K Number
K250802

Validate with FDA (Live)

Date Cleared
2025-04-14

(31 days)

Product Code
Regulation Number
880.6860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STERRAD 100NX Sterilizer with ALLClear Technology is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.

The STERRAD 100NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer STANDARD cycle:

  • Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.

Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer FLEX Scope cycle:

  • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 1065 mm or shorter. A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.

Note: With the exception of the 1 x 1065 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 1065 mm flexible endoscopes were validated without any additional load.

The STERRAD 100NX EXPRESS Cycle is designed for surface sterilization of both metal and nonmetal medical devices at low temperatures.

  • It can sterilize instrument surfaces and instruments having diffusion-restricted spaces, such as the hinged portion of forceps and scissors
  • It can sterilize rigid and semi-rigid endoscopes without lumens.

Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs on the bottom shelf.

The STERRAD 100NX DUO Cycle is designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions:

  • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter
  • Accessory devices that are normally connected to a flexible endoscope during use
  • Flexible endoscopes without lumens

Note 1: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.

  • One Dual channel flexible endoscope with two polyethylene and Teflon (polytetrafluoroethylene) lumens that each have an inside diameter of 1 mm or larger and a length of 885 mm or shorter.

Note 2: The validation studies for DUO Cycle was performed using a validation load consisting of 1 dual channel endoscope with 1 mm diameter x 885 mm long and ≥2.2 mm diameter x ≤845 mm long PTFE/PE lumens and 1 dual channel endoscope with ≥1 mm diameter x ≤670 mm long and ≥2.2 mm diameter x ≤845 mm long PTFE/PE lumens

The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following:

  • Medical devices, including multi-channel endoscopes (including duodenoscopes), with no more than 4 channels, with lumen dimensions of ≥1mm x ≤1500mm in length, and ≥2mm x ≤1630mm in length.
  • One flexible endoscope per tray, and no more than two flexible endoscopes per cycle

Note1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf.

Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD User Guide "STERRAD Sterilizer Cycle Selection table" for ULTRA GI compatible duodenoscopes.

Device Description

The STERRAD 100NX Sterilization System with ALLClear Technology is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas plasma using electrical energy. The sterilization process is a multiphase process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to achieve sterilization. The hardware for the STERRAD 100NX Sterilizer consists of a sterilizer chamber and a variety of instruments and components which are housed in a covered frame. The sterilizer also uses accessories such as reusable instrument trays, biological indicator, chemical indicator products and sterilant cassettes.

The STERRAD 100NX Sterilizer has five cleared sterilization cycles: STANDARD, FLEX, EXPRESS, DUO and ULTRA GI Cycles.

The STERRAD 100NX Sterilizer with ALLClear Technology described within this submission expands the indications of the DUO Cycle of the STERRAD 100NX Sterilizer to include one dual-channel flexible endoscope with two polyethylene and Teflon (polytetrafluoroethylene) lumens that each have an inside diameter of 1 mm or larger and a length of 885 mm or shorter and with the weight of 3.4 kg (7.4 lbs). There are no other changes to the indications for use for the DUO cycle.

AI/ML Overview

This document is a 510(k) clearance letter for a medical device (STERRAD® 100NX Sterilizer with ALLClear™ Technology). It is NOT a study describing the acceptance criteria and performance of an AI/ML-based device.

Therefore, the requested information regarding AI/ML-specific testing (e.g., sample size, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, training set details) cannot be extracted from this document because it is irrelevant to a sterilization device clearance.

The document describes the acceptance criteria and test results for a physical medical device (a sterilizer), not a software or AI/ML diagnostic tool.

However, I can extract the information relevant to the sterilizer's performance and acceptance criteria as presented:


Acceptance Criteria and Device Performance for the STERRAD® 100NX Sterilizer with ALLClear™ Technology (DUO Cycle Claims Expansion)

This submission focuses on expanding the indications of the DUO Cycle of the STERRAD 100NX Sterilizer. The study demonstrates the safety and effectiveness of the device for sterilization of medical instruments.

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance CriteriaReported Device Performance
Sterilization VerificationSAL (Sterility Assurance Level) of 10⁻⁶ shall be demonstrated.Pass
Surface SterilizationAll test samples show no growth.Pass
Mated Surface SterilizationAll test samples show no growth.Pass
Growth InhibitionNo growth inhibition shall be indicated for processed samples.Pass
BiocompatibilityThe biological safety of materials shall be demonstrated following exposure to the sterilant agent.Pass
Simulated Use TestMicrobial performance should be demonstrated under simulated conditions.Pass
Device Functionality and Material CompatibilityDual channel flexible endoscopes shall remain within established functional specifications post processing.Pass

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not explicitly state the exact number of samples (e.g., number of sterilization cycles, number of instruments) used for each test. Instead, it refers to "test samples" and provides details on the type and dimensions of the medical devices tested.
    • DUO Cycle Validation Load: "The validation studies for DUO Cycle was performed using a validation load consisting of 1 dual channel endoscope with 1 mm diameter x 885 mm long and 2.2 mm diameter x 835 mm long PTFE/PE lumens, and 1 dual channel endoscope with 1 mm diameter x 670 mm long and 2.2 mm diameter x 845 mm long PTFE/PE lumens." This implies specific endoscopes were used for the validation tests.
    • Weight of Load: For the DUO Cycle, the validation load consisted of two flexible endoscopes with accessory devices weighing a total of 13.2 lbs.
  • Data Provenance: The data provenance (e.g., country of origin) is not specified. The studies are described as "non-clinical testing" and are conducted to support a 510(k) submission to the U.S. FDA, implying they were conducted in a controlled lab setting, likely by the manufacturer. The tests are retrospective in the sense that they are completed prior to submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable in the context of this device. Ground truth for sterilization efficacy is typically established through microbiological methods (e.g., spore kill, sterility assurance levels, no growth) following established sterilization standards and protocols (e.g., AAMI, ISO standards), rather than expert consensus on interpretation.

4. Adjudication method for the test set:

  • Not applicable. Results are quantitative (e.g., growth/no growth, SAL calculation, functional specifications) and determined by laboratory testing, not human interpretation requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is a sterilization device, not an AI/ML diagnostic tool used by human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a physical sterilizer. Its performance is evaluated in its designed operation, which is "standalone" in the sense that it performs the sterilization process automatically.

7. The type of ground truth used:

  • The ground truth for sterilization efficacy is based on microbiological testing (sterility/no growth) and physical performance parameters of the device (e.g., maintenance of function of the processed instruments). This involves challenging the sterilizer with biological indicators (known number of resilient microorganisms) and assessing the complete kill, as well as verifying the endoscopes' functionality after processing.

8. The sample size for the training set:

  • Not applicable. This is a physical device, not an AI/ML model that requires training data in the traditional sense. The "training" of the device is its design, engineering, and manufacturing in compliance with sterilization standards.

9. How the ground truth for the training set was established:

  • Not applicable. As above, this is a physical device. The "ground truth" for its development would be established through engineering specifications, previous sterilizer designs, and compliance with relevant industry standards for sterilization processes and validation.

FDA 510(k) Clearance Letter - STERRAD® 100NX Sterilizer with ALLClear™ Technology

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 14, 2025

Advanced Sterilization Products Inc.
Ayse Erkan
Sr. Regulatory Affairs Program Lead
33 Technology Drive
Irvine, California 92618

Re: K250802
Trade/Device Name: STERRAD® 100NX Sterilizer with ALLClear™ Technology (10104)
Regulation Number: 21 CFR 880.6860
Regulation Name: Ethylene Oxide Gas Sterilizer
Regulatory Class: Class II
Product Code: MLR
Dated: March 14, 2025
Received: March 14, 2025

Dear Ayse Erkan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250802 - Ayse Erkan Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250802 - Ayse Erkan Page 3

Sincerely,

Stephen A. Anisko -S
Digitally signed by Stephen A. Anisko -S
Date: 2025.04.14 14:44:09 -04'00'

for: Christopher K. Dugard
Division Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4


DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026

Indications for Use

See PRA Statement below.

510(k) Number (if known)
K250802

Device Name
STERRAD® 100NX Sterilizer with ALLClear™ Technology (10104)

Indications for Use (Describe)

The STERRAD 100NX Sterilizer with ALLClear Technology is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue.

The STERRAD 100NX Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer STANDARD cycle:

  • Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle.

Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer FLEX Scope cycle:

  • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 1065 mm or shorter. A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load.

Note: With the exception of the 1 x 1065 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 1065 mm flexible endoscopes were validated without any additional load.

The STERRAD 100NX EXPRESS Cycle is designed for surface sterilization of both metal and nonmetal medical devices at low temperatures.

  • It can sterilize instrument surfaces and instruments having diffusion-restricted spaces, such as the hinged portion of forceps and scissors
  • It can sterilize rigid and semi-rigid endoscopes without lumens.

Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs on the bottom shelf.

The STERRAD 100NX DUO Cycle is designed for sterilization of medical devices including most flexible endoscopes, with the following materials and dimensions:

  • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter
  • Accessory devices that are normally connected to a flexible endoscope during use
  • Flexible endoscopes without lumens

Note 1: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs.

  • One Dual channel flexible endoscope with two polyethylene and Teflon (polytetrafluoroethylene) lumens that each have an inside diameter of 1 mm or larger and a length of 885 mm or shorter.

Note 2: The validation studies for DUO Cycle was performed using a validation load consisting of 1 dual channel endoscope with 1 mm diameter x 885 mm long and ≥2.2 mm diameter x ≤845 mm long PTFE/PE lumens and 1 dual channel endoscope with ≥1 mm diameter x ≤670 mm long and ≥2.2 mm diameter x ≤845 mm long PTFE/PE lumens

The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following:

  • Medical devices, including multi-channel endoscopes (including duodenoscopes), with no more than 4 channels, with lumen dimensions of ≥1mm x ≤1500mm in length, and ≥2mm x ≤1630mm in length.
  • One flexible endoscope per tray, and no more than two flexible endoscopes per cycle

Note1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs),

FORM FDA 3881 (8/23) Page 1 of 2 PSC Publishing Services (301) 443-6740 EF

Page 5

one endoscope per shelf.

Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD User Guide "STERRAD Sterilizer Cycle Selection table" for ULTRA GI compatible duodenoscopes.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 2 of 2 PSC Publishing Services (301) 443-6740 EF

Page 6

510(k) Summary for K250802

Advanced Sterilization Products, Inc.

STERRAD™ 100NX Sterilizer with ALLClear™ Technology

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

General Information

Submitter Name: Advanced Sterilization Products, Inc.

Address: 33 Technology Drive
Irvine, CA 92618

Contact Person: Ayse Erkan
Senior Regulatory Affairs Program Lead
Telephone: (714)580-2880
Email: ayse.erkan@asp.com

Date Prepared: April 10, 2025

Devices Names

Proprietary Name:STERRAD™ 100NX Sterilizer with ALLClear™ Technology (10104)
Device Common Name:Hydrogen Peroxide Gas Plasma Sterilization System
Classification Name:Ethylene Oxide Gas Sterilizer
Product Code:MLR
Device Class:Class II
CFR Section:21 CFR 880.6860

Predicate Devices

STERRAD 100NX Sterilizer with ALLClear Technology K234082 cleared July 5, 2024.

Device Description

The STERRAD 100NX Sterilization System with ALLClear Technology is a self-contained stand-alone system of hardware and software designed to sterilize medical instruments and devices using a patented hydrogen peroxide gas plasma process. Hydrogen peroxide vapor is generated by injecting aqueous hydrogen peroxide into the vaporizer where the solution is heated and vaporized, introducing the vapor into the chamber under sub-ambient pressure and transforming the vapor into a gas plasma using electrical energy. The sterilization process is a multiphase process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to achieve sterilization. The hardware for the STERRAD 100NX Sterilizer consists of a sterilizer chamber and a variety of instruments and components which are housed in a covered frame. The sterilizer also uses accessories such as reusable instrument trays, biological indicator, chemical indicator products and sterilant cassettes.

Page 7

The STERRAD 100NX Sterilizer has five cleared sterilization cycles: STANDARD, FLEX, EXPRESS, DUO and ULTRA GI Cycles.

The STERRAD 100NX Sterilizer with ALLClear Technology described within this submission expands the indications of the DUO Cycle of the STERRAD 100NX Sterilizer to include one dual-channel flexible endoscope with two polyethylene and Teflon (polytetrafluoroethylene) lumens that each have an inside diameter of 1 mm or larger and a length of 885 mm or shorter and with the weight of 3.4 kg (7.4 lbs). There are no other changes to the indications for use for the DUO cycle.

Intended Use/Indications for Use

The intended use of the subject STERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI has not changed as a result of expanded indications for DUO Cycle. Refer to Table 1 below for a comparison of intended use and indications for use of the predicate and proposed devices.

Technological Characteristics

The technological characteristics associated with the sterilization process for the proposed STERRAD 100NX Sterilizer with ALLClear Technology with expanded indications for the Duo Cycle are equivalent to those of the previously cleared STERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI cycle; there are no modifications being introduced to alter existing sterilization cycles and other physical features of the subject device.

The predicate device continues to have the same technological characteristics, sterilization performance, and physical traits as the predicate device, STERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI cycle.

Table 1. Comparison between the subject device and predicate devices

FeatureSTERRAD 100NX Sterilizer with ALLClear Technology DUO cycle claims expansion Subject DeviceSTERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI Cycle K234082 Predicate DeviceComparison
Intended UseDesigned for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.Designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture.Identical
Indications for UseThe STERRAD 100NX Sterilizer with ALLClear Technology is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue. The STERRAD 100NX Sterilizer can sterilize instruments which have diffusion-The STERRAD 100NX Sterilizer with ALLClear Technology is designed for sterilization of both metal and nonmetal medical devices at low temperatures. The STERRAD sterilization process is a multiphase sterilization process that utilizes a combination of exposure to hydrogen peroxide vapor and plasma to safely sterilize medical instruments and materials without leaving toxic residue. The STERRAD 100NX Sterilizer can sterilize instruments which have diffusion-Similar, includes expanded claims for DUO cycle

Page 8

FeatureSTERRAD 100NX Sterilizer with ALLClear Technology DUO cycle claims expansion Subject DeviceSTERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI Cycle K234082 Predicate DeviceComparison
restricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer STANDARD cycle: • Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle. Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer FLEX Scope cycle: • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 1065 mm or shorter. A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load. Note: With the exception of the 1 x 1065 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 1065 mm flexible endoscopes were validated without any additional load. The STERRAD 100NX EXPRESS Cycle is designed for surface sterilization of both metal and nonmetal medical devices at low temperatures. • It can sterilize instrument surfaces and instruments having diffusion-restricted spaces, such as the hinged portion of forceps and scissors • It can sterilize rigid and semi-rigid endoscopes without lumens. Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs placed on the bottom shelf. The STERRAD 100NX DUO Cycle is designed for sterilization of medical devices including most flexiblerestricted spaces, such as the hinged portion of forceps and scissors. Medical devices with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer STANDARD cycle: • Single channel stainless steel lumens with an inside diameter of 0.7 mm or larger and a length of 500 mm or shorter. A maximum of ten single channel stainless steel lumens, five per tray per sterilization cycle. Medical devices, including most flexible endoscopes, with the following materials and dimensions can be processed in the STERRAD 100NX Sterilizer FLEX Scope cycle: • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and length of 1065 mm or shorter. A maximum of two flexible endoscopes, one per tray per sterilization cycle. No additional load. Note: With the exception of the 1 x 1065 mm flexible endoscopes, the validation studies were performed using a validation load consisting of two instrument trays each weighing 10.7 lbs. The 1 x 1065 mm flexible endoscopes were validated without any additional load. The STERRAD 100NX EXPRESS Cycle is an additional optional cycle designed for surface sterilization of both metal and nonmetal medical devices at low temperatures. • It can sterilize instrument surfaces and instruments having diffusion-restricted spaces, such as the hinged portion of forceps and scissors • It can sterilize rigid and semi-rigid endoscopes without lumens. Note: The validation studies for EXPRESS Cycle were performed using a validation load consisting of a single instrument tray weighing 10.7 lbs placed on the bottom shelf. The STERRAD 100NX DUO Cycle is an additional optional cycle designed for sterilization of medical devices including most flexible endoscopes,

Page 9

FeatureSTERRAD 100NX Sterilizer with ALLClear Technology DUO cycle claims expansion Subject DeviceSTERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI Cycle K234082 Predicate DeviceComparison
endoscopes, with the following materials and dimensions: • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter o Accessory devices that are normally connected to a flexible endoscope during use o Flexible endoscopes without lumens Note 1: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs. • One Dual channel flexible endoscope with two polyethylene and Teflon (polytetrafluoroethylene) lumens that each have an inside diameter of 1 mm or larger and a length of 885 mm or shorter. Note 2: The validation studies for DUO Cycle was performed using a validation load consisting of 1 dual channel endoscope with 1 mm diameter x 885 mm long and 2.2 mm diameter x 835 mm long PTFE/PE lumens, and 1 dual channel endoscope with 1 mm diameter x 670 mm long and 2.2 mm diameter x 845 mm long PTFE/PE lumens. The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following: • Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions having an inside diameter (ID) of ≥1mm x ≤1500mm in length, or ≥2mm ID x ≤1630mm in length.with the following materials and dimensions: • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter • Accessory devices that are normally connected to a flexible endoscope during use • Flexible endoscopes without lumens Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs. The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following: • Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions having an inside diameter (ID) of ≥1mm x ≤1500mm in length, or ≥2mm ID x ≤1630mm in length. • One flexible duodenoscope per tray, and no more than two

Page 10

FeatureSTERRAD 100NX Sterilizer with ALLClear Technology DUO cycle claims expansion Subject DeviceSTERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI Cycle K234082 Predicate DeviceComparison
endoscopes, with the following materials and dimensions: • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter o Accessory devices that are normally connected to a flexible endoscope during use o Flexible endoscopes without lumens Note 1: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs. • One Dual channel flexible endoscope with two polyethylene and Teflon (polytetrafluoroethylene) lumens that each have an inside diameter of 1 mm or larger and a length of 885 mm or shorter. Note 2: The validation studies for DUO Cycle was performed using a validation load consisting of 1 dual channel endoscope with 1 mm diameter x 885 mm long and 2.2 mm diameter x 835 mm long PTFE/PE lumens, and 1 dual channel endoscope with 1 mm diameter x 670 mm long and 2.2 mm diameter x 845 mm long PTFE/PE lumens. The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following: • Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions having an inside diameter (ID) of ≥1mm x ≤1500mm in length, or ≥2mm ID x ≤1630mm in length.with the following materials and dimensions: • Single channel polyethylene and Teflon (polytetrafluoroethylene) flexible endoscopes with an inside diameter of 1 mm or larger and a length of 875 mm or shorter • Accessory devices that are normally connected to a flexible endoscope during use • Flexible endoscopes without lumens Note: The validation studies for DUO Cycle were performed using a validation load consisting of two flexible endoscopes with their accessory devices weighing a total of 13.2 lbs. The STERRAD 100NX Sterilizer ULTRA GI Cycle is designed for sterilization of the following: • Hydrogen peroxide compatible flexible multi-channel duodenoscopes, with no more than 4 channels, with lumen dimensions having an inside diameter (ID) of ≥1mm x ≤1500mm in length, or ≥2mm ID x ≤1630mm in length. • One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf. Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD Sterilizer Cycle Selection table for ULTRA GI Cycle compatible duodenoscopes
Sterilization CyclesULTRA GI STANDARD FLEX EXPRESS DUOULTRA GI STANDARD FLEX EXPRESS DUOIdentical
DimensionsOne Door: 30.5" W x 70.9" H x 41.5" D (775 mm x 1800 mm x 1055 mm) Two Door: 30.5" W x 70.9" H x 43.1" D (775 mm x 1800 mm x 1095 mm)One Door: 30.5" W x 70.9" H x 41.5" D (775 mm x 1800 mm x 1055 mm) Two Door: 30.5" W x 70.9" H x 43.1" D (775 mm x 1800 mm x 1095 mm)Identical
H2O2 Concentration by Weight59-94% Depending on Cycle59-94% Depending on CycleIdentical
Number Of Half Cycles22Identical
Chamber Volume152L152LIdentical
Key Critical Process ParametersPressure, temperature & H2O2 concentration, H2O2 Dose & Exposure TimePressure, temperature & H2O2 concentration, H2O2 Dose & Exposure TimeIdentical
Use of Pre-Conditioning StepYesYesIdentical
Use of Secondary Step to Remove Residual H2O2 (technology)Yes (plasma)Yes(plasma)Identical

Page 11

FeatureSTERRAD 100NX Sterilizer with ALLClear Technology DUO cycle claims expansion Subject DeviceSTERRAD 100NX Sterilizer with ALLClear Technology and ULTRA GI Cycle K234082 Predicate DeviceComparison
• One flexible duodenoscope per tray, and no more than two flexible duodenoscope per cycle Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf. Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD User Guide "STERRAD Sterilizer Cycle Selection table" for ULTRA GI compatible duodenoscopesflexible duodenoscope per cycle Note 1: The STERRAD 100NX Sterilizer ULTRA GI Cycle was validated using a load weight of 15.4 lbs (2 x 7.7 lbs), one endoscope per shelf. Note 2: Only duodenoscopes that have been cleared as compatible with vaporized hydrogen peroxide are acceptable. Check STERRAD Sterilizer Cycle Selection table for ULTRA GI Cycle compatible duodenoscopes

Cycle Summary of Non-Clinical Testing

The following performance testing was conducted to verify the DUO Cycle device functionality.

Table 2. Performance Test Results – STERRAD 100NX Sterilizer

TestAcceptance CriteriaResults
Sterilization VerificationSAL of 10⁻⁶ shall be demonstrated.Pass
Surface SterilizationAll test samples show no growth.Pass
Mated Surface SterilizationAll test samples show no growth.Pass
Growth InhibitionNo growth inhibition shall be indicated for processed samples.Pass
BiocompatibilityThe biological safety of materials shall be demonstrated following exposure to the sterilant agent.Pass
Simulated Use TestMicrobial performance should be demonstrated under simulated conditions.Pass
Device Functionality and Material CompatibilityDual channel flexible endoscopes shall remain within established functional specifications post processing.Pass

Cycle Clinical Testing

No clinical data was generated in support of this submission.

Summary

The subject device, STERRAD 100NX Sterilizer with ALLClear Technology with DUO cycle claims expansion, and its predicate device utilize the same technology, sterilization cycles, and sterilization validation methods to sterilize medical devices. Based on the results of the performance testing, the change to the indications of the DUO Cycle does not raise any new questions of safety or effectiveness. ASP believes the subject STERRAD 100NX Sterilizer with ALLClear Technology with DUO cycle claims expansion, to be as safe and effective as the predicate device and reference devices.

Conclusion

Based on the intended use, technological characteristics, and non-clinical performance data; the subject device, STERRAD 100NX with ALLClear Technology with DUO Cycle claims expansion is as safe, as effective, and performs as well as the legally marketed predicate device cleared via K234082.

§ 880.6860 Ethylene oxide gas sterilizer.

(a)
Identification. An ethylene gas sterilizer is a nonportable device intended for use by a health care provider that uses ethylene oxide (ETO) to sterilize medical products.(b)
Classification. Class II (performance standards).