K Number
K250841

Validate with FDA (Live)

Date Cleared
2025-08-13

(146 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation)) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.

Device Description

Electrotherapy Electrodes provide a means for establishing electrical contact between the lead connected to a TENS, EMS or NMES stimulation device and the skin, and are multi-layer, flexible structures composed of laminated materials commonly used in the application:

1st layer: Insulation material: Tan fabric
2nd layer: Double sides adhesive tape
3rd layer: Conductive film (Carbon Film)
4th layer: Self-adhesive conductive hydrogel
5th layer: Plastic release film
Connection: Leadwire/Snap button

The electrodes are designed for single-patient & single application use. Because of the adhesive nature of the conductive hydrogel, no securing materials are required to secure the device to the patient's skin.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the Electrotherapy Electrodes (K250841) does not describe a study involving AI or human reader performance. Instead, it focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing.

Therefore, many of the requested details about acceptance criteria for AI and human performance studies, expert involvement, adjudication, MRMC studies, standalone AI performance, and ground truth establishment cannot be found in this document.

However, I can extract the acceptance criteria and performance data related to the non-clinical tests that were performed to demonstrate substantial equivalence for this device.


Acceptance Criteria and Device Performance (Non-Clinical)

The device, Electrotherapy Electrodes (K250841), is a Class II cutaneous electrode. The study described focuses on demonstrating its substantial equivalence to a predicate device (ZMI Self-Adhesive Electrodes, K180865) primarily through non-clinical performance and biocompatibility testing. No AI component or human reader study is discussed.

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

Acceptance Criteria (Metric)Acceptance LimitReported Device Performance (Result)Study Description
Electrical Performance
A.C. Impedance≤ 160 ohmsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Electrical performance was evaluated under IFU (Instructions for Use) normal use conditions. This likely involved measuring the A.C. Impedance of the electrodes. The acceptance limit for the predicate device was <200 ohms, and the subject device's limit is ≤ 160 ohms, indicating a stricter passing criterion for the new device which it demonstrably met to claim SE.
Physical/Mechanical Performance
Adhesive PerformancePassed skin adhesion testNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Evaluated by performing a skin adhesion test.
Electrode StabilityPassedNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Evaluated by assessing electrical performance under IFU normal use conditions.
Packaging LabelingMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test packaging labeling.
Product AppearanceMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test product appearance.
DimensionMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test dimension.
Connection between electrode and electrode wireMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test the connection.
Cross-sectional area of electrode connection wireMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test the cross-sectional area.
Wire lengthMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test wire length.
Plug insertion and unplugging forceMet requirementsNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Method developed to test plug insertion and unplugging force.
Safety and Biocompatibility
Electrical SafetyComplies with IEC 60601-1Not explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Testing performed per IEC 60601-1 and IEC 60601-2-2 Edition 6.0 2017-03.
Biocompatibility (Cytotoxicity)Complies with ISO 10993-5Not explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Testing performed per ISO 10993-5 (in vitro cytotoxicity).
Biocompatibility (Skin Sensitization)Complies with ISO 10993-10Not explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Testing performed per ISO 10993-10 (skin sensitization).
Biocompatibility (Irritation)Complies with ISO 10993-23Not explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Testing performed per ISO 10993-23 (irritation).
Shipping and Shelf Life
Performance Testing of Shipping Containers and SystemsComplies with ASTM D4169-23Not explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Testing performed per ASTM D4169-23.
Shelf LifeDemonstratedNot explicitly stated, but "All tests were verified to meet acceptance criteria." implies it passed.Shelf life testing was performed.

Based on the provided document, the following information is NOT available as the submission does not involve an AI or human reader performance study:

  1. Sample sizes used for the test set and the data provenance: Not applicable to this type of non-clinical device testing.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these tests is based on objective measurements against engineering and safety standards.
  3. Adjudication method for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this device, "ground truth" refers to compliance with established industry standards (e.g., IEC, ISO, ASTM) and internal performance specifications.
  7. The sample size for the training set: Not applicable (no AI model).
  8. How the ground truth for the training set was established: Not applicable (no AI model).

FDA 510(k) Clearance Letter - Electrotherapy Electrodes

Page 1

August 13, 2025

Baisheng Medical Co., Ltd.
Kaimin Chen
RA Specialist
No.11, Fusheng Road, Xinhui District
Jiangmen
Guangdong, 529100
China

Re: K250841
Trade/Device Name: Electrotherapy Electrodes
Regulation Number: 21 CFR 882.1320
Regulation Name: Cutaneous Electrode
Regulatory Class: Class II
Product Code: GXY
Dated: July 14, 2025
Received: July 14, 2025

Dear Kaimin Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250841 - Kaimin Chen Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250841 - Kaimin Chen Page 3

Sincerely,

Tushar Bansal -S

Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250841

Device Name
Electrotherapy Electrodes

Indications for Use (Describe)
Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation)) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

510(K) SUMMARY

510(K) Number: K250841

I. GENERAL INFORMATION

Applicant:
Baisheng Medical Co., Ltd.
No.11,Fusheng Road,Xinhui District,Jiangmen,Guangdong.China. P.C:529100
Chen Xi
Tel : 86 - 0750 - 6628113
E - mail: renzheng@bs0750.com

Contact Person:
Chen Kaimin
Phone:+86-15626286758
E - mail: venus@bs0750.com

Date Prepared: Aug 13 ,2025

II. DEVICE INFORMATION

Trade Name: Electrotherapy Electrodes
Model : LBS-Φ2, LBS-Φ3, LBS-Φ5, LBS-Φ7, LBS-Φ9, LBS-2×2, LBS-4×4, LBS-5×5, LBS-2×3, LBS-5×3.5, LBS-6×4, LBS-6×5, LBS-8×4, LBS-9×6, LBS-10×7, LBS-12×8, LBS-12×10, LBS-8.5×5, LBS-12×7, LBS-15×5, LBS-Φ2L, LBS-Φ3L, LBS-Φ5L, LBS-Φ7L, LBS-Φ9L, LBS-2×2L, LBS-4×4L, LBS-5×5L, LBS-2×3L, LBS-5×3.5L, LBS-6×4L, LBS-6×5L, LBS-8×4L, LBS-9×6L, LBS-10×7L, LBS-12×8L, LBS-12×10L, LBS-8.5×5L, LBS-12×7L, LBS-15×5L
Common Name : Cutaneous Electrode
Classification Name: Cutaneous Electrode
Regulation Number: 21 CFR 882.1320
Regulatory Class: Class II
Product Code: GXY
**Classification Panel :**Neurology
Type of 510(k) submission: Traditional

510(k) Summary Vol 002 Doc 01 Page 1 of 8

Page 6

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

PREDICATE DEVICE INFORMATION

PREDICATE DEVICE INFORMATION

Predicate device:
510(K) Number:K180865
Company Name: ZMI Electronics, Ltd.
Trade Name: ZMI Self-Adhesive Electrodes
Common Name: Cutaneous Electrode
Product Code: GXY

III. DEVICE DESCRIPTION

Electrotherapy Electrodes provide a means for establishing electrical contact between the lead connected to a TENS, EMS or NMES stimulation device and the skin, and are multi-layer, flexible structures composed of laminated materials commonly used in the application:

1st layer: Insulation material: Tan fabric
2nd layer: Double sides adhesive tape
3rd layer: Conductive film (Carbon Film)
4th layer: Self-adhesive conductive hydrogel
5th layer: Plastic release film
Connection: Leadwire/Snap button

The electrodes are designed for single-patient & single application use. Because of the adhesive nature of the conductive hydrogel, no securing materials are required to secure the device to the patient's skin.

510(k) Summary Vol 002 Doc 01 Page 2 of 8

Page 7

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

IV. INDICATIONS FOR USE

Electrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.

510(k) Summary Vol 002 Doc 01 Page 3 of 8

Page 8

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

V. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

The following tables shows similarities and differences of use, design, and material between our device and the predicate devices.

Comparison to predicate device:

DescriptionSubject DevicePredicate DeviceComparison
510K NumberK250841K180865-
Product CodeGXYGXYSE
Proprietary NameCutaneous ElectrodeCutaneous Electrode-
ManufacturerBaisheng Medical Co., Ltd.ZMI Electronics, Ltd.-
Intended useElectrotherapy Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. Electrotherapy Electrodes are intended to be used with marketed Electrical Stimulators, (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.ZMI Self-adhesive electrodes are intended for use as a reusable, conductive adhesive interface between the patient's skin and the marketed electrical stimulators (i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) or PGF (Pulsed Galvanic Stimulation) for transmitting electrical current. The electrodes are for OTC (Over-The-Counter) or Prescription use.SE

510(k) Summary Vol 002 Doc 01 Page 4 of 8

Page 9

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

Target PopulationAdultAdultSE
Design feature and materialFive layers: 1st layer: Insulation material: Tan fabric 2nd layer: Double sides adhesive tape 3rd layer: Conductive film (Carbon Film) 4th layer: Self-adhesive conductive hydrogel 5th layer: Plastic release film Connection: Leadwire/Snap buttonThree layers: 1. Insulation backing material: Fabric/Foam/ Tan fabric 2. Conductive film: Aluminum foil film /Carbon film/Carbon film coated with silver/ 3. Conductive hydrogelSE
Electrical ConnectionLeadwire Snap buttonLeadwire Snap button Magnetic buttonSE
Lead Wire connectorLeadwire connector .080" (2mm) female socket connectorLeadwire connector .080" (2mm) female socket connectorSE
Non-sterileNon-sterileNon-sterileSE
Disposable/ ReusableDisposableReusable-
Adhesive TypeSelf-adhesiveSelf-adhesiveSE
Environment of useFor OTC (Over-The-Counter) or Prescription use.For OTC (Over-The-Counter) or Prescription use.SE
BiocompatibilityComplies with ISO 10993Complies with ISO 10993SE
Single Patient UseYesYesSE

510(k) Summary Vol 002 Doc 01 Page 5 of 8

Page 10

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

DescriptionSubject DevicePredicate DeviceComparison
ShapeRectangular, RoundRound, Rectangular, Square, Oval, and ButterflySE The shape and size are related to the treatment area. Smaller area may cause higher current density, which leads to discomfort and burns. The current densities for any shaped electrodes exceeding 2mA r.m.s /cm2 may require the special attention of the user. The color is mainly for identification and marketing purpose, and does not raise any new concerns of safety or effectiveness.
ColorTanWhite, Red, Black, and Tan
A.C. Impedance≤ 160 ohms<200 ohmsSE

Discussion:
Our device is essentially identical to the predicate device in terms of indication for use, design feature and material, electrical connection, labeling, adhesive type, environment of use, biocompatibility and A.C. Impedance. The minor differences do not raise any safety and effectiveness concerns.

510(k) Summary Vol 002 Doc 01 Page 6 of 8

Page 11

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

VI. Summary of Non-Clinical Testing

Electrotherapy Electrodes are composed of Tan fabric, Double sided adhesive tape, Conductive film (Carbon Film), self-adhesive conductive hydrogel, Plastic release film, Leadwire/Snap button. Adhesive performance was evaluated by performing a skin adhesion test. Electrode stability was evaluated by assessing electrical performance under IFU normal use conditions.

Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use. The following tests were performed on the device.

  • Electrical safety per IEC 60601-1
  • IEC 60601-2-2 Edition 6.0 2017-03 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories, section 201.15.101.7
  • ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
  • Shelf life

In addition to the available device-specific guidelines, we have developed a method to test the packaging labeling, product appearance, impedance, dimension, connection between electrode and electrode wire, the cross-sectional area of the electrode connection wire, wire length and plug insertion and unplugging force surrounding the Electrotherapy Electrodes.

510(k) Summary Vol 002 Doc 01 Page 7 of 8

Page 12

Baisheng Medical Co., Ltd.
[Electrotherapy Electrodes] 510(k) Submission K250841

Biocompatibility testing was performed to verify the equivalent safety of the materials that are used. According to ISO 10993-1:2018 , we evaluated and conducted the compatibility test for the proposed device. The following tests were identified and done.

  • ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
  • ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation

All tests were verified to meet acceptance criteria.

VII. Clinical Test Conclusion

No clinical study is included in this submission

VIII. Conclusion

Verification testing of the Electrotherapy Electrodes included electrical and mechanical tests to show that it meets its target specifications over a range of operating and storage conditions. Verification and performance testing further demonstrated that it meets user needs as reflected in the functional specification.

The submitted Electrotherapy Electrodes have the same intended use and similar technological characteristics as the predicate devices. Moreover, information contained in the submission supplied demonstrates that any differences in their characteristics do not raise new concerns of safety or effectiveness. Thus, the submitted adhesive electrodes are substantially equivalent to the predicate devices.

510(k) Summary Vol 002 Doc 01 Page 8 of 8

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).