial equivalence to a predicate device that is already legally marketed for the same indication. The "performance" assessment is based on the similarity of technical specifications and the successful completion of non-clinical tests to applicable standards, ensuring safety and basic functionality.

ItemAcceptance Criteria (Implied)Reported Device Performance
Clinical EfficacySubstantially equivalent indications for use as predicate (hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin, with permanent hair reduction defined as a long-term, stable reduction at 6, 9, and 12 months post-treatment.)The proposed device has the same Indications for Use as the predicate devices, K232709 and K241642. The document states that the safety of the product is no problem despite fluence differences, and effectiveness is considered similar to a reference device with comparable fluence. No direct performance data is provided in terms of hair reduction percentage or success rates from a clinical trial in this document.
Technical DesignLaser Type, Classification, Wavelength, Frequency, Pulse Duration, Power Supply, Dimensions, Weight to be comparable or justified if different.- Laser Type, Classification, Wavelength, Frequency, Pulse Duration, Power Supply: Marked as "SAME" as the predicate. - Spot Size: Different, but argued not to affect safety and effectiveness, only treatment area. - Fluence: Different from the primary predicate but similar to a reference device (K241642), ensuring effectiveness is "not affected." Safety argued to be no problem due to lower max fluence than the predicate. - Dimension/Weight: Different but argued not to affect safety/effectiveness and comply with mechanical performance standards.
Safety TestingCompliance with recognized safety standards.The device passed non-clinical tests verifying compliance with IEC 60601-1, IEC 60825-1, IEC 60601-2-22, ISO 10993-10, ISO 10993-23, ISO 10993-5, and IEC 60601-1-2. These standards cover general safety, laser safety, specific laser equipment safety, biocompatibility (skin sensitization, irritation, cytotoxicity), and electromagnetic compatibility.

2. Sample size used for the test set and the data provenance

The document explicitly states: "No clinical study is included in this submission."

Therefore:

  • Sample size for the test set: Not applicable (no clinical test set was used for this submission).
  • Data provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical study or test set with expert-established ground truth was part of this submission. The "ground truth" for clearance in this context is established by the FDA's acceptance of substantial equivalence to an already cleared device, which presumably had clinical data at the time of its own clearance.

4. Adjudication method for the test set

Not applicable, as no clinical test set requiring adjudication was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a laser for hair removal, not an AI-assisted diagnostic or imaging device that would typically involve human readers or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device (diode laser), not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the purpose of this 510(k) submission, the "ground truth" for the device's indications for use and general safety/effectiveness is implicitly the regulatory clearance and established performance of the predicate devices. The applicant is demonstrating that their device is sufficiently similar to these predicates to be considered equally safe and effective.

8. The sample size for the training set

Not applicable, as no clinical study was conducted. There is no "training set" in the context of this 510(k) submission for efficacy.

9. How the ground truth for the training set was established

Not applicable, as no clinical study or training set was used.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.08.00
Silver Spring, MD 20993
www.fda.gov

August 8, 2025

Hebei Newangie Technology Co., Ltd
℅ Wang Ray
General Manager
Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd.
FangShan District
Beijing, 102401
China

Re: K251801
Trade/Device Name: Diode laser device (BM091)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: June 12, 2025
Received: June 12, 2025

Dear Wang Ray:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K251801 - Wang Ray Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K251801 - Wang Ray Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251801

Please provide the device trade name(s).

Diode laser device (BM091)

Please provide your Indications for Use below.

The Diode laser device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI),including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

Please select the types of uses (select one or both, as applicable).
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Diode laser device Page 9 of 35

Page 5

510(k) Summary

The assigned 510(k) Number: K251801

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

  1. Date of Preparation: 2025.06.11

  2. Sponsor Identification

Hebei Newangie Technology Co., Ltd
No. 509, Taihang South Street, High-tech Zone, Shijiazhuang City, Hebei Province Room 402, Building A-5, Guoxietang Industrial Park

Contact Person: Ning Wang
Position: Management Representative
Tel: +86-18633029508
Email: 892674402@qq.com

  1. Designated Submission Correspondent

Mr. Ray Wang
Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing, 102401, China
Tel: +86-18910677558
Fax: +86-10-56335780
Email: information@believe-med.com

Page 6

  1. Identification of Proposed Device

Trade Name: Diode laser deviece
Model: BM091
Common Name: Powered Laser Surgical Instrument

Regulatory Information
Classification Name: Powered Laser Surgical
Instrument Classification: II
Product Code: GEX
Regulation Number: 878.4810
Review Panel: General & Plastic Surgery

  1. Identification of Predicate Device(s)

510(k) Number: K232709
Product Name: Diode laser therapy device
Manufacturer: Hebei Zhemai Technology Co., Ltd

Reference Device:
510(k) Number: K241642
Product Name: Diode Laser Hair Removal System (MBT-Diode Laser)
Manufacturer: Beijing Mega Beauty Technology Co.,Ltd.

  1. Device Description

Diode laser device consists of a main unit, a handheld, a foot switch and a power cord.

Device composition and handle characteristics: The device includes a handle with a single-band 808nm semiconductor laser, and the semiconductor lasers of the handle are powered by the laser power supply in the host to emit the laser of the corresponding wavelength.

Working principle of semiconductor laser: Semiconductor laser uses semiconductor materials of different doping types as laser working substance, uses natural cleavage surface to form resonant cavity for laser oscillation and amplification, and adds forward voltage to PN junction area of semiconductor laser to form particle number inversion of non-equilibrium carrier between conduction band and valence band of semiconductor substance. When a large number of electrons and holes in particle number inversion state are recombined, excess energy will be released, and these energies will be expressed in the form of photons, that is, laser is formed. Due to resonance amplification of cleavage surface resonant cavity, stimulated feedback is realized, so that laser can be directionally emitted and output from semiconductor laser.

  1. Indication For Use Statement:

Diode laser device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI),including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months

Page 7

after the completion of a treatment regime.

  1. Substantially Equivalent (SE) Comparison

Tab 1 General Comparison

ItemProposed DevicePredicate Device K232709Reference Device K241642Remark
Device NameDiode laser deviceDiode laser therapy deviceDiode Laser Hair Removal System
Classification Regulation21 CFR 878.481021 CFR 878.481021 CFR 878.4810SAME
ClassIIIIIISAME
Product CodeGEXGEXGEXSAME
Indication for useThe Diode laser device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.The Diode laser therapy device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI),including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.SAME

Tab 2 Performance Comparison

ItemProposed DevicePredicate Device K232709Reference Device K241642Remark
Laser TypeDiode LaserDiode LaserDiode LaserSAME
Laser ClassificationClass IVClass IVClass IVSAME
Laser wavelength808 nm808 nm808 nmSAME

Page 8

Spot Size12mm x 20mm12mm × 35mm15mm × 15mmDifferent1
Fluence2-60j/cm25-100 J/cm21-55.96j/cm2Different2
Frequency1-10HZ1-10HZ1-10HZSAME
Pulse Duration5-200ms5-200ms2-240msSAME
Power SupplyAC100-240V/50-60HzAC110-240V/60HzAC110-240V/50-60HzSAME
Dimension375mm440mm1170mm55cm55cm123cm650mm580mm1210mmDifferent3
Weight63kg62kg68kgDifferent4

Analysis

Different1-Spot Size
The proposed device is different in Spot Size from the predicate, Spot size only affects the area of treatment, not affect the therapeutic effect. Therefore, this difference will not affect the safety and effectiveness.

Different2-Fluence
The proposed device has different Fluence from the predicate device.

For the difference on fluence between the predicate and proposed device, The Fluence of Predicate Device K232709 is 5-100J/cm2, while the propose device is 2-60 J/cm2.It is lower than 100 J/cm2, the safety of the product is no problem. However, the effectiveness cannot be determined, so we selected Reference Device K241642 with the same intended use for Fluence comparison,The Fluence of the Reference Device is 1-55.96J/cm2, which is similar to the Fluence of the proposed device, so the effectiveness of the product is not affected. And the proposed device has passed the IEC60601-1 test, IEC60601-1-2 test, IEC60601-2-22 test, IEC60825-1 test and performance test, the safety and performance of the product can be ensured.

Different3/4- Dimension/ Weight
The proposed device is different in dimension and weight from the predicate device.

However, the dimension and weight difference are just in physical specification and this difference will not raise any issues in safety and effectiveness. By complying with AAMI/ANSI/ES 60601-1, the mechanical performance of the proposed device is determined to be accepted. Therefore, this difference will not affect safety and effectiveness of the proposed device.

  1. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • IEC 60601-1 Edition 3.2 2020-08, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification, and requirements
  • IEC 60601-2-22: 2012 Medical electrical equipment - Part 2-22: Particular requirements for

Page 9

basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment

  • ISO 10993-10 Fourth edition 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
  • ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation
  • ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
  • IEC 60601-1-2 Edition 4.1 2020-09, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Test
  1. Clinical Test Conclusion

No clinical study is included in this submission.

  1. Conclusion

Based on the non-clinical tests performed, the subject device is as safe, as effective, and performance as well as the legally marketed predicate device, Diode laser therapy device leared under K232709.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.