AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 15cm-48cm (approx. 5.9"-18.9").

Device Description

Fully Automatic Electronic Blood Pressure Monitor (BP-300C, BP-300CV, BP-300V, BPM1, BPX1, KD-338N, KD-553, KD-557BR, KD-558, KD-558BR, KD-595, KD-5031N, KD-5810, KD-5810B, KD-5811, KD-5811A, KD-5811V, KD-5815, KD-5920, KD-5920L, KD-5920TL, KD-5923, KN-550LT) is designed and manufactured according to IEC 80601-2-30.

The operational principle is based on Oscillo-metric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure and display the result. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

AI/ML Overview

The provided document is a 510(k) clearance letter for various blood pressure monitors. It outlines the regulatory approval process and compares the new devices to a predicate device. However, it does not contain the detailed acceptance criteria and study results in the format typically used for AI/software devices.

Specifically, this document describes validation against standards for medical electrical equipment (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11) and automated non-invasive sphygmomanometers (IEC 80601-2-30, ISO 81060-2). It focuses on the substantial equivalence of the physical blood pressure monitors and their underlying oscillometric and pressure sensor technology, rather than the performance of an AI algorithm based on a test set, ground truth, and expert interpretations.

Therefore, many of the specific questions about AI/software device validation (e.g., sample size for the test set, data provenance, number of experts for ground truth, MRMC studies, standalone performance, training set details) cannot be answered from this document.

However, I can extract information related to the performance of the blood pressure monitors themselves, based on the included standards.


Acceptance Criteria and Device Performance (for Blood Pressure Monitor functionality, not AI):

Since this is a blood pressure monitor, the primary performance criteria relate to its accuracy in measuring blood pressure and pulse rate, and compliance with relevant safety and performance standards for automated non-invasive sphygmomanometers.

Acceptance CriteriaReported Device Performance
Accuracy (ISO 81060-2): "Meeting criteria 1 and criteria 2 of ISO 81060-2"Stated as "verified by meeting criteria 1 and criteria 2 of ISO 81060-2". (Specific numerical values for mean difference and standard deviation are not provided in this summary but are implicitly met by passing the standard.)
Pulse rate range40-180 times/min
Pulse rate accuracyLess than 60: ±3bpmMore than 60 (incl.): ±5%
Systolic Range60-260 mmHg
Diastolic Range40-199 mmHg
Pressure AccuracyWithin ±3 mmHg
Cuff pressure Range0-300 mmHg
Over pressure Limit300 mmHg
Compliance with: - IEC 60601-1:2005+AMD1: 2012+AMD2: 2020- IEC 60601-1-2:2014+AMD1: 2020- IEC 60601-1-11: 2015+AMD1: 2020- IEC 80601-2-30: 2018All listed standards were met, demonstrating basic safety, essential performance, EMC, and home healthcare environment compliance.

Unable to Answer from Document (Common for AI/Software Device Submissions, but not for this type of device):

The following questions are not applicable or cannot be answered from this 510(k) summary because the device described is a physical blood pressure monitor, not an AI/software device that interprets medical images or other complex data requiring expert adjudication, training sets, or MRMC studies.

  • Sample size used for the test set and the data provenance:
    • Test Set Size: "A total of 231 patients (107 males and 124 females) were enrolled in the study." This is the clinical study population for blood pressure measurement accuracy.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The study is described as a "clinical study," which implies prospective data collection for the purpose of the study.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    • Not applicable. Ground truth for blood pressure measurement is established through a standard auscultation method (manual measurement by medical professionals using a stethoscope and sphygmomanometer), not by interpretation of images by experts.
  • Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    • Not applicable. Ground truth is direct measurement by a reference method.
  • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    • Not applicable. This is not an AI-assisted diagnostic device.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    • The device is a standalone blood pressure monitor. No human-in-the-loop interaction for interpretation (as in AI devices) is relevant. Its performance is its direct measurement accuracy.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    • Ground Truth: "Standard auscultation method was used as the reference blood pressure monitor measuring." This is the established clinical standard for direct comparison.
  • The sample size for the training set:
    • Not applicable. This is not an AI/machine learning device requiring a training set.
  • How the ground truth for the training set was established:
    • Not applicable.

FDA 510(k) Clearance Letter - K251113

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

August 4, 2025

Andon Health Co., Ltd.
Liu Yi, President
No. 3 Jin Ping Street
Ya An Road, Nankai District
Tianjin, Tianjin 300190
China

Re: K251113

Trade/Device Name:

  • iHealth Compare Wireless Blood Pressure Monitor (BP-300C)
  • iHealth Compare Pro Wireless Blood Pressure Monitor (BP-300CV)
  • iHealth Compare S Wireless Blood Pressure Monitor (BP-300V)
  • iHealth Wireless Blood Pressure Monitor (BPX1)
  • iHealth Blood Pressure Monitor (KD-595)
  • iHealth Track Pro Connected Blood Pressure Monitor (KN-550LT)
  • Semi Automatic Blood Pressure Monitor (KD-388N)
  • Arm Blood Pressure Monitor(KD-553)
  • Arm Blood Pressure Monitor(KD-557BR)
  • Arm Blood Pressure Monitor(KD-558)
  • Arm Blood Pressure Monitor(KD-558BR)
  • Arm Blood Pressure Monitor (KD-5031N)
  • Arm Blood Pressure Monitor (KD-5810)
  • Arm Blood Pressure Monitor (KD-5810B)
  • Arm Blood Pressure Monitor (KD-5811)
  • Arm Blood Pressure Monitor (KD-5811A)
  • Arm Blood Pressure Monitor (KD-5811V)
  • Arm Blood Pressure Monitor (KD-5815)
  • Arm Blood Pressure Monitor (KD-5920)
  • Arm Blood Pressure Monitor (KD-5920L)
  • Arm Blood Pressure Monitor (KD-5920TL)
  • Arm Blood Pressure Monitor (KD-5923)
  • iHealth Clear Wireless Blood Pressure Monitor (BPM1)

Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: July 7, 2025
Received: July 7, 2025

Page 2

K251113 - Liu Yi
Page 2

Dear Liu Yi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Page 3

K251113 - Liu Yi
Page 3

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

K251113
Page 1 of 1

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251113

Please provide the device trade name(s).

iHealth Compare Wireless Blood Pressure Monitor (BP-300C);
iHealth Compare Pro Wireless Blood Pressure Monitor (BP-300CV);
iHealth Compare S Wireless Blood Pressure Monitor (BP-300V);
iHealth Wireless Blood Pressure Monitor (BPX1);
iHealth Blood Pressure Monitor (KD-595);
iHealth Track Pro Connected Blood Pressure Monitor (KN-550LT);
Semi Automatic Blood Pressure Monitor (KD-388N);
Arm Blood Pressure Monitor (KD-553);
Arm Blood Pressure Monitor (KD-557BR);
Arm Blood Pressure Monitor (KD-558);
Arm Blood Pressure Monitor (KD-558BR);
Arm Blood Pressure Monitor (KD-5031N);
Arm Blood Pressure Monitor (KD-5810);
Arm Blood Pressure Monitor (KD-5810B);
Arm Blood Pressure Monitor (KD-5811);
Arm Blood Pressure Monitor (KD-5811A);
Arm Blood Pressure Monitor (KD-5811V);
Arm Blood Pressure Monitor (KD-5815);
Arm Blood Pressure Monitor (KD-5920);
Arm Blood Pressure Monitor (KD-5920L);
Arm Blood Pressure Monitor (KD-5920TL);
Arm Blood Pressure Monitor (KD-5923);
iHealth Clear Wireless Blood Pressure Monitor (BPM1)

Please provide your Indications for Use below.

Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a noninvasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 15cm-48cm (approx. 5.9"-18.9").

Please select the types of uses (select one or both, as applicable).
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 1 of 1

iHealth Compare Wireless Blood Pressure Monitor
Page 8 of 52

Page 5

510(k) SUMMARY

(In accordance with 21 CFR 807.92)

1.0 Applicant Information

Name: Andon Health Co., Ltd.
Address: No 3, Jinping Street, Ya An Road, Nankai District, Tianjin, 300190, P.R. China
Phone Number: 86-22-87611660
Fax Number: 86-22-87612379
Contact: Mr. Liu Yi
Date of Preparation: 04/11/2025

2.0 Device Information

Device Name:

  • BP-300C: iHealth Compare Wireless Blood Pressure Monitor
  • BP-300CV: iHealth Compare Pro Wireless Blood Pressure Monitor
  • BP-300V: iHealth Compare S Wireless Blood Pressure Monitor
  • BPX1: iHealth Wireless Blood Pressure Monitor
  • KD-595: iHealth Blood Pressure Monitor
  • KN-550LT: iHealth Track Pro Connected Blood Pressure Monitor
  • KD-388N: Semi Automatic Blood Pressure Monitor
  • KD-553, KD-557BR, KD-558, KD-558BR, KD-5031N, KD-5810, KD-5810B, KD-5811, KD-5811A, KD-5811V, KD-5815, KD-5920, KD-5920L, KD-5920TL, KD-5923: Arm Blood Pressure Monitor
  • BPM1: iHealth Clear Wireless Blood Pressure Monitor

K251113 Page 1 of 10

Page 6

Common Name: Blood Pressure Monitor

Classification Name: Non-Invasive Blood Pressure Measurement System

3.0 Classification

Product Code: DXN
Regulation Number: 21 CFR 870.1130
Classification: II
Review Panel: 870 Cardiovascular

4.0 Predicate Device Information

Manufacturer: Andon Health Co., Ltd.
Device Name: Fully Automatic Electronic Blood Pressure Monitor
510(k) Number: K234041
Classification: II
Product Code: DXN

5.0 Intended Use

Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 15cm - 48cm (approx.5.9″~18.9″).

6.0 Device Description

Fully Automatic Electronic Blood Pressure Monitor (BP-300C, BP-300CV,

K251113 Page 2 of 10

Page 7

BP-300V, BPM1, BPX1, KD-338N, KD-553, KD-557BR, KD-558, KD-558BR, KD-595, KD-5031N, KD-5810, KD-5810B, KD-5811, KD-5811A, KD-5811V, KD-5815, KD-5920, KD-5920L, KD-5920TL, KD-5923, KN-550LT) is designed and manufactured according to IEC 80601-2-30.

The operational principle is based on Oscillo-metric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure and display the result. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user.

7.0 Comparison of Technological Characteristics with Predicate Device

The following table is the summary of the technological characteristics of the proposed subject device and predicate device.

ItemSubject DevicePredicate Device Model: KD-5811BT 510(k) No.: K234041Comparison Result
Name and ModeliHealth Compare Wireless Blood Pressure MonitoriHealth Track Pro Connected Blood Pressure MonitorSemi Automatic Blood Pressure MonitorArm Blood Pressure MonitorBPM1: iHealth Clear Wireless Blood Pressure MonitorFully Automatic Electronic Blood Pressure Monitor--
Intended UseFully Automatic Electronic Blood Pressure Monitor is for use byFully Automatic Electronic Blood Pressure Monitor is forThe cuff range of subject device is different from the predicate device, but the

K251113 Page 3 of 10

Page 8

medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 15cm-48cm (approx. 5.9"-18.9").use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 17cm-42cm (approx. 6 11/16"-16 17/32")clinical report submitted demonstrates that the subject devices are as safe and effective as the predicate device.
Rx or OTCOTCOTCSame
PopulationAdultAdultSame
Cuff LocationUpper armUpper armSame

Physical Attributes

Weight (exclude batteries and cuff)About 328gAbout 290gAbout 330gAbout 200gAbout 220gAbout 300gAbout 211gAbout 166gAbout 235gAbout 221gAbout 232gAbout 237gAbout 239gAbout 223gAbout 189gAbout 225gAbout 233gAbout 191gAbout180gAbout 350gAbout 239gThe electrical safety report and EMC report submitted demonstrate that the subject devices are as safe and effective as the predicate device.
Dimensions144.5mm×115mm×60.7mm136mm×106mm×61mm98mm×98mm×46mm153mm×108mm×60mm136mm×94mm×52mm165mm×96mm×65mm138mm×98mm×48mm107mm×80mm×52mm150mm×95mm×41mm139.4mm×93.8mm×43.4139.4mm×93.8mm×43.4mmThe electrical safety report and EMC report submitted demonstrate that the subject devices are as safe and effective as the predicate device.

K251113 Page 4 of 10

Page 9

mm138mm×98mm×48.9mm136mm×120mm×61mm136mm×100mm×64mm139.4mm×93.8mm×41.8mm119mm×118mm×51mm
Memory2×360 times1×99 times120 times1×60 times1×90 times2×60 times4×30 times2×120 times2×200 times2×120 times2×120 timesThe software validation report submitted demonstrates the subject device are as safe and effective as the predicate device.
Displayed Calculated ParametersSYSDIAPulseIHBSYSDIAPulseIHBSame
Display componentLCD with backlightLCD without backlightLED without backlightLCD With backlightThe electrical safety report submitted demonstrates that the subject devices are as safe and effective as the predicate device.
Average functionThe average value of the last three results in the current user memory zoneThe average value in the morning of this week. It can display the morning average data from the 1 to 10 weeks ago.The average value of all the readings for the selected userThe average value of all the readings for the selected userThe average value of all the results which are measured from 5 o'clock to 9 o'clock in the last 7 daysThe average value of all the results which are measured from 18 o'clock to 20 o'clock in the last 7 daysNo Average function1. The average value of the last three results in the current user memory zone2. The average value of all the readings for the selected user3. The average value of all the results which are measured from 5 o'clock to 9 o'clock in the last 7 days4. The average value of all the results which are measured from 18 o'clock to 20 o'clock in the last 7 daysAverage functions are not core functions and key performance. The subject devices are as safe and effective as the predicate device in achieving its intended use.
Other display informationAll Models:Battery usageBattery usageSame
Date/ TimeDate/ TimeDisplay "Date/ Time" are not the core functions and key

K251113 Page 5 of 10

Page 10

Noneperformance. The subject devices are as safe and effective as the predicate device in achieving its intended use.
Blood pressure levelNoneBlood pressure levelDisplay "Blood pressure level" are not the core functions and key performance. The subject devices are as safe and effective as the predicate device in achieving its intended use.
Cuff wrap "OK" symbolCuff wrap "loose" symbolNoneCuff wrap "OK" symbolShow the cuff wrap symbol in different with predicate device have been validated, the software validation report submitted demonstrates the subject device are as safe and effective as the predicate device.Display "Cuff wrap symbol" are not the core functions and key performance. The subject devices which do not display this information are as safe and effective as the predicate device in achieving its intended use.
Body movement symbolNoneBody movement symbolDisplay "Body movement symbol" are not the core functions and key performance. The devices do not display this information are as safe and effective as the predicate device in achieving its intended use.
Bluetooth symbolNoneBluetooth symbolThe other models do not have Bluetooth function do not display this information.
Status MarkingNoneNoneThe models which have added status marking display compared with predicate device have been software validated. The software validation report submitted demonstrates these devices are as safe and effective as the predicate device.
Wireless FunctionBluetoothWi-FiNoneBluetoothThe model which has a different wireless function with predicate device have been tested. The software validation report and wireless coexistence testing report submitted demonstrates the device is as safe and effective

K251113 Page 6 of 10

Page 11

as the predicate device.Wireless function are not the core functions and key performance. The models do not have wireless function are as safe and effective as the predicate device in achieving its intended use.
Sound indication FunctionSound indication functionNoneNoneThe models which added sound indication function compared with predicate device have been tested. The software validation report submitted demonstrates these devices are as safe and effective as the predicate device.

Electrical Power

DC Mains5V6V:No Medical AC adapter5VThe models whose DC Mains are different with the predicate device have been tested. The electrical safety test report and EMC test report demonstrate these devices are as safe and effective as the predicate device.
Battery4×1.5V SIZE AA4×1.5V SIZE AAA3×1.5V SIZE AA13.7V Li-ion 1200mAh13.7V Li-ion 2200mAh1*3.7V Li-ion 1800mAh4×1.5V SIZE AAThe models whose battery are different with predicate device have been tested. The electrical safety test report and EMC test report demonstrate the subject device are as safe and effective as the predicate device.

Operation Limitation

Temperature5~40℃5~40℃Same
Humidity≤85% RH≤85% RHSame

Storage Limitation

Temperature-20℃ to 55℃-20℃ to 55℃Same
Humidity≤90% RH≤90% RHSame

Performance Parameter

Pulse rate40-180 times/min40-180 times/minSame
Pulse rate accuracyLess than 60: ±3bpmMore than 60 (incl.): ±5%Less than 60: ±3bpmMore than 60 (incl.): ±5%Same
Technique MethodOscillo-metricOscillo-metricSame
Measure processMeasure during deflatingMeasure during inflatingMeasure during deflatingThe models whose measure process are different from the predicate device have been tested, the clinical

K251113 Page 7 of 10

Page 12

investigation report submitted demonstrates that the subject devices are as safe and effective as the predicate device.
Systolic Range60-260 mmHg60-260 mmHgSame
Diastolic Range40-199 mmHg40-199 mmHgSame
Pressure AccuracyWithin ±3 mmHgWithin ±3 mmHgSame
Cuff pressure Range0-300 mmHg0-300 mmHgSame
Over pressure Limit300 mmHg300 mmHgSame
AlgorithmOscillometricOscillometricSame

There are no significant differences between the subject device and the predicate device. They are identical in terms of intended use, materials, design, and manufacturing methods.

8.0 Discussion of Non-Clinical Testing

Non-clinical tests were conducted to verify that the subject device met all design specifications. The test results demonstrated that the subject device complies with the following standards:

• IEC 60601-1:2005+AMD1: 2012+AMD2: 2020, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.

• IEC 60601-1-2:2014+AMD1: 2020, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests.

K251113 Page 8 of 10

Page 13

• IEC 60601-1-11: 2015+AMD1: 2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

• IEC 80601-2-30: 2018, Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.

The tests above demonstrate that the subject devices do not raise new questions of safety and effectiveness as compared to the predicate device.

9.0 Discussion of Clinical Testing

Comparison and evaluation are carried out between the subject device and cleared devices, and it shows that:

The clinical study was conducted in accordance with ISO 81060-2:2018+Amd-1:2020. A total of 231 patients (107 males and 124 females) were enrolled in the study, standard auscultation method was used as the reference blood pressure monitor measuring, the same arm sequential method was chosen. Accuracy of the blood pressure monitors was verified by meeting criteria 1 and criteria 2 of ISO 81060-2.

10.0 Comparison to the Predicate Device and Conclusion

K251113 Page 9 of 10

Page 14

The conclusion drawn from the nonclinical tests and clinical test demonstrate that the subject devices are substantially equivalent to the predicate device.

The subject devices are very similar to the predicate device in the intended use, design principle, materials, performance and applicable standards. Their appearance, memory capacity, display information, Electrical Power are different. However, the tests in this submission demonstrate that these differences do not raise any new questions of safety and effectiveness. And the subject device is as safe, effective and performs as well or better than the legally marketed predicate device.

Page 10 of 10

K251113

10 / 10

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).