(176 days)
The Omnipod 5 algorithm is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. The Omnipod 5 algorithm is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years of age and older. The Omnipod 5 algorithm is intended for single patient use and requires a prescription.
The Omnipod 5 algorithm is a software-only device that enables automated insulin delivery by taking in glucose inputs from a connected iCGM and calculating insulin micro-bolus outputs for delivery by a connected ACE Pump.
The Omnipod 5 algorithm is part of the Omnipod 5 Automated Insulin Delivery System, which also includes the Omnipod 5 ACE Pump and the SmartBolus Calculator regulated devices. The Omnipod 5 ACE Pump and the SmartBolus Calculator are functionally independent from the Omnipod 5 algorithm. The Omnipod 5 algorithm is intended to be digitally connected to a third party iCGM, the Omnipod 5 ACE Pump, and the SmartBolus Calculator.
The Omnipod 5 algorithm software is installed on both the Omnipod 5 Pod and Omnipod 5 App (either installed on a compatible smartphone or on the Insulet provided Controller) to comprise the Omnipod 5 System. As all regulated devices are required for the Omnipod 5 System, a prescription for the Omnipod 5 System and Omnipod 5 Pods includes:
• Omnipod Pods with Omnipod 5 algorithm software and
• Omnipod 5 Application with Omnipod 5 algorithm and SmartBolus Calculator software provided on Insulet provided Controller and when installed on the user's personal mobile device.
The Omnipod 5 System is a hybrid closed loop system and therefore can operate in either open loop (Manual Mode; Omnipod 5 algorithm disabled) or closed loop (Automated Mode; Omnipod
5 algorithm enabled). When Automated Mode is turned on, the Omnipod 5 algorithm (installed on the Pod) controls insulin delivery based on recent CGM values.
N/A
FDA 510(k) Clearance Letter - Omnipod 5 Algorithm
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.02
December 3, 2025
Insulet Corporation
Alexander Hamad
Manager, Regulatory Affairs
100 Nagog Park
Acton, Massachusetts 01720
Re: K251779
Trade/Device Name: Omnipod 5 algorithm
Regulation Number: 21 CFR 862.1356
Regulation Name: Interoperable Automated Glycemic Controller
Regulatory Class: Class II
Product Code: QJI
Dated: June 9, 2025
Received: June 10, 2025
Dear Alexander Hamad:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K251779 - Alexander Hamad
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
Page 3
K251779 - Alexander Hamad
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSHUA BALSAM -S
Joshua M. Balsam, Ph.D.
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
Please provide the device trade name(s).
Omnipod 5 algorithm
Please provide your Indications for Use below.
The Omnipod 5 algorithm is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. The Omnipod 5 algorithm is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years of age and older. The Omnipod 5 algorithm is intended for single patient use and requires a prescription.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Omnipod 5 algorithm
Page 8 of 41
Page 5
510(k) Summary
Page 1 of 6
| Submission Number | K251779 |
|---|---|
| Date prepared: | November 28, 2025 |
| Submitter Name: | Insulet Corporation |
| Submitter Address: | 100 Nagog ParkActon, MA 01720 |
| FDA Establishment Owner/Operator Number: | 9056196 |
| FDA Establishment Registration Number: | 3014585508 |
| Primary Contact Person | Alexander HamadManager, Regulatory Affairs |
| Phone: | 978-600-2432 |
| Fax: | 978-600-0120 |
| Device Trade / Proprietary Name: | Omnipod 5 algorithm |
| Device Common Name: | Interoperable Automated Glycemic Controller |
| Review Panel (s): | Clinical Chemistry |
| Product Code(s): | QJI |
| Regulation Numbers: | 21 CFR 862.1356 |
| Submission Type: | Traditional 510(k) |
| Device Class: | Class II |
| Device Predicate: | K241777 (SmartAdjust™ technology) |
Page 6
Page 2 of 6
1. Intended Use and Indications for Use
The Omnipod 5 algorithm is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. The Omnipod 5 algorithm is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years of age and older. The Omnipod 5 algorithm is intended for single patient use and requires a prescription.
2. Device Description
The Omnipod 5 algorithm is a software-only device that enables automated insulin delivery by taking in glucose inputs from a connected iCGM and calculating insulin micro-bolus outputs for delivery by a connected ACE Pump.
The Omnipod 5 algorithm is part of the Omnipod 5 Automated Insulin Delivery System, which also includes the Omnipod 5 ACE Pump and the SmartBolus Calculator regulated devices. The Omnipod 5 ACE Pump and the SmartBolus Calculator are functionally independent from the Omnipod 5 algorithm. The Omnipod 5 algorithm is intended to be digitally connected to a third party iCGM, the Omnipod 5 ACE Pump, and the SmartBolus Calculator.
The Omnipod 5 algorithm software is installed on both the Omnipod 5 Pod and Omnipod 5 App (either installed on a compatible smartphone or on the Insulet provided Controller) to comprise the Omnipod 5 System. As all regulated devices are required for the Omnipod 5 System, a prescription for the Omnipod 5 System and Omnipod 5 Pods includes:
• Omnipod Pods with Omnipod 5 algorithm software and
• Omnipod 5 Application with Omnipod 5 algorithm and SmartBolus Calculator software provided on Insulet provided Controller and when installed on the user's personal mobile device.
The Omnipod 5 System is a hybrid closed loop system and therefore can operate in either open loop (Manual Mode; Omnipod 5 algorithm disabled) or closed loop (Automated Mode; Omnipod
Page 7
Page 3 of 6
5 algorithm enabled). When Automated Mode is turned on, the Omnipod 5 algorithm (installed on the Pod) controls insulin delivery based on recent CGM values.
3. Summary of Technological Characteristics Compared to Predicate
Please refer to Table 1 below for a substantial equivalence table, which includes a comparison of the technological characteristics between the subject device and its predicate.
Table 1. Omnipod 5 Algorithm Substantial Equivalence Comparison
| Element of Comparison | Subject Device: The Omnipod 5 algorithm | Predicate Device: SmartAdjust technology (K241777) |
|---|---|---|
| Intended use/indications for use | The Omnipod 5 algorithm is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. The Omnipod 5 algorithm is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years or age and older. The Omnipod 5 algorithm is intended for single patient use and requires a prescription | SmartAdjust™ technology is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and pause delivery of insulin based on current and predicted glucose values. SmartAdjust™ technology is intended for the management of type 1 diabetes mellitus in persons 2 years of age and older and type 2 diabetes mellitus in persons 18 years or age and older. SmartAdjust™ technology is intended for single patient use and requires a prescription |
| Device Type | Interoperable automated glycemic controller | Interoperable automated glycemic controller |
| Environment of use | Ambulatory use | Ambulatory use |
| Specific Drug/Biologic Use | U-100 InsulinSystem has been tested with NovoLog®, Humalog®, and Admelog® | U-100 InsulinSystem has been tested with NovoLog®, Humalog®, and Admelog® |
Page 8
Page 4 of 6
| Element of Comparison | Subject Device: The Omnipod 5 algorithm | Predicate Device: SmartAdjust technology (K241777) |
|---|---|---|
| Prescription Status | Prescription Device | Prescription Device |
| Principles of Operation | Algorithmic software device intended to automatically increase, decrease, and pause delivery of insulin based on iCGM readings and predicted glucose values | Algorithmic software device intended to automatically increase, decrease, and pause delivery of insulin based on iCGM readings and predicted glucose values |
| Communication and Pairing | Bluetooth Low Energy (BLE) wireless technology | Bluetooth Low Energy (BLE) wireless technology |
| Device Hosting Controller | Omnipod 5 ACE Pump | Omnipod 5 ACE Pump |
| Digitally Connected Devices | Cleared iCGM and ACE Pump, which includes a display device (Insulet-provided Controller phone or user's compatible smartphone) | Cleared iCGM and ACE Pump, which includes a display device (Insulet-provided Controller phone or user's compatible smartphone) |
| System Functionality Modes | • The Omnipod 5 algorithm enabled (Automated Mode): closed-loop, automatically increase, decrease, and suspend delivery of insulin based on current and predicted glucose values• The Omnipod 5 algorithm not enabled (Manual Mode): open loop, basal delivery based on user-defined programs | • The Omnipod 5 algorithm enabled (Automated Mode): closed-loop, automatically increase, decrease, and suspend delivery of insulin based on current and predicted glucose values• The Omnipod 5 algorithm not enabled (Manual Mode): open loop, basal delivery based on user-defined programs |
| Alarms/Alerts | • Out of Range Alert• Low Alert (ACE Pump)• Maximum Insulin Delivery Alert | • Out of Range Alert• Low Alert (ACE Pump)• Maximum Insulin Delivery Alert |
| Alert/Alarm Display | ACE Pump alarms/alerts will be displayed and communicated to the user via visual, audible, and/or vibratory cues. | ACE Pump alarms/alerts will be displayed and communicated to the user via visual, audible, and/or vibratory cues. |
| Target Glucose | 100-150 mg/dl, user-customizable | 110-150 mg/dl, user-customizable |
Page 9
Page 5 of 6
| Element of Comparison | Subject Device: The Omnipod 5 algorithm | Predicate Device: SmartAdjust technology (K241777) |
|---|---|---|
| Control Range | ||
| Mechanism of Software Update | Firmware over the Air | Firmware over the Air |
| History Storage | Up to 90 days (user insulin history) | Up to 90 days (user insulin history) |
| Labeling | Package Labels, User's Guide (contains Instructions for Use), Technical User Guide | Package Labels, User's Guide (contains Instructions for Use), Technical User Guide |
4. Summary of Non-Clinical Performance Testing
• Software V&V: software verification and validation testing was performed in accordance with IEC 62304:2015 and FDA's guidance document General Principles of Software Validation – issued January 11, 2002. Software documentation was provided in accordance with FDA guidance document Content of Premarket Submissions for Device Software Functions – issued June 14, 2023.
- Interoperability: Interoperability documentation was provided as it relates to changes to the Omnipod 5 algorithm according to the FDA guidance document Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices - Guidance for Industry and Food and Drug Administration Staff – Issued September 6, 2017.
• In Silico Clinical Study: Performance Validation was conducted via an in silico clinical trial. Details of the study were provided in accordance with the FDA guidance document Assessing the Credibility of Computational Modeling and Simulation in Medical Device Submissions – issued November 17, 2023. The study concludes that the simulated performance of Omnipod 5 with the 100 mg/dL target as compared to the real-world performance of Omnipod 5 with the 110 mg/dL target meets the acceptance criteria for non-inferiority and demonstrates substantial equivalence to the predicate device.
• Special Controls: evaluation of the Special Controls for this device (regulation 21 CFR 862.1356) supports the safety and effectiveness of the device.
Page 10
Page 6 of 6
5. Conclusion
The Omnipod 5 algorithm has the same intended use and indications for use as the predicate device. The Omnipod 5 algorithm has the same technological characteristics and principles of operation as the predicate device. The software has been updated to add 100 mg/dL as a target glucose value input to the Algorithm and to modify the ADR alert to improve user experience and allow users to remain in Automated Mode after acknowledging the Alert. These modifications to the design do not raise different questions of safety and effectiveness. The changes to the software are supported by in silico clinical validation and design verification and validation which demonstrate that the subject device and predicate device are substantially equivalent. Product labeling has been updated to reflect the changes to the software, add a summary of the in silico clinical trial, and update the name of the device. The differences between the predicate and subject devices do not raise any different questions about safety and effectiveness, therefore, the Omnipod 5 algorithm is substantially equivalent to its predicate.
§ 862.1356 Interoperable automated glycemic controller.
(a)
Identification. An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect (
e.g., more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events.
(ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices.
(iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to (
e.g., specification of glucose sensor performance) or accept commands from (e.g., specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices.
(vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place (
e.g., validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (e.g., commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller.
(2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures to address device system hazards;
(ii) Design decisions related to how the risk control measures impact essential performance; and
(iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to connected devices;
(ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices;
(iii) Sharing of necessary state information between the controller and any connected devices (
e.g., battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified;
(v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and
(vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance.
(4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include:
(i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices;
(ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices (
e.g., infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and
(iv) Connectivity events (
e.g., establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section.
(6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section.
(7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable.
(8) The labeling required under § 809.10(b) of this chapter must include:
(i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part.
(ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe.
(iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate.
(iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section.
(v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate.
(vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section.
(vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.