K Number
K250866

Validate with FDA (Live)

Date Cleared
2025-05-20

(57 days)

Product Code
Regulation Number
888.3075
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sovereign Posterior Cervical System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.

The Sovereign Posterior Cervical System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, the Sovereign Posterior Cervical System may be connected to the Xultan 5.5 system using rod-to-rod connectors or transition rods. Please refer to the individual system's Instructions for Use for a list of indications for use for each system.

Device Description

The Met One Technologies Sovereign Posterior Cervical System is an implant device designed for the posterior stabilization of the cervical and upper thoracic spine. The system includes implants such as polyaxial screws, blocker screws, rods, crosslinks, lateral connectors, and rod to rod connectors to form a unique construct to match patient anatomy.

Polyaxial screws are offered in a range of sizes in both standard and reduction tulips for ease of use. All screws are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136).

Straight, pre-bent, and transition rods are offered in 3.5mm and 4.0mm diameters and in two different materials, titanium alloy (Ti-6Al-4V ELI per ASTM F136) and cobalt chrome (Co-Cr28-Mo6 per ASTM F1537). Transition rods are fully compatible with Met One Technologies Xultan 5.5 Pedicle Screw System line of implants.

A variety of connectors and crosslinks are offered for surgeons to create the desired construction. Rod to rod crosslinks improve the torsional stability of the construct. Lateral connectors are provided in open and closed configurations, 3.5mm and 4.0mm diameters, and in long and short lengths. Rod to rod connectors are offered in several configurations to join rods together. All connectors are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136).

AI/ML Overview

I am sorry, but the provided text is an FDA 510(k) clearance letter for a Sovereign Posterior Cervical System, which is a physical medical device (spinal implant system), not an AI/Software as a Medical Device (SaMD).

Therefore, the document does not contain any information about:

  • Acceptance criteria for an AI/SaMD's performance.
  • A study proving the device meets AI/SaMD acceptance criteria.
  • A test set (sample size, provenance).
  • Experts establishing ground truth.
  • Adjudication methods.
  • MRMC comparative effectiveness studies.
  • Standalone algorithm performance.
  • Types of ground truth.
  • Training set details.

The non-clinical testing described (ASTM F1717 and ASTM F1798) refers to mechanical testing of the physical implant's strength and durability, not software performance.

As such, I cannot fulfill your request for information related to AI/SaMD acceptance criteria and study details based on the provided input.

FDA 510(k) Clearance Letter - Sovereign Posterior Cervical System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 20, 2025

Met One Technologies, LLC
Liberato Aguilar, COO
4519 Osbourne Drive Suite C
El Paso, Texas 79922

Re: K250866
Trade/Device Name: Sovereign Posterior Cervical System
Regulation Number: 21 CFR 888.3075
Regulation Name: Posterior Cervical Screw System
Regulatory Class: Class II
Product Code: NKG
Dated: May 19, 2025
Received: May 19, 2025

Dear Liberato Aguilar:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Page 2

K250866 - Liberato Aguilar
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250866 - Liberato Aguilar
Page 3

Sincerely,

EILEEN CADEL -S

for
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.

510(k) Number (if known): K250866

Device Name: Sovereign Posterior Cervical System

Indications for Use (Describe)

The Sovereign Posterior Cervical System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.

The Sovereign Posterior Cervical System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, the Sovereign Posterior Cervical System may be connected to the Xultan 5.5 system using rod-to-rod connectors or transition rods. Please refer to the individual system's Instructions for Use for a list of indications for use for each system.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250866 Page 1 of 2

510(k) Summary

510(k)#: K250866
Prepared on: 2025-05-16

Contact Details

21 CFR 807.92(a)(1)

FieldInformation
Applicant NameMet One Technologies, LLC
Applicant Address4519 Osborne Drive Suite C El Paso TX 79922 United States
Applicant Contact Telephone915-373-3855
Applicant ContactMr. Evan Carbonell
Applicant Contact Emailevan@met1tech.com
Correspondent NameMet One Technologies, LLC
Correspondent Address4519 Osbourne Drive Suite C El Paso TX 79922 United States
Correspondent Contact Telephone915-892-6044
Correspondent ContactMr. Liberato Aguilar
Correspondent Contact Emaillibe@met1tech.com

Device Name

21 CFR 807.92(a)(2)

FieldInformation
Device Trade NameSovereign Posterior Cervical System
Common NamePosterior cervical screw system
Classification NamePosterior Cervical Screw System
Regulation Number888.3075
Product Code(s)NKG

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K131833Centurion POCT SystemNKG
K162575Z-LINKPC SystemNKG
K142838Synapse Occipital-Cervical-Thoracic (OCT) SystemNKG
K192013VERTICALE Cervical SystemNKG
K170108Xultan 5.5 Pedicle Screw SystemNKB

Device Description Summary

21 CFR 807.92(a)(4)

The Met One Technologies Sovereign Posterior Cervical System is an implant device designed for the posterior stabilization of the

Page 6

K250866 Page 2 of 2

cervical and upper thoracic spine. The system includes implants such as polyaxial screws, blocker screws, rods, crosslinks, lateral connectors, and rod to rod connectors to form a unique construct to match patient anatomy.

Polyaxial screws are offered in a range of sizes in both standard and reduction tulips for ease of use. All screws are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136).

Straight, pre-bent, and transition rods are offered in 3.5mm and 4.0mm diameters and in two different materials, titanium alloy (Ti-6Al-4V ELI per ASTM F136) and cobalt chrome (Co-Cr28-Mo6 per ASTM F1537). Transition rods are fully compatible with Met One Technologies Xultan 5.5 Pedicle Screw System line of implants.

A variety of connectors and crosslinks are offered for surgeons to create the desired construction. Rod to rod crosslinks improve the torsional stability of the construct. Lateral connectors are provided in open and closed configurations, 3.5mm and 4.0mm diameters, and in long and short lengths. Rod to rod connectors are offered in several configurations to join rods together. All connectors are manufactured from titanium alloy (Ti-6Al-4V ELI per ASTM F136).

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The Sovereign Posterior Cervical System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine (T1-T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.

The Sovereign Posterior Cervical System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

In order to achieve additional levels of fixation, the Sovereign Posterior Cervical System may be connected to the Xultan 5.5 system using rod-to-rod connectors or transition rods. Please refer to the individual system's Instructions for Use for a list of indications for use for each system.

Indications for Use Comparison

21 CFR 807.92(a)(5)

The Sovereign Posterior Cervical System has identical or similar indications for use as the primary and additional predicate devices.

Technological Comparison

21 CFR 807.92(a)(6)

The Sovereign Posterior Cervical System has similar technological characteristics, surgical approach, materials, range of sizes, design, and principles of operation as the predicate devices. Any differences in technological characteristics do not raise different questions of safety or effectiveness. Performance testing demonstrates the device has mechanical performance substantially equivalent to that of the predicate devices.

Non-Clinical and/or Clinical Tests Summary & Conclusions

21 CFR 807.92(b)

Mechanical testing was performed to demonstrate substantial equivalence using ASTM F1717 static compression bending, dynamic compression bending, and static torsion testing and ASTM F1798 static axial gripping capacity and static torsional gripping capacity.

The results of the mechanical testing showed that the worst-case constructs were substantially equivalent to legally marketed predicate devices.

§ 888.3075 Posterior cervical screw system.

(a)
Identification. Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (
e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.(b)
Classification. Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
e.g., CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.