(117 days)
Phoenix Sinus Tarsi Stent is an implant stabilization device used in the treatment of talotarsal joint instability in adult and pediatric patients four years of age and older. The stent is designed to stabilize the talus to prevent excessive anterior, and/or medial, and/or plantarflexion of the talus, while allowing normal talotarsal joint motion.
The Phoenix Sinus Tarsi Stent is a non-pyrogenic type IIB extra-osseous talotarsal stabilization (EOTTS) device used in the treatment of talotarsal joint instability per Graham Et al. EOTTS classification system. The stent system consists of an implant designed to be inserted into the sinus tarsi and has corresponding instrumentation to facilitate the insertion. All stents are manufactured from Ti6Al-4V ELI per ASTM F136 and have 5 sizes with varying diameter.
The Phoenix Sinus Tarsi Stent System is a medical device, and the provided FDA 510(k) clearance letter focuses on its substantial equivalence to predicate devices based on non-clinical performance testing. This type of submission does not typically involve the kind of AI/ML performance evaluation criteria and studies that would address human-in-the-loop performance, multi-reader multi-case studies, or detailed ground truth establishment as requested.
The provided document describes non-clinical performance testing to demonstrate the device's physical and biological properties. It does not involve AI or algorithms, nor does it refer to human interpretation of medical images or data. Therefore, many of the requested categories (such as human readers, AI assistance, ground truth for AI, etc.) are not applicable to this specific device and its regulatory submission.
However, I can extract the acceptance criteria and study information that is present in the document regarding the device's physical and material performance.
Acceptance Criteria and Study for the Phoenix Sinus Tarsi Stent System (Non-Clinical Performance)
The Phoenix Sinus Tarsi Stent System's acceptance criteria and performance are established through a series of non-clinical tests designed to demonstrate its safety and functionality as a physical implant. These tests do not involve AI/ML components or human interpretation studies.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria Met (Yes/No) | Reported Device Performance |
|---|---|---|
| Mechanical Performance | ||
| Screw Pullout | Yes | Performed as intended per ASTM F543 |
| Cantilever Bending | Yes | Performed as intended per ASTM 2193 |
| Biocompatibility | ||
| Endotoxin Testing | Yes | Met predetermined criteria per AAMI ST72 and USP <85> |
| Cytotoxicity Testing | Yes | Met predetermined criteria per ISO 10993-5 |
| Biocompatibility Risk Assessment | Yes | Concluded positively (details not specified, but implied met criteria) |
| Sterilization & Packaging | ||
| Sterilization Testing | Yes | Met predetermined criteria per ISO 11137-1, ISO 11137-2 |
| Packaging Shelf-Life Performance Testing | Yes | Met predetermined criteria per ISO 11607-1, ASTM F88/F88M, ASTM F2096, ASTM F1886/1886M |
Note: The document explicitly states: "All testing showed the subject device performed as intended. All testing met applicable predetermined acceptance criteria."
2. Sample Size for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated for each test (e.g., number of stents tested for pullout, bending, etc.). For physical and material tests, sample sizes are typically determined by relevant ISO/ASTM standards.
- Data Provenance: The tests are non-clinical, implying laboratory-based testing of the device itself (e.g., physical specimens of the stent and its materials), not patient data. Therefore, concepts like "country of origin of the data" or "retrospective/prospective" studies are not applicable in this context.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not Applicable. For non-clinical performance testing of a physical implant, "ground truth" is established by adherence to recognized international standards (e.g., ASTM, ISO, AAMI, USP) and predefined pass/fail criteria for material and mechanical properties. There are no human "experts" establishing ground truth in the sense of medical diagnosis or interpretation for this type of testing.
4. Adjudication Method for the Test Set:
- Not Applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation of data where consensus on ground truth is required. For the non-clinical tests described, outcomes are typically objectively measured against predefined criteria specified in the test standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. An MRMC study is designed to evaluate the performance of diagnostic devices, especially those involving human interpretation (e.g., radiologists reading images), often with and without AI assistance. The Phoenix Sinus Tarsi Stent System is a physical implant, not a diagnostic device involving human interpretation; thus, MRMC studies are not applicable.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This device does not incorporate any AI or algorithm component. Its performance is purely based on its physical and material properties.
7. The Type of Ground Truth Used:
- Scientific Standards and Predetermined Criteria: For mechanical tests (e.g., screw pullout, cantilever bending), ground truth is defined by the requirements and methodologies outlined in the applicable ASTM standards (e.g., ASTM F543, ASTM 2193), with specific performance thresholds (e.g., minimum pullout strength, maximum deflection).
- Biological/Material Standards: For biocompatibility (Endotoxin, Cytotoxicity) and sterilization, ground truth is based on meeting the specifications and acceptable limits defined by international standards (e.g., AAMI ST72, USP <85>, ISO 10993-5, ISO 11137-1/2).
- Packaging Integrity: For shelf-life, ground truth is meeting parameters defined by ISO 11607-1 and relevant ASTM standards regarding package integrity.
8. The Sample Size for the Training Set:
- Not Applicable. This device does not use AI/ML, therefore, there is no "training set."
9. How the Ground Truth for the Training Set was Established:
- Not Applicable. There is no training set for this device.
FDA 510(k) Clearance Letter - Phoenix Sinus Tarsi Stent System
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.00
August 27, 2025
Astra OrthoMed, Inc.
℅ Justin Gracyalny
Regulatory Affairs Program Manager
Secure BioMed Evaluations
7828 Hickory Flat Hwy
Woodstock, Georgia 30188
Re: K251382
Trade/Device Name: Phoenix Sinus Tarsi Stent System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC
Dated: July 28, 2025
Received: July 28, 2025
Dear Justin Gracyalny:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
August 27, 2025
Astra OrthoMed, Inc.
℅ Justin Gracyalny
Regulatory Affairs Program Manager
Secure BioMed Evaluations
7828 Hickory Flat Hwy
Woodstock, Georgia 30188
Re: K251382
Trade/Device Name: Phoenix Sinus Tarsi Stent System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC
Dated: July 28, 2025
Received: July 28, 2025
Dear Justin Gracyalny:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Page 2
K251382 - Justin Gracyalny Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K251382 - Justin Gracyalny Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K251382
Please provide the device trade name(s).
Phoenix Sinus Tarsi Stent System
Please provide your Indications for Use below.
Phoenix Sinus Tarsi Stent is an implant stabilization device used in the treatment of talotarsal joint instability in adult and pediatric patients four years of age and older. The stent is designed to stabilize the talus to prevent excessive anterior, and/or medial, and/or plantarflexion of the talus, while allowing normal talotarsal joint motion.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Phoenix Sinus Tarsi Stent System Page 10 of 49
Page 5
K251382 510(k) SUMMARY: Phoenix Sinus Tarsi Stent System
Date Prepared: August 21, 2025
Sponsor:
Astra OrthoMed
330 Franklin Road Suite 135A-213
Brentwood, TN 37027
586-604-9462
510(k) Contact:
Secure BioMed Evaluations
Justin Gracyalny, MSE
7828 Hickory Flat Highway, Suite 120
Woodstock, GA 30188
770-837-2681
Regulatory@SecureBME.com
Trade Name: Phoenix Sinus Tarsi Stent System
Common Name: Sinus Tarsi Implant
Code –Classification: HWC - Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040)
Primary Predicate: K142534 GraMedica LLC. HyProCure II
Reference Device: K042030 GraMedica LLC. HyProCure Subtalar Implant System
K111834 Trilliant Surgical LTD Disco Subtalar Implant
Device Description: The Phoenix Sinus Tarsi Stent is a non-pyrogenic type IIB extra-osseous talotarsal stabilization (EOTTS) device used in the treatment of talotarsal joint instability per Graham Et al. EOTTS classification system. The stent system consists of an implant designed to be inserted into the sinus tarsi and has corresponding instrumentation to facilitate the insertion. All stents are manufactured from Ti6Al-4V ELI per ASTM F136 and have 5 sizes with varying diameter.
Indications for Use Statement: Phoenix Sinus Tarsi Stent is an implant stabilization device used in the treatment of talotarsal joint instability. The stent is designed to stabilize the talus to prevent excessive anterior, and/or medial, and/or plantarflexion of the talus, while allowing normal talotarsal joint motion.
Comparison of Technological Characteristics
The Phoenix Sinus Tarsi Stent System is substantially equivalent to the legally marketed predicate devices in terms of intended use, design, materials of construction, mechanical safety and performance. The subject device indications for use include minor differences for clarification. Any differences in terminology do not create a change in the device and are simply clarification of terminology as compared to predicate device.
There are no significant technological differences between the subject and predicate device. The subject device uses the same material as the predicate and reference devices, offers sizes within the ranges offered by the predicate and reference devices, and all devices are manufactured using subtractive techniques. Any minor geometric differences are supported by performance testing.
Page 1 of 2
Page 6
Non-Clinical Performance Testing Summary
Performance testing for Phoenix Sinus Tarsi Stent System include:
- Screw Pullout Testing per ASTM F543
- Cantilever Bending Testing per ASTM 2193
- Endotoxin Testing per AAMI ST72 and USP <85>
- Sterilization Testing per ISO 11137-1, ISO 11137-2
- Cytotoxicity Testing per ISO 10993-5
- Biocompatibility Risk Assessment
- Packaging Shelf-Life Performance Testing per ISO 11607-1, ASTM F88/F88M, ASTM F2096, ASTM F1886/1886M
All testing showed the subject device performed as intended. All testing met applicable predetermined acceptance criteria. All testing supports the subject device is substantially equivalent to the predicate device.
Conclusions
Based on the similarities of the intended use/indications for use, technological and functional characteristic, and the results of the non-clinical performance testing, the subject device is substantially equivalent to the legally marketed predicate device.
Page 2 of 2
N/A