K Number
K251015

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-04-30

(28 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APRO 45 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

The Alembic Aspiration Tubing is intended to connect the APRO 45 Catheter to the aspiration pump.

Device Description

The APRO 45 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 45 Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 45 Catheter targets aspiration from a suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 45 Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 45 Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 45 Catheter is available in lengths of 115, 125, 146, and 160 cm and is provided with an introducer sheath. The APRO 45 Catheter contains animal derived materials made from tallow derivatives.

The Alembic Aspiration Tubing connects the APRO 45 Catheter to the aspiration pump. The flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

AI/ML Overview

The provided document is a 510(k) clearance letter and summary for the APRO 45 Catheter and Alembic Aspiration Tubing. It primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. As such, it does not contain information typically found in studies proving a device meets acceptance criteria related to clinical performance, such as sample sizes for test sets, data provenance, expert ground truth establishment, or multi-reader multi-case studies.

Therefore, many of the requested sections will state that the information is not available or not applicable based on the provided document.


Acceptance Criteria and Reported Device Performance

The study supporting this 510(k) clearance relied on non-clinical bench testing to demonstrate that the APRO 45 Catheter and Alembic Aspiration Tubing meet predefined specifications and perform as intended. The acceptance criteria and reported device performance are detailed in Table 2 (Summary of Non-Clinical Bench Test Results) of the 510(k) Summary.

Table of Acceptance Criteria and Reported Device Performance (from Table 2):

TestAcceptance CriteriaReported Device Performance (Conclusion)
Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications.Introducer Sheath meets the visual and dimensional specifications.The APRO 45 Catheter met the acceptance criteria.The Introducer Sheath met the acceptance criteria.
ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 45 Catheter met the acceptance criteria.
Vacuum IntegrityCatheter is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 45 Catheter met the acceptance criteria.
Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 45 Catheter met the acceptance criteria.
Catheter Hub LeakageCatheter does not leak air into hub assembly with methods specified in ISO 10555-1, Annex D.The APRO 45 Catheter met the acceptance criteria.
Hub CompatibilityCatheter meets the requirements specified in ISO 80369-7.The APRO 45 Catheter met the acceptance criteria.
Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 45 Catheter met the acceptance criteria.
Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 45 Catheter met the acceptance criteria.
Liquid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 45 Catheter met the acceptance criteria.
Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 45 Catheter met the acceptance criteria.
Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 45 Catheter met the acceptance criteria.
3-Point BendCatheter 3-Point Bend force must be acceptable. Forces were compared to a predicate.The APRO 45 Catheter met the acceptance criteria.
Simulated Use including Clot RetrievalWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications. Clot retrieval from various locations of the anatomical model was performed using the subject device system.The APRO 45 Catheter and Alembic Aspiration Tubing met the acceptance criteria.
Corrosion ResistanceCatheter shall show no signs of corrosion when tested in accordance with ISO 10555-1, Annex A.The APRO 45 Catheter met the acceptance criteria.

Additional Requested Information:

  1. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify exact sample sizes for each bench test conducted. It refers to "the subject device and predicate device" for comparisons, implying relevant samples were used for laboratory testing.
    • Data Provenance: The data is from non-clinical bench testing and simulated use in anatomical neurovascular models, conducted by Alembic, LLC. It is by nature prospective experimental data, not patient-derived.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This was a non-clinical bench study. The "ground truth" was established by predefined engineering specifications and international standards (e.g., ISO), not by expert human interpretation of medical data.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. As a non-clinical bench study, expert adjudication of medical results was not performed. Test results were assessed against predetermined engineering acceptance criteria.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. An MRMC comparative effectiveness study was not performed. This is a medical device (catheter and tubing), not an AI diagnostic algorithm, and its clearance was based on non-clinical performance and substantial equivalence to a predicate device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is a physical medical device, not an algorithm. Therefore, "standalone" performance in the context of AI algorithms is not relevant. The device's performance was evaluated through bench testing.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For bench tests, the "ground truth" was defined by established engineering and material specifications, international standards (e.g., ISO 10555-1, ISO 80369-7, ISO 11135, ISO 10993-1), and acceptable performance in simulated models (e.g., clot retrieval from an anatomical neurovascular model). This is analogous to a "reference standard" from engineering specifications rather than clinical ground truth like pathology or outcomes.
  7. The sample size for the training set:

    • Not Applicable. This submission is for a physical medical device. There is no "training set" in the context of machine learning. The device development process involves design, prototyping, and iterative testing, not AI model training.
  8. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for an AI algorithm mentioned, the establishment of ground truth for such a set is not relevant.

FDA 510(k) Clearance Letter - APRO 45 Catheter and Alembic Aspiration Tubing

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 30, 2025

Alembic, LLC
Stina Almaleh
Regulatory and Quality Engineer
627 National Avenue
Mountain View, California 94043

Re: K251015
Trade/Device Name: APRO 45 Catheter and Alembic Aspiration Tubing
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: NRY
Dated: April 1, 2025
Received: April 2, 2025

Dear Stina Almaleh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K251015 - Stina Almaleh
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K251015 - Stina Almaleh
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan -S

Naira Muradyan, Ph.D.
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K251015

Device Name: APRO 45 Catheter and Alembic Aspiration Tubing

Indications for Use (Describe)

The APRO 45 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

The Alembic Aspiration Tubing is intended to connect the APRO 45 Catheter to the aspiration pump.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

APRO® 45 Catheter and Alembic Aspiration Tubing 510(k) Summary

510(k) Number: K251015
Page 1 of 6

510(k) Summary

This 510(k) Summary is provided in accordance with the requirements of 21 CFR §807.92.

1) Submitter information

Submitter: Alembic, LLC
627 National Ave.
Mountain View, CA 94043

Contact: Stina Almaleh
Regulatory and Quality Engineer
Email: salmaleh@alembicllc.com
Telephone Number: (650) 388-5088

Date Prepared: April 28, 2025

2) Device Name and Classification

Trade/Proprietary Name: APRO® 45 Catheter and Alembic Aspiration Tubing
Common Name: Catheter, Thrombus Retriever
Classification Name: Percutaneous Catheter, 21 CFR 870.1250
Regulatory Class: Class II
Product Codes: NRY
Review Panel: Neurology

3) Legally Marketed Predicate Device

Predicate Device: K232971 APRO® 55 Catheter and Alembic Aspiration Tubing

4) Device Description

The APRO 45 Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 45 Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 45 Catheter targets aspiration from a suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 45 Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 45 Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 45 Catheter is available in lengths of 115, 125, 146, and 160 cm and is provided with an introducer sheath. The APRO 45 Catheter contains animal derived materials made from tallow derivatives.

The Alembic Aspiration Tubing connects the APRO 45 Catheter to the aspiration pump. The

Page 6

APRO® 45 Catheter and Alembic Aspiration Tubing
510(k) Number: K251015
Page 2 of 6

flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

5) Indications for Use

The APRO 45 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

The Alembic Aspiration Tubing is intended to connect the APRO 45 Catheter to the aspiration pump.

6) Technological Characteristics Comparison

Alembic has demonstrated that the APRO 45 Catheter and Alembic Aspiration Tubing are substantially equivalent to the predicate device APRO 55 Catheter and Alembic Aspiration Tubing based on the similarity in materials, similarity in design concept, and the same fundamental operating principles. A comparison of the APRO 45 Catheter with the predicate is summarized in Table 1 below. There are no changes to the Alembic Aspiration Tubing compared to that included as part of the K232971 submission.

Table 1. APRO 45 Catheter and Alembic Aspiration Tubing Comparison with the Predicate Device

CategorySubject DeviceAPRO 45 Catheter and Alembic Aspiration TubingPredicate DeviceAPRO 55 Catheter and Alembic Aspiration Tubing
510(k) NumberK251015K232971
Regulatory ClassIdentical to predicate.Class II, 21 CFR 870.1250, NRY
Indications for UseThe APRO 45 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.The Alembic Aspiration Tubing is intended to connect the APRO 45 Catheter to the aspiration pump.The APRO 55 Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t PA) or who fail IV t-PA therapy are candidates for treatment.The Alembic Aspiration Tubing is intended to connect the APRO 55 Catheter to the aspiration pump.
Principles of OperationSame as predicate.Using conventional catheterization techniques under fluoroscopic guidance, the catheter is advanced into the target vessel over an appropriate neurovascular guidewire. The catheter is positioned proximal to the thrombus to aspirate.

Page 7

APRO® 45 Catheter and Alembic Aspiration Tubing
510(k) Number: K251015
Page 3 of 6

Table 1. APRO 45 Catheter and Alembic Aspiration Tubing Comparison with the Predicate Device

CategorySubject DeviceAPRO 45 Catheter and Alembic Aspiration TubingPredicate DeviceAPRO 55 Catheter and Alembic Aspiration Tubing
Materials
HubSame as predicate.Nylon
AdhesiveSame as predicate.Cyanoacrylate
Strain ReliefSame as predicate.Santoprene (thermoplastic elastomer)
LinerPolytetrafluoroethylene (PTFE)/Pebax compositePTFE/Tecoflex composite
Shaft Coil or BraidSame as predicate.304V stainless steel braid304V stainless steel coil
ExtrusionsPolyether block amideThermoplastic polyurethanes, thermoplastic elastomer
Marker BandSame as predicate.Platinum/ iridium
CoatingSame as predicate.Hydrophilic coating
Dimensions
Proximal Outer Diameter0.060 inch0.066 inch
Distal Outer Diameter0.058 inch0.066 inch
Inner Diameter0.045 inch0.055 inch
Effective Lengths115, 125, 146, 160 cm125, 137 cm
Coated Length60 cm90, 102 cm
Tip ShapeSame as predicate.Straight
Accessories
Introducer SheathSame as predicate.Yes
Packaging Materials
PouchSame as predicate.Nylon/polyethylene/ Tyvek
Packaging TubeN/APolyethylene
Packaging Hoop CoilPolyethyleneN/A
Packaging CardSame as predicate.Polyethylene
Shelf CartonSame as predicate.Solid bleached sulfate paperboard
Other
SterilizationSame as predicate.Ethylene oxide
Shelf Life1 year6 months
Use ConditionsSame as predicate.Sterile, single use, disposable
Alembic Aspiration Tubing
Aspiration TubingSame as predicate.110-inch lengthTubing ID = 0.110 inch

7) Performance Data

Alembic performed non-clinical bench, sterility, shelf-life, and biocompatibility testing. The results demonstrate substantial equivalence of the subject APRO 45 Catheter and Alembic Aspiration Tubing to the legally marketed predicate device.

A. Design Verification Testing – Non-Clinical Bench

Bench performance testing was conducted to support the APRO 45 Catheter submission. The results of the design verification and validation testing confirm that the APRO 45 Catheter conforms to the predefined specifications and performs as intended. Included in Table 2 is a description of bench performance tests that were conducted to support

Page 8

APRO® 45 Catheter and Alembic Aspiration Tubing
510(k) Number: K251015
Page 4 of 6

substantial equivalence determination.

Table 2. Summary of Non-Clinical Bench Test Results

TestAcceptance CriteriaConclusion
Visual and Dimensional CharacteristicsCatheter meets the visual and dimensional specifications.Introducer Sheath meets the visual and dimensional specifications.The APRO 45 Catheter met the acceptance criteria.The Introducer Sheath met the acceptance criteria.
ParticulateCatheter meets the acceptance criteria. Subject device was evaluated with a predicate device under the same test conditions.The APRO 45 Catheter met the acceptance criteria.
Vacuum IntegrityCatheter is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 45 Catheter met the acceptance criteria.
Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 45 Catheter met the acceptance criteria.
Catheter Hub LeakageCatheter does not leak air into hub assembly with methods specified in ISO 10555-1, Annex D.The APRO 45 Catheter met the acceptance criteria.
Hub CompatibilityCatheter meets the requirements specified in ISO 80369-7.The APRO 45 Catheter met the acceptance criteria.
Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 45 Catheter met the acceptance criteria.
Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 45 Catheter met the acceptance criteria.
Liquid LeakageCatheter must withstand pressure with methods specified in ISO 10555-1, Annex C.The APRO 45 Catheter met the acceptance criteria.
Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 45 Catheter met the acceptance criteria.
Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 45 Catheter met the acceptance criteria.
Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 45 Catheter met the acceptance criteria.
3-Point BendCatheter 3-Point Bend force must be acceptable. Forces were compared to a predicateThe APRO 45 Catheter met the acceptance criteria.
Simulated Use including Clot RetrievalWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications. Clot retrieval from various locations of the anatomical model was performed using the subject device system.The APRO 45 Catheter and Alembic Aspiration Tubing met the acceptance criteria.
Corrosion ResistanceCatheter shall show no signs of corrosion when tested in accordance with ISO 10555-1, Annex A.The APRO 45 Catheter met the acceptance criteria.

B. Design Verification Testing – Animal

Substantial equivalence was established based on non-clinical bench performance data. Animal testing data were not deemed necessary.

Page 9

APRO® 45 Catheter and Alembic Aspiration Tubing
510(k) Number: K251015
Page 5 of 6

C. Sterilization and Shelf-Life

The APRO 45 Catheter and Alembic Aspiration Tubing are sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10⁻⁶ in accordance with ISO 11135. Aging studies for the APRO 45 Catheter and Alembic Aspiration Tubing have established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and the device functionality were performed and met the acceptance criteria.

D. Biocompatibility

Biocompatibility testing has been completed for the APRO 45 Catheter as an externally communicating device with circulating blood contact for a limited (≤ 24 hours) duration in accordance with ISO 10993-1 and the FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process"."

Table 3. Summary of Biocompatibility Test Results

TestResultsConclusions
Sensitization (Guinea Pig Maximization)The APRO 45 Catheter did not elicit a sensitization response.Non-sensitizing
Irritation/Intracutaneous ReactivityThe APRO 45 Catheter demonstrated no evidence of irritation.Non-irritant
Cytotoxicity (MEM Elution, L929 cells)The APRO 45 Catheter did not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
Hemolysis - IndirectThere were no significant differences between the test article extract and negative control article results.Non-hemolytic
Hemolysis - DirectThere were no differences between the hemolytic index of the test article and the negative control.Non-hemolytic
Thrombogenicity – Partial Thromboplastin Time (PTT)The average clotting time of the test article was not significantly different from the vehicle and negative control.Non-thrombogenic
Thrombogenicity – Platelet Leukocyte CountThe test article performed similar to comparator and negative controls.
Thrombogenicity – Comparative Surface and Geometry AssessmentComparative surface assessment using 40x optical microscopy at representative locations of the subject device and the predicate device did not find any differences in the roughness or presence of any defects. The subject device has similar surface morphology as the predicate device APRO 55 Catheter.
SC5b9 Complement ActivationThe Sc5b9 concentration of the test article was statistically less than the positive control and was not statistically higher than the negative control.Acceptable
Acute Systemic ToxicityNo weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice.Non-toxic
Material-Mediated PyrogenicityAll individual rabbits for both the test article and negative control showed a total rise in temperature of < 0.5ºC and were determined to be nonpyrogenic.Non-pyrogenic

Page 10

APRO® 45 Catheter and Alembic Aspiration Tubing
510(k) Number: K251015
Page 6 of 6

E. Clinical Testing

Substantial equivalence was established based on non-clinical performance data. Human clinical data were not deemed necessary.

8) Conclusion

Based on the comparison of the technological characteristics and the non-clinical testing, the subject device is found to be substantially equivalent to the predicate device. The differences in technological characteristics do not raise new questions of safety and effectiveness. Testing was conducted to demonstrate that the subject device meets the specifications and performs as intended.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).