K Number
K251305

Validate with FDA (Live)

Date Cleared
2025-08-26

(120 days)

Product Code
Regulation Number
884.6180
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).
Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes(MII).

Device Description

The Ultra-Fast Vitri and Ultra-Fast Warm is composed of a set of three media to vitrify and warm oocytes for assisted reproductive technology (ART) procedures.

The Ultra-Fast Vitri includes two components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol and dimethyl sulfoxide. There are two vitrification procedures to choose from. During the vitrification process, oocytes are first exposed to ES then in VS within several minutes. Using this methodology, permeating cryoprotectants can replace water in the oocytes prior to vitrification and storage in liquid nitrogen. The Ultra-Fast Vitri comes prepackaged with 1.5 mL vial or 4 mL vial of ES, three 1.5 mL vials or three 4 mL vials of VS.

Ultra-Fast Warm is composed of one media used for warming and removing cryoprotectants from vitrified oocytes. It is composed of Thawing Solution (TS). The Ultra-Fast Warm comes pre-packaged with four 4.0 ml vials of TS.

All the media in the Ultra-Fast Vitri and Ultra-Fast Warm contain Gentamicin. The media undergoes aseptic filtration, while the vials are sterilized by radiation.

AI/ML Overview

Based on the provided FDA 510(k) Clearance Letter, here's a description of the acceptance criteria and the study that proves the device meets them:

Device: Ultra-Fast Vitri; Ultra-Fast Warm
Description: A set of three media used for the vitrification (cryopreservation) and warming of oocytes (MII) for Assisted Reproductive Technology (ART) procedures.


Acceptance Criteria and Reported Device Performance

The core acceptance criteria for this device, as demonstrated through non-clinical and clinical performance data, revolve around its biological compatibility and effectiveness in preserving and recovering oocytes without compromising their viability or subsequent reproductive outcomes.

Acceptance Criteria CategorySpecific Metric (Unit)Acceptance CriteriaReported Device Performance (Ultra-Fast Vitri/Warm)
Non-Clinical Performance
Color/AppearanceVisual inspectionAcceptable appearancePassed
pH TestingpH value (range)7.20 – 7.60Passed (7.20 – 7.60)
Endotoxin TestingEndotoxin level (EU/mL)<0.25 EU/mL (LAL methodology)Passed (<0.25 EU/mL LAL methodology)
Osmolality TestingOsmolality value(Not explicitly stated in document, but implied as conforming to media standards)Passed
Sterility TestingSterilityPasses USP <71>Passed (Passes USP <71>)
Gentamicin TestGentamicin presence/level(Not explicitly stated, but implied as conforming to specification)Passed
Initial Media Dispensing ValidationFunctional dispensing/packaging(Not explicitly stated, but implied as successful)Passed
Mouse Embryo Assay (MEA)One-cell embryo development (96 hours)>80%Passed (>80%)
BiocompatibilityBiocompatibility with cellsPassesPasses
Storage StabilityTemperature range (°C)2 – 8°C2 – 8°C
Shelf LifeDuration (months)12 months12 months
Clinical Performance
Oocyte Survival Rate%(Implied to be comparable to conventional protocol)100.0% (Ultra-Fast) vs. 90.9% (Conventional)
Clinical Pregnancy Rate%(Implied to be comparable to conventional protocol)65.2% (Ultra-Fast) vs. 54.3% (Conventional)
Live Birth Rate%(Implied to be comparable to conventional protocol)56.5% (Ultra-Fast) vs. 52.2% (Conventional)

Study Proving Device Meets Acceptance Criteria

The provided document describes both non-clinical (bench) and clinical performance studies to demonstrate the safety and effectiveness of the Ultra-Fast Vitri and Ultra-Fast Warm device and its substantial equivalence to the predicate device.

1. Non-Clinical Performance Data (Bench Testing):

  • Description: A series of laboratory tests conducted directly on the media to confirm its physical, chemical, and biological properties.
  • Specific Tests: Color/Appearance, pH Testing, Endotoxin testing, Osmolality Testing, Sterility Testing, Gentamicin Test, Initial Media Dispensing Validation, Mouse Embryo Assay (MEA), and Biocompatibility.
  • Proof of Concept: The device passed all these tests, including achieving >80% one-cell development in the Mouse Embryo Assay, which is a critical biological performance indicator for reproductive media as per FDA guidance.

2. Clinical Performance Data:

  • Study Design: A comparative study referenced from literature that evaluated the effectiveness of the ultra-fast vitrification and warming protocols using the subject device against conventional protocols (presumably using the predicate or similar conventional vitrification/warming solutions).
  • Study Objective: To demonstrate comparable outcomes (oocyte survival, clinical pregnancy, live birth rates) between the ultra-fast protocol and conventional protocols.

Here's a breakdown of the specific requested information about the clinical study:

  • 2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 1,077 mature oocytes in total.
      • 519 oocytes for the conventional vitrification and warming protocols group.
      • 558 oocytes for the ultra-fast protocols (subject device) group.
    • Data Provenance: The document states "The referenced literature used Kitazato's vitrification and warming solutions (K171748 and K160864)".
      • It does not explicitly state the country of origin of the data.
      • It does not explicitly state if the study was retrospective or prospective. However, given it's a "study" comparing protocols, it's typically prospective, but this cannot be confirmed from the text.
  • 3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This information is not provided in the document. The study focuses on clinical outcomes (survival, pregnancy, live birth rates) rather than human interpretation of images or other subjective assessments that would require expert consensus for ground truth.
  • 4. Adjudication Method for the Test Set:

    • This information is not applicable/not provided. Adjudication methods (like 2+1, 3+1) are typically relevant for studies involving human interpretation where reviewer disagreement needs to be resolved (e.g., radiology studies). This study measures biological/clinical outcomes.
  • 5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    • No, an MRMC study was not done. MRMC studies are used to assess the impact of a device (often AI) on human reader performance, typically in diagnostic imaging. This study evaluated the direct clinical effectiveness of the media itself.
    • Therefore, an effect size of how much human readers improve with AI vs. without AI assistance is not applicable.
  • 6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:

    • Yes, in a sense, the clinical study assessed the "standalone" performance of the media with its associated protocols. It compared the outcomes from using the media (with its specific ultra-fast protocol) to conventional media/protocols. There isn't an "algorithm" in the traditional sense for this device; it's a chemical formulation and protocol. The "performance" is the biological outcome achieved by the oocytes.
  • 7. The Type of Ground Truth Used:

    • The ground truth was based on clinical outcomes data:
      • Oocyte survival rate after vitrification and thawing.
      • Clinical pregnancy rate (following embryo transfer resulting from these oocytes).
      • Live birth rate (following clinical pregnancy).
    • These are considered objective biological and clinical endpoints.
  • 8. The Sample Size for the Training Set:

    • This information is not applicable/not provided. This device is a media (consumable), not an AI algorithm that requires a separate "training set" of data. The "development" of the media and protocols would be based on laboratory research and refinement rather than a data training paradigm.
  • 9. How the Ground Truth for the Training Set Was Established:

    • This information is not applicable/not provided for the same reasons as #8.

FDA 510(k) Clearance Letter - Ultra-Fast Vitri; Ultra-Fast Warm

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 26, 2025

Kitazato Corporation
℅ Mei Li
Quality and Regulatory Affairs Consultant
Emergo Global Consulting, LLC
2500 Bee Cave Road Building 1, Suite 300
Austin, Texas 78746

Re: K251305
Trade/Device Name: Ultra-Fast Vitri; Ultra-Fast Warm
Regulation Number: 21 CFR 884.6180
Regulation Name: Reproductive Media And Supplements
Regulatory Class: Class II
Product Code: MQL
Dated: March 27, 2025
Received: July 29, 2025

Dear Mei Li:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251305 - Mei Li
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific

Page 3

K251305 - Mei Li
Page 3

regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia -S

Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251305

Device Name: Ultra-Fast Vitri; Ultra-Fast Warm

Indications for Use (Describe):
Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).
Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes(MII).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary – K251305

Ultra-Fast Vitri; Ultra-Fast Warm

1. Submission Sponsor

Kitazato Corporation
100-10 Yanagishima,
Fuji Shizuoka, 416-0932
Japan
Ms. Kyoko Izumi
Quality Assurance Manager
Phone Number: (81) 545 65 7122

2. Submission Correspondent

Emergo by UL
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327-9997
Email: LST.US.EmergoFDASubmissions@ul.com
Contact: Mei Li
Title: Quality and Regulatory Affairs Consultant

3. Date Prepared

August 26, 2025

4. Device Identification

Trade/Proprietary Name: Ultra-Fast Vitri; Ultra-Fast Warm
Common Name: Vitrification and warming media for vitrified oocytes
Regulation Name: Reproductive media and supplements
Regulation Number: 884.6180
Product Code: MQL, Reproductive Media and Supplements
Class: Class II
Classification Panel: Obstetrics/Gynecology

5. Legally Marketed Predicate Device(s)

Device name: Vitrification Kit and Thawing Kit
510(k) number: K171748

Page 6

K251305
Page 2 of 5

Manufacturer: Kitazato Corporation

The predicate device has not been subject to a design-related recall.

6. Device Description

The Ultra-Fast Vitri and Ultra-Fast Warm is composed of a set of three media to vitrify and warm oocytes for assisted reproductive technology (ART) procedures.

The Ultra-Fast Vitri includes two components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol and dimethyl sulfoxide. There are two vitrification procedures to choose from. During the vitrification process, oocytes are first exposed to ES then in VS within several minutes. Using this methodology, permeating cryoprotectants can replace water in the oocytes prior to vitrification and storage in liquid nitrogen. The Ultra-Fast Vitri comes prepackaged with 1.5 mL vial or 4 mL vial of ES, three 1.5 mL vials or three 4 mL vials of VS.

Ultra-Fast Warm is composed of one media used for warming and removing cryoprotectants from vitrified oocytes. It is composed of Thawing Solution (TS). The Ultra-Fast Warm comes pre-packaged with four 4.0 ml vials of TS.

All the media in the Ultra-Fast Vitri and Ultra-Fast Warm contain Gentamicin. The media undergoes aseptic filtration, while the vials are sterilized by radiation.

7. Indication for Use Statement

Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).
Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes (MII).

8. Substantial Equivalence Discussion

The following table compares the Ultra-Fast Vitri and Ultra-Fast Warm to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any different questions of safety or effectiveness as compared to the predicate device.

Comparison of Characteristics

FeatureSubject Device: Ultra-Fast Vitri and Ultra-Fast WarmPredicate: Vitrification Kit and Thawing Kit (K171748)
Device ClassificationClass II Media, ReproductiveClass II Media, Reproductive
Product CodeMQLMQL, MQK
Regulation No.21 CFR 884.618021 CFR 884.6180
ManufacturerKitazato CorporationKitazato Corporation
Intended UseReproductive media and supplementsReproductive media and supplements

Page 7

K251305
Page 3 of 5

FeatureSubject Device: Ultra-Fast Vitri and Ultra-Fast WarmPredicate: Vitrification Kit and Thawing Kit (K171748)
Indications for UseUltra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).The Vitrification Kit is indicated for use in the preparation, vitrification and storage of oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.
Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes (MII).The Thawing Kit is indicated for use in the preparation and thawing of vitrified oocytes (Mll), pronuclear (PN) zygotes through day 3 cleavage stage embryos and blastocyst stage embryos.
Neither Ultra-Fast Vitri not Ultra-Fast Warm makes direct contact with the patient, only with the oocyte.This product does not make direct contact with the patient, only with the gametes/embryos: Oocyte, PN through Blastocyst.
Embryo StageUltra-Fast Vitri: OocyteVitrification kit: Oocyte, PN through Blastocyst
Ultra-Fast Warm: OocyteThawing kit: Oocyte, PN through Blastocyst
Principle of OperationUltra-Fast Vitri: Provides users with the ability to cryopreserve supernumerary oocytes created during the in vitro fertilization procedureVitrification kit: Provides users with the ability to cryopreserve supernumerary oocytes or embryos created during the in vitro fertilization procedure
Ultra-Fast Warm: Provides users with the ability to warm vitrified oocytes for use at a future point in timeThawing kit: Provides users with the ability to warm vitrified oocytes or embryos for use at a future point in time
Vitrification FormulationIn a Medium 199 HEPES buffered MediumIn a Medium 199 HEPES buffered Medium
Ethylene glycolEthylene glycol
DMSODMSO
TrehaloseTrehalose
Hydroxypropyl Cellulose (HPC)Hydroxypropyl Cellulose (HPC)
GentamicinGentamicin
Vitrification ProtocolProcedures provided for both Repro and Standard plates. Faster overall procedure with fewer steps.Procedure provided for Repro plate. More steps and time to the procedure.
Vitrification Steps2 Procedures, each procedure with 2 Steps. The first step consists of 1 min equilibration in ES. The second step consists of 1 min vitrification in VS.2 Steps. The first step consists of a 15-minute equilibration involving BS and ES. The second step consists of 1-1.5 min vitrification involving VS.

Page 8

K251305
Page 4 of 5

FeatureSubject Device: Ultra-Fast Vitri and Ultra-Fast WarmPredicate: Vitrification Kit and Thawing Kit (K171748)
The difference between the procedure using repro plates vs standard dishes relates to the volumes used (procedure on standard dish uses drops, this dish does not have wells) and the plate/dish in which the procedure is carried out.
Vitrification Media ComponentUltra-Fast Vitri devices can be composed on the following based on the reference number:Basic Solution (BS) 1.5mL x 1 Vial, Equilibration Solution (ES) 1.5mL x 1 Vial, and Vitrification Solution (VS) 1.5mL x 2 Vials
VT601USUF (91268): 1×1.5mL vial ES, 3×1.5mL vial VS
VT601USUF-4 (91269): 1×4mL vial ES, 3×4mL vial VS
VT601USUF-ESx4 (91260): 4×1.5mL vial ES
VT601USUF-VSx4 (91261): 4×1.5mL vial VS
VT601USUF-ES4x4 (91266): 4×4mL vial ES
VT601USUF-VS4x4 (91267): 4×4mL vial VS
Thawing FormulationIn a Medium 199 HEPES buffered MediumIn a Medium 199 HEPES buffered Medium
Hydroxypropyl Cellulose (HPC) (v/v)Hydroxypropyl Cellulose (HPC) (v/v)
GentamicinGentamicin
TrehaloseTrehalose
Thawing ProtocolFaster overall procedure with fewer steps.More steps and time to the procedure.
Thawing Steps1 Step consisting of placing the oocyte 1 min in TS.Multiple Steps involving three different solutions (TS, DS and WS).
Thawing Media ComponentTS solution 4.0mL x 4 VialsThawing Solution (TS)4.0mL x 2 Vials, Diluent Solution (DS) 4.0mL x 1 Vial, Washing Solution (WS)4.0mLx1 Vial
PackagingSolutions are packed in plastic vials. 4 vials are packed in a cardboard box.Solutions are packed in plastic vials. 4 vials are packed in a cardboard box.
SterileSolutions sterilized using aseptic processing techniques through filtration. Vial containers are sterilized via radiationSolutions sterilized using aseptic processing techniques through filtration Vial containers are sterilized via radiation
EndotoxinEndotoxin by LAL methodology <0.25 EU/mL (LAL)Endotoxin by LAL methodology <0.25 EU/mL (LAL)
Mouse Embryo Assay>80% one-cell 96 hours>80% one-cell 96 hours
Sterility TestingPasses USP <71>Passes USP <71>
pH Test7.20 – 7.607.20 – 7.60
BiocompatibilityPassesPasses

Page 9

K251305
Page 5 of 5

FeatureSubject Device: Ultra-Fast Vitri and Ultra-Fast WarmPredicate: Vitrification Kit and Thawing Kit (K171748)
Storage2 – 8°C2 – 8°C
Shelf Life12 months12 months

9. Non-Clinical Performance Data

Bench testing included the following:

  • Color/Appearance
  • pH Testing
  • Endotoxin testing
  • Osmolality Testing
  • Sterility Testing
  • Gentamicin Test
  • Initial Media Dispensing Validation

To demonstrate safety and effectiveness of Ultra-Fast Vitri and Ultra-Fast Warm and to show substantial equivalence to the predicate device, the Mouse Embryo Assay (MEA) was conducted using the subject device with revised protocols to align with the ultra-fast procedures. The Ultra-Fast Vitri and Ultra-Fast Warm passed the testing in accordance with FDA guidance shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:

  • Mouse Embryo Assay (MEA) – Passed
    • Mouse Embryo Assay for Assisted Reproduction Technology Devices, issued January 5, 2021

10. Clinical Performance Data

Clinical performance data to support the safety and performance of the subject device was included in this submission. The referenced literature used Kitazato's vitrification and warming solutions (K171748 and K160864) and compared the effectiveness of vitrification and warming using the conventional protocol to the ultra-fast protocols of the subject device (i.e., vitrification: 1 minute in equilibration solution and 1 minutes in vitrification solutions; thawing: 1 minute in thawing solution at 37˚C). The study was conducted using 1,077 mature oocytes (n=519 for the conventional vitrification and warming protocols and n=558 for the ultra-fast protocols). The results from the study showed that oocyte survival rate after vitrification and thawing (100.0% vs. 90.9%), clinical pregnancy rate (65.2% vs. 54.3%), and live birth rate (56.5% vs. 52.2%) obtained with the ultra-fast vitrification and warming and with the conventional vitrification and warming protocols, respectively, were comparable.

11. Statement of Substantial Equivalence

The results of the performance testing described above demonstrate that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.6180 Reproductive media and supplements.

(a)
Identification. Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.