K Number
K251042

Validate with FDA (Live)

Date Cleared
2025-06-06

(64 days)

Product Code
Regulation Number
878.4810
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • Blue light: Treatment of mild to moderate inflammatory acne.
  • Red+IR light: Treatment of facial wrinkles.
  • Amber light: Treatment of facial wrinkles.
  • Mixed light: Treatment of mild to moderate inflammatory acne.
Device Description

The Light Therapy System is an Over-the-Counter (OTC), home-use, wearable light based device, and intended for the use in treating full-face wrinkles and mild to moderate inflammatory acne. Light radiates from the inner surface of the device onto the face. This light is generated by LED with four different spectrum wavelengths: blue (460nm), amber (605nm), red (630nm) and infrared (850nm), and the device has four light treatment, that is, blue light and mixed light mode to treat acne, while amber light mode and red+ IR mode to treat facial wrinkles.

The device is powered by external power supply by connecting with the power adapter and controlled by the controller that should be connected to the main unit. The device's power-on/off, mode switch and treatment time adjustment can be operated by the controller. The device will automatically shut down when the treatment time is over.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the "Light Therapy System (M500, L6)" does not contain information regarding a clinical study or performance data that would typically include acceptance criteria for clinical effectiveness.

The document focuses heavily on non-clinical testing for safety and electrical performance, and a comparison of technological characteristics with predicate devices to establish substantial equivalence.

Therefore, based solely on the provided text, it is not possible to describe the acceptance criteria and the study that proves the device meets the acceptance criteria in the manner requested (clinical effectiveness).

The "Conclusions" section on page 12 states: "Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device Light Therapy System is as safe, as effective, and performs as well as the legally marketed predicate devices."

This statement regarding "effectiveness" is likely based on the technological characteristics comparison with predicate devices (Section VI) rather than a direct clinical performance study of the subject device. The clearance is based on substantial equivalence to predicate devices that have established clinical effectiveness, and the assumption is that similar technological characteristics (wavelengths, intensities, treatment times for similar indications) will yield similar clinical outcomes.

If information about clinical performance and acceptance criteria were present, it would typically be found under a "Clinical Testing" or "Performance Data" section that is absent in this excerpt for clinical efficacy.

To answer your specific points based on the absence of clinical study details in this document:


1. A table of acceptance criteria and the reported device performance

  • Cannot be provided from the given document for clinical performance. The document lists non-clinical performance (biocompatibility, electrical safety, EMC, software validation) and states they "passed," implying acceptance criteria were met for these aspects, but no specific metrics are given beyond "passed."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable for clinical performance based on this document. No clinical test set or data provenance is detailed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable for clinical performance based on this document. No experts or ground truth establishment for a clinical test set are described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable for clinical performance based on this document. No adjudication method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a light therapy system, not an AI-assisted diagnostic device, and no MRMC study is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a light therapy system, not an algorithm. Its performance is inherent to its physical light emission.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable for clinical performance based on this document. For the non-clinical tests, the ground truth is derived from the "passed" status against established engineering and biological safety standards (e.g., IEC standards, ISO standards).

8. The sample size for the training set

  • Not applicable for clinical performance based on this document. This applies to machine learning models, not a light therapy device's direct clinical performance study for a 510(k) substantial equivalence.

9. How the ground truth for the training set was established

  • Not applicable for clinical performance based on this document. (See point 8).

Summary of what the document does provide regarding "acceptance criteria" and "study":

The "study" in this context refers to non-clinical testing and comparison to predicate devices to demonstrate safety and substantial equivalence.

Acceptance Criteria & Reported Performance (Non-Clinical):

Non-Clinical Test CategoryAcceptance Criteria (Implied: Pass)Reported Performance
Biocompatibility TestingConformity to ISO 10993-1, 10993-5, 10993-10, 10993-23 standards."Passed" for all listed tests (In Vitro Cytotoxicity, Skin Sensitization, Irritation).
Electrical Safety & EMCConformity to IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-83, IEC 62471 standards."Passed" for all listed standards.
Software Verification & ValidationSoftware documentation consistent with Basic Documentation level of concern. All software requirement specifications met. All software hazards mitigated to acceptable risk levels."Demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels."

Data Provenance (Non-Clinical):

  • These tests are typically conducted by the manufacturer or third-party labs on prototypes/production samples. The country of origin for these tests is not specified in the document, but the manufacturer is based in China. This is "prospective" in the sense that the tests were performed on the device intended for clearance.

Type of Ground Truth (Non-Clinical):

  • The ground truth for non-clinical tests is established by international standards (ISO, IEC) and regulatory guidance documents (FDA recognized standards). Testing laboratories verify compliance with these predefined standards.

The clearance is based on the argument that the device's technological characteristics are substantially equivalent to legally marketed predicate devices, which themselves would have demonstrated efficacy through previous clinical data or established scientific principles. The non-clinical tests confirm the current device's safety and performance align with those precedents.

FDA 510(k) Clearance Letter - Light Therapy System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 6, 2025

Guangzhou Ahead Intelligent Technology Co., Ltd.
℅ Riley Chan
Registration engineer
Feiying Drug & Medical Consulting Technical Service Group
Rm 2401 Zhenge International Business Center, No. 3101-90
Qianhai Road
Shenzhen, Guangdong 518052
China

Re: K251042
Trade/Device Name: Light Therapy System (M500, L6)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: March 31, 2025
Received: April 3, 2025

Dear Riley Chan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K251042 - Riley Chan Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

Page 3

K251042 - Riley Chan Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by TANISHA L. HITHE -S
Date: 2025.06.06 13:34:52 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K251042

Please provide the device trade name(s).
Light Therapy System (M500, L6)

Please provide your Indications for Use below.

  • Blue light: Treatment of mild to moderate inflammatory acne.
  • Red+IR light: Treatment of facial wrinkles.
  • Amber light: Treatment of facial wrinkles.
  • Mixed light: Treatment of mild to moderate inflammatory acne.

Please select the types of uses (select one or both, as applicable).
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

Light Therapy System Page 9 of 41

Page 5

K251042 - 510(k) Summary

"510(k) Summary" as required by 21 CFR Part 807.92.

I. Submitter

Guangzhou Ahead Intelligent Technology Co., Ltd.
Room 505, Building 2, No.10, Industrial 1st Road, Zhicun, Dashi Street, Panyu District, Guangzhou, Guangdong, China
Post code: 511430

Leo Zheng
General manager
Tel: +86 18664649916
Email: 187308650@qq.com

Prepared date: 2025-6-4

II. Device

Name of Device: Light Therapy System
Model(s): M500, L6
Common or Usual Name: Light Based Over The Counter Wrinkle Reduction
Over-The-Counter Powered Light Based Laser For Acne
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHS, OLP
Regulation Number: 21 CFR 878.4810

III. Predicate Device & Reference Device

Predicate device

ManufacturerTrade Name510(k) NumberCleared Date
Shenzhen SUNGPO HI-TECH Electronic Co.,LtdLED Facial Mask/ MZ-01, NEWKEY01, SP-FM-01K230351May 5, 2023
Hunan Guangye Biotechnology Co., Ltd.Beauty LED Mask/ Model:KFB265K221151July 19, 2022
Guangdong Newdermo Biotech Co.,LtdLED light therapy mask (FM-01, FM-02, FM-03)K223544Feb. 23, 2023
Biophotas IncBiophotas Celluma3K171323Sep. 1, 2017
Ningbo Hesi Electric Co., LtdFlexible LED light therapy (Model: HK209)K200983July 14, 2021
Shenzhen Aozemei Technology Co., LTDMicro-current Facial Beauty DeviceK241718Oct. 28, 2024

Reference device

Page 6

ManufacturerTrade Name510(k) NumberCleared Date
Marci Beauty IncInfrared Red Blue LED Light Heat Beauty MachineK210545May 20, 2022
Shenzhen Borria Technology Co., LtdLED Therapy Mask (MN1, M226)K242385April 7, 2025
Galactic Beauty, LLCMMSphereK190443June 24, 2019

IV. Device Description

The Light Therapy System is an Over-the-Counter (OTC), home-use, wearable light based device, and intended for the use in treating full-face wrinkles and mild to moderate inflammatory acne. Light radiates from the inner surface of the device onto the face. This light is generated by LED with four different spectrum wavelengths: blue (460nm), amber (605nm), red (630nm) and infrared (850nm), and the device has four light treatment, that is, blue light and mixed light mode to treat acne, while amber light mode and red+ IR mode to treat facial wrinkles.

The device is powered by external power supply by connecting with the power adapter and controlled by the controller that should be connected to the main unit. The device's power-on/off, mode switch and treatment time adjustment can be operated by the controller. The device will automatically shut down when the treatment time is over.

V. Indications for Use

  • Blue light: Treatment of mild to moderate inflammatory acne.
  • Red+IR light: Treatment of facial wrinkles.
  • Amber light: Treatment of facial wrinkles.
  • Mixed light: Treatment of mild to moderate inflammatory acne.

VI. Comparison of Technological Characteristics With the Predicate Device

The Light Therapy System has the same intended use as the predicate devices. The technological characteristics such as wavelength, intensity, are similar to the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the Light Therapy System may be found substantially equivalent to its predicate devices.

Table1: Substantial Equivalence Comparison for M500

Page 7

ItemsSubject devicePrimary Predicate devicePredicate device 1Predicate device 2Predicate device 3Reference device 1Reference device 2Difference discussion
K numberK251042K230351K221151K223544K241718K210545K190443/
Device nameLight Therapy System (Model: M500)LED Facial Mask/ MZ-01, NEWKEY01, SP-FM-01Beauty LED Mask/ Model: KFB265LED light therapy mask (FM-01, FM-02, FM-03)Micro-current Facial Beauty Device, Model: AM-810B, AM-810W, AM-812B, AM-812WInfrared Red Blue LED Light Heat Beauty MachineMMSphere™/
Product CodeOHS, OLPOHS, OLPOHS, OLPOHS, OLP, ILYOHS, OLPOHS, OLPOHS, OLPSame
Intended use/ Indications for UseBlue light: Treatment of mild to moderate inflammatory acne. Red+IR light: Treatment of facial wrinkles. Amber light: Treatment of facial wrinkles. Mixed light: Treatment of mild to moderate inflammatory acne.LED Facial Mask is an over the counter device that is intended to use LED light for the treatment of wrinkles and mild to moderate acne.The device is intended to use LED light for the treatment of wrinkles and mild to moderate acne.Red light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne.Micro-current Facial Beauty Device is intended for the treatment of facial wrinkles, and mild to moderate inflammatory acne.The Infrared Red Blue LED Light Heat Beauty Machine is an hand-held over-the-counter phototherapy device, the red and infrared light is intended for the use in treating wrinkles on the face and the blue light is intended for the treatment of the mild to moderate inflammatory acne. This device is indicated for adults only.MMSphere™ Light Therapy Device emits energy in the red, blue and amber regions of the spectrum, specifically indicated to treat wrinkles and/or mild to moderate acne. The MMSphere™ is designed to be used for 20 minute treatments three to seven times per week.Same
Prescription /OTCOTCOTCOTCOTCOTCOTCOTCSame
Software/Firmware/Microprocessor Control?YesYesYesYesYesYesYesSame
Power supplyExternal adapter 100-240 V ~ 50/60 HzExternal adapter Input:AC100-240V 50-60Hz 0.2A Output: DC 12V 0.5AInput: 100-240 V~, 50/60 Hz Output: 5V 1AInput: 100-240V~, 50/60Hz, 0,25A Output: DC 5 V, 500 mAAdapter input: 5V 0.5A Internal battery: 3.7V/600mAh900mAh, Rechargeable Li-Ion batteriesUnknownSame

Page 8

ItemsSubject devicePrimary Predicate devicePredicate device 1Predicate device 2Predicate device 3Reference device 1Reference device 2Difference discussion
Shape designM500: MaskMaskMaskMaskHand-held TypeHand-held TypeHandheld and stationarySimilar
Intended location of useFaceFaceFaceFace and bodyFaceEntire FaceFaceSame
Energy typeLight emitting diodes (LEDs)Light emitting diodes (LEDs)Light emitting diodes (LEDs)LEDLEDLEDLEDSame
WavelengthBlue: 460nm±5nm Red: 630nm±5nm Amber: 605nm±5nm Near-infrared: 850nm±5nmBlue: 465nm±5nm Red: 625nm±5nm Amber: 605nm±5nmRed (637nm±5nm)and IR (854nm±5nm); Blue (465±5nm)Red: 620nm Blue: 460nm Infrared: 850nm Mixed: 620nm and 850nm and 460nmBlue: 415±10nm Amber: 605±10nm Red: 630±10nmBlue: 465nm Red: 620-630nm IR: 845-855nm605nm 625nm 465nmSimilar
Intensity (mW/cm²)M500: Blue: 40mW/cm² Amber: 10mW/cm² Red+IR: 20mW/cm² Mixed mode: 12mW/cm²Blue: 1563mW/cm2 Red: 3175mW/cm2 Amber: 15~42mW/cm²Red+IR: 25.5mW/cm² Blue: 1.36mW/cm²Red light: 2.03.0mW/cm2 Blue light: 2.04.0mW/cm2 Infrared light: 2.04.0mW/cm2 Mixed light: 9.012.0mW/cm2Red: 2.5mW/cm2 Amber: 15mW/cm2 Blue: 1.4mW/cm2Blue light: 5.4mW/cm2 Red+IR light: 7.2mW/cm2Red 2.45mW/cm2 Blue 1.33mW/cm2 Amber: unpublishedSimilar
Treatment time10 minutes each time, 3~5 times a week10 minutes/day, 3 times per week10min each timeManual Mode: 15minutes each time. Automatic Mode: 10minutes each time. 3-4 treatment a week, reduce to 1-2 treatment a week once the results shown.3-5 minutes at a time /2-3 times a week.3 times a week for 30 min. 4weeks20mins/day, 120daysSimilar
Main materialsNylon+spandex, PVC leather, ABS plasticABS and PC plasticUnknownUnknownABS, PCABS Plastic and Aluminum HeadUnknownDifferent, but solved by biocompatibility test

Page 9

Table 2: Substantial Equivalence Comparison for L6

ItemsSubject devicePrimary Predicate devicePredicate device 1Predicate device 2Predicate device 3Predicate device 4Reference device 1Reference device 2Difference discussion
K numberK251042K171323K230351K200983K223544K241718K242385K190443/
Device nameLight Therapy System (Model: L6)Biophotas Celluma3LED Facial Mask/ MZ-01, NEWKEY01, SP-FM-01Flexible LED light therapy (Model: HK209)LED light therapy mask (FM-01, FM-02, FM-03)Micro-current Facial Beauty Device, Model: AM-810B, AM-810W, AM-812B, AM-812WLED Therapy Mask (MN1, M226)MMSphere™/
Product CodeOHS, OLPOHSOHS, OLPOHS, OLP, ILYOHS, OLP, ILYOHS, OLPOHS, OLP, ILYOHS, OLPSame
Intended use/ Indications for UseBlue light: Treatment of mild to moderate inflammatory acne. Red+IR light: Treatment of facial wrinkles. Amber light: Treatment of facial wrinkles. Mixed light: Treatment of mild to moderate inflammatory acne.The BioPhotas Celluma3 is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full face wrinkles.LED Facial Mask is an over the counter device that is intended to use LED light for the treatment of wrinkles and mild to moderate acne.The FLEXIBLE LED LIGHT THERAPY (Model: HK209) is intended to: - The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris. -The device emitting energy in the red and infrared spectrum is intended for the treatment of full-face wrinkles -The device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation ofRed light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne.Micro-current Facial Beauty Device is intended for the treatment of facial wrinkles, and mild to moderate inflammatory acne.Red light: Treatment of full-face wrinkles. Blue light (only suitable model M226 and MN1 LED Facial Mask): Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.MMSphere™ Light Therapy Device emits energy in the red, blue and amber regions of the spectrum, specifically indicated to treat wrinkles and/or mild to moderate acne. The MMSphere™ is designed to be used for 20 minute treatments three to seven times per week.Same

Page 10

ItemsSubject devicePrimary Predicate devicePredicate device 1Predicate device 2Predicate device 3Predicate device 4Reference device 1Reference device 2Difference discussion
Prescription /OTCOTCOTCOTCOTCOTCOTCOTCOTCSame
Software/Firmware/Microprocessor Control?YesYesYesYesYesYesYesYesSame
Power supplyExternal adapter 100-240 V ~ 50/60 HzAn AC adaptor for 110-120VExternal adapter Input:AC100-240V 50-60Hz 0.2A Output: DC 12V 0.5AInput: 100–240Vac, 2.0A, 50/60HzInput: 100-240V~, 50/60Hz, 0,25A Output: DC 5V, 500mAAdapter input: 5V 0.5A Internal battery: 3.7V/600mAhInput: DC 5V, 1A Built-in rechargeable lithium battery: DC 3.7V 2500mAhUnknownSame
Shape designL6: PanelPanelMaskPanelMaskHand-held TypeMaskHandheld and stationarySimilar
Intended location of useFaceWhole faceFaceEntire Face and bodyFace and bodyFaceFace and neckFaceSame
Energy typeLight emitting diodes (LEDs)Light emitting diodes (LEDs)Light emitting diodes (LEDs)LEDLEDLEDLEDsLEDSame
WavelengthBlue: 460nm±5nm Red: 630nm±5nm Amber: 605nm±5nm Near-infrared: 850nm±5nmRed:640nm+/-25nm NIR:880nm+/-50nmBlue: 465nm±5nm Red: 625nm±5nm Amber: 605nm±5nm465nm, 640nm, 880nmRed: 620nm Blue: 460nm Infrared: 850nm Mixed: 620nm and 850nm and 460nmBlue: 415±10nm Amber: 605±10nm Red: 630±10nmRed: 630nm±5nm Blue: 470nm±5nm Near-Infrared: 850nm±5nm Mixed light: 630nm and 850nm605nm 625nm 465nmSimilar
Intensity (mW/cm²)At 1cm: Blue: 1.63.0mW/cm² Amber: 10.511.5mW/cm² Red+IR: 1.12.0mW/cm² Mixed: 1011mW/cm² At 5cm: Blue: 1.32.5mW/cm² Amber: 10.011.0mW/cm² Red+IR: 0.91.5mW/cm² Mixed: 910mW/cm²6.5mW/cm²Blue: 1563mW/cm² Red: 3175mW/cm² Amber:15~42mW/cm²6.5mW/cm²Red light: 2.03.0mW/cm² Blue light: 2.04.0mW/cm² Infrared light: 2.04.0mW/cm² Mixed light: 9.012.0mW/cm²Red: 2.5mW/cm² Amber: 15mW/cm² Blue: 1.4mW/cm²Red: 0.892.55W/cm² Blue: 1.444.09mW/cm² Near-Infrared: 1.833.05mW/cm² Mixed light: 0.952.64mW/cm²Red 2.45mW/cm² Blue 1.33mW/cm² Amber: unpublishedSimilar

Page 11

ItemsSubject devicePrimary Predicate devicePredicate device 1Predicate device 2Predicate device 3Predicate device 4Reference device 1Reference device 2Difference discussion
The distance between the LEDs to treatment surface1~5cmAs closed to the skinAs closed to the skin10cmAs closed to the skinAs closed to the skinAs closed to the skinUnknownSimilar Note 4
Treatment time10 minutes each time, 3~5 times a week3 treatments per week (1800 seconds) 4 weeks10 minutes/day, 3 times per week3 times a week for 30 min. 4 weeksManual Mode: 15minutes each time. Automatic Mode: 10minutes each time. 3-4 treatment a week, reduce to 1-2 treatment a week once the results shown.3-5 minutes at a time /2-3 times a week.10 minutes each time20mins/day, 120 daysSimilar
Main materialsNylon+spandex, PVC leather, ABS plasticUnknownABS and PC plasticUnknownUnknownABS, PCUnknownUnknownDifferent, but solved by biocompatibility test

Page 12

VII. Non-Clinical Testing

The following performance data were provided in support of the substantial equivalence determination.

1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the Light Therapy System was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process', Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including:

  • ISO 10993-5: 2009, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity
  • ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
  • ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation

2) Electrical Safety and EMC

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

  • IEC 60601-1:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2:2020, Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility
  • IEC 60601-1-11:2020, Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
  • IEC 60601-2-83: 2022, Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
  • IEC 62471: 2006, Photobiological safety of lamps and lamp systems

3) Software Verification and Validation

Software documentation consistent with Basic Documentation level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

VIII. Conclusions

Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device Light Therapy System is as safe, as effective, and performs as well as the legally marketed predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.