K Number
K251602

Validate with FDA (Live)

Date Cleared
2025-10-10

(136 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

αEvolve Imaging is an imaging chain intended for adults, with Artificial Intelligence Denoising (AID) designed to reduce noise in real-time fluoroscopic images and signal enhancement algorithm, Multi Frequency Processing (MFP).

Device Description

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging, is an interventional X-ray system with a floor mounted C-arm as its main configuration. An optional ceiling mounted C-arm is available to provide a bi-plane configuration where required. Additional units include a patient table, X-ray high-voltage generator and a digital radiography system. The C-arms can be configured with designated X-ray detectors and supporting hardware (e.g. X-ray tube and diagnostic X-ray beam limiting device). The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging includes αEvolve Imaging, an imaging chain intended for adults, with Artificial Intelligence Denoising (AID) designed to reduce noise in real-time fluoroscopic images and signal enhancement algorithm, Multi Frequency Processing (MFP).

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based solely on the provided FDA 510(k) summary:

Overview of the Device and its New Feature:

The device is the Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging. It's an interventional X-ray system. The new feature, αEvolve Imaging, includes Artificial Intelligence Denoising (AID) to reduce noise in real-time fluoroscopic images and a signal enhancement algorithm, Multi Frequency Processing (MFP). The primary claim appears to be improved image quality (noise reduction, sharpness, contrast, etc.) compared to the previous version's (V9.5) "super noise reduction filter (SNRF)."


1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary does not explicitly state "acceptance criteria" with numerical thresholds for each test. Instead, it describes various performance evaluations and their successful outcomes. For the clinical study, the success criteria are clearly defined.

Acceptance Criteria (Inferred/Stated)Reported Device Performance
Bench Testing (Image Quality)
1. Change in Image Level, Noise & Structure: AID to be better at preserving mean image intensity, improved denoising, and image structure preservation compared to SNRF.AID determined to be better at preserving mean image intensity and suggested to have improved denoising and image structure preservation (using student's t-test).
2. Signal-to-Variance Ratio (SVR) and Signal-to-Noise Ratio (SNR): AID to show improved ability to preserve image signal while decreasing image noise compared to SNRF.AID determined to have improved ability to preserve image signal while decreasing image noise (using student's t-test).
3. Modulation Transfer Function (MTF): Improved performance for low-to-mid frequencies and similar high-frequency region compared to SNRF.Results showed improved performance for low-to-mid frequencies in all test cases, and high-frequency region of MTF curve was similar for AID and SNRF in majority of cases (using student's t-test).
4. Robustness to Detector Defects: Detector defects to be sufficiently obvious to inform clinician of service need, and image quality outside the defect area to remain visually unaffected, facilitating procedure completion.Detector defects were sufficiently obvious, and image quality outside the area of the detector defect remained visually unaffected, facilitating sufficient image quality to finish the procedure.
5. Normalized Noise Power Spectrum (NNPS): AID to have smaller noise magnitude in the frequency range of ~0.1 cycles/mm to 1.4 cycles/mm, with negligible differences above 1.4 cycles/mm.AID had a smaller noise magnitude in the frequency range of ~0.1 cycles/mm to 1.4 cycles/mm. Noise magnitudes above 1.4 cycles/mm were very small and differences considered negligible.
6. Image Lag Measurement: AID to perform better in reducing image lag compared to SNRF.AID determined to perform better in reducing image lag (using student's t-test).
7. Contrast-to-Noise Ratio (CNR) of Low Contrast Object: AID to show significantly higher CNR for low-contrast elements compared to SNRF.AID had a significantly higher CNR than images processed with SNRF for all elements and test cases (using student's t-test).
8. Contrast-to-Noise Ratio (CNR) of High Contrast Object: AID to show significantly higher CNR for high-contrast objects (guidewire, vessels) compared to SNRF.AID had a significantly higher vessel and guidewire CNR than images processed with SNRF for all test cases (using student's t-test).
Clinical Study (Reader Study)
Overall Preference (Binomial Test): Image sequences denoised by AID chosen significantly more than 50% of the time over SNRF.The Binomial test found that image sequences denoised by AID were chosen significantly more than 50% of the time (indicating overall preference).
Individual Image Quality Metrics (Wilcoxon Signed Rank Test): Mean score of AID images significantly higher than SNRF for sharpness, contrast, confidence, noise, and absence of image artifacts.The mean score of AID imaging chain images was significantly higher than that of the SNRF imaging chain for sharpness, contrast, confidence, noise, and the absence of image artifacts.
Generalizability: Algorithm to demonstrate equivalent or improved performance compared to the predicate with diverse clinical data.Concluded that the subject algorithm demonstrated equivalent or improved performance, compared to the predicate device, as demonstrated by the results of the above testing.

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary provides the following information about the clinical test set:

  • Clinical Dataset Source: Patient image sequences were acquired from three hospitals:
    • Memorial Hermann Hospital (Houston, Texas, USA)
    • Waikato Hospital (Hamilton, New Zealand)
    • Saiseikai Kumamoto Hospital (Kumamoto, Japan)
  • Data Provenance: The study used retrospective "patient image sequences" for side-by-side comparison. The summary does not specify if the acquisition itself was prospective or retrospective, but the evaluation of pre-existing sequences makes it a retrospective study for the purpose of algorithm evaluation.
  • Sample Size: The exact number of patient image sequences or cases used in the clinical test set is not specified in the provided document. It only mentions that the sequences were split into four BMI subgroups.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: The document states the clinical comparison was "reviewed by United States board-certified interventional cardiologists." The exact number of cardiologists is not specified.
  • Qualifications: "United States board-certified interventional cardiologists." No mention of years of experience or other specific qualifications is provided.

4. Adjudication Method for the Test Set

The document describes a "side-by-side comparison" reviewed by experts in the clinical performance testing section. For the overall preference and individual image quality metrics, statistical tests (Wilcoxon signed rank test and Binomial test) were used. This implies that the experts rated or expressed preference for both AID and SNRF images, and these individual ratings/preferences were then aggregated and analyzed.

The exact adjudication method (e.g., 2+1, 3+1 consensus) for establishing a ground truth or a final decision on image quality aspects is not explicitly stated. It seems each expert provided their assessment, and these assessments were then statistically analyzed for superiority rather than reaching a consensus for each image pair.


5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: Yes, a type of MRMC comparative study was conducted. The clinical performance testing involved multiple readers (US board-certified interventional cardiologists) evaluating multiple cases (patient image sequences).

  • Effect Size of Human Readers' Improvement with AI Assistance: The study directly compared AID-processed images to SNRF-processed images in a side-by-side fashion. It doesn't measure how much humans improve with AI assistance in a diagnostic task (e.g., how much their accuracy or confidence improves when using AI vs. not using AI). Instead, it measures the perceived improvement in image quality of the AI-processed images when evaluated by human readers.

    • The study determined: "the mean score of the AID imaging chain images was significantly higher than that of the SNRF imaging chain with regard to sharpness, contrast, confidence, noise, and the absence of image artifacts."
    • And for overall preference, "the Binomial test found that the image sequences denoised by AID were chosen significantly more than 50% of the time."

    This indicates a statistically significant preference for and higher perceived image quality in AID-processed images by readers. However, it does not quantify diagnostic performance improvement with AI assistance, as it wasn't a study of diagnostic accuracy but rather image quality assessment. The "confidence" metric might hint at improved reader confidence using AID images, but it's not a direct measure of diagnostic effectiveness.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, extensive standalone performance testing of the AID algorithm was conducted through "Performance Testing – Bench" and "Image Quality Evaluations." This involved objective metrics and phantom studies without human subjective assessment.

Examples include:

  • Change in Image Level, Noise and Structure
  • Signal-to-Variance Ratio (SVR) and Signal-to-Noise Ratio (SNR)
  • Modulation Transfer Function (MTF)
  • Robustness to Detector Defects (visual comparison, but the algorithm's output is purely standalone)
  • Normalizes Noise Power Spectrum (NNPS)
  • Image Lag Measurement
  • Contrast-to-Noise Ratio of a Low Contrast Object
  • Contrast-to-Noise Ratio of a High Contrast Object

7. The Type of Ground Truth Used

  • For Bench Testing: The ground truth for bench tests was primarily established through physical phantoms and objective image quality metrics. For example, the anthropomorphic chest phantom, low-contrast phantom, and flat field fluoroscopic images provided known characteristics against which AID and SNRF performance were measured using statistical tests.
  • For Clinical Study: The ground truth for the clinical reader study was established by expert opinion/subjective evaluation (preference and scores for sharpness, contrast, noise, confidence, absence of artifacts) from "United States board-certified interventional cardiologists." There is no mention of a more objective ground truth like pathology or outcomes data for the clinical image evaluation.

8. The Sample Size for the Training Set

The document does not provide any information about the sample size used for the training set of the Artificial Intelligence Denoising (AID) algorithm.


9. How the Ground Truth for the Training Set was Established

The document does not provide any information about how the ground truth for the training set was established. It describes the AID as "Artificial Intelligence Denoising (AID) designed to reduce noise," implying a machine learning approach, but details on its training are missing from this summary.

FDA 510(k) Clearance Letter - Canon Medical Systems

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

October 10, 2025

Canon Medical Systems Corporation
℅ Jonathan Toy
Manager, Regulatory Affairs
Canon Medical Systems, USA
2441 Michelle Drive
TUSTIN, CA 92780

Re: K251602
Trade/Device Name: Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging
Regulation Number: 21 CFR 892.1650
Regulation Name: Image-Intensified Fluoroscopic X-Ray System
Regulatory Class: Class II
Product Code: OWB
Dated: August 28, 2025
Received: August 28, 2025

Dear Jonathan Toy:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251602 - Jonathan Toy Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251602 - Jonathan Toy Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251602

Device Name: Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging

Indications for Use (Describe):
This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

αEvolve Imaging is an imaging chain intended for adults, with Artificial Intelligence Denoising (AID) designed to reduce noise in real-time fluoroscopic images and signal enhancement algorithm, Multi Frequency Processing (MFP).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) SUMMARY

This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

1. SUBMITTER'S NAME

Canon Medical Systems Corporation
1385 Shimoishigami
Otawara-Shi, Tochigi 324-8550, Japan

2. OFFICIAL CORRESPONDENT

Junichiro Araoka
Senior Manager, Quality Assurance Department

3. CONTACT PERSON, U.S. AGENT and ADDRESS

Contact Person
Jonathan Toy
Manager, Regulatory Affairs
Canon Medical Systems USA
2441 Michelle Drive, Tustin, CA 92780
Phone: (562) 709-0291
Fax: (714) 730-1310
jtoy@us.medical.canon

Official Correspondent/U.S. Agent
Orlando Tadeo, Jr.
Director, Regulatory Affairs
Canon Medical Systems USA
2441 Michelle Drive, Tustin, CA 92780
Phone: (714) 483-1551
Fax: (714) 730-1310
otadeo@us.medical.canon

4. MANUFACTURING SITE

Canon Medical Systems Corporation (CMSC)
1385 Shimoishigami
Otawara-shi, Tochigi 324-8550, Japan

5. ESTABLISHMENT REGISTRATION

9614698

6. DATE PREPARED

May 27, 2025

7. TRADE NAME(S)

Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging

8. COMMON NAME

Interventional Fluoroscopic X-ray System

9. CLASSIFICATION PANEL

Radiology

K251602

Page 6

10. DEVICE CLASSIFICATION

a) Classification Name: Image-Intensified Fluoroscopic X-ray System
b) Regulation Number: 21 CFR 892.1650
c) Regulation Class: Class II

11. PRODUCT CODE

OWB

12. PERFORMANCE STANDARD

This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

13. PREDICATE DEVICE

Trade NameAlphenix, INFX-8000V/B, INFX-8000V/S, V9.5
Marketed byCanon Medical Systems USA, Inc.
510(k) NumberK233107
Clearance DateAugust 30, 2024
Common NameInterventional Fluoroscopic X-ray System
Classification NameImage-Intensified Fluoroscopic X-ray System
Regulation Number21 CFR 892.1650
Regulation ClassClass II
Product CodeOWB

14. REASON FOR SUBMISSION

Modification of a cleared device

15. SUBMISSION TYPE

Traditional 510(k)

16. DEVICE DESCRIPTION

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging, is an interventional X-ray system with a floor mounted C-arm as its main configuration. An optional ceiling mounted C-arm is available to provide a bi-plane configuration where required. Additional units include a patient table, X-ray high-voltage generator and a digital radiography system. The C-arms can be configured with designated X-ray detectors and supporting hardware (e.g. X-ray tube and diagnostic X-ray beam limiting device). The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging includes αEvolve Imaging, an imaging chain intended for adults, with Artificial Intelligence Denoising (AID) designed to reduce noise in real-time fluoroscopic images and signal enhancement algorithm, Multi Frequency Processing (MFP).

Page 7

17. INDICATIONS FOR USE

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

αEvolve Imaging is an imaging chain intended for adults, with Artificial Intelligence Denoising (AID) designed to reduce noise in real-time fluoroscopic images and signal enhancement algorithm, Multi Frequency Processing (MFP).

18. SUBSTANTIAL EQUIVALENCE

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging is substantially equivalent to the Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5, which received premarket clearance under K233107, marketed by Canon Medical Systems. The intended use of the Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging is the same as that of the predicate device. A comparison of the technological characteristics between the subject and the predicate device is included below.

Predicate DeviceSubject Device
Device Name, Model NumberAlphenix, INFX-8000V/B, INFX-8000V/S, V9.5Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging
510(k) NumberK233107This submission
Software to support αEvolve ImagingNot AvailableAvailable
Stylish Design Cover XGSD-120H/B1 (/BU for upgrade)Not AvailableAvailable
Deep Learning Server XIDF-DLS801 for αEvolve ImagingNot AvailableAvailable

19. SAFETY

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC 60601-2-43, IEC60601-2-28, and IEC TR 60601-4-2. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, are met.

LIST OF APPLICABLE STANDARDS

  • IEC 60601-1:2005+A1:2012+A2:2020
  • IEC 60601-1-2:2014 + A1:2020
  • IEC 60601-1-3:2008+A1:2013+A2:2021
  • IEC 60601-1-6:2010+A1:2013+A2:2020
  • IEC 60601-2-28:2017
  • IEC 60601-2-43:2010+A1:2017+A2:2019
  • IEC 62304:2006+A1:2015
  • IEC 62366-1:2015 + A1:2020
  • IEC 81001-5-1:2021
  • ISO 17664-2:2021
  • IEC TR 60601-4-2:2016

Page 8

20. TESTING

Performance Testing – Bench

Image Quality Evaluations

Image quality assessments were performed, utilizing phantom and clinical datasets, to evaluate the image quality of the artificial intelligence denoising (AID) algorithm compared to the predicate device, super noise reduction filter (SNRF). The following image quality performance tests were conducted:

1. Change in Image Level, Noise and Structure

  • Raw, AID, and SNRF image sequences of an anthropomorphic chest phantom, acquired at various settings, were compared for noise reduction, change in image level, and preservation of structural similarity index measurement (SSIM) using a student's t-test. The test determined AID to be better at preserving mean image intensity and suggested AID to have improved denoising and image structure preservation.

2. Signal-to-Variance Ratio (SVR) and Signal-to-Noise Ratio (SNR)

  • Anthropomorphic chest phantom images processed with AID and SNRF were compared for SVRs and SNRs measurements using a student's t-test. This test determined the improved ability of AID to preserve image signal while decreasing image noise.

3. Modulation Transfer Function (MTF)

  • The MTF of images denoised by AID and SNRF were visually compared at three key points using a student's t-test. The results of this test showed improved performance for low—to mid-frequencies in all test cases, and the high-frequency region of the MTF curve was found to be similar for AID and SNRF in the majority of cases. The test suggests AID has an improved ability to transfer contrast at various resolutions.

4. Robustness to Detector Defects

  • A comparison of standard images (reference) and images with detector defects, processed with AID, was visually compared for bad pixels and the presence of readout lines that were sufficiently obvious to inform a clinician to call a service technician. The test determined the detector defects were sufficiently obvious to inform the clinician that the detector needed service. Additionally, the image quality outside the area of the detector defect remained visually unaffected, facilitating sufficient image quality to finish the procedure.

Page 9

5. Normalizes Noise Power Spectrum (NNPS)

  • Flat field fluoroscopic images, denoised using AID and SNRF, were compared for the magnitude and shape of the NNPS. The results of this test found that AID had a smaller noise magnitude in the frequency range of ~0.1 cycles/mm to 1.4 cycles/mm. While AID had a larger noise magnitude than SNRF above 1.4 cycles/mm, both algorithms had very small noise magnitudes at these frequencies and the difference was considered negligible. Additionally, AID had increased noise magnitudes at frequencies below ~0.1 cycles/mm and a different noise texture. This tradeoff may be attributed to these frequencies improved spatial resolution (i.e., MTF), which was evaluated using a reader study.

6. Image Lag Measurement

  • The measured mean image intensity of RAW, AID, and SRNF images was compared to the mean background image intensity using a student's t-test to determine whether there was a significant difference. This test determined that AID performed better in reducing image lag.

7. Contrast-to-Noise Ratio of a Low Contrast Object

  • Fluoroscopic image sequences of the low-contrast phantom were acquired using AID and SNRF processing, and the CNR of each low-contrast element in the phantom was compared using a student's t-test. The results showed AID had a significantly higher CNR than images processed with SNRF for all elements and test cases before and after post-processing with the remainder of the imaging chain. This test supports that AID significantly improves the CNR of low-contrast objects.

8. Contrast-to-Noise Ratio of a High Contrast Object

  • Fluoroscopic image sequences of the anthropomorphic chest phantom were acquired using AID and SNRF processing and the CNR of the guide wire tip and vessels in the phantom was compared using a student's t-test. The results showed AID had a significantly higher vessel and guidewire CNR than images processed with SNRF for all test cases before and after post-processing with the remainder of the imaging chain. This test supports that AID significantly improves the CNR of high-contrast objects.

Additionally, testing with a diverse clinical data set was performed to demonstrate the generalizability of the algorithm. It was concluded that the subject algorithm demonstrated equivalent or improved performance, compared to the predicate device, as demonstrated by the results of the above testing.

Page 10

Performance Testing – Clinical Images

Patient image sequences from the AID and SNRF imaging chains were acquired from Memorial Hermann Hospital (Houston, Texas), Waikato Hospital (Hamilton, New Zealand), and Saiseikai Kumamoto Hospital (Kumamoto, Japan). The patient image sequences were then split into four body mass index (BMI) subgroups for side-by-side comparison and reviewed by United States board-certified interventional cardiologists. This test was considered successful if the Wilcoxon signed rank test found superior performance of the AID imaging chain compared to SNRF. For the overall preference, the test was successful if the Binomial test found that the image sequences denoised by AID were chosen significantly more than 50% of the time. The reader study determined that the mean score of the AID imaging chain images was significantly higher than that of the SNRF imaging chain with regard to sharpness, contrast, confidence, noise, and the absence of image artifacts.

Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems.

Software Documentation for a Basic Documentation Level, per the FDA guidance document, "Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, was determined appropriate. This documentation includes justification for the Basic Documentation Level determination as well as testing which demonstrates that the verification and validation requirements have been met.

Cybersecurity documentation followed FDA cybersecurity premarket guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" issued on September 27, 2023.

Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level.

21. CONCLUSION

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6 with αEvolve Imaging, performs in a manner similar to and is intended for the same use as the predicate device, as indicated in the product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management, and design controls, it is concluded that the subject device has demonstrated substantial equivalence to the predicate device and is as safe and effective for its intended use.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.