K Number
K251378

Validate with FDA (Live)

Device Name
CoolTone
Date Cleared
2025-07-14

(73 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CoolTone is indicated to be used for:

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen.
  • Strengthening, Toning and Firming of buttocks and thighs.
Device Description

The CoolTone is a non-invasive therapeutic device. The device produces an electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the CoolTone helps to strengthen, tone and firm abdomen, buttocks and thighs.

The device housing protects the patient from electrical shock and mechanical injuries. The device is a mobile standalone equipment with four wheels.

Two large applicators are connected to the control unit and can be used simultaneously depending on the treatment. The device is a medical equipment that generates a magnetic field by applying a strong current to an applicator. The CoolTone is equipped with the securement system which is designed to maintain applicator position throughout treatment.

A large color touch screen facilitates the use of the device. The on-screen information guides the user step-by-step through the entire treatment process. The treatment is operated through parameters such as frequency, time and intensity. Three pre-set treatment options are available for users to choose from: Abdomen, buttocks and thighs.

AI/ML Overview

The provided FDA 510(k) clearance letter pertains to the CoolTone device (K251378). It states that this device is substantially equivalent to a previously cleared CoolTone device (K220601).

Crucially, the document explicitly states: "No clinical testing was required for this change." This means that for the K251378 submission, no new studies were performed to prove the device meets acceptance criteria. Instead, the clearance relies on its substantial equivalence to the predicate device, which presumably had clinical data supporting its initial clearance.

Therefore, many of the requested information points regarding acceptance criteria and study details cannot be directly extracted from this specific 510(k) clearance letter. The information below reflects what can be inferred or directly cited from the provided text about K251378, with an acknowledgement that the core "proof" resides within the documentation for the predicate device (K220601).


1. Table of Acceptance Criteria and Reported Device Performance

As stated, no new clinical study was conducted for this specific 510(k) submission (K251378). Therefore, there are no new acceptance criteria or reported device performance for this updated device. The clearance is based on the identical technological characteristics and indications for use to the predicate device (K220601). The "acceptance criteria" for K251378 are implicitly met by demonstrating substantial equivalence to the predicate device, meaning it performs as safely and effectively for the stated indications.

Acceptance Criteria (Inferred from Predicate Device's Clearance)Reported Device Performance (Inherited from Predicate)
Improvement of abdominal toneExpected to provide improved abdominal tone
Strengthening of abdominal musclesExpected to provide strengthened abdominal muscles
Development of firmer abdomenExpected to provide a firmer abdomen
Strengthening of buttocks and thighsExpected to provide strengthened buttocks and thighs
Toning of buttocks and thighsExpected to provide toned buttocks and thighs
Firming of buttocks and thighsExpected to provide firmer buttocks and thighs

2. Sample size used for the test set and the data provenance

Not applicable for K251378 as no new clinical testing was performed for this submission. The substantial equivalence argument relies on prior data for the predicate device (K220601).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable for K251378 as no new clinical testing was performed for this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for K251378 as no new clinical testing was performed for this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable for K251378. The CoolTone device is a powered muscle stimulator, not an AI-assisted diagnostic or imaging device that would typically involve human readers or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable for K251378. The CoolTone is a physical device for muscle stimulation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable for K251378 as no new clinical testing was performed. For the predicate device (K220601), clinical trials would likely have used objective measurements (e.g., imaging to assess muscle thickness, physical measurements for circumference changes, or patient/physician reported outcomes on tone/firmness) to establish the ground truth for muscle strengthening, toning, and firming.

8. The sample size for the training set

Not applicable for K251378 as no new clinical testing or algorithm training was performed for this submission.

9. How the ground truth for the training set was established

Not applicable for K251378 as no new clinical testing or algorithm training was performed for this submission.

FDA 510(k) Clearance Letter - CoolTone Device

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

July 14, 2025

Zimmer MedizinSysteme GmbH
℅ Scott Blood
Principal Consultant
QARA Consulting
151 Gleasondale Road
Stow, Massachusetts 01775

Re: K251378
Trade/Device Name: CoolTone
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: NGX
Dated: June 16, 2025
Received: June 16, 2025

Dear Scott Blood:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K251378 - Scott Blood

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K251378 - Scott Blood

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K251378

Device Name
CoolTone

Indications for Use (Describe)

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen.
  • Strengthening, toning and firming of buttocks and thighs.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K251378 510(K) SUMMARY

SUBMITTER: Zimmer MedizinSysteme GmbH
Junkersstrasse 9
89231 Neu-Ulm, Germany
Establishment Registration: 8010720

CONTACT: Mrs. Ute Killet
Manager, Regulatory Affairs
Phone: +49-731-9761-216
Fax: +49-731-9761-118
E-mail: u.killet@zimmer.de

DATE PREPARED: May 30, 2025

TRADE NAME: CoolTone

COMMON NAME: Powered Muscle Stimulator

CLASSIFICATION NAME: Stimulator, Muscle, Powered, For Muscle Conditioning

DEVICE CLASSIFICATION: Class II, 21 CFR §890.5850

PRODUCT CODE: NGX

PREDICATE DEVICE: CoolTone (K220601)

DEVICE DESCRIPTION:

The CoolTone is a non-invasive therapeutic device. The device produces an electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the CoolTone helps to strengthen, tone and firm abdomen, buttocks and thighs.

The device housing protects the patient from electrical shock and mechanical injuries. The device is a mobile standalone equipment with four wheels.

Two large applicators are connected to the control unit and can be used simultaneously depending on the treatment. The device is a medical equipment that generates a magnetic field by applying a strong current to an applicator. The CoolTone is equipped with the

Page 6

securement system which is designed to maintain applicator position throughout treatment.

A large color touch screen facilitates the use of the device. The on-screen information guides the user step-by-step through the entire treatment process. The treatment is operated through parameters such as frequency, time and intensity. Three pre-set treatment options are available for users to choose from: Abdomen, buttocks and thighs.

INDICATIONS FOR USE:

CoolTone is indicated to be used for:

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen.
  • Strengthening, Toning and Firming of buttocks and thighs.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The technological characteristics of the subject device and the predicate device are identical.

Table 1: Technological Similarities between Proposed and Predicate Device

Technological CharacteristicsSUBJECT DEVICE Zimmer MedizinSyteme GmbH CoolTone (K251378)PREDICATE DEVICE Zimmer MedizinSyteme GmbH CoolTone (K220601)
Product Code and RegulationPhysical Medicine 21 CFR 890.5850 NGX – Stimulator Muscle, Powered, Muscle ConditioningPhysical Medicine 21 CFR 890.5850 NGX – Stimulator Muscle, Powered, Muscle Conditioning
Indications for UseThe CoolTone is indicated to be used for: • Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen. • Strengthening, Toning and Firming of buttocks and thighs.The CoolTone is indicated to be used for: • Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen. • Strengthening, Toning and Firming of buttocks and thighs.
Primary FunctionMuscle stimulationMuscle stimulation

Page 7

Technological CharacteristicsSUBJECT DEVICE Zimmer MedizinSyteme GmbH CoolTone (K251378)PREDICATE DEVICE Zimmer MedizinSyteme GmbH CoolTone (K220601)
Principle of ActionInitiating action potential of nerves results in muscle contractionInitiating action potential of nerves results in muscle contraction
Electrical ProtectionClass I, BFClass I, BF
User InterfaceTouch screenTouch screen
Touch screen size12"12"
Positioning of ApplicatorSecurement systemSecurement system
Firmware ControlledYesYes
Type of EnergyMagnetic fieldMagnetic field
Magnetic Field IntensityLarge applicator 0.5 – 1.35T +/- 20%Large applicator 0.5 – 1.35T +/- 20%
Number of outputs22
Number of Magnetic Coils in the Applicator11
Number of applicatorsUp to two large applicators can be operational at same timeUp to two large applicators can be operational at same time
Applicator connectionDetachable from the control unitDetachable from the control unit
Total Induced Current in Tissue (mA)327327
Type of OperationContinuousContinuous
Pulse Repetition Rate1 – 150 Hz1 – 150 Hz
Pulse DurationLarge applicator 370 µs +/- 20%Large applicator 370 µs +/- 20%
Pulse Amplitude0 – 100 %0 – 100 %
Selection of parameters (Intensity, Time)YesYes
Treatment TimeUp to 30 minUp to 30 min
Shape of Stimulation PulseSymmetrical Biphasic Sine WaveSymmetrical Biphasic Sine Wave
Energy Source100 - 240 VAC, 50-60 Hz100 – 240 VAC, 50–60 Hz
System Dimensions (WxHxD)600 x 1100x 600 mm600 x 1100x 600 mm
Operating Ambient Temperature10°C to 28 °C10°C to 28 °C
Environmental SpecificationsFor indoor use onlyFor indoor use only

Page 8

Table 2: Technological Similarities between Proposed and Predicate Device per FDA Guidance for Industry for Powered Muscle Stimulators for 510(k)s (June 9, 1999)

Sections 2,3 Technological CharacteristicsSUBJECT DEVICE Zimmer MedizinSyteme GmbH CoolTone (K251378)PREDICATE DEVICE Zimmer MedizinSyteme GmbH CoolTone (K220601)Comments
4. Power Source(s) - Method of Line Current Isolation - Patient Leakage Current Normal condition Single fault condition60601 compliant < 100 µA 3.9 µA60601 compliant < 100 µA 3.9 µA60601 Compliant
5. Number of Output Modes11
6. Number of Output Channels - Synchronous or Alternating? - Method of Channel Isolation2 Synchronous N/A2 Synchronous N/ANo electrodes - applicators are not connected to patient
7. Regulated Current or Regulated Voltage?VoltageVoltageControlled voltage to the coil
8. Software/Firmware/Microprocessor Control?YesYes
9. Automatic Overload Trip?N/AN/ANo electrodes - applicators are not connected to patient
10. Automatic No-Load Trip?N/AN/ANo electrodes - applicators are not connected to patient
11. Automatic Shut Off?YesYesUnit shuts off with specified timer

Page 9

Sections 2,3 Technological CharacteristicsSUBJECT DEVICE Zimmer MedizinSyteme GmbH CoolTone (K251378)PREDICATE DEVICE Zimmer MedizinSyteme GmbH CoolTone (K220601)Comments
12. Patient Override Control?NoNoTreatment is delivered by health care provider
13. Indicator Display: - On/Off Status? - Low Battery? - Voltage/Current Level?Yes N/A NoYes N/A No
14. Timer Range (minutes)Up to 30 MinUp to 30 Min
15. Compliance with Voluntary Standards?YesYes
16. Compliance* with 21 CFR 898? (*Becomes mandatory beginning May 9, 2000)YesYes
17. Weight80 Kg80 Kg
19. Housing Materials and ConstructionSteel and Injection Molded PlasticsSteel and Injection Molded Plastics
Waveform (e.g., pulsed monophasic, biphasic)Symmetrical Biphasic Sine WaveSymmetrical Biphasic Sine Wave
Shape (e.g., rectangular, spike, rectified sinusoidal)SinusoidalSinusoidal
Maximum Output Voltage (specify units)N/AN/ANo electrodes - applicators are not connected to patient
Maximum Output Current (specify units)N/AN/ANo electrodes - applicators are not connected to patient
Pulse Width (specify units)370 µs +/- 20%250-400 µs +/- 20%
Frequency (Hz)1-150 Hz1-150 Hz

Page 10

Sections 2,3 Technological CharacteristicsSUBJECT DEVICE Zimmer MedizinSyteme GmbH CoolTone (K251378)PREDICATE DEVICE Zimmer MedizinSyteme GmbH CoolTone (K220601)Comments
For interferential modes only: - Beat Frequency (Hz)N/AN/A
For multiphasic waveforms only: - Symmetrical phases?YesYesBiphasic
Phase Duration (include units) (state range, if applicable) (both phases, if asymmetrical)370 µs +/- 20%370 µs +/- 20%
Net Charge (mC per pulse) (If zero, state method of achieving zero net charge.)N/AN/ANo electrodes – applicators not connected to the patient
Maximum Phase Charge, (mC)N/AN/ANo electrodes – applicators not connected to the patient
Maximum Current Density, (mA/cm²)N/AN/ANo electrodes – applicators not connected to the patient – see SAR report
Maximum Power Density, (W/cm²)(using smallest electrode conductive surface area)N/AN/ANo electrodes – applicators not connected to the patient
Burst Mode7 (i.e., pulse trains) a. Pulses per burst b. Bursts per second c. Burst duration (seconds) d. Duty Cycle [Line (b) x Line (c)]N/AN/A
ON Time (seconds)N/AN/A

Page 11

Sections 2,3 Technological CharacteristicsSUBJECT DEVICE Zimmer MedizinSyteme GmbH CoolTone (K251378)PREDICATE DEVICE Zimmer MedizinSyteme GmbH CoolTone (K220601)Comments
OFF Time (seconds)N/AN/A
Additional Features (if applicable)N/AN/A

Product Labeling

This premarket submission included changes in product labeling. The user's manual was updated to capture information obtained in post-market surveillance activity. The labeling changes described in the submission do not affect the intended use, performance or risk profile of the device.

Clinical Study

No clinical testing was required for this change.

Substantial Equivalence Determination

a. Intended Use: There is no change to the indications for use for this submission between the subject and predicate device (K220601).

b. Technological Characteristics: The CoolTone has the same technological characteristics as the predicate device (K220601).

c. Labeling: The changes in labeling described in this submission does not create any new questions of safety and performance as compared to the predicate device.

The CoolTone is substantially equivalent to the predicate K220601.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).