K Number
K251146

Validate with FDA (Live)

Date Cleared
2025-10-17

(186 days)

Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended Use:
The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.

Indications for Use
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

The monitors are only for use on one patient at a time.

The monitors are not therapeutic devices.

The monitors are for prescription use only.

The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.

BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

The IntelliVue NMT is intended to be used as an objective neuromuscular transmission monitor that measures the muscle response to electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.

The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

The Edwards FloTrac solution offers continuous assessment of hemodynamic parameters. It is indicated to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is indicated for use in adult critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. The Edwards FloTrac solution is indicated to be used in the operating room, intensive care unit, and emergency room.

The monitors are intended for use by trained healthcare professionals in a hospital environment.
They are not intended for home use.
The monitors are additionally intended for use in transport situations within hospital environments.

Device Description

The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management.

The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and non-invasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, EEG, BIS, NMT, and gas analysis.

The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals. The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display. The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - IntelliVue Patient Monitors

Page 1

October 17, 2025

Philips Medizin Systeme Boeblingen GmbH
Vasiliki Begkou
Principal Regulatory Affairs Specialist
Hewlett-Packard-Strasse 2
Boeblingen, BW 71034
Germany

Re: K251146
Trade/Device Name: IntelliVue Patient monitors MX400, MX450, MX500, MX550
Regulation Number: 21 CFR 870.1025
Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm)
Regulatory Class: Class II
Product Code: MHX, MLD, DSJ, DXN, DPT, DQA, MSX, CCK, FLL
Dated: September 19, 2025
Received: September 19, 2025

Dear Vasiliki Begkou:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251146 - Vasiliki Begkou Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251146 - Vasiliki Begkou Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JENNIFER W. SHIH -S

Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 2

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K251146

Device Name: IntelliVue Patient monitors MX400, MX450, MX500, MX550

Indications for Use (Describe)

Intended Use:
The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.

Indications for Use
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

The monitors are only for use on one patient at a time.

The monitors are not therapeutic devices.

The monitors are for prescription use only.

The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.

BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

The IntelliVue NMT is intended to be used as an objective neuromuscular transmission monitor that measures the muscle response to electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

Page 5

FORM FDA 3881 (6/20) Page 2 of 2

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.

The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory.

The Edwards FloTrac solution offers continuous assessment of hemodynamic parameters. It is indicated to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is indicated for use in adult critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. The Edwards FloTrac solution is indicated to be used in the operating room, intensive care unit, and emergency room.

The monitors are intended for use by trained healthcare professionals in a hospital environment.
They are not intended for home use.
The monitors are additionally intended for use in transport situations within hospital environments.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 510(k) Summary

K251146
Philips Hospital Patient Monitoring
Page 1 of 9

510(k) Summary

1.1 Submitter

Date PreparedApril 14, 2025
Submitter/OwnerPhilips Medizin Systeme Böblingen GmbHFDA Establishment Number 9610816Hewlett-Packard-Str. 271034 BöblingenGermany
Key ContactVasiliki BegkouPrincipal Regulatory Affairs SpecialistVasiliki.begkou@philips.comPhone: +49 15110207311
510(k) Submission TypeTraditional 510(k)

1.2 Device

Trade NameIntelliVue Patient monitors MX400, MX450, MX500, MX550
Common NameMultiparameter Patient Monitor
Classification NamePanel & Name: Cardiovascular DevicesSubpart & Division: 21 CFR §870.1025Regulatory Class: IIProduct Code: MHX

1.3 Predicate Device

510(k) No.CompanyDevice NameProduct Code
K223574PhilipsIntelliVue Patient Monitor MX400IntelliVue Patient Monitor MX450IntelliVue Patient Monitor MX500IntelliVue Patient Monitor MX550MHX
K180881Edwards LifesciencesHemoSphere Advanced Monitor,HemoSphere Swan-Ganz Module,HemoSphere Oximetry Cable, HemoSphere Pressure Cable, Acumen Hypotension Prediction Index featureDQK

Page 7

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 2 of 9

The subject devices are substantially equivalent to the legally marketed predicate devices.

1.4 Device Description

IntelliVue Patient Monitors MX400, MX450, MX500, MX550 – description of the device per 21 CFR 807.92(a) (4)

The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management.

The monitors support multiple non-invasive and invasive measurements such as ECG, arrhythmia, ST, QT, SpO2, respiration rate, pulse rate, heart rate, invasive and non-invasive blood pressure, temperature, CO2, tcpO2/ tcpCO2, C.O., CCO, intravascular SO2, SvO2, ScvO2, EEG, BIS, NMT, and gas analysis.

The monitors offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They are located at the patient bedside vicinity and can also be used during patient transport inside hospitals. The monitors have a color display with touchscreen as a primary input device. They also support a specialized remote control, keyboard and pointing devices such as a mouse. External displays can be connected to a built-in video port to provide an adaptive duplicate image of the primary display. The monitors can interact with several compatible external measuring and auxiliary devices locally at the bedside or in transport situations and with the Central Station via LAN or wireless link.

1.5 Intended Use and Indication for Use

Intended Use as required per 21 CFR 807.92(a)(5)

Intended Use:
The devices are intended to be used for monitoring and recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics, and neonates.

Indications for Use:
The monitors are indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.

The monitors are only for use on one patient at a time.

The monitors are not therapeutic devices.

The monitors are for prescription use only.

The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

Page 8

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 3 of 9

ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The transcutaneous gas measurement (tcGas) with the M1018A plug-in module is restricted to neonatal patients only.

BIS is intended for use under the direct supervision of a licensed health care practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

The SSC Sepsis Protocol, in the ProtocolWatch clinical decision support tool, is intended for use with adult patients only.

The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

The derived measurement Pulse Pressure Variation (PPV) is intended for use with sedated patients receiving controlled mechanical ventilation and mainly free from cardiac arrhythmia. The PPV measurement has been validated only for adult patients.

The IntelliVue NMT is intended to be used as an objective neuromuscular transmission monitor that measures the muscle response to electrical stimulation of a peripheral nerve. The NMT Module is intended to be used with adult and pediatric patients.

The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

The non-invasive Masimo O3 Regional Oximeter System and accessories are indicated for use as an adjunct monitor of absolute and trended regional hemoglobin oxygen saturation of blood (rSO2) in the cerebral region under the sensors. The Masimo O3 Regional Oximeter System and accessories are indicated for use on adults ≥40 kg and on pediatrics ≥5 kg and <40 kg, in healthcare environments.

Page 9

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 4 of 9

The SedLine Sedation Monitor is intended to monitor the state of the brain by real-time data acquisition and processing of EEG signals. The system includes the Patient State Index (PSI), a proprietary computed EEG variable that is related to the effect of anesthetic agents. The agents include: Alfentanil, Desflurane, Fentanyl, Isoflurane, Nitrous Oxide, Propofol, Remifentanil, and Sevoflurane. The SedLine Sedation Monitor is intended for use with adult patients (18 years of age and older) in the operating room (OR), intensive care unit (ICU), and clinical research laboratory. The Edwards FloTrac solution offers continuous assessment of hemodynamic parameters. It is indicated to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is indicated for use in adult critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. The Edwards FloTrac solution is indicated to be used in the operating room, intensive care unit, and emergency room. The monitors are intended for use by trained healthcare professionals in a hospital environment. They are not intended for home use. The monitors are additionally intended for use in transport situations within hospital environments.

1.6 Comparison of Intended Uses for Subject Device and Predicate

The IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 have the same intended use as their predicate device.

Additionally, the MX500 and MX550 models inherited the Hemosphere Advanced Monitor indication:

  • The Edwards FloTrac solution offers continuous assessment of hemodynamic parameters. It is indicated to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. It is indicated for use in adult critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. The Edwards FloTrac solution is indicated to be used in the operating room, intensive care unit, and emergency room.

Page 10

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 5 of 9

1.7 Comparison of Technological Characteristics with Predicate Device

Item of ComparisonDescription/Rationale
Similarities
Device Design- The hardware design of the subject device is the same as that of the predicate device.- device design of added Edwards Lifesciences external measurement devices composed of the FloTrac Pressure Cable-Philips and FloTrac sensor is equivalent to their predicate devices- Subject and predicate device run on IntelliVue Software (version R.00.), which is also the software platform available in the predicate device (version P.01.).
Energy Source- Subject device is powered by AC mains or from built-in battery, as the predicate device.- Subject and predicate device do not deliver energy to patients- added external Edwards Lifesciences devices do not have their own power supply; it is supplied by the host monitor. This is the same as for their predicate devices
Software/Hardware Features- Proposed modification does not introduce any new technological hardware features.- Measurements offered by the subject device are also offered by the predicate device. Subject and predicate devices have the same interfaces.- Software features introduced for the proposed IntelliVue Software version R.00.01, rely on a well-established software platform and architecture which was iteratively developed from the version available in the predicate device.- Hardware and Software of the added external Edwards Lifesciences devices are equivalent to their predicates- measurements of physiological parameters are the same as in the predicate devices

Page 11

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 6 of 9

1.7 Comparison of Technological Characteristics with Predicate Device

Item of ComparisonDescription/Rationale
Physiological Parameters- The subject device supports the same measurement modules as the predicate device, except for: - The subject device (MX500/MX550 only) additionally supports the new plug-in module for FloTrac (with P/N 867409)- Measurement Modules M1019A (Gas Module), M1013A (Gas Module), M1014A (Spirometry Module) are no longer supported by the subject device. Previous versions (incl. predicate device) of MX500/MX500 were compatible with these modules.- existing physiological parameters of the predicate devices remain unchanged- the physiological parameters of the added external Edwards Lifesciences devices are the same as those of their respective predicate devices
Operating Principle and Mechanism of Action- Operating principle and mechanism of action from subject device is unchanged from the predicate device.
Human Interface- The human interface of the IntelliVue MX400, MX450, MX500 and MX550 remain the same. Data presentation and user control of the new plug-in module for FloTrac (with P/N 867409) are done through the Patient Monitors (MX500/MX550 only). This is unchanged from how the predicate device supports and operates with other plug-in measurements (K223574). The addition of the user interfaces for the FloTrac module do not influence the patient monitors´ user interface or the user interaction with the patient monitors.- The human interfaces of added Edwards Lifesciences devices are unchanged to their predicates
Measurement Accessories- Accessories applied to the predicate device are compatible with the subject devices.- all accessories of the predicate devices - IntelliVue Patient Monitor MX400 - IntelliVue Patient Monitor MX450 - IntelliVue Patient Monitor MX500 - IntelliVue Patient Monitor MX550as well as the Edwards Lifesciences added accessories are equivalent to their predicates

Page 12

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 7 of 9

1.7 Comparison of Technological Characteristics with Predicate Device

Item of ComparisonDescription/Rationale
Differences
Compatibility with FloTrac Plug-in Module (867409)- The new plug-in module FloTrac (P/N 867409) acts as a communication interface between the IntelliVue Patient Monitors MX500/MX550 and external Edwards Lifesciences pressure cable and sensor.- the new plug-in module does not alter or modify the contents of the transferred data
Software- The software of the subject devices was slightly changed when compared to its predicate device, to: - further enhance some of the existing functionalities, i.e enabling fast access of applications, recording more alarm logging information, disabling the configuration of the "AlarmsOffAtStart" feature, modifying the factory default for the NBP, enhancing the existing visual indications in the case of speaker malfunction, transmitting alarm limits of more numerics to the central station, introducing the clinical analytics framework into the Software infrastructure. - extend its capability, so that the MX500 and MX550 patient monitors accept the physiological data provided by the Edwards FloTrac as compatible and display the FloTrac measurement on these patient monitors' screen, i.e compatibility with new FloTrac Plug-in Module.

1.8 Substantial Equivalence Summary

Operational and technological characteristics form the basis for the determination of substantial equivalence of the subject devices with the legally marketed predicate devices. The subject devices are substantially equivalent to the predicate devices.

1.9 Performance Data

Clinical Tests – Harmonized Standards

The subject devices have passed all safety tests for demonstrated compliance with the recognized standards below.

StandardFDA Recognition #Title #

Page 13

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 9 of 9

StandardFDA Recognition #Title
IEC 60601-1:2005 incl. AMD1:2012 and AMD2:202019-49Medical Electrical Equipment – Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2:2014 incl. AMD1:202019-36Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests.
IEC 60601-1-6:2010 incl. AMD1:2013 and AMD2:20205-132Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability.
IEC 60601-1-8:2006 incl. AMD1:2012 and AMD2:20205-131Medical electrical equipment – Part 1-8: General requirements for basic safety and essential performance – Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.
IEC 62304:2006 incl. AMD1: 201513-79Medical device software Software life-cycle processes

Page 14

IntelliVue Patient Monitors MX400, MX450, MX500 and MX550
510(k) Summary K251146
Philips Hospital Patient Monitoring Page 9 of 9

Non-clinical Bench Tests

No new issues of safety or effectiveness are introduced with the changes proposed for the IntelliVue Patient Monitors MX400, MX450, MX500 and MX550.

Clinical Studies

The subject devices, like the primary predicate devices, did not require clinical trials.

1.10 Conclusion

Substantial equivalence assessment, based on non-clinical bench testing, electrical safety, electromagnetic compatibility, software verification and validation, human factors and interoperability testing, demonstrate that the changes proposed for the IntelliVue Patient Monitors MX400, MX450, MX500 and MX550 do not raise different questions of safety and effectiveness when compared to their predicate devices (previous version of the same device cleared with K223574).

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.