K Number
K250879

Validate with FDA (Live)

Device Name
Paradigm System
Manufacturer
Date Cleared
2025-06-04

(72 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.

Device Description

The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons. The Paradigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.

The Paradigm System encompasses non-sterile operating room equipment components as well as sterilizable, reusable instruments, and off-the-shelf sterile-packaged, single-use accessories.

The Paradigm System is used with an external monitor.

AI/ML Overview

The provided FDA 510(k) Clearance Letter for the Paradigm System (K250879) indicates substantial equivalence to a predicate device (K243258). While it mentions that "Verification and Validation activities have been conducted to provide assurance that the device meets the performance requirements," and lists various standards used in testing, it does not provide specific acceptance criteria or the detailed results of a study that proves the device meets those criteria.

The document primarily focuses on demonstrating substantial equivalence based on the device's intended use, indications for use, fundamental scientific technology, and technical characteristics being the same as the predicate. It states that "Performance, safety and usability testing demonstrate that the differences between the subject device and the predicate device do not raise new risks of safety and effectiveness," but does not present the granular data of what those performance, safety, and usability tests entailed or their quantitative outcomes against defined acceptance criteria.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and the reported device performance: This information is not present in the provided text.
  2. Sample sizes used for the test set and the data provenance: Not specified.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified.
  4. Adjudication method for the test set: Not specified.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size: Not mentioned.
  6. If a standalone (algorithm only without human-in-the-loop performance) was done: Not specified.
  7. The type of ground truth used: Not specified.
  8. The sample size for the training set: Not specified.
  9. How the ground truth for the training set was established: Not specified.

The provided document is a 510(k) clearance letter, which summarizes the FDA's finding of substantial equivalence based on information submitted by the manufacturer. It typically does not contain the detailed raw data, study designs, or complete results of every verification and validation test performed by the manufacturer. That detailed information would be found in the manufacturer's submission to the FDA, not in the clearance letter itself.

The document lists standards used, such as ASTM F2554-22 "Standard Practice For Measurement Of Positional Accuracy Of Computer Assisted Surgical Systems." This standard likely defines methods for measuring positional accuracy, which would be a key performance metric for a stereotaxic navigation system. However, the document does not report the specific results of applying this standard or the acceptance criteria for those results.

FDA 510(k) Clearance Letter - Paradigm System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 4, 2025

Proprio, Inc.
Shannon Eubanks
Vice President, Systems Engineering & Regulatory
111 W. John St., Suite 308
Seattle, Washington 98119

Re: K250879
Trade/Device Name: Paradigm System
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: April 4, 2025
Received: April 7, 2025

Dear Shannon Eubanks:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250879 - Shannon Eubanks Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

Page 3

K250879 - Shannon Eubanks Page 3

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250879

Device Name
Paradigm System

Indications for Use (Describe)
The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary Statement

Submitter: Proprio, Inc.

Contact Person: Shannon Eubanks
VP of Systems Engineering and Regulatory
Phone: (425) 802-6063
Email: seubanks@propriovision.com

Trade Name: Paradigm System

Common Name: Orthopedic stereotaxic instrument

Classification: Class II

Product Code: OLO

Regulation: 21 CFR 882.4560

Predicate Device(s): The subject device is equivalent to the following device: Paradigm System (K243258)

Device Description: The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons. The Paradigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.

The Paradigm System encompasses non-sterile operating room equipment components as well as sterilizable, reusable instruments, and off-the-shelf sterile-packaged, single-use accessories.

The Paradigm System is used with an external monitor.

Indications For Use: The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.

K250879 Page 1 of 4

Page 6

Paradigm System (Subject Device)Paradigm System (Predicate Device)Comment
Device Overview
510(k) Number Decision DateK250879K243258 February 21, 2025
ManufacturerProprio, Inc.Proprio, Inc.
ClassificationClass IIClass II
Product CodeOLOOLO
Regulation21 CFR 882.456021 CFR 882.4560
Medical SpecialtyNeurologyNeurology
Indications For UseThe Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.
Principle of operationParadigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement. The Paradigm System enables surgical navigation of instruments relative to the patient spinal anatomy by combining preoperative imaging segmentation and intraoperative registration of the patient's live anatomy. Preoperative CT data is uploaded to the console and the system software generates a 3D model of the patient's operative vertebrae. The 3D model is used for preoperative planning such as implant placements or instrument trajectories. During the procedure, once the patient's anatomy is exposed, the Prism Sensor Array acquires imaging to match the topography of the live anatomy to the 3D model. This intraoperative registration of live anatomy to the 3D model creates a common coordinate space between theParadigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement. The Paradigm System enables surgical navigation of instruments relative to the patient spinal anatomy by combining preoperative imaging segmentation and intraoperative registration of the patient's live anatomy. Preoperative CT data is uploaded to the console and the system software generates a 3D model of the patient's operative vertebrae. The 3D model is used for preoperative planning such as implant placements or instrument trajectories. During the procedure, once the patient's anatomy is exposed, the Prism Sensor Array acquires imaging to match the topography of the live anatomy to the 3D model. This intraoperative registration of live anatomy to the 3D model creates a common coordinate space between the

K250879 Page 2 of 4

Page 7

Paradigm System (Subject Device)Paradigm System (Predicate Device)Comment
Device Overview
510(k) Number Decision DateK250879K243258 February 21, 2025
ManufacturerProprio, Inc.Proprio, Inc.
ClassificationClass IIClass II
Product CodeOLOOLO
Regulation21 CFR 882.456021 CFR 882.4560
Medical SpecialtyNeurologyNeurology
Indications For UseThe Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.
Principle of operationParadigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.Paradigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.
The Paradigm System enables surgical navigation of instruments relative to the patient spinal anatomy by combining preoperative imaging segmentation and intraoperative registration of the patient's live anatomy. Preoperative CT data is uploaded to the console and the system software generates a 3D model of the patient's operative vertebrae. The 3D model is used for preoperative planning such as implant placements or instrument trajectories. During the procedure, once the patient's anatomy is exposed, the Prism Sensor Array acquires imaging to match the topography of the live anatomy to the 3D model. This intraoperative registration of live anatomy to the 3D model creates a common coordinate space between theThe Paradigm System enables surgical navigation of instruments relative to the patient spinal anatomy by combining preoperative imaging segmentation and intraoperative registration of the patient's live anatomy. Preoperative CT data is uploaded to the console and the system software generates a 3D model of the patient's operative vertebrae. The 3D model is used for preoperative planning such as implant placements or instrument trajectories. During the procedure, once the patient's anatomy is exposed, the Prism Sensor Array acquires imaging to match the topography of the live anatomy to the 3D model. This intraoperative registration of live anatomy to the 3D model creates a common coordinate space between the

K250879 Page 2 of 4

Page 8

Paradigm System (Subject Device)Paradigm System (Predicate Device)Comment
preoperative CT data and the patient anatomy. This enables virtual information to be overlaid on the live anatomy during the procedure for intraoperative guidance.preoperative CT data and the patient anatomy. This enables virtual information to be overlaid on the live anatomy during the procedure for intraoperative guidance.
Technical ComparisonStereotaxic image guided surgical navigation system during spine surgery. Preoperative CT data is segmented and a 3D model is generated. Preoperative planning can be performed. Intraoperative image capture using reflective markers on spinal instruments and registers the patient's anatomy to the preoperative 3D model for intraoperative guidance during the procedure.Stereotaxic image guided surgical navigation system during spine surgery. Preoperative CT data is segmented and a 3D model is generated. Preoperative planning can be performed. Intraoperative image capture using reflective markers on spinal instruments and registers the patient's anatomy to the preoperative 3D model for intraoperative guidance during the procedure.
ComponentsCart, Arm, Sensor Array, External Monitor, Surgical Instruments, SoftwareCart, Arm, Sensor Array, External Monitor, Surgical Instruments, Software
InstrumentsDynamic Reference Frames, Awl, Pedicle Probes Drill Guide, Drill Bits, Depth Cannulas, TrocarDynamic Reference Frames, Awl, Pedicle Probes
Energy Source120V120V
Sterilization Method – Surgical InstrumentsSteamSteam
Minimum SAL1 x 10⁻⁶1 x 10⁻⁶
Required AccessoriesReflective Markers (Northern Digital K033621)Reflective Markers (Northern Digital K033621)
Optional AccessoriesParadigm DrapeParadigm Drape
Sterilization Method – drapeEthylene OxideEthylene Oxide
Minimum SAL1 x 10⁻⁶1 x 10⁻⁶

K250879 Page 3 of 4

Page 9

Functional and Safety Testing:

Verification and Validation activities have been conducted to provide assurance that the device meets the performance requirements under the indications for use conditions and to establish substantial equivalence with the predicate. Proprio performed the following testing to ensure the safety and effectiveness of the Paradigm System:

  • Non-Clinical Verification Tests
  • Non-Clinical Design Validation
  • Compliance Conformity Assessments

The following standards were used in testing:

  • IEC 62366-1: 2015/AMD1:2020 Medical Devices – Part 1: Application of Usability Engineering to Medical Devices, including Amendment 1
  • ISO 10993-1:2018 Biological evaluation of medical devices: Part 1: Evaluation and testing within a risk management process
  • ISO 14971:2019 Medical devices – Application of risk management to medical devices
  • ISO 17665-1:2006 Sterilization of health care products – Moist heat – Part:1 Requirements for the development, validation and routine control of a sterilization process for medical device
  • ANSI/AAMI ST79:2017 Comprehensive guide to steam sterilization and sterility assurance in health care facilities
  • ASTM F2554-22 Standard Practice For Measurement Of Positional Accuracy Of Computer Assisted Surgical Systems
  • ANSI/AAMI ST98:2022 Cleaning validation of healthcare products – Requirements for development and validation of a cleaning process for medical devices
  • IEC 62304 Edition 1.1 2015-06 Medical device software – Software life cycle processes

Conclusion: The Paradigm System intended use, indications for use, and fundamental scientific technology is the same as the predicate device. Performance, safety and usability testing demonstrate that the differences between the subject device and the predicate device do not raise new risks of safety and effectiveness.

K250879 Page 4 of 4

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).