K Number
K251527

Validate with FDA (Live)

Device Name
Brain WMH
Manufacturer
Date Cleared
2025-09-25

(129 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Brain WMH is intended for automatic labeling, visualization, and volumetric quantification of WMH from T2w FLAIR MR images. The output consists of segmentations, visualizations and volumetric measurements of WMH. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting.

Device Description

Brain WMH is a software as a medical device (SaMD) that provides automatic quantification of white matter hyperintensities (WMHs) based on magnetic resonance (MR) images to assist trained medical professionals. The device takes fluid-attenuated inversion recovery (FLAIR) MR images as input. Its output consists of a report in DICOM encapsulated pdf and DICOM secondary capture format, and DICOM secondary captures of the segmentations as color overlay on the input image.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Quantib Brain WMH device, based on the provided FDA 510(k) clearance letter:


Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific Metric/DescriptionAcceptance CriteriaReported Device Performance
WMH SegmentationDice coefficient (standalone)Within the range of interobserver variability0.58 ± 0.21
Anatomical Location LabelingAccuracy (standalone)Within the range of interobserver performanceWithin the range of interobserver performance
Longitudinal ValidationCorrectly labeled WMH across scans from same patientNot explicitly stated, but implies high accuracy97.1% correctly labeled
Scan-Rescan ReproducibilityConsistency of WMH volumes between short-interval, same-subject study pairsConsistency/high agreementShowed consistent volumes

Note: The document indicates that the device's standalone segmentation performance (Dice coefficient) was higher than the predicate device's standalone performance, further suggesting it met or exceeded expectations.


Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: 110 studies from 90 patients.
    • Data Provenance: Multiple clinical sites, with the majority of data acquired in the United States. The data was retrospective, collected from ethnically diverse male and female adult patients.
    • Scan-Rescan Reproducibility Test Set: A separate, unspecified number of short-interval, same-subject scan pairs.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Ground Truth Experts: Three experts.
    • Adjudicator: One expert.
    • Qualifications: The document states "Three experts served as truthers and one expert as an adjudicator," but does not specify their qualifications (e.g., "radiologist with 10 years of experience").
  3. Adjudication Method for the Test Set:

    • The document states: "The ground truth process included multiple experts. Three experts served as truthers and one expert as an adjudicator." This implies a 3+1 adjudication method, where three experts initially define the ground truth, and a fourth expert adjudicates any discrepancies or provides a final decision.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • An MRMC study was not explicitly mentioned for evaluating human reader improvement with AI assistance. The performance testing section primarily focuses on standalone performance and comparison to the predicate device's standalone performance.
    • No effect size for human reader improvement with AI vs. without AI assistance is reported.
  5. Standalone Performance (Algorithm-Only) Study:

    • Yes, a standalone performance study was done. The document explicitly states: "The standalone performance of Brain WMH segmentation, as measured by Dice coefficient (0.58 ± 0.21) was higher than the standalone performance of the predicate device and fell within the range of interobserver variability."
    • Further, "the anatomical location labeling performance was also within the range of the interobserver performance."
  6. Type of Ground Truth Used:

    • Expert Consensus/Labeling: The ground truth for the test set was established through "multiple experts," specifically "Three experts served as truthers and one expert as an adjudicator." This indicates an expert consensus approach to annotation.
  7. Sample Size for the Training Set:

    • Training Set Sample Size: 474 T2-weighted FLAIR scans.
  8. How the Ground Truth for the Training Set was Established:

    • The document states that the WMH segmentation model was "trained with 474 T2-weighted FLAIR scans." However, it does not explicitly describe the method used to establish the ground truth for this training set. It only mentions the data was "collected from different geographical areas, with the majority of the data acquired in the United States."

FDA 510(k) Clearance Letter - Brain WMH

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 25, 2025

Quantib BV
Nathan Hunt
Head of Quality, Regulatory, and Compliance
Westblaak 130
Rotterdam, Zuid-Holland 3012 KM
Netherlands

Re: K251527
Trade/Device Name: Brain WMH
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: August 19, 2025
Received: August 19, 2025

Dear Nathan Hunt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"


U.S. FOOD & DRUG ADMINISTRATION

September 25, 2025

Quantib BV
Nathan Hunt
Head of Quality, Regulatory, and Compliance
Westblaak 130
Rotterdam, Zuid-Holland 3012 KM
Netherlands

Re: K251527
Trade/Device Name: Brain WMH
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: August 19, 2025
Received: August 19, 2025

Dear Nathan Hunt:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"


U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 2

K251527 - Nathan Hunt Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

Page 3

K251527 - Nathan Hunt Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak. Ph.D.
Assistant Director
Magnetic Resonance and Nuclear Medicine Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K251527

Device Name
Brain WMH

Indications for Use (Describe)

Intended Use

Brain WMH is intended for automatic labeling, visualization, and volumetric quantification of WMH from T2w FLAIR MR images. The output consists of segmentations, visualizations and volumetric measurements of WMH. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting.

Intended User Population

The intended users of Brain WMH are trained medical professionals.

Intended Patient Population

Brain WMH is intended for post-processing of brain magnetic resonance imaging (MRI) data from patients 22 years of age or older, without pathologies (e.g. tumor or stroke), and treatment induced features (e.g. from prior surgery or radiation treatment), and without contrast enhancement.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

510(k) Number (if known)
K251527

Device Name
Brain WMH

Indications for Use (Describe)

Intended Use

Brain WMH is intended for automatic labeling, visualization, and volumetric quantification of WMH from T2w FLAIR MR images. The output consists of segmentations, visualizations and volumetric measurements of WMH. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting.

Intended User Population

The intended users of Brain WMH are trained medical professionals.

Intended Patient Population

Brain WMH is intended for post-processing of brain magnetic resonance imaging (MRI) data from patients 22 years of age or older, without pathologies (e.g. tumor or stroke), and treatment induced features (e.g. from prior surgery or radiation treatment), and without contrast enhancement.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary

Quantib BV
Brain WMH

In accordance with 21 CFR 807.92 the following summary of information is provided, on this date, September 10th, 2025:

1. 510(k) Submitter

Quantib BV
Westblaak 130
3012 KM Rotterdam
Tel: +31 10 841 1749

Contact Person:
B. Nathan Hunt
Head of Quality, Regulatory, and Compliance
Email: regulatory@deephealth.com

Date prepared:
September 10th, 2025

2. Device

Trade Name of Device: Brain WMH
Common or Usual Name: Medical Image Software
Classification Names: Medical image management and processing system (892.2050)
Regulation Class: II
Product Code: QIH

3. Predicate Device

Trade Name of Device: Quantib ND (K213737)
Common or Usual Name: Medical Image Software
Classification Names: Medical image management and processing system (892.2050)
Regulation Class: II
Product Code: LLZ

K251527


Quantib

K251527

510(k) Summary

Quantib BV
Brain WMH

In accordance with 21 CFR 807.92 the following summary of information is provided, on this date, September 10th, 2025:

1. 510(k) Submitter

Quantib BV
Westblaak 130
3012 KM Rotterdam
Tel: +31 10 841 1749

Contact Person:
B. Nathan Hunt
Head of Quality, Regulatory, and Compliance
Email: regulatory@deephealth.com

Date prepared:
September 10th, 2025

2. Device

Trade Name of Device: Brain WMH
Common or Usual Name: Medical Image Software
Classification Names: Medical image management and processing system (892.2050)
Regulation Class: II
Product Code: QIH

3. Predicate Device

Trade Name of Device: Quantib ND (K213737)
Common or Usual Name: Medical Image Software
Classification Names: Medical image management and processing system (892.2050)
Regulation Class: II
Product Code: LLZ

Page 6

Quantib

4. Device Description

Brain WMH is a software as a medical device (SaMD) that provides automatic quantification of white matter hyperintensities (WMHs) based on magnetic resonance (MR) images to assist trained medical professionals. The device takes fluid-attenuated inversion recovery (FLAIR) MR images as input. Its output consists of a report in DICOM encapsulated pdf and DICOM secondary capture format, and DICOM secondary captures of the segmentations as color overlay on the input image.

5. Indications for Use

Brain WMH is intended for automatic labeling, visualization, and volumetric quantification of WMH from T2w FLAIR MR images. The output consists of segmentations, visualizations and volumetric measurements of WMH. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting.

Intended User Population

The intended users of Brain WMH are trained medical professionals.

Intended Patient Population

Brain WMH is intended for post-processing of brain magnetic resonance imaging (MRI) data from patients 22 years of age or older, without pathologies (e.g. tumor or stroke), and treatment induced features (e.g. from prior surgery or radiation treatment), and without contrast enhancement.

6. Predicate Device Comparison

The subject device and the predicate device are both Class II medical devices regulated under 21 CFR 892.2050. As main functionality, the devices provide segmentation and quantification of WMHs and support longitudinal analysis. Furthermore, the devices have similar indications for use, identical intended users, and target the same anatomical site.

There are minor technical differences between the subject device and the predicate device:

Brain WMH employs an updated artificial intelligence-based algorithm for WMH segmentation, whereas the predicate device uses a different machine learning approach. Both methodologies are functionally similar in purpose and provide segmentation capabilities. Both devices undergo validation using comparable performance metrics and evaluation processes.

Brain WMH requires only T2 FLAIR imaging as input, while the predicate device requires both FLAIR and T1 scans. This streamlined input requirement enhances clinical workflow efficiency by reducing imaging requirements while improving segmentation performance.

Brain WMH operates as a fully automatic system, whereas the predicate device includes manual adjustment functionality. This fully automatic approach streamlines clinical workflow while maintaining the same core WMH segmentation and quantification function. Both devices are intended for use by trained medical professionals who review and interpret results within established clinical workflows.

Page 7

Quantib

Brain WMH segments WMHs and labels them according to their anatomical location, while the predicate performs segmentation of WMHs without anatomical location labeling.

Brain WMH provides specialized WMH segmentation and quantification, while the predicate device offers both brain structure segmentation and WMH analysis. This focused approach allows Brain WMH to deliver a specialized WMH assessment.

Despite these technical differences, both devices maintain equivalent core functionality for WMH segmentation and volumetric quantification. The technical modifications represent improvements in workflow efficiency and algorithmic approach rather than fundamental changes in device purpose or safety profile. Both devices generate automatic segmentations and volumetric data for clinical assessment by trained medical professionals, ensuring equivalent clinical utility and safety within established medical workflows.

Clinical performance testing demonstrates that the subject device performs comparably to the predicate device.

7. Performance Data

7.1. Quality and safety

The design and development of Brain WMH followed the following FDA recognized standards and guidance documents:

  • ISO 14971:2019 – Medical Devices – Application of Risk Management to Medical Devices (#5-125)
  • IEC 62304:2015 – Medical Device Software – Software Life Cycles Processes (#13-79)
  • NEMA PS3 – Digital Imaging and Communications in Medicine (DICOM) Set (#12-300)
  • Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May, 2005)
  • Guidance for Industry and FDA Staff: Software as a Medical Devices (SaMD): Clinical Evaluation (December 2017)

Brain WMH is a software only device. Verification testing included software unit testing, software integration testing, and system testing. Testing confirmed that the software, as designed and implemented, satisfied the software requirements and has no unintentional differences from the predicate device.

Page 8

Quantib

7.2. Training dataset

The WMH segmentation model was trained with 474 T2-weighted FLAIR scans collected from different geographical areas, with the majority of the data acquired in the United States. The data includes studies from 1.5 T and 3.0 T scanners from GE, Philips, and Siemens.

7.3. Performance testing

Validation of the model and software was performed using standalone performance testing, including a comparison of the subject device performance to the predicate device performance on the same dataset. The test dataset included 110 studies from 90 patients with T2-weighted FLAIR images acquired from multiple clinical sites, with the majority of data acquired at sites in the United States. The dataset represents 1.5 T (32 studies) and 3.0 T (78 studies) scanners from GE (60 studies), Philips (39 studies) and Siemens (11 studies), and various scan settings. Scans were collected from ethnically diverse and representative male and female adult patients with age ranging from 22 to 90 years. The clinical indication for imaging was either multiple sclerosis or cognitive impairment. Scan-rescan reproducibility was conducted on a separate test set of short-interval, same-subject scan pairs. To ensure independence of the test data, the test sets were quarantined datasets not used to train or tune the model, only for software validation following internal processing aligned with good machine learning practices.

The ground truth process included multiple experts. Three experts served as truthers and one expert as an adjudicator.

Performance testing concluded that the subject device met the pre-specified performance criteria. The standalone performance of Brain WMH segmentation, as measured by Dice coefficient (0.58 ± 0.21) was higher than the standalone performance of the predicate device and fell within the range of interobserver variability. Similarly, the anatomical location labeling performance was also within the range of the interobserver performance. The longitudinal validation demonstrated high accuracy in WMH labeling across scans acquired from the same patient (97.1% correctly labelled). Additionally, scan-rescan reproducibility showed consistent volumes between short-interval, same-subject study pairs.

8. Conclusion

The verification and validation testing conducted to support this submission confirms that the subject device performs comparably to the predicate device. The minor differences between the subject and predicate device do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. Therefore, the information provided in this submission demonstrates that Brain WMH is substantially equivalent to the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).