K Number
K250977

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-04-28

(28 days)

Product Code
Regulation Number
878.4840
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Suture-TOOL System is intended for abdominal wall closure after laparotomy in patients 18 years old and older.

Device Description

The Suture-TOOL System is a single-use, sterile, suture applicator with a polydioxanone resorbable suture pre-mounted to the needle. The polydioxanone monofilament synthetic absorbable sutures are prepared from polyester, poly(p-dioxanone). The empirical molecular formula of the polymer is as follows: -(O-CH2-CH2-O-CH2-CO)n- The sutures are dyed with D&C Violet No. 2 (21CFR§ 74.3602). The Suture-TOOL System includes the following: • Polydioxanone resorbable suture pre-mounted on the needle – 1055S, 1020S (Suture&Needle). • Suture-TOOL – 1018S (suture applicator). Before use, the user must follow the instructions to attach the needle, preloaded with the suture, to the applicator.

AI/ML Overview

The provided FDA 510(k) clearance letter pertains to the Suture-TOOL System, which is a surgical suture device, not an AI or software-based medical device. Therefore, the information requested regarding acceptance criteria and study proving performance for an AI/software device (sample size, ground truth, expert adjudication, MRMC studies, standalone performance) is largely not applicable to this submission.

However, I can extract the relevant acceptance criteria and performance study information for this physical medical device based on the provided text.


Acceptance Criteria and Study for Suture-TOOL System (Physical Device)

1. A table of acceptance criteria and the reported device performance

The acceptance criteria for the Suture-TOOL System are based on established USP (United States Pharmacopeia) standards for surgical sutures. The performance is reported as meeting or exceeding these criteria.

Acceptance Criterion (USP Standard)Reported Device Performance
USP 41-NF36:2018 <881> Tensile StrengthMet or exceeded USP criteria
USP 41-NF36:2018 <861> Sutures - DiameterMet or exceeded USP criteria
USP 41-NF36:2018 <871> Sutures - Needle AttachmentMet or exceeded USP criteria
Material/Component Specific:Reported Performance:
Polydioxanone suture (USP 0 size) tensile strength residual at 4 weeks71.7%
Polydioxanone suture (USP 0 size) tensile strength residual at 8 weeks16.5%
Absorption time~ 180-210 days post-implantation

2. Sample size used for the test set and the data provenance

The document does not explicitly state the sample sizes used for the physical and mechanical testing of the sutures (e.g., number of sutures tested for tensile strength, diameter, or needle attachment).

  • Sample Size: Not explicitly stated (e.g., "n=" not provided for each test). The text only mentions "Physical testing was performed on the poly(p-dioxanone) synthetic absorbable suture PDO USP 0 to USP 29".
  • Data Provenance: Not specified regarding country of origin. The data is derived from physical performance testing of the manufactured device components. It is inherently "prospective" in the sense that the tests are conducted on manufactured products to verify design.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable to this physical device submission. Ground truth, in the context of AI/software, refers to expert interpretations or labels. For a surgical suture, the "ground truth" is defined by the objective, quantitative measurements against established USP physical and mechanical standards. No human expert interpretation is involved in establishing the "ground truth" for suture diameter, tensile strength, or needle attachment.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically used for expert consensus in labeling or interpreting data, common in AI/software performance studies. For physical testing of a medical device like a suture, results are quantitative and objective, eliminating the need for adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. MRMC studies are specific to evaluating the clinical impact of AI or imaging devices on human reader performance. This submission is for a physical surgical suture, which does not involve "human readers" in the context of interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical suture, not an algorithm.

7. The type of ground truth used

The "ground truth" for the Suture-TOOL System's performance is based on pre-defined, objective, quantitative physical and mechanical standards set by the United States Pharmacopeia (USP) for surgical sutures. Specifically, USP standards <861> for Suture Diameter, <871> for Suture Attachment, and <881> for Tensile Strength.

8. The sample size for the training set

Not applicable. As a physical medical device, there is no "training set" in the machine learning sense. The device is manufactured and tested based on established engineering and materials science principles.

9. How the ground truth for the training set was established

Not applicable. There is no concept of "ground truth for the training set" for this physical device. The device's design and manufacturing rely on established material properties and engineering standards.


Summary regarding the nature of this 510(k) submission:

This 510(k) submission for the Suture-TOOL System is for a physical medical device (surgical suture). The "study" proving it meets acceptance criteria for FDA clearance predominantly involves physical and mechanical performance testing against recognized industry standards (USP). The questions related to AI/software performance studies (human experts, ground truth adjudication for interpretative tasks, MRMC, standalone algorithm performance, training sets) are therefore not relevant to this specific device and the information provided in the clearance letter. The substantial equivalence claim is primarily based on the device being identical to a previously cleared predicate device, with the addition of a new size (USP 0) that also demonstrated conformance to the same physical and mechanical standards.

FDA 510(k) Clearance Letter - Suture-TOOL System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

April 28, 2025

Suturion AB
℅ Adam Harris
Senior Director, Regulatory and Strategic Development
Target Health
110 West 40th Street, 8th Floor
New York, New York 10018

Re: K250977
Trade/Device Name: Suture-TOOL System
Regulation Number: 21 CFR 878.4840
Regulation Name: Absorbable Polydioxanone Surgical Suture
Regulatory Class: Class II
Product Code: NEW, GAB
Dated: March 28, 2025
Received: March 31, 2025

Dear Adam Harris:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.


April 28, 2025

Suturion AB
℅ Adam Harris
Senior Director, Regulatory and Strategic Development
Target Health
110 West 40th Street, 8th Floor
New York, New York 10018

Re: K250977
Trade/Device Name: Suture-TOOL System
Regulation Number: 21 CFR 878.4840
Regulation Name: Absorbable Polydioxanone Surgical Suture
Regulatory Class: Class II
Product Code: NEW, GAB
Dated: March 28, 2025
Received: March 31, 2025

Dear Adam Harris:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 2

K250977 - Adam Harris Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250977 - Adam Harris Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TEK N. LAMICHHANE -S

Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250977

Device Name: Suture-TOOL System

Indications for Use (Describe): The Suture-TOOL System is intended for abdominal wall closure after laparotomy in patients 18 years old and older.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: 07/31/2026

Indications for Use

See PRA Statement below.

510(k) Number (if known): K250977

Device Name: Suture-TOOL System

Indications for Use (Describe): The Suture-TOOL System is intended for abdominal wall closure after laparotomy in patients 18 years old and older.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

Suture-TOOL System-K250977 Suturion AB

510(k) Summary (K250977)

1. APPLICANT/CONTACT

Suturion AB

Contact Person

Adam Harris, MM, RAC
Senior Director, Regulatory and Strategic Development
Target Health
110 West 40th Street, 8th Floor
New York, NY 10018
Telephone: (646) 479-2437
aharris@targethealth.com

Date of Summary: April 28, 2025

2. DEVICE

Device Proprietary NameSuture-TOOL System
Classification nameAbsorbable polydioxanone surgical suture
Product CodeNEW
Regulation21 CFR 878.4840
Device ClassificationClass II
Review PanelGeneral & Plastic Surgery

3. PREDICATE

| Predicate Device | K242835 Suture-TOOL System Absorbable polydioxanone surgical suture |

4. DESCRIPTION OF DEVICE

The Suture-TOOL System is a single-use, sterile, suture applicator with a polydioxanone resorbable suture pre-mounted to the needle. The polydioxanone monofilament synthetic absorbable sutures are prepared from polyester, poly(p-dioxanone). The empirical molecular formula of the polymer is as follows:

-(O-CH2-CH2-O-CH2-CO)n-

Page 6

Suture-TOOL System-K250977 Suturion AB

The sutures are dyed with D&C Violet No. 2 (21CFR§ 74.3602).

The Suture-TOOL System includes the following:

• Polydioxanone resorbable suture pre-mounted on the needle – 1055S, 1020S (Suture&Needle).

• Suture-TOOL – 1018S (suture applicator)

Before use, the user must follow the instructions to attach the needle, preloaded with the suture, to the applicator.

5. INDICATIONS FOR USE

The Suture-TOOL System is intended for abdominal wall closure after laparotomy in patients 18 years old and older.

6. COMPARISON WITH THE PREDICATE DEVICE - SUMMARY

The Subject and Predicate devices are identical. The Subject device adds the size USP 0 suture which has the same material, design, intended use and technological characteristics as the predicate device K242835. The Subject and Predicate devices both provide sterile, single-use applicators with the same manufacturing process. The device characteristics comparing the Subject device to the Predicate device are summarized below.

Table 1: REGULATORY INFORMATION COMPARISON

Predicate DeviceSubject Device
K242835 Suture-TOOL SystemK250977 Suture-TOOL System
Absorbable polydioxanone surgical sutureSame
NEWSame
21 CFR 878.4840Same
Class IISame
General & Plastic SurgerySame

Page 7

Suture-TOOL System-K250977 Suturion AB

Table 2: OVERALL SUBSTANTIAL EQUIVALENCE COMPARISON

Subject DevicePredicate Device
Suture-TOOL System (Suture and Applicator)Suture-TOOL System (Suture and Applicator)
ManufacturerSuturion, ABSuturion, AB
510(k) NumberK250977K242835
Absorbable polydioxanone surgical sutureSame
Product CodeNEWSame
Regulation21 CFR 878.4840Same
Device ClassIISame
Suture applicatorSame
Product CodeGABSame
Regulation21 CFR 878.4800Same
Device ClassClass 1Same
Intended UseThe Suture-TOOL System is intended for abdominal wall closure after laparotomy in patients 18 years old and older.Same
Device DescriptionThe Suture-TOOL System is a single-use, sterile, suture applicator with a polydioxanone resorbable suture pre-mounted to the needle. The polydioxanone monofilament synthetic absorbable sutures are prepared from the polyester, poly(p-dioxanone).The Suture-TOOL System includes the following:1) Polydioxanone resorbable suture pre-mounted on the needle – 1055S, 1020S (Suture&Needle) and2)2) Suture-TOOL – 1018S (suture applicator).Before use, the user must follow the instructions to attach the needle, preloaded with the suture, to the applicator.The Suture-TOOL System is a single-use, sterile, suture applicator with a polydioxanone resorbable suture pre-mounted to the needle. The polydioxanone monofilament synthetic absorbable sutures are prepared from the polyester, poly(p-dioxanone).The Suture-TOOL System includes the following:1) the Suture-TOOL suture applicator (1018S), and2) a polydioxanone resorbable suture pre- mounted on the needle (1020S).Before use, the user must follow the instructions to attach the needle, preloaded with the suture, to the applicator

Page 8

Suture-TOOL System-K250977 Suturion AB

Table 2: OVERALL SUBSTANTIAL EQUIVALENCE COMPARISON

Subject DevicePredicate Device
Suture-TOOL System (Suture and Applicator)Suture-TOOL System (Suture and Applicator)
Suture materialResorbableSame
ChemicalPoly(p-dioxanone)Same
USP sizes2-0, 02-0
Approx. Tensile strength residual at time:4 weeks 71.7%8 weeks 16.5%Same
Absorption~ 180-210 days post-implantationSame
NeedleDouble-pointed AISI 316 LSame
SterilityEthylene Oxide gasSame

7. PERFORMANCE TESTING

Physical testing was performed on the poly(p-dioxanone) synthetic absorbable suture PDO USP 0 to USP 29, including <861> suture diameter, <871> suture attachment, <881>tensile strength.

• 6-415 USP 41-NF36:2018 <881> Tensile Strength
• 6-416 USP 41-NF36:2018 <861> Sutures - Diameter
• 6-417 USP 41-NF36:2018 <871> Sutures - Needle Attachment

In all cases, the USP criteria were met or exceeded.

Design Verification/Validation

Verification testing performed with the Suture-TOOL System assessed design control to confirm the specification requirements established. The testing demonstrated that the Suture-TOOL System meets all specifications according to device requirements.

Page 9

Suture-TOOL System-K250977 Suturion AB

8. CONCLUSION

The Subject and Predicate devices are identical, that is, as there are no differences in technological characteristics or performance characteristics between the Subject and Predicate devices. Thus, the Sponsor has determined that the Suture-Tool System does not raise new questions of safety and effectiveness in comparison to the Predicate device and is substantially equivalent to the Predicate device.

§ 878.4840 Absorbable polydioxanone surgical suture.

(a)
Identification. An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.