K Number
K251053

Validate with FDA (Live)

Date Cleared
2025-07-15

(102 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Shinetell Plus™ Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

Device Description

Shinetell Plus™ Digital Early Pregnancy Test is used for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine, and is designed to be tested in dip or midstream mode. The test device consists of a single test strip assembled in a plastic housing, with an absorbent tip. The device is in a ready-to-use format.

Shinetell Plus™ Digital Early Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. The test would detect the light intensity by using the LED as the light source. After that, the result can be displayed on the display screen.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Shinetell Plus™ Digital Early Pregnancy Test, based on the provided 510(k) clearance letter:

1. Table of Acceptance Criteria and Reported Device Performance

The FDA clearance letter does not explicitly state "acceptance criteria" as a separate, quantitative table. However, we can infer the acceptance criteria from the performance studies presented, particularly the sensitivity and agreement rates. The primary criterion is demonstrating a sensitivity of 10 mIU/mL and high agreement with professional results and the predicate device.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Analytical Sensitivity (LOD)10 mIU/mL (based on predicate and stated sensitivity)Midstream & Dip Testing: 100% positive at 10 mIU/mL hCG.
Cross-ReactivityNo false positives with related hormones (hLH, hFSH, hTSH)No cross-reactivity observed with 1000 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 μIU/mL hTSH. Not affected by hCG β-core fragment up to 500,000 pmol/L.
InterferenceNo interference from common substances or urine propertiesNo interference observed at specified concentrations for various substances (Acetaminophen, Acetylsalicylic acid, Ascorbic acid, Atropine, Caffeine, Gentisic acid, Glucose, Hemoglobin, Tetracycline, Ampicillin, Albumin, β-hydroxybutyrate, Ephedrine, Phenylpropanolamine, Phenothiazine, EDTA, Salicyclic Acid, Benzoylecgonine, Cannabinol, Codeine, Ethanol, Bilirubin, Pregnanediol, Thiophene, Ketone). No effect from urine pH (4-9) or density (1.000-1.035).
Hook EffectNo hook effect up to clinically relevant high concentrationsNo hook effect observed up to 500,000 mIU/mL hCG.
Clinical Agreement (vs. Predicate)High agreement (e.g., >95%) with predicate deviceDip Testing: 100% conformity (49 Pos, 51 Neg vs Predicate). Midstream Testing: 100% conformity (48 Pos, 52 Neg vs Predicate).
Lay Person Readability & InterpretationHigh agreement (e.g., >95%) between lay user and professional results; ease of use; clear labelingIn-Stream Method: 100% positive and 100% negative conformity with professional results (48 Pos, 52 Neg). Dip Method: 100% positive and 100% negative conformity with professional results (49 Pos, 51 Neg). Lay person results for spiked samples showed 98-100% agreement with professional results. Questionnaires indicated ease of use and clear labeling.
Early Detection PerformanceDetect pregnancy early (e.g., 5 days before EMP)Detected 67.7% positive hCG five days before the Expected Menstrual Period (EMP). Detected 100% positive hCG on the day of EMP.
StabilityStable for a reasonable shelf life24 months at 35.6-86°F (based on real-time stability study).

2. Sample Sizes and Data Provenance

  • Analytical Performance (Precision/Reproducibility/Sensitivity):

    • Test Set Sample Size: For each hCG concentration (0, 3, 5, 7.5, 8, 10, 25, 50 mIU/mL), 10 replicates were tested per day for 5 days for each of 3 device lots, by 3 different operators. This sums to (8 concentrations * 10 replicates/day * 5 days * 3 lots) = 1200 tests for Midstream and another 1200 tests for Dip Testing.
    • Data Provenance: Not explicitly stated, but typically these are laboratory-controlled, prospective studies with spiked samples. The hCG standard is traceable to the 5th WHO.
  • Specificity (Non-pregnant females):

    • Test Set Sample Size: 300 urine samples (100 from pre-menopausal, 100 from peri-menopausal, 100 from post-menopausal women).
    • Data Provenance: Not explicitly stated, but implies prospectively collected urine samples from healthy non-pregnant females.
  • Method Comparison Study (vs. Predicate):

    • Test Set Sample Size: Urine samples from 200 women suspected to be pregnant in the early stage (<5 weeks).
    • Data Provenance: Collected from women presenting to test for pregnancy, implying a prospective, clinical setting, likely in an unspecified country (as no country of origin is mentioned).
  • Lay Person Study:

    • Test Set Sample Size: 200 women for self-testing. For spiked samples, 200 lay persons tested the samples.
    • Data Provenance: Individuals with varying educational and occupational backgrounds from three sites. This is a prospective, user study.
  • Early Pregnancy Test Study:

    • Test Set Sample Size: 650 urine samples from 65 pregnant women (65 characterized cycle segments of conceptive cycles).
    • Data Provenance: Collected from pregnant women. This is a prospective, clinical study.

3. Number of Experts and Qualifications for Ground Truth for Test Set

  • Analytical Performance: The ground truth for hCG concentrations was established by spiking negative urine with a known hCG standard (Traceable to the 5th WHO). No human experts were involved in establishing this ground truth, as it's an analytical measurement.
  • Specificity (Non-pregnant females): The ground truth was based on the healthy, non-pregnant status of the female participants. No specific experts or their qualifications for this judgment are mentioned, but it's assumed to be based on medical history or clinical assessment.
  • Method Comparison Study: The ground truth was the result obtained from the predicate device (Wondfo One Step HCG Urine Pregnancy Test Midstream/Strip/Cassette, K150022). No human experts were involved in establishing a separate ground truth.
  • Lay Person Study (vs. Professional Results): The "Professional Result" served as the ground truth. The qualifications of these "professionals" are not specified.
  • Early Pregnancy Test Study: The "Expected Menstrual Period (EMP)" and "pregnant women" served as the ground truth, implying clinical determination of pregnancy and cycle dating. No specific experts or their qualifications for this determination are mentioned.

4. Adjudication Method for the Test Set

  • No explicit adjudication method (like 2+1, 3+1) is mentioned for any of the studies.
  • For analytical performance, the ground truth was established by precise spiking of known concentrations.
  • For method comparison and lay person studies, direct comparison was made to the predicate device or "professional results" without specifying expert adjudication process if discrepancies arose.
  • For the early pregnancy test study, the ground truth seems to be clinical data (EMP, confirmed pregnancy) rather than expert review of the test results themselves.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance.
  • This device is an Over-The-Counter (OTC) pregnancy test, where the "reader" is typically the lay user interpreting a digital display. The studies focused on the device's standalone performance, comparison to a predicate, and lay user comprehension/accuracy.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Yes, a standalone study was done. The entire analytical performance section (sensitivity, hook effect, cross-reactivity, interfering substances) and the early pregnancy test study evaluated the device's technical performance to detect hCG and display a result ("Pregnant" or "Not Pregnant") without human interpretation beyond reading the digital display.
  • The "Professional Results" in the Lay Person study also reflect the device's standalone performance when used by trained individuals.

7. Type of Ground Truth Used

  • Analytical Performance: Known concentrations of hCG spiked into negative urine (traceable to WHO International Standard). This is an analytical gold standard.
  • Specificity: Healthy, non-pregnant status of study participants. This is an outcome/status-based ground truth related to the absence of the target analyte.
  • Method Comparison: Results from a legally marketed predicate device.
  • Lay Person Study: "Professional Results" using the Shinetell Plus™ device. This indicates a reference test ground truth, where the "professionals" are presumably highly accurate users.
  • Early Pregnancy Test Study: Clinically determined pregnancy and cycle dating relative to the Expected Menstrual Period (EMP). This is outcomes data/clinical status ground truth.

8. Sample Size for the Training Set

  • The document does not explicitly state a training set size or methodology for the Shinetell Plus™ Digital Early Pregnancy Test. This is common for this type of immunoassay device, which relies on established chemical and optical detection principles rather than machine learning algorithms that typically require large training datasets. The device's "learning" is inherent in its design and manufacturing.
  • The testing described is primarily for validation (verification and validation), not for training a predictive model.

9. How the Ground Truth for the Training Set Was Established

  • As a training set is not explicitly mentioned or implied for an AI/machine learning model, this question is not directly applicable based on the provided document. The device operates on an immunoassay principle with "light reflection" and "photodiode" detection, suggesting a pre-programmed algorithm based on known HCG detection thresholds, rather than a learned model.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

July 15, 2025

Hangzhou AllTest Biotech Co., Ltd.
℅ Jenny Xia
Director
LSI International Inc.
504 E Diamond Ave., Suite H
Gaithersburg, Maryland 20877

Re: K251053
Trade/Device Name: Shinetell Plus™ Digital Early Pregnancy Test
Regulation Number: 21 CFR 862.1155
Regulation Name: Human Chorionic Gonadotropin (HCG) Test System
Regulatory Class: Class II
Product Code: LCX
Dated: June 3, 2025
Received: June 3, 2025

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251053 - Jenny Xia
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251053 - Jenny Xia
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph Kotarek
Branch Chief, Toxicology Branch
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251053

Device Name: Shinetell Plus™ Digital Early Pregnancy Test

Indications for Use (Describe):
Shinetell Plus™ Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

Type of Use (Select one or both, as applicable):

  • Prescription Use (Part 21 CFR 801 Subpart D)
  • Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) SUMMARY

K251053

1. Date: June 29, 2025

1. Submitter:
Hangzhou AllTest Biotech Co., Ltd.
No. 550 Yinhai Street
Hangzhou, China

2. Contact person:
Jenny Xia
LSI International Inc.
504 East Diamond Ave., Suite H
Gaithersburg, MD 20877
Telephone: 301-525-6856
Fax: 301-916-6213
Email: jxia@lsi-consulting.org

3. Device Name: Shinetell Plus™ Digital Early Pregnancy Test

Classification: Class II
Product Code: LCX
CFR: 21 CFR 862.1155

4. Predicate Devices:
Wondfo One Step HCG Urine Pregnancy Test Midstream, Wondfo One Step HCG Urine Pregnancy Test Strip, Wondfo One Step HCG Urine Pregnancy Test Cassette, K150022

5. Intended Use

Shinetell Plus™ Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

6. Device Description

Shinetell Plus™ Digital Early Pregnancy Test is used for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine, and is designed to be tested in dip or midstream mode. The test device consists of a single test strip assembled in a plastic housing, with an absorbent tip. The device is in a ready-to-use format.

Shinetell Plus™ Digital Early Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. The test would detect the light intensity by using the LED as the light source. After that, the result can be displayed on the display screen.

Page 6

7. Substantial Equivalence Information

Similarities

ItemCandidate devicePredicate device
Intended useA rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.Same
SpecimenUrineUrine
Assay technicalImmunochromatographic assayImmunochromatographic assay
Sensitivity10 mIU/mL10 mIU/mL
ResultsQualitativeQualitative
Target userOver the counter useOver the counter use

Differences

ItemDevicePredicate
ReadoutDigital/LCD screenVisual Testing Line
Time to result3 minutes5 minutes
FormatMidstreamStrip, cassette, midstream

8. Test Principle

Human chorionic gonadotropin (HCG) is a glycoprotein produced by the placenta during pregnancy. Shinetell Plus™ Digital Early Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. After the appropriate urine sample is added to the absorbent tip, the clock symbol will appear and blink on the Display Screen after urine is applied, indicating the test is working. The HCG in urine specimen will react with the anti-β-HCG antibody-colloidal gold conjugate and form a compound. As the liquid flows to the Test area of the test strip, the compound will be captured by the anti-α-HCG antibody immobilized on the Test area, then a colored line will be formed on the Test area.

The test would detect the light intensity by using the LED as the light source. The

Page 7

background of the LED lighted zone will be darker due to the color of colloidal gold conjugate, and the intensity of the reflected light received by the photodiode decreases significantly. By detecting the signal parameter of the reflected light intensity, which is decreasing significantly, the test can determine that the sample is added properly. The LED light will irradiate the test strip and then be reflected to the photodiode and the photodiode can respond to the different light intensity value by testing the intensity of reflected light.

In approximately 3 minutes, the result will be shown on the display screen. If test value is over the preset threshold, the test result is positive and "Pregnant" is displayed on the screen. Otherwise, the result is negative and "Not Pregnant" is displayed on the screen.

If the sample is added improperly, the background color of the LED lighted zone does not decrease significantly. By detecting the signal parameter of the reflected light intensity, which isn't decreasing significantly, the test can determine that the sample is added improperly, and the test will be invalid and "?" is displayed on the screen.

9. Performance Characteristics

A. Analytical performance

a. Precision/Reproducibility/Sensitivity

Negative female urine was spiked with hCG standard (Traceable to the 5th WHO) to hCG concentrations of 0, 3, 5, 7.5, 8, 10, 25 and 50 mIU/mL. Each sample was tested by both dip and midstream methods in 10 replicates per day for 5 days for each device lot. Total of three device lots were tested. Tests were performed by three different operators for each sample concentration.

The results are summarized in the table below:

Midstream Testing

hCG Concentration (mIU/mL)Lot 1Lot 2Lot 3Total result% Negative% Positive
-+-+-+-+
050050050015001000
350050050015001000
550050050015001000
7.53515351536141064470.729.3
8252524262426737748.751.3
1005005005001500100
2505005005001500100
5005005005001500100

Dip Testing

Page 8

hCG Concentration (mIU/mL)Lot 1Lot 2Lot 3Total result% Negative% Positive
-+-+-+-+
050050050015001000
350050050015001000
550050050015001000
7.53515361436141074371.328.7
824262426242672784852
1005005005001500100
2505005005001500100
5005005005001500100

Shinetell Plus™ Digital Early Pregnancy Test exhibited reproducible results. Based on the above results, the sensitivity of Shinetell Plus™ Digital Early Pregnancy Test is demonstrated to be 10 mIU/mL.

b. Linearity/assay reportable range:

Linearity is not applicable since this is a qualitative test.

c. Hook effect test:

Negative urine samples were spiked with varying hCG concentrations (6,250 mIU/mL, 12,500 mIU/mL, 25,000 mIU/mL, 50,000 mIU/mL, 100,000 mIU/mL, 200,000 mIU/mL and 500,000 mIU/mL). All tested concentrations gave a positive result. The results demonstrated that no hook effect was observed at hCG concentration up to 500,000 mIU/mL.

d. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability:
Shinetell Plus™ Digital Early Pregnancy Test is calibrated against reference material traceable to WHO International Standard 5th edition, NIBSC code 07/364.

Stability:
Products in sealed foil pouch are stable for 24 months at 35.6-86°F, based on the real time stability study.

e. Specificity and cross reactivity

To evaluate specificity, 300 urine samples were collected from healthy, non-pregnant female in pre-menopausal (ages 1840 years old), peri-menopausal (4155 years old) and post-menopausal (>55 years old) groups. 100 people for each age group. Both dip and midstream testing are evaluated. No false positive results were observed for any of the age groups.

Page 9

To evaluate cross-reactivity, negative and positive urine samples (0, 5 and 10 mIU/mL hCG) were spiked with potential cross reactants (1000 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 μIU/mL hTSH). No cross-reactivity was observed at tested concentration.

To evaluate the effect of the hCG β-core fragment, Negative urine samples (0 and 5 mIU/mL hCG) and positive urine samples (10 and 20,000 mIU/mL hCG) were spiked with hCG β-core fragment (hCGβcf) at concentrations of 50,000 pmol/L, 125,000 pmol/L, 250,000pmol/L and 500,000pmol/L. The performance of Shinetell Plus™ Digital Early Pregnancy Test is not affected by hCG β-core fragment concentrations up to 500,000 pmol/L.

f. Interfering substance

To evaluate potential interferers with Shinetell Plus™ Digital Early Pregnancy Test, urine samples containing 0, 5 and 10 mIU/mL hCG were spiked with the interfering substance to obtain the certain desired test concentration. No interference effect was observed at the tested concentration shown in table below:

SubstanceConcentration
Acetaminophen20 mg/dL
Acetylsalicylic acid20 mg/dL
Ascorbic acid20 mg/dL
Atropine20 mg/dL
Caffeine20 mg/dL
Gentisic acid20 mg/dL
Glucose2 g/dL
Hemoglobin20 mg/dL
Tetracycline20 mg/dL
Ampicillin20 mg/dL
Albumin20 mg/dL
β-hydroxybutyrate2000 mg/dL
Ephedrine20 mg/dL
Phenylpropanolamine20 mg/dL
Phenothiazine20 mg/dL
EDTA80 mg/dL
Salicyclic Acid20 mg/dL
Benzoylecgonine10 mg/dL
Cannabinol10 mg/dL
Codeine6ug/dL
Ethanol1.0%
Bilirubin2mg/dL
Pregnanediol1500μg/dL
Thiophene20 mg/dL
Ketone20 mg/dL

Page 10

To evaluate the effect of urine pH on the results of Shinetell Plus™ Digital Early Pregnancy Test, urine samples containing 0, 5 and 10 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. The results indicated that urine pH ranges between 4 and 9 does not affect the performance of Shinetell Plus™ Digital Early Pregnancy Test.

To evaluate the effect of urine density on the results of Shinetell Plus™ Digital Early Pregnancy Test, urine samples containing 0, 5 and 10 mIU/mL hCG were tested at density values of 1.000, 1.005, 1.011, 1.015, 1.019, 1.024, 1.029 and 1.035. The results indicated that urine with a relative density of 1.000 to 1.035 does not affect the performance of Shinetell Plus™ Digital Early Pregnancy Test.

B. Method comparison study

Method comparison with predicate device
The performance of the new device was compared to the predicate test. Urine samples were collected from 200 women presenting to test for pregnancy. All of these women were suspected to be pregnant in the early stage of less than 5 weeks. All samples were tested with candidate and predicate devices at three POC sites.

Dip Testing

Pregnancy Test (Shinetell Plus™)Predicate Test (Wondfo test)Total
Positive (+)Negative (-)
Positive (+)49049
Negative (-)05151
Total4951100

Midstream Testing

Pregnancy Test (Shinetell Plus™)Predicate Test (Wondfo test)Total
Positive (+)Negative (-)
Positive (+)48048
Negative (-)05252
Total4852100

The conformity between Shinetell Plus™ Digital Early Pregnancy Test and the predicate device is 100%.

C. Lay person study

200 women's individual pregnancy status was self-tested using either in-stream method or dip method. Individuals with varying educational and occupational

Page 11

backgrounds from three sites were chosen for the study. Each subject tested her own urine sample using the device according to the package insert and provided a sample for professional testing.

Summary

In-Stream Method

Lay user ResultProfessional ResultTotal
PositiveNegative
Positive48048
Negative05252
Total4852100

Dip Method

Lay user ResultProfessional ResultTotal
PositiveNegative
Positive49049
Negative05151
Total4951100

From the above tables, the lay person results showed 100% positive and 100% negative conformity with the professional results.

Spiked urine samples were also tested by lay person. Urine samples were prepared at 5mIU/ml, 7.5mIU/ml, 8mIU/ml and 10mIU/ml hCG concentrations by spiking hCG into negative pooled urine specimens. Each sample was aliquoted into individual containers and blind-labeled. These samples were tested by 200 lay persons.

Lay person vs Professional

hCG ConcentrationsLay person resultsProfessional resultsPercent No. of Agreement
No. of NegativeNo. of PositiveNo. of NegativeNo. of Positive
5 mIU/ml10001000100%
7.5 mIU/ml7129693198%
8 mIU/ml5050524898%
10 mIU/ml01000100100%

Each lay person was given a questionnaire to assess the readability of the labeling. The results of the questionnaire reflected that the consumers found the test easy to use and that they did not have trouble understanding the labeling and interpreting

Page 12

the results.

D. Early Pregnancy Test Study

In this study, total 650 urine samples from 65 characterized cycle segments of conceptive cycles were collected from 65 pregnant women. All samples were masked and randomized. Each sample was tested both in-stream and dip methods using three lots of the device. The new device detected 67.7% positive hCG five days before the Expected Menstrual Period (EMP), and 100% positive hCG on the day of EMP. No differences were observed between different test methods. The following table is the summary of the data.

Day relative to Expected Menstrual Period(EMP)Number of PositiveNumber of NegativeNumber of Total% Positive
EMP-8362654.6%
EMP-78576512.3%
EMP-623426535.3%
EMP-544216567.7%
EMP-45696586.2%
EMP-36236595.4%
EMP-26416598.5%
EMP-165065100.0%
EMP65065100.0%
EMP+165065100.0%

10. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the device, it's concluded that Shinetell Plus™ Digital Early Pregnancy Test is substantially equivalent to the predicate.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.