K Number
K250957

Validate with FDA (Live)

Date Cleared
2025-05-30

(60 days)

Product Code
Regulation Number
874.4680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DISPOSABLE BIOPSY FORCEPS FB-211D/221D/231D/241D are intended to be used to collect tissue within the tracheobronchial tree in combination with a bronchoscope

Device Description

The Disposable Biopsy Forceps FB-211D, FB-221D, FB-231D, and FB-241D (also known as "EndoJaw") have been designed to collect tissue within the tracheobronchial tree in combination with a flexible bronchoscope. The Disposable Biopsy Forceps are inserted into the channel of a bronchoscope to collect tissue by biting the surface of the lesion with a pair of forceps located at the distal end of the device. The Disposable Biopsy Forceps are withdrawn from the channel and the tissue sample is collected.

The Disposable Biopsy Forceps consists of a handle and an insertion portion. The handle consists of a handle body and slider. The insertion portion consists of a sheath, metallic coil (operating wire), cups, and forceps cap. The slider is connected to the operating wire in the insertion portion and the cups are opened/closed by advancing/retreating the slider with the function of the connection parts. The subject device is available in multiple distal end cup shapes; FB-211D – Alligator jaw-step, FB-221D – Alligator jaw-step with needle, FB-231D – Oval type, and FB-241D – Oval type with needle.

AI/ML Overview

This response is based solely on the provided FDA 510(k) Clearance Letter for the Olympus Disposable Biopsy Forceps.

Analysis of the Provided Document:

The provided document is an FDA 510(k) Clearance Letter for a medical device: Disposable Biopsy Forceps. This type of clearance is for devices that are "substantially equivalent" to legally marketed predicate devices. The key takeaway from this document is that the clearance is based on comparing the new device to a predicate device through non-clinical bench testing, not on clinical performance studies involving human subjects or AI-assisted diagnostic accuracy.

Therefore, the questions regarding ground truth, expert adjudication, MRMC studies, standalone algorithm performance, and sample sizes for training/test sets related to diagnostic accuracy (which would be typical for an AI/software as a medical device) are not applicable to this specific 510(k) clearance documentation.

The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the engineering and performance specifications tested during the non-clinical bench testing to demonstrate substantial equivalence to the predicate device.


Responses based on the provided FDA 510(k) Clearance Letter:

1. A table of acceptance criteria and the reported device performance

The document states: "All test samples passed pre-defined acceptance criteria." However, the specific quantitative acceptance criteria for each test and the detailed reported performance metrics are not explicitly provided in this summary. The document only lists the types of performance tests conducted.

Test Performed (Non-Clinical Bench Testing)Acceptance Criteria (Not Explicitly Stated in Document)Reported Device Performance
Insertion Performance into the Bronchoscope(Implied: Smooth insertion without obstruction/damage)Passed acceptance criteria
Withdrawal Performance from the Bronchoscope(Implied: Smooth withdrawal without obstruction/damage)Passed acceptance criteria
Opening and Closing of the Forceps Cups(Implied: Consistent and complete opening/closing)Passed acceptance criteria
Visual Inspection of the Insertion Portion(Implied: No defects, damage, or manufacturing flaws)Passed acceptance criteria
Performance After Repeated Insertion and Withdrawal Operations(Implied: Maintained performance and integrity over repetitions)Passed acceptance criteria
Performance After Repeated Open/Close Operations(Implied: Maintained performance and integrity over repetitions)Passed acceptance criteria
Connection Strength Between the S-Cover and the Coil Sheath(Implied: Sufficient strength to prevent separation)Passed acceptance criteria
Visual Inspection for Defects of the Needle and Forceps Teeth(Implied: No visual defects on critical components)Passed acceptance criteria
Biocompatibility Testing(Implied: Meets ISO 10993-1 standards for cytotoxicity, sensitization, irritation, acute systemic toxicity, material mediated pyrogenicity)Passed acceptance criteria
Sterilization Validation (SAL 10⁻⁶)(Implied: Achieves required sterility assurance level)Passed acceptance criteria
Shelf-Life Testing (5 years)(Implied: Maintains packaging integrity and product performance over 5 years, per ASTM F1980-21, ISO 11607-1/2)Passed acceptance criteria
Human Factors Evaluation(Implied: Safe for intended users, uses, and use environments)Determined safe

2. Sample sizes used for the test set and the data provenance

  • Sample Size: The document mentions "All test samples" but does not specify the numerical sample sizes used for each of the non-clinical bench tests.
  • Data Provenance: The tests were conducted by Olympus. The document does not specify country of origin for data as this is a device clearance based on manufacturing and performance testing, not patient data from clinical studies. The testing was prospective in the sense that the tests were designed and executed to evaluate the new device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as this 510(k) clearance is for a physical medical device (biopsy forceps) and is based on engineering performance demonstration through non-clinical bench testing, not on clinical diagnostic accuracy or interpretation of medical images/data (which would require expert ground truth).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable for the reasons stated in point 3. Adjudication methods are relevant for subjective image interpretation or clinical outcomes, not for objective bench test measurements of a physical device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable as the clearance is for a physical biopsy forceps and does not involve AI or human readers for diagnostic interpretation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the clearance is for a physical biopsy forceps and does not involve an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

This question is not applicable as the device is cleared based on non-clinical performance characteristics (e.g., insertion, withdrawal, force of jaws, sterility, shelf-life) which are evaluated against predefined engineering specifications and physical measurements, not against clinical ground truth like pathology or outcomes data.

8. The sample size for the training set

This question is not applicable as this is a physical medical device clearance, not an AI/machine learning device that requires a "training set."

9. How the ground truth for the training set was established

This question is not applicable for the reasons stated in point 8.

FDA 510(k) Clearance Letter - Olympus Disposable Biopsy Forceps

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

May 30, 2025

Olympus Medical Systems Corporation
℅ Susan Lewandowski
Manager, Program Regulatory Affairs
Olympus Surgical Technologies of America
800 West Park Drive
Westborough, Massachusetts 01581

Re: K250957
Trade/Device Name: Disposable Biopsy Forceps (FB-211D); Disposable Biopsy Forceps (FB-221D); Disposable Biopsy Forceps (FB-231D); Disposable Biopsy Forceps (FB-241D)
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (flexible or rigid) and accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: March 30, 2025
Received: March 31, 2025

Dear Susan Lewandowski:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250957 - Susan Lewandowski Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250957 - Susan Lewandowski Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ethan L. Nyberg -S

Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.K250957
Please provide the device trade name(s).

Disposable Biopsy Forceps (FB-211D);
Disposable Biopsy Forceps (FB-221D);
Disposable Biopsy Forceps (FB-231D);
Disposable Biopsy Forceps (FB-241D)

Please provide your Indications for Use below.

The DISPOSABLE BIOPSY FORCEPS FB-211D/221D/231D/241D are intended to be used to collect tissue within the tracheobronchial tree in combination with a bronchoscope

Please select the types of uses (select one or both, as applicable).☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Disposable Biopsy Forceps Page 10 of 34

Page 5

Disposable Biopsy Forceps FB-series

510(k) Summary

1. General Information

Date Prepared: May 27, 2025

510(K) submitter: Olympus Medical Systems Corporation
2951 Ishikawa-cho, Hachioji-shi, Tokyo Japan 192-8507

Contact: Seiko Yunoki

Correspondent: Olympus Surgical Technologies of America
800 West Park Drive, Westborough, MA 01581

Primary Contact: Susan Lewandowski
Email: susan.lewandowski@olympus.com

2. Device Information

Device Name:

  • Disposable Biopsy Forceps FB-211D
  • Disposable Biopsy Forceps FB-221D
  • Disposable Biopsy Forceps FB-231D
  • Disposable Biopsy Forceps FB-241D

Common Name: Biopsy Forceps
Classification: 874.4680 Bronchoscope (flexible or rigid) and accessories
Regulatory Class: II
Product Code: EOQ
Device Panel: Ear, Nose, & Throat

3. Predicate Device Information

Olympus Single Use Biopsy Forceps FB-433D – K172726

4. Device Description

The Disposable Biopsy Forceps FB-211D, FB-221D, FB-231D, and FB-241D (also known as "EndoJaw") have been designed to collect tissue within the tracheobronchial tree in combination with a flexible bronchoscope. The Disposable Biopsy Forceps are inserted into the channel of a bronchoscope to collect tissue by biting the surface of the lesion with a pair of forceps located at the distal end of the device. The Disposable Biopsy Forceps are withdrawn from the channel and the tissue sample is collected.

The Disposable Biopsy Forceps consists of a handle and an insertion portion. The handle consists of a handle body and slider. The insertion portion consists of a sheath, metallic coil

Page 6

Disposable Biopsy Forceps FB-series

(operating wire), cups, and forceps cap. The slider is connected to the operating wire in the insertion portion and the cups are opened/closed by advancing/retreating the slider with the function of the connection parts. The subject device is available in multiple distal end cup shapes; FB-211D – Alligator jaw-step, FB-221D – Alligator jaw-step with needle, FB-231D – Oval type, and FB-241D – Oval type with needle.

The subject device has the same technological characteristics and similar design as the applicable predicate device.

5. Indications for Use

The DISPOSABLE BIOPSY FORCEPS FB-211D/FB-221D/FB-231D/FB-241D are intended to be used to collect tissue within the tracheobronchial tree in combination with a bronchoscope.

6. Predicate Comparison

DescriptionSubject Device (SD)Single Use Biopsy ForcepsFB-211D, FB-221D, FB-231D, and FB-241DPredicate Device (PD)Single Use Biopsy ForcepsFB-433D(K172726)Comparison
Design of Insertion portion of ForcepsThe insertion portion is a stainless steel coil covered by a polyethylene sheath.The insertion portion is a stainless steel coil.DIFFERENTNon-clinical bench testing was performed to demonstrate that this difference in technology does not impact the effectiveness of the device.
Cup ShapeFB-211DAlligator Jaw-stepFB-221DAlligator jaw-step with needleFB-231DOval typeFB-241DStandard (Oval type), FenestratedSIMILARNon-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device.

Page 7

Disposable Biopsy Forceps FB-series

DescriptionSubject Device (SD)Single Use Biopsy ForcepsFB-211D, FB-221D, FB-231D, and FB-241DPredicate Device (PD)Single Use Biopsy ForcepsFB-433D(K172726)Comparison
Oval type with Needle
Compatible EndoscopesBronchoscopes with channel diameter 2.0 mm and working length 600 mm or less (exclude ultrasound endoscopes).Bronchoscopes with channel diameter 1.7 mm and working length 600 mm or less (exclude ultrasound endoscopes).SIMILARThere is a difference in the compatible endoscope channel diameter between the SD and PD. Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device.
Product SpecificationsMaximum insertion portion diameter: 1.9 mmWorking length: 1150 mmMaximum insertion portion diameter: 1.5 mmWorking length: 1150 mmSIMILARThere is a difference in the maximum insertion portion channel diameter between the SD and PD. Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device.
Single Use/ReusableSingle UseSingle UseSAME
SterilizationGammaEthylene Oxide (EtO)DIFFERENTThere is a difference in sterilization methods between the SD and PD. Sterilization validation testing was performed to demonstrate the efficacy of the sterilization dose.
Shelf-LifeFive yearsThree yearsDIFFERENTThere is a difference in shelf-life expiry dating between the SD and PD.

Page 8

Disposable Biopsy Forceps FB-series

DescriptionSubject Device (SD)Single Use Biopsy ForcepsFB-211D, FB-221D, FB-231D, and FB-241DPredicate Device (PD)Single Use Biopsy ForcepsFB-433D(K172726)Comparison
Oval type with NeedleStability / Shelf-life testing was performed to demonstrate the difference does not impact the effectiveness of the subject device.
Compatible EndoscopesBronchoscopes with channel diameter 2.0 mm and working length 600 mm or less (exclude ultrasound endoscopes).Bronchoscopes with channel diameter 1.7 mm and working length 600 mm or less (exclude ultrasound endoscopes).SIMILARThere is a difference in the compatible endoscope channel diameter between the SD and PD. Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device.
Product SpecificationsMaximum insertion portion diameter: 1.9 mmMaximum insertion portion diameter: 1.5 mmSIMILAR
Working length: 1150 mmWorking length: 1150 mmThere is a difference in the maximum insertion portion channel diameter between the SD and PD. Non-clinical bench testing was performed to demonstrate that design differences between the subject and predicate devices do not impact the performance of the subject device.
Single Use/ReusableSingle UseSingle UseSAME
SterilizationGammaEthylene Oxide (EtO)DIFFERENTThere is a difference in sterilization methods between the SD and PD. Sterilization validation testing was performed to demonstrate the efficacy of the sterilization dose.
Shelf-LifeFive yearsThree yearsDIFFERENTThere is a difference in shelf-life expiry dating between the SD and PD.

7. Non-Clinical/Clinical Tests Summary and Conclusion

Olympus performed bench testing to demonstrate substantial equivalence to the predicate device. Test samples were final, finished devices subjected to the full manufacturing process including sterilization.

The following performance side-by-side bench tests were conducted to demonstrate substantial equivalence between the subject and predicate devices. All test samples passed pre-defined acceptance criteria.

  • Insertion Performance into the Bronchoscope
  • Withdrawal Performance from the Bronchoscope
  • Opening and Closing of the Forceps Cups
  • Visual Inspection of the Insertion Portion
  • Performance After Repeated Insertion and Withdrawal Operations
  • Performance After Repeated Open/Close Operations
  • Connection Strength Between the S-Cover and the Coil Sheath
  • Visual Inspection for Defects of the Needle and Forceps Teeth

8. Biocompatibility Testing

Biocompatibility testing for the Disposable Biopsy Forceps was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." The biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity.

9. Sterilization and Shelf Life

The devices are sterilized via gamma radiation. Sterilization validation (VDmax25) was conducted in accordance with ISO 11137-1:2006/Amd1:2013 and Amd2:2018 Sterilization of Health Care Product – Radiation Part 1 – Requirements for development, validation and routine control of a sterilization process for medical devices and ISO11137-2:2013 Sterilization of Health Care Product – Radiation Part 2 – Establishing the sterilization dosage. The sterility assurance level is 10⁻⁶.

Page 9

Disposable Biopsy Forceps FB-series

The shelf-life of the subject device is five years. Olympus conducted stability testing to demonstrate packaging integrity and product performance for the stated expiration date. The accelerated aging test was conducted in accordance with ASTM F1980-21. Package integrity testing was conducted in accordance with ASTM F1980-21, ISO 11607-1:2019 and ISO 11607-2:2019 after accelerated aging and simulated distribution. Package Integrity testing included Visual Inspection of Package, Peel Strength of Package, and Integrity of the Package. All samples successfully met the acceptance criteria and passed testing.

10. Human Factors Testing

Olympus conducted a Human Factors Evaluation and determined that the Disposable Biopsy Forceps FB-series are safe for intended users, uses, and use environments.

11. Conclusion

Based on the comparison to the predicate device, intended use, technological characteristics, and performance testing, the Disposable Biopsy Forceps FB-211D, FB-221D, FB-231D, and FB-241D are substantially equivalent to the predicate device.

§ 874.4680 Bronchoscope (flexible or rigid) and accessories.

(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.