K Number
K250785

Validate with FDA (Live)

Date Cleared
2025-05-06

(53 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Single-use Flexible Ureteroscope is designed for use with Vathin Display Units, endotherapy accessories and other ancillary devices for the endoscopy and endoscopic surgery within urinary tract and kidney in adults.

Device Description

The Single-use Flexible Ureteroscope is intended to be used with Vathin Digital Video Monitor. During diagnosis and treatment, the Single-use Flexible Ureteroscope is inserted into the ureter or into the kidney or renal pelvis through the ureter, and the image sensor (CMOS) at the end of the Single-use Flexible Ureteroscope converts the received mucosal reflected light signals into electrical signals, transmitted to the Digital Video Monitor through the cable, the Digital Video Monitor receives the image signal from the endoscope and processes it, converts it into an image signal that can be displayed on the display screen, and finally presents it on the screen of the display.

The Single-use Flexible Ureteroscope is provided sterile (sterilized by EO) and intended to be single-use. It is for use in professional Healthcare Facility Environment.

AI/ML Overview

The provided FDA 510(k) clearance letter and its associated summary for the "Single-use Flexible Ureteroscope" does not contain information regarding a study with specific acceptance criteria and detailed device performance metrics in the way one might expect for an AI-enabled device or a diagnostic accuracy study. This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance data.

Therefore, the requested information cannot be fully extracted from the provided text. I will provide the acceptance criteria as implied by the non-clinical tests conducted to prove substantial equivalence, and note where specific information is missing based on your request.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria for each bench test, nor does it provide specific numerical performance results. Instead, it states that the device "complies with" or "met its design specifications" or "has been validated according to" relevant standards.

Acceptance Criterion (Implied)Reported Device Performance
Biocompatibility:
No CytotoxicityNo Cytotoxicity
No Irritation to SkinNo Irritation to Skin
No significant sensitizationNo significant evidence of sensitization
No Pyrogenicity(Test performed)
No Acute Systemic Toxicity(Test performed)
Electrical Safety:
Compliance with IEC 60601-1Complies with IEC 60601-1
Compliance with IEC 60601-2-18Complies with IEC 60601-2-18
Electromagnetic Compatibility (EMC):
Compliance with IEC 60601-1-2Complies with IEC 60601-1-2
Compliance with IEC 60601-4-2Complies with IEC 60601-4-2
Sterilization:
Validation according to ISO 11135Validated according to ISO 11135
Shelf Life & Packaging:
Packaging integrity validationValidated
Product performance after accelerated aging per ASTM F1980-21Carried out and presumably passed
Product performance after simulated transportation per ASTM D4169-22Carried out and presumably passed
Mechanical Performance:
Compliance with ISO 8600-1Performed in accordance with
Compliance with ISO 8600-4Performed in accordance with
Optical Performance:
Field of view (120°)(Tested to meet design specifications)
Direction of view (0°)(Tested to meet design specifications)
Depth of field(Tested to meet design specifications)
Geometric distortion(Tested to meet design specifications)
Image intensity uniformity(Tested to meet design specifications)
Color performance(Tested to meet design specifications)
Signal-To-Noise Ratio(Tested to meet design specifications)
Dynamic Range(Tested to meet design specifications)
Photobiological Safety:
Compliance with IEC 62471:2006Tested according to IEC 62471:2006

Missing Information Details:

  1. Sample size used for the test set and the data provenance: This information is not provided in the document. The studies were non-clinical bench tests and biocompatibility tests, not clinical trials with patient data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This is not applicable as the studies were non-clinical bench tests and material safety tests, not studies requiring expert interpretation of clinical images or data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for non-clinical bench tests.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document describes a medical instrument, not an AI-enabled device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For non-clinical tests, the "ground truth" is typically defined by the specified performance limits or standards outlined in the applicable international standards (e.g., ISO, IEC, ASTM). For biocompatibility, it's the lack of adverse biological reactions.
  7. The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
  8. How the ground truth for the training set was established: Not applicable, as this is not an AI/machine learning device.

In summary, the provided FDA document details the non-clinical testing performed to demonstrate the safety and effectiveness of a physical medical device (a single-use flexible ureteroscope) for substantial equivalence to a predicate device. It does not contain information related to diagnostic accuracy studies, AI performance, or clinical trials with human readers or patient data.

FDA 510(k) Clearance Letter - Single-use Flexible Ureteroscope

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 6, 2025

Hunan Vathin Medical Instrument Co.Ltd
Charlene Du
RA Manager
1/F, Building 12,Innovation and Entrepreneurship Service Ctr
No 9 Chuanqi west road, Jiuhua Economic Development Zone
Xiangtan, Hunan 411100
CHINA

Re: K250785
Trade/Device Name: Single-use Flexible Ureteroscope
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: II
Product Code: FGB
Dated: March 14, 2025
Received: March 14, 2025

Dear Charlene Du:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

May 6, 2025

Hunan Vathin Medical Instrument Co.Ltd
Charlene Du
RA Manager
1/F, Building 12,Innovation and Entrepreneurship Service Ctr
No 9 Chuanqi west road, Jiuhua Economic Development Zone
Xiangtan, Hunan 411100
CHINA

Re: K250785
Trade/Device Name: Single-use Flexible Ureteroscope
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: II
Product Code: FGB
Dated: March 14, 2025
Received: March 14, 2025

Dear Charlene Du:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 3

K250785 - Charlene Du Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

Page 4

K250785 - Charlene Du Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark R. Kreitz -S

for Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 5

K250785

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250785

Device Name
Single-use Flexible Ureteroscope

Indications for Use (Describe)
The Single-use Flexible Ureteroscope is designed for use with Vathin Display Units, endotherapy accessories and other ancillary devices for the endoscopy and endoscopic surgery within urinary tract and kidney in adults.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 6

Premarket Notification 510(k) Submission

K250785
Page 1 of 5

510(k) Summary

I Submitter

Device submitter: Hunan Vathin Medical Instrument Co., Ltd.

Address: 1/F, Building 12, Innovation and Entrepreneurship Service Center, No 9 Chuanqi west road, Jiuhua Economic Development Zone, 411100 Xiangtan, Hunan, China

Contact person: Du Jing
Title: RA Manager
Phone: +86-18915069265
E-mail: charlene@vathin.com

Date prepared: March 12, 2025

II Device

Name of Device: Single-use Flexible Ureteroscope
Common name: Single-use Flexible Ureteroscope
Models: US-S2602, US-E2602, US-S2602R, US-E2602R
Classification name: Ureteroscope and Accessories, Flexible/rigid (21 CFR 876.1500)
Regulatory class: Class II
Product code: FGB

III Predicate Device

Trade name: Uscope
Classification name: Ureteroscope and Accessories, Flexible/rigid (21 CFR 876.1500)
Regulatory class: Class II
Product code: FGB
Submitter: Zhuhai Pusen Medical Technology Co., Ltd.
510(k) number: K172098

IV Device description

The Single-use Flexible Ureteroscope is intended to be used with Vathin Digital Video Monitor. During diagnosis and treatment, the Single-use Flexible Ureteroscope is inserted into the ureter or into the kidney or renal pelvis through the ureter, and the image sensor (CMOS) at the end of the Single-use Flexible Ureteroscope converts the received mucosal reflected light signals into electrical signals,

Page 7

Premarket Notification 510(k) Submission

K250785
Page 2 of 5

transmitted to the Digital Video Monitor through the cable, the Digital Video Monitor receives the image signal from the endoscope and processes it, converts it into an image signal that can be displayed on the display screen, and finally presents it on the screen of the display.

The Single-use Flexible Ureteroscope is provided sterile (sterilized by EO) and intended to be single-use. It is for use in professional Healthcare Facility Environment.

V Indications for use

The Single-use Flexible Ureteroscope is designed for use with Vathin Display Units, endotherapy accessories and other ancillary devices for the endoscopy and endoscopic surgery within urinary tract and kidney in adults.

VI Comparison of technological characteristics with the predicate devices

The Single-Use Flexible Ureteroscope is similar to the predicate device in the following areas:

  • Intended use
  • Operation Principle
  • Design and performance specifications
  • Digital video technology and illumination source
  • It allows for irrigation
  • It is single-use and delivered sterile

The Single-Use Flexible Ureteroscope is different to the predicate device in the following areas:

  • Larger bending angle, minimum inner diameter of instrument channel
  • Longer working length
  • Smaller maximum outer diameter

The differences between the Single-Use Flexible Ureteroscope and predicate device do not alter suitability of the subject device for its intended use.

Device featureSubject DevicePredicate Device
Trade NameSingle-Use Flexible UreteroscopeUscope (K172098)
Classification NameEndoscope and accessoriesEndoscope and accessories
Product CodeFGBFGB
Regulation Number21 CFR 876.150021 CFR 876.1500
Intended useThe Single-use Flexible Ureteroscope is designed for use with Vathin Display Units, endotherapy accessories andThis instrument has been designed to be used with endo-therapy accessories such as a biopsy forceps and other

Page 8

Premarket Notification 510(k) Submission

K250785
Page 3 of 5

Device featureSubject DevicePredicate Device
other ancillary devices for the endoscopy and endoscopic surgery within urinary tract and kidney in adults.ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney.
Compatible Display UnitDigital Video Monitor, DVM-B1, K231135Eview, UTV100
Application fieldThe device is for use in a hospital or qualified medical institution.The Medical Video Endoscope system is for use in a hospital or qualified medical institution.
Intended userThe device is only to be used by skilled medical staff trained in clinical endoscopic techniques and procedures.The system is only to be used by skilled physicians trained in clinical endoscopic techniques and procedures.
Patient populationAdultsAdults
Scope typeFlexibleFlexible
Field of view120°120°
Direction of view
Bending angleUp: 285° Down: 285°Up: 270° Down: 270°
Maximum outer diameter of insertion part (mm)2.83.2
Minimum inner diameter of instrument channel (mm)1.21.0
Working length (mm)700650
Digital video technologyCMOSCMOS
Illumination sourceLEDLED
Single-useYesYes
BiocompatibilityNo CytotoxicityNo Cytotoxicity

Page 9

Premarket Notification 510(k) Submission

K250785
Page 4 of 5

Device featureSubject DevicePredicate Device
No Irritation to Skin No significant evidence of sensitizationNo Irritation to Skin No significant evidence of sensitization
SterilizationEOEO

VII Summary of Non-clinical tests:

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the Single-Use Flexible Ureteroscope was conducted in accordance with "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'" and International Standard ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The Single-use Flexible Ureteroscope is considered for mucosal membrane contacting for a duration of less than 24 hours. The following tests were performed as recommended: Cytotoxicity, Irritation, Sensitization, Pyrogenicity and Acute systemic toxicity.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the device. The device complies with the IEC 60601-1 and IEC60601-2-18 standards for safety and the IEC 60601-1-2, IEC 60601-4-2 standards for EMC.

Sterilization validation

EO sterilization of the Single-Use Flexible Ureteroscope has been validated according to ISO 11135.

Shelf life and Packaging testing

Shelf life and sterile barrier system of the Single-Use Flexible Ureteroscope have been validated. Packaging integrity testing and product performance testing were carried out following the accelerated aging of the final devices per ASTM F1980-21 and simulated transportation distribution according to ASTM D4169-22.

Bench performance testing

Bench testing was performed to ensure that the device met its design specifications and is substantially equivalent to the predicate device. The following performance testing was conducted.

Page 10

Premarket Notification 510(k) Submission

K250785
Page 5 of 5

Bench testing - Mechanical Performance

The mechanical performance testing was conducted in accordance with applicable parts of ISO 8600-1, ISO 8600-4.

Bench testing – Optical Performance

  • Field of view
  • Direction of view
  • Depth of field
  • Geometric distortion
  • Image intensity uniformity
  • Color performance
  • Signal-To-Noise Ratio
  • Dynamic Range

Bench testing – Photobiological Safety

The subject device was tested according to FDA recognized standards IEC 62471:2006.

VIII Conclusion

The device is substantially equivalent to predicate device. The non-clinical performance testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.