(285 days)
The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40kg. In addition, it is intended for use in healthy adult patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the supervision of a healthcare professional.
External counter-pulsation (ECP) as a general technology involves the application of pressure to the patient's legs using inflatable cuffs or sleeves to enhance perfusion at the heart. ECP using the Wave PRO involves wrapping the patient's calves in sleeves containing inflatable bladders, which compress the calves in sync with the patient's cardiac cycle. During diastole, a series of five bladders within the sleeves inflate sequentially from the patient's ankles towards their knees. This compression results in diastolic augmentation, which is observable from the pulse data obtained using a finger-tip PPG sensor. During the diastolic phase, the heart is relaxed, allowing for coronary perfusion to the cardiac muscle. By compressing the inflatable bladders during this portion of the cardiac cycle, aortic retrograde flow is increased, allowing for enhanced coronary perfusion.
The Pression Wave PRO System consists of three primary subsystems: 1) the Compression Control System, which controls the flow of compressed air to the sleeves, measures the phase of the cardiac cycle, and provides the graphical user interface (GUI); 2) Patient-Worn Calf Sleeves, which direct the compressive force of compressed air, delivered comfortably and consistently, to the patient's calf muscles; and 3) the Compression Timing Algorithm (Software), which determines when the patient's heart is in diastole, using a photoplethysmography (PPG) sensor, and controls the timed application of compression to the patient's calf muscles.
N/A
FDA 510(k) Clearance Letter - Pression Wave PRO External Counter-Pulsation System
Page 1
December 22, 2025
Pression, Inc.
℅ Julie Broderick
Principal Consultant
Broderick Regulatory Consulting, LLC
7 Kendall Street
Winchester, Massachusetts 01890
Re: K250756
Trade/Device Name: Pression Wave PRO External Counter-Pulsation System
Regulation Number: 21 CFR 870.5225
Regulation Name: External counter-pulsating device
Regulatory Class: Class II
Product Code: DRN
Dated: October 10, 2025
Received: October 14, 2025
Dear Julie Broderick:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K250756 - Julie Broderick Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Page 3
K250756 - Julie Broderick Page 3
Sincerely,
Nicole M. Gillette -S
Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support,
Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
Please provide the device trade name(s).
Pression Wave™ PRO External Counter-Pulsation System
Please provide your Indications for Use below.
The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40kg. In addition, it is intended for use in healthy adult patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the supervision of a healthcare professional.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Pression Wave™ PRO External Counter-Pulsation System Page 9 of 69
Page 5
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
I. Submitter Information
Company: Pression, Inc.
190 West Lincoln Highway
Coatesville, PA 19320 USA
Contact: Adam Salamon
President & CEO
Pression, Inc.
Phone No.: +1 215-326-9525
Email: adam@pression.com
Date Prepared: 22 December 2025
510(k) No.: K250756
II. Name of Device
Device Trade Name: Pression Wave™ PRO External Counter-Pulsation System
Model No.: WP-01
Classification Name: Device, Counter-Pulsating, External
Product Code: DRN
Regulation Number: 21 CFR 870.5225, External counter-pulsating device
Device Class: II
Panel Identification: Cardiovascular
III. Predicate Devices
All predicate and reference devices listed below are in the same Product Code:
Classification Name: Device, Counter-Pulsating, External
Product Code: DRN
Regulation Number: 21 CFR 870.5225, External counter-pulsating device
Device Class: II
Panel Identification: Cardiovascular
Predicate Device
Manufacturer: Applied Cardiac Systems, Inc. (ACS)
Trade Name: ACS Model NCP-2 External Counterpulsation System
510(k) No.: K042413
Page 6
Reference Devices
Manufacturer: Vamed Medical Instrument Co., Ltd
Trade Name: External Counterpulsation System, ECP MC-3
510(k) No.: K190683
Manufacturer: CardioMedics, Inc.
Trade Name: CardiAssist Counterpulsation System, Mark 3000
510(k) No.: K022107
Manufacturer: Renew Group Pte Ltd. / Stage 2 Innovations
Trade Name: Renew NCP-5 External Counterpulsation System
510(k) No.: K152115
IV. Device Description
External counter-pulsation (ECP) as a general technology involves the application of pressure to the patient's legs using inflatable cuffs or sleeves to enhance perfusion at the heart. ECP using the Wave PRO involves wrapping the patient's calves in sleeves containing inflatable bladders, which compress the calves in sync with the patient's cardiac cycle. During diastole, a series of five bladders within the sleeves inflate sequentially from the patient's ankles towards their knees. This compression results in diastolic augmentation, which is observable from the pulse data obtained using a finger-tip PPG sensor. During the diastolic phase, the heart is relaxed, allowing for coronary perfusion to the cardiac muscle. By compressing the inflatable bladders during this portion of the cardiac cycle, aortic retrograde flow is increased, allowing for enhanced coronary perfusion.
The Pression Wave PRO System consists of three primary subsystems: 1) the Compression Control System, which controls the flow of compressed air to the sleeves, measures the phase of the cardiac cycle, and provides the graphical user interface (GUI); 2) Patient-Worn Calf Sleeves, which direct the compressive force of compressed air, delivered comfortably and consistently, to the patient's calf muscles; and 3) the Compression Timing Algorithm (Software), which determines when the patient's heart is in diastole, using a photoplethysmography (PPG) sensor, and controls the timed application of compression to the patient's calf muscles.
V. Indications for Use
The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40kg. In addition, it is intended for use in healthy adult patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use under the supervision of a healthcare professional.
Page 7
VI. Comparison of Technological Characteristics
| Characteristic | New DeviceWave PRO External Counter-Pulsation System | Predicate DeviceModel NCP-2 External Counterpulsation System | Reference DeviceModel ECP-MC3 External Counterpulsation System | Reference DeviceCardiAssist Counterpulsation System, Mark 3000 | Reference DeviceRenew NCP-5 External Counterpulsation System | Substantial Equivalence |
|---|---|---|---|---|---|---|
| Manufacturer | Pression, Inc. | Applied Cardiac Systems, Inc. | Vamed Medical Instrument Co., Ltd. | CardioMedics, Inc. | Renew Group Pte Ltd. / Stage 2 Innovations, Inc. | N/A |
| 510(k) Number | K250756 | K042413 | K190683 | K022107 | K152115 | N/A |
| Regulation No. | 21 CFR 870.5225 | 21 CFR 870.5225 | 21 CFR 870.5225 | 21 CFR 870.5225 | 21 CFR 870.5225 | Substantially EquivalentSame as predicates |
| Regulation Name | External counter-pulsating device | External counter-pulsating device | External counter-pulsating device | External counter-pulsating device | External counter-pulsating device | Substantially EquivalentSame as predicates |
| Regulatory Class | Class II | Class II | Class II | Class II | Class II | Substantially EquivalentSame as predicates |
| Product Classification Code | DRN | DRN | DRN | DRN | DRN | Substantially EquivalentSame as predicates |
| Intended Use / Indications for Use | The Pression Wave PRO External Counter-Pulsation System is indicated for the treatment of chronic stable angina (CSA) that is refractory to optimal anti-anginal medical therapy and without options for revascularization in adult patients that weigh greater than 40 kg. In addition, it is intended for use in healthy adult patients to provide improvement in | ACS Model NCP-2 External Counterpulsation Device is a noninvasive external Counterpulsation device intended for the use in the treatment of patients with stable or unstable angina pectoris, acute myocardial infarction, cardiogenic shock or congestive heart failure. | The ECP-MC3 device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvements in vasodilation, and increased blood flow. It is intended for use under the | The CardioMedics, Inc., CardiAssist Counterpulsation System, Mark 3000, is intended to provide external counterpulsation (ECP) for the treatment of ischemic heart disease by increasing perfusion during diastole in people with chronic angina pectoris, congestive heart failure, myocardial infarction and cardiogenic shock. Use of this | The Renew NCP-5 External Counterpulsation System is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvement in vasodilation, increased VO2, and increased blood flow. It is intended for use | Substantially EquivalentSubstantially the same as predicates¹ |
¹ The ACS Model NCP-2 External Counterpulsation System and CardiAssist Counterpulsation System Mark 3000 were cleared prior to FDA's reclassification of ECP systems for indications other than chronic stable angina to class III (PMA) in December 2013. Both device indication/intended use includes chronic or stable angina pectoris within the cleared indication, which is the same as the Pression Wave device (chronic stable angina).
Page 8
| Characteristic | New DeviceWave PRO External Counter-Pulsation System | Predicate DeviceModel NCP-2 External Counterpulsation System | Reference DeviceModel ECP-MC3 External Counterpulsation System | Reference DeviceCardiAssist Counterpulsation System, Mark 3000 | Reference DeviceRenew NCP-5 External Counterpulsation System | Substantial Equivalence |
|---|---|---|---|---|---|---|
| vasodilation, increased VO2, and increased blood flow.It is intended for use under the supervision of a healthcare professional. | oversight of a healthcare professional. | device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function. | under the oversight of a healthcare professional. | |||
| Device Description / Operating Principle | The application of pressure to the patient's legs using inflatable cuffs or sleeves surrounding the calves, inflated sequentially from the ankle to the knee and synchronized to the diastolic phase of the cardiac cycle, results in diastolic augmentation and systolic unloading, which is observable from the pulse data obtained using a finger-tip PPG sensor. During the diastolic phase, the heart is relaxed, allowing for coronary perfusion to the cardiac muscle. By compressing the inflatable sleeves during this portion of the cardiac cycle, aortic retrograde flow is increased, allowing for enhanced coronary perfusion and systolic unloading. | When the heart is in its relaxed state during the diastolic period, pressure is applied sequentially from the calves, to the lower thighs, to the upper thighs and buttocks, forcing blood back to the heart, increasing coronary perfusion pressure and coronary blood flow (diastolic augmentation), as well as venous return. Immediately before the heart begins to eject blood during the next systolic phase, the Cuffs are rapidly deflated and all externally applied pressure is eliminated. The vasculature in the lower extremities re-conforms and is able to receive the output of the heart with lessened resistance, thereby reducing systolic pressure and the workload of the heart (decreased after-load).Although not specifically mentioned in the predicate 510(k) summary, systolic unloading is a known hemodynamic effect of ECP. | The ECP-MC3 system is computer-controlled system that inflates and deflates three pairs of air cuffs in synchronization with the patient's cardiac cycle. The three pairs of cuffs are wrapped around the calves, lower thighs, and upper thighs/buttocks of the patients. As diastole begins, the cuffs inflate separately in sequence from the calves, to the lower thighs, to the upper thighs including the lower buttocks. This inflation sequence generates and impels a counter-pulsation wave, increasing diastolic blood pressure (diastolic augmentation), coronary perfusion pressure and coronary blood flow, and raising cardiac output.Although not specifically mentioned in the predicate 510(k) summary, systolic unloading is a known hemodynamic effect of ECP | The CardioMedics CardiAssist ECP System is a non-invasive circulatory assist device which provides increased circulation via external counterpulsation (ECP) for the treatment of ischemic heart disease including congestive heart failure, chronic angina pectoris, acute myocardial infarction and cardiogenic shock. External counterpulsation therapy improves cardiac function by enhancing the perfusion of the coronary vasculature, the development of coronary collateral circulation, and by reducing the workload of the heart.The CardiAssist ECP System consists of the portable console containing the computer and pumps with a touch screen for user interface, an integral strip chart recorder, and leg cuffs and hoses. Additional components provided with the System include a finger plethysmograph and 3-lead ECG cable and leads.This CardiAssist ECP System sequentially compresses the legs from the calves, thighs and | The Renew™ NCP-5 External Counterpulsation System (ECS) is comprised of the following major components, a switch panel, laptop PC computer, a foldable treatment bed, main unit, and a set of patient cuffs. The device is a microprocessor-controlled system that sequentially inflates then deflates three pairs of air cuffs, which compress vascular beds in the muscles of the calves, thighs, and buttocks in synchrony with the heart cycle to achieve the desired therapy. | Substantially EquivalentSame as predicates except that the Wave PRO utilizes sleeves on the calves only, whereas the predicate devices utilize cuffs on the calves, thighs and buttocks/hips.Clinical data included in the 510(k) demonstrate that the Wave PRO System produces the same or better hemodynamic effect as the predicate device.The Wave PRO is portable and substantially equivalent in this feature to the CardiAssist and Renew ECP reference devices. |
Page 9
| Characteristic | New DeviceWave PRO External Counter-Pulsation System | Predicate DeviceModel NCP-2 External Counterpulsation System | Reference DeviceModel ECP-MC3 External Counterpulsation System | Reference DeviceCardiAssist Counterpulsation System, Mark 3000 | Reference DeviceRenew NCP-5 External Counterpulsation System | Substantial Equivalence |
|---|---|---|---|---|---|---|
| buttocks, 40 milliseconds apart, by inflating three sets of flexible fabric cuffs during diastole. This results in movement of blood from the legs to the heart and entire upper body. Pressure, up to 310 mmHg, is applied with the timing and duration of each pulse, synchronized with the patient's ECG. When properly triggered, the pressure pulses applied to the vascular bed of the legs and buttocks transmit retrograde pressure through the entire vascular system. At the aorta, the aortic valve prevents retrograde flow into the left ventricle. Thus, a peak pulse of diastolic pressure occurs at or above systolic levels which increases the driving pressure in the coronary vasculature | ||||||
| Device Description / Operating Principle | The application of pressure to the patient's legs using inflatable cuffs or sleeves surrounding the calves, inflated sequentially from the ankle to the knee and synchronized to the diastolic phase of the cardiac cycle, results in diastolic augmentation and systolic unloading, which is observable from the pulse data obtained using a finger-tip PPG sensor. During the diastolic phase, the heart is relaxed, allowing for coronary perfusion to the cardiac muscle. By compressing the inflatable sleeves during this portion of the cardiac cycle, aortic retrograde flow is increased, allowing for enhanced coronary perfusion and systolic unloading. | When the heart is in its relaxed state during the diastolic period, pressure is applied sequentially from the calves, to the lower thighs, to the upper thighs and buttocks, forcing blood back to the heart, increasing coronary perfusion pressure and coronary blood flow (diastolic augmentation), as well as venous return. Immediately before the heart begins to eject blood during the next systolic phase, the Cuffs are rapidly deflated and all externally applied pressure is eliminated. The vasculature in the lower extremities re-conforms and is able to receive the output of the heart with lessened resistance, thereby reducing systolic pressure and the workload of the heart (decreased after-load).Although not specifically mentioned in the predicate 510(k) summary, systolic unloading is a known hemodynamic effect of ECP. | The ECP-MC3 system is computer-controlled system that inflates and deflates three pairs of air cuffs in synchronization with the patient's cardiac cycle. The three pairs of cuffs are wrapped around the calves, lower thighs, and upper thighs/buttocks of the patients. As diastole begins, the cuffs inflate separately in sequence from the calves, to the lower thighs, to the upper thighs including the lower buttocks. This inflation sequence generates and impels a counter-pulsation wave, increasing diastolic blood pressure (diastolic augmentation), coronary perfusion pressure and coronary blood flow, and raising cardiac output.Although not specifically mentioned in the predicate 510(k) summary, systolic unloading is a known hemodynamic effect of ECP | The CardioMedics CardiAssist ECP System is a non-invasive circulatory assist device which provides increased circulation via external counterpulsation (ECP) for the treatment of ischemic heart disease including congestive heart failure, chronic angina pectoris, acute myocardial infarction and cardiogenic shock. External counterpulsation therapy improves cardiac function by enhancing the perfusion of the coronary vasculature, the development of coronary collateral circulation, and by reducing the workload of the heart.The CardiAssist ECP System consists of the portable console containing the computer and pumps with a touch screen for user interface, an integral strip chart recorder, and leg cuffs and hoses. Additional components provided with the System include a finger plethysmograph and 3-lead ECG cable and leads.This CardiAssist ECP System sequentially compresses the legs from the calves, thighs and buttocks, 40 milliseconds apart, by inflating three sets of flexible fabric cuffs during diastole. This results in movement of blood from the legs to the heart and entire upper body. Pressure, up to 310 mmHg, is applied with the timing and duration of each pulse, synchronized with the patient's ECG. When properly triggered, the pressure pulses applied to the vascular bed of the legs and buttocks transmit retrograde pressure through the entire vascular system. At the aorta, the aortic valve prevents retrograde flow into the left ventricle. Thus, a peak pulse of diastolic pressure occurs at or above systolic levels which increases the driving pressure in the coronary vasculature | The Renew™ NCP-5 External Counterpulsation System (ECS) is comprised of the following major components, a switch panel, laptop PC computer, a foldable treatment bed, main unit, and a set of patient cuffs. The device is a microprocessor-controlled system that sequentially inflates then deflates three pairs of air cuffs, which compress vascular beds in the muscles of the calves, thighs, and buttocks in synchrony with the heart cycle to achieve the desired therapy. | Substantially EquivalentSame as predicates except that the Wave PRO utilizes sleeves on the calves only, whereas the predicate devices utilize cuffs on the calves, thighs and buttocks/hips.Clinical data included in the 510(k) demonstrate that the Wave PRO System produces the same or better hemodynamic effect as the predicate device.The Wave PRO is portable and substantially equivalent in this feature to the CardiAssist and Renew ECP reference devices. |
Page 10
| Characteristic | New DeviceWave PRO External Counter-Pulsation System | Predicate DeviceModel NCP-2 External Counterpulsation System | Reference DeviceModel ECP-MC3 External Counterpulsation System | Reference DeviceCardiAssist Counterpulsation System, Mark 3000 | Reference DeviceRenew NCP-5 External Counterpulsation System | Substantial Equivalence |
|---|---|---|---|---|---|---|
| Portability | Portable (34 kg) | Stationary | Stationary | Portable (~400 kg) | Portable (~250 kg) | Substantially EquivalentSame as portable reference devices but much lighter |
| Treatment pressure range | 41-50 kPa (6-7.25 PSI)(automatically set) | 13-41 kPa (1.9-6 PSI)(manually set) | 0-50 kPa (0-7.25 PSI)(manually set) | 11-41 kPa (1.6-6 PSI)(manually set) | 0-41 kPa (0-6 PSI)(manually set) | Substantially EquivalentSame as ECP-MC3 reference predicate |
| Maximum safety pressure | 55.2 kPa (8 PSI) | 55.2 kPa (8 PSI) | 55 kPa (8 PSI) | Unreported | Unreported | Substantially EquivalentSame as predicate and ECP-MC3 reference device |
Page 11
| Characteristic | New DeviceWave PRO External Counter-Pulsation System | Predicate DeviceModel NCP-2 External Counterpulsation System | Reference DeviceModel ECP-MC3 External Counterpulsation System | Reference DeviceCardiAssist Counterpulsation System, Mark 3000 | Reference DeviceRenew NCP-5 External Counterpulsation System | Substantial Equivalence |
|---|---|---|---|---|---|---|
| Software validation | Per FDA guidance, Content of Premarket Submissions for Device Software Functions, June 2023 | Per FDA guidance, General Principles of Software Validation, January 2002 | Per FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (version unknown) | Unknown | The software and firmware used in the system was validated per FDA's guidance document. In all instances, the NCP-5 system functioned as intended. | Substantially EquivalentSame as predicate |
| Electrical safety | IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance | Per IEC 601-1-4, Medical Electrical Equipment Part 1: General requirements for safety. Collateral Standard: Programmable electrical medical systems, 1996-05 | IEC60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety, 2005+A1:2012 | Unknown | Electrical safety and EMC compatibility testing per IEC 60601 was performed to ensure appropriate electrical performance. | Substantially EquivalentSame as reference devices, using most recent standard |
| Electromagnetic compatibility (EMC) | IEC EN 60601-1-2, Edition 4.1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | Unknown | IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, 2014 | Unknown | Electrical safety and EMC compatibility testing per IEC 60601 was performed to ensure appropriate electrical performance. | Substantially EquivalentSame as reference devices, using most recent standard |
| Usability | Per FDA Guidance for Applying Human Factors and Usability Engineering to Medical Devices, February 2016 | Unknown | IEC 60601-1-6 Edition 3.1 Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability | Unknown | Unknown | SimilarSimilar to reference predicate, using most recent FDA guidance; achieves same goal |
| Clinical testing | A clinical investigation of the Wave PRO has demonstrated substantial equivalence to the predicate device. | Yes, 510(k) summary states:An Independent Review Board has overseen the clinical investigation of the External Counterpulsation Device. Results have demonstrated substantial equivalence to the predicate device. | Clinical performance is not deemed necessary. | Unknown | A literature review regarding the use of external counterpulsation in healthy patients was provided to support this additional indication. | Substantially EquivalentSame as predicate |
Page 12
VII. Performance or Clinical Testing
Bench performance testing to verify the design of the Pression Wave PRO System, all of which PASSED, included:
- Compression and Sensor Accuracy
- Product Safety & Electrical Testing
- EMC Testing
- Software Verification
- End of Line / Performance Test
- Label Verification
- Cycle Verification
- Service Life Verification (system and compression sleeves)
- Reprocessing Validation (compression sleeves)
- Summative Usability Testing
- Biocompatibility Assessment
The following standards were applied:
- IEC 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 62304, Software lifecycle processes
- ISO 10993-1, Biological evaluation of medical devices – Part 1: evaluation and testing within a risk management process
- ISO 14971, Medical devices — Application of risk management to medical devices
- ISO 15223-1, Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements
The WaveCRESST clinical study demonstrated that the Wave PRO ECP System is non-inferior to the predicate device, ACS Model NCP-02 ECP System, with respect to the primary effectiveness endpoint, diastolic augmentation (DA) ratio. Details of the study are as follows:
- Level of Evidence: Randomized crossover-control study comparing Wave PRO ECP to ACS Model NCP-02 ECP (sequence of device use was randomized).
- Study Population: Volunteer participants at least 35 years of age with a history of mild to moderate coronary artery disease, but otherwise healthy.
- Location of Study: United States (one site)
- Safety Endpoints: Rates of device-related adverse events and device-related serious adverse events
Page 13
-
Primary Effectiveness Endpoint: DA ratio during the 8th minute of compressions, as calculated by the blinded Independent Waveform Evaluator. The difference of the average DA ratio attained using the Wave PRO and the predicate device was calculated for all randomized study participants with paired ECP data., with randomized assignment of ECP device sequence and unbiased (blinded) waveform scoring of primary effectiveness endpoint, DA ratio, and secondary endpoint systolic unloading.
-
Patient Accountability:
| Randomization | |||
|---|---|---|---|
| Total | Wave PRO First | ACS First | |
| Screened | 60 | ||
| Screen fail | 8 | ||
| Randomized (Intent to Treat) | 52 | 26 | 26 |
| Modified Intent to Treat Cohort & Safety Cohort (mITT)(received any ECP compressions) | 51 | 25 | 26 |
| Completed Study | 48 | 22 | 26 |
| Per Protocol (PP) Cohort(received any compressions from both ECP treatments with no major protocol deviations) | 48 | 22 | 26 |
-
Baseline Demographic Information (mITT population, N=51):
- Sex (% female): 47.1%
- Age at procedure (mean ± SD, range): 60.9 ± 9.1 (41.0, 78.0)
- Race: 2% American Indian/Alaska Native; 2.0% Asian; 47.1% Black/African American; 45.1% White; 2.0% Other; 2.0% Unknown
- Ethnicity (% Hispanic or Latino): 2.0%
- BMI (kg/m^2) (mean ± SD, range): 32.6 ± 6.8 (17.7, 50.5)
- Fitzpatrick Skin Type: type I = 5.9%; type II = 19.6%; type III = 19.6%; type IV = 7.8%; type V = 33.3%; type VI = 13.7%
-
Results, Primary Effectiveness Endpoint: The WaveCRESST study met its primary endpoint. In the mITT population, LS mean DA ratio was 0.83 (95% CI: 0.77, 0.88) for the Wave PRO device and 0.57 (95% CI: 0.52, 0.62) for the predicate ACS device (p<.0001). The Wave PRO device exceeded performance with the difference of 0.26 compared to the predicate device, far exceeding the non-inferiority margin of -0.15. The Wave PRO System was therefore determined to be non-inferior to the predicate ACS.
-
Results, Safety: There were no adverse events, serious or otherwise, that occurred during the WaveCRESST study.
-
Conclusions: The WaveCRESST study met its primary endpoint and demonstrated that the Wave PRO System was non-inferior to the predicate device with respect to effectiveness. There were no adverse events. The Wave PRO System was therefore shown to be safe and effective in this study.
VIII. Special Controls
The Pression Wave PRO External Counter-Pulsation System meets the special controls specified in 21 CFR 870.5225, External counter-pulsating device.
IX. Conclusions and Benefit/Risk Assessment
Based on the nonclinical and clinical testing summarized in this 510(k), the Pression Wave PRO External Counter-Pulsation System is as safe, performs as well as, and is at least as effective as the predicate device. Data from the WaveCRESST clinical study indicates that the Wave PRO System is non-inferior to the predicate device with respect to effectiveness based on DA ratio. There were no adverse events in the clinical study, and nonclinical testing and risk management have determined that the risk of the Wave PRO System has been reduced as far as possible. The Wave PRO System is at least as safe as the predicate device.
The clinical benefits of ECP treatment in patients with chronic stable angina are well documented in the literature. Given that the Wave PRO System has been shown safe and effective in nonclinical and clinical testing, the expected benefits of the device outweigh any risks.
Pression concludes that the Pression Wave PRO External Counter-Pulsation System is substantially equivalent to the predicate device, K042413, and the reference devices, K190683, K022107 and K152115. The Pression Wave PRO External Counter-Pulsation System does not raise any new issues or concerns of safety or effectiveness and is at least as safe and effective as the predicate device.
§ 870.5225 External counter-pulsating device.
(a)
Identification. An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.(b)
Classification. (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.