K Number
K251555

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-11-06

(169 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pace Surgical Ultra™ Compression Screw System is intended for the internal fixation and stabilization of arthrodeses, osteotomies, fractures, and nonunions of the foot and ankle. The System is specifically indicated for:

Fractures and Osteotomies

  • Fractures of the tarsals, metatarsals and other fractures of the foot (i.e. LisFranc)
  • Avulsion fractures and fractures of the 5th metatarsal (i.e. Jones Fracture)
  • Talar fractures
  • Ankle fractures
  • Navicular fractures
  • Fractures of the fibula, malleolus, and calcaneus
  • Metatarsal and phalangeal osteotomies
  • Weil osteotomy
  • Calcaneal osteotomy

Fusion resulting from neuropathic osteoarthopathy (Charcot), example:

  • Medial and lateral column
  • Subtalar, talonavicular, and calcaneocuboid

Arthrodesis and Deformity Correction

  • 1st MTP arthrodesis
  • Metatarsal deformity correction
  • Tarsometatarsal joint arthrodesis
  • Naviculocuneiform joint arthrodesis
  • Talonavicular arthrodesis
  • Subtalar joint arthrodesis
  • Triple arthrodesis
  • Medial column arthrodesis
  • Subtalar joint distraction arthrodesis
  • Ankle arthrodesis
  • Lateralizing calcaneal osteotomy
  • Lateral column lengthening
  • Hammertoe

Hallux Valgus Correction

  • Fixation of osteotomies (i.e. Akin, Scarf, Chevron)
  • Interphalangeal (IP) arthrodesis
  • Proximal, midshaft, or distal osteotomy
  • Lapidus arthrodesis
Device Description

The Ultra Compression Screw System is intended for the internal fixation and stabilization of arthrodeses, osteotomies, fractures, and nonunions of the foot and ankle. The system consists of various solid and cannulated screws in diameters ranging from 2-8mm with partial and fully threaded variants. Screws are available in headed and headless variants. Compatible washers are also included in the system. All screws and washers are manufactured from titanium alloy per ASTM F136.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Ultra™ Compression Screw System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

November 6, 2025

Pace Surgical
℅ Danielle Besal
Regulatory Consultant
MRC Global
9085 E. Mineral Circle, Suite 110
Centennial, Colorado 80112

Re: K251555
Trade/Device Name: Ultra™ Compression Screw System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC, HTN
Dated: October 10, 2025
Received: October 10, 2025

Dear Danielle Besal:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251555 - Danielle Besal Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251555 - Danielle Besal Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K251555

Please provide the device trade name(s).

Ultra™ Compression Screw System

Please provide your Indications for Use below.

The Pace Surgical Ultra™ Compression Screw System is intended for the internal fixation and stabilization of arthrodeses, osteotomies, fractures, and nonunions of the foot and ankle. The System is specifically indicated for:

Fractures and Osteotomies

  • Fractures of the tarsals, metatarsals and other fractures of the foot (i.e. LisFranc)
  • Avulsion fractures and fractures of the 5th metatarsal (i.e. Jones Fracture)
  • Talar fractures
  • Ankle fractures
  • Navicular fractures
  • Fractures of the fibula, malleolus, and calcaneus
  • Metatarsal and phalangeal osteotomies
  • Weil osteotomy
  • Calcaneal osteotomy

Fusion resulting from neuropathic osteoarthopathy (Charcot), example:

  • Medial and lateral column
  • Subtalar, talonavicular, and calcaneocuboid

Arthrodesis and Deformity Correction

  • 1st MTP arthrodesis
  • Metatarsal deformity correction
  • Tarsometatarsal joint arthrodesis
  • Naviculocuneiform joint arthrodesis
  • Talonavicular arthrodesis
  • Subtalar joint arthrodesis
  • Triple arthrodesis
  • Medial column arthrodesis
  • Subtalar joint distraction arthrodesis
  • Ankle arthrodesis
  • Lateralizing calcaneal osteotomy
  • Lateral column lengthening
  • Hammertoe

Hallux Valgus Correction

  • Fixation of osteotomies (i.e. Akin, Scarf, Chevron)
  • Interphalangeal (IP) arthrodesis
  • Proximal, midshaft, or distal osteotomy
  • Lapidus arthrodesis

Ultra Compression Screw System Page 8 of 43

Page 5

Please select the types of uses (select one or both, as applicable).

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Ultra Compression Screw System Page 9 of 43

Page 6

510(k) Summary

K251555
510(k)#: K251555
Prepared on: 2025-10-10

Contact Details

21 CFR 807.92(a)(1)

FieldValue
Applicant NamePace Surgical
Applicant Address7 Great Valley Parkway, Suite 295 Malvern PA 19355 United States
Applicant Contact Telephone610-952-0462
Applicant ContactBill Rhoda
Applicant Contact Emailrhoda@runwayhc.com
Correspondent NameMRC Global
Correspondent Address9085 E. Mineral Circle, Suite 110 Centennial CO 80112 United States
Correspondent Contact Telephone9018278670
Correspondent ContactDanielle Besal
Correspondent Contact Emaildanielle.besal@askmrcglobal.com

Device Name

21 CFR 807.92(a)(2)

FieldValue
Device Trade NameUltra Compression Screw System
Common NameSmooth or threaded metallic bone fixation fastener
Classification NameScrew, Fixation, Bone
Regulation Number888.3040
Product Code(s)HWC, HTN

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K203011Monster Screw SystemHWC

Device Description Summary

21 CFR 807.92(a)(4)

The Ultra Compression Screw System is intended for the internal fixation and stabilization of arthrodeses, osteotomies, fractures, and nonunions of the foot and ankle. The system consists of various solid and cannulated screws in diameters ranging from 2-8mm with partial and fully threaded variants. Screws are available in headed and headless variants. Compatible washers are also included in the system. All screws and washers are manufactured from titanium alloy per ASTM F136.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Pace Surgical Ultra™ Compression Screw System is intended for the internal fixation and stabilization of arthrodeses, osteotomies, fractures, and nonunions of the foot and ankle. The System is specifically indicated for:

Page 1 of 2

Page 7

K251555 Page 2 of 2

Fractures and Osteotomies

  • Fractures of the tarsals, metatarsals and other fractures of the foot (i.e. LisFranc)
  • Avulsion fractures and fractures of the 5th metatarsal (i.e. Jones Fracture)
  • Talar fractures
  • Ankle fractures
  • Navicular fractures
  • Fractures of the fibula, malleolus, and calcaneus
  • Metatarsal and phalangeal osteotomies
  • Weil osteotomy
  • Calcaneal osteotomy

Fusion resulting from neuropathic osteoarthopathy (Charcot), example:

  • Medial and lateral column
  • Subtalar, talonavicular, and calcaneocuboid

Arthrodesis and Deformity Correction

  • 1st MTP arthrodesis
  • Metatarsal deformity correction
  • Tarsometatarsal joint arthrodesis
  • Naviculocuneiform joint arthrodesis
  • Talonavicular arthrodesis
  • Subtalar joint arthrodesis
  • Triple arthrodesis
  • Medial column arthrodesis
  • Subtalar joint distraction arthrodesis
  • Ankle arthrodesis
  • Lateralizing calcaneal osteotomy
  • Lateral column lengthening
  • Hammertoe

Hallux Valgus Correction

  • Fixation of osteotomies (i.e. Akin, Scarf, Chevron)
  • Interphalangeal (IP) arthrodesis
  • Proximal, midshaft, or distal osteotomy
  • Lapidus arthrodesis

Indications for Use Comparison

21 CFR 807.92(a)(5)

The subject and predicate devices have the same intended use of bone fixation and stabilization. The minor differences in the indications for use statement does not impact its intended use. In addition, the specific indication examples are identical to the predicate.

Technological Comparison

21 CFR 807.92(a)(6)

The subject device is similar in technological characteristics to the predicate device. The materials are similar to the predicate and the geometry and sizing are within the predicate device offerings.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Mechanical testing per ASTM F543 was conducted on the subject devices.

Clinical data were not applicable.

The subject devices met the performance criteria specified in FDA Guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers - Performance Criteria for Safety and Performance Based Pathway;" therefore, can be considered substantially equivalent to the predicate device.

Page 2 of 2

N/A