K Number
K250714
Manufacturer
Date Cleared
2025-11-21

(256 days)

Product Code
Regulation Number
872.3260
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Treatment of dental hypersensitivity and reduction of post-operative sensitivity.

Device Description

The FluoroDose Varnish is a fluoride varnish available in mint flavor. FluoroDose varnish is sold in single-use Lollitrays, containing a 0.3ml dose and a Benda Brush applicator brush. The device is not sold as sterile and does not require sterilization prior to use. The product is offered in a 2.1% sodium fluoride variation and a 5% sodium fluoride variation. The device is used as a dentin desensitizer to be used by healthcare professionals including dentists and dental hygienists.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - FluoroDose Varnish

Page 1

November 21, 2025

Centrix, Inc.
℅ Joseph Azary
Regulatory Consultant
Aztech Regulatory & Quality LLC
543 Long Hill Avenue
Shelton, Connecticut 06484

Re: K250714

  • Trade/Device Name: FluoroDose Varnish
  • Regulation Number: 21 CFR 872.3260
  • Regulation Name: Cavity Varnish
  • Regulatory Class: Class II
  • Product Code: LBH
  • Dated: October 23, 2025
  • Received: October 23, 2025

Dear Joseph Azary:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250714 - Joseph Azary Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250714 - Joseph Azary Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MICHAEL E. ADJODHA -S

Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

FieldInformation
Marketing application/submission numberK250714
Device trade name(s)FluoroDose Varnish
Indications for UseTreatment of dental hypersensitivity and reduction of post-operative sensitivity.
Types of uses☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

510(k) Summary

CENTRIX
K250714
FLUORODOSE VARNISH

1. SUBMITTER/510(K) HOLDER

Centrix Inc
770 River Road
Shelton, CT 06484
FDA Registration# 1281412

Contact Name: Joseph Azary (Aztech Regulatory & Quality LLC)
Email: jazary@erols.com
Telephone: (203) 242-6670
Date Revised: November 21, 2025

2. DEVICE NAME

  • Proprietary Name: Centrix FluoroDose Varnish
  • Common/Usual Name: Cavity Varnish
  • Classification Name: Cavity Varnish
  • Classification Regulation: 21 CFR 872.3260
  • Product code: LBH
  • Classification: Class 2
  • Medical Specialty (Panel): Dental

3. PREDICATE DEVICES

  • Primary Predicate: Varnish America (Young Dental / MPL) K040098
  • Secondary Predicate: Elevate SmartCoat 2.5% (FluoriMax) K131376
  • Reference Predicate: 3M Clinpro 2.1% K231338

Page 6

4. DEVICE DESCRIPTION

The FluoroDose Varnish is a fluoride varnish available in mint flavor. FluoroDose varnish is sold in single-use Lollitrays, containing a 0.3ml dose and a Benda Brush applicator brush. The device is not sold as sterile and does not require sterilization prior to use. The product is offered in a 2.1% sodium fluoride variation and a 5% sodium fluoride variation. The device is used as a dentin desensitizer to be used by healthcare professionals including dentists and dental hygienists.

5. INDICATIONS FOR USE

Treatment of dental hypersensitivity and reduction of post-operative sensitivity.

6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

Modifying the dentin surface to occlude the tubules is accepted as a useful approach to preventing dentin hypersensitivity. FluoroDose varnish contains suspensions of sodium fluoride in a rosin base.

The varnishes have identical regulatory classification, product codes, contraindications, mode of action, dispensing method, use of applicators, flavoring, shelf life, and being sold non-sterile.

While there may be some minor variations in wording the indications for use are equivalent in that all of the varnishes are for the treatment of hypersensitive teeth.

The fluoride amount and fluoride ions are equivalent between the 5% varnishes and the 2.1% varnishes.

The materials are equivalent in that all of the varnishes are based on the usage of rosin, sodium fluoride, beeswax, xylitol and flavor. There are some minor differences in the material formulations but are not considered significant concerns of safety or effectiveness.

The testing specifically included comparison to a predicate device to compare to a lower dose varnish and found that test results were equivalent.

Page 7

DeviceCentrix Fluorodose Subject Device K250714K040098 MPL VarnishAmerica 5% Primary PredicateSmartCoat 2.5% NaF Varnish (K131376) Reference Device3M Clinpro 2.1% (Reference, K231338) Reference DeviceComparison
Product CodeLBHLBHLBHLBHIDENTICAL
Regulation21 CFR.872.3260 Cavity varnish21 CFR.872.3260 Cavity varnish21 CFR.872.3260 Cavity varnish21 CFR.872.3260 Cavity varnishIDENTICAL
Indications For Use – 510(k)Treatment of dental hypersensitivity and Reduction of PostOperative Sensitivity.Treatment of dental hypersensitivity and Reduction of PostOperative Sensitivity.Intended for use on sensitive teeth, overexposed dentin and root sensitivity and under temporary restorations or cements where postoperative sensitivity is of concern.Treatment of hypersensitive teethAll three devices are indicated for use in the treatment of hypersensitive teeth.
ContraindicationUlcerative gingivitis and stomatitis Known allergic reactions to fluoride Avoid ingestion duringUlcerative gingivitis and stomatitis Known allergic reactions to fluoride Avoid ingestion during applicationNone known per 510(k) Summary IFU includes ulcerative gingivitis and stomatitis.None known per 510(k) Summary IFU includes ulcerative gingivitis and stomatitis.IDENTICAL between subject device and primary predicate. SIMILAR between subject device and reference predicate

Page 8

DeviceCentrix Fluorodose Subject Device K250714K040098 VarnishAmerica Primary Predicate3M Clinpro 2.1% (K231338) Reference PredicateSmartCoat 2.5% NaF Varnish (K131376) Secondary PredicateComparison
Product Code RegulationLBH 21 CFR.872.3260 Cavity varnishLBH 21 CFR.872.3260 Cavity varnishLBH 21 CFR.872.3260 Cavity varnishLBH 21 CFR.872.3260 Cavity varnishIDENTICAL IDENTICAL
Indications For Use – 510(k)Treatment of dental hypersensitivity and Reduction of PostOperative Sensitivity.Treatment of dental hypersensitivity and Reduction of PostOperative Sensitivity.Intended for use on sensitive teeth, overexposed dentin and root sensitivity and under temporary restorations or cements where postoperative sensitivity is of concern.Treatment of hypersensitive teethAll three devices are indicated for use in the treatment of hypersensitive teeth.
ContraindicationUlcerative gingivitis and stomatitis Known allergic reactions to fluoride Avoid ingestion during application Avoid ingestion duringUlcerative gingivitis and stomatitis Known allergic reactions to fluoride Avoid ingestion during applicationNone known per 510(k) Summary IFU includes ulcerative gingivitis and stomatitis.None known per 510(k) Summary IFU includes ulcerative gingivitis and stomatitis.IDENTICAL between subject device and primary predicate. SIMILAR between subject device and reference predicate

Substantial Equivalency – Technological Characteristics

DeviceCentrix FluorodoseK040098 VarnishAmerica Primary Predicate3M Clinpro 2.1% (K231338) Reference PredicateSmartCoat 2.5% NaF Varnish (K131376) Secondary PredicateComparison
Mode of ActionDentin Tubule OcclusionDentin Tubule OcclusionDentin Tubule OcclusionDentin Tubule OcclusionIDENTICAL
Dispensing FormSingle use of 0.3ml dose using supplied applicator brushSingle use of 0.4ml dose using supplied applicator brushSingle-use, unit-amount in LPopSingle Use Ampules of 1ml using supplied applicator Also available in Multi-use, multi-unit bottle 8ml or 12ml based on information available on company websiteSubject device identical to primary predicate and 3M reference. The SmartCoat / FluoriMax is available for multiple use
ApplicatorDisposable brush applicatorDisposable brush applicatorDisposable brush applicatorDisposable brush applicatorIDENTICAL
Fluoride Compound, AmountSodium fluoride (NaF), 2.1% and 5.0%Sodium fluoride (NaF), 5%Sodium fluoride (NaF), 2.1% (wt/wt)Sodium fluoride (NaF), 2.5% (wt/wt)IDENTICAL Subject device and Varnish America are both offered in 5% versions. SIMILAR Subject device and ClinPro and SmartCoat / FluoriMax have are range between 2.1% and 2.5%.
FlavoringAvailable in MintAvailable in MintAvailable in MintAvailable in MintIDENTICAL
Amount of fluoride Ion22,600 ppm for 5% version22,600 ppm for 5% version9,500 ppm for 2.1% version11,300 ppm for 2.5% versionThe fluoride ions are identical application
Intended UserDental professionalDental professionalDental professionalDental professionalIDENTICAL

Page 9

DeviceCentrix FluorodoseK040098 VarnishAmerica Primary Predicate3M Clinpro 2.1% (K231338) Reference PredicateSmartCoat 2.5% NaF Varnish (K131376) Secondary PredicateComparison
Mode of ActionDentin Tubule OcclusionDentin Tubule OcclusionDentin Tubule OcclusionDentin Tubule OcclusionIDENTICAL
Dispensing FormSingle use of 0.3ml dose using supplied applicator brushSingle use of 0.4ml dose using supplied applicator brushSingle-use, unit-amount in LPopSingle Use Ampules of 1ml using supplied applicator Also available in Multi-use, multi-unit bottle 8ml or 12ml based on information available on company websiteSubject device identical to primary predicate and 3M reference. The SmartCoat / FluoriMax is available for multiple use
ApplicatorDisposable brush applicatorDisposable brush applicatorDisposable brush applicatorDisposable brush applicatorIDENTICAL
Fluoride Compound, AmountSodium fluoride (NaF), 2.1% and 5.0%Sodium fluoride (NaF), 5%Sodium fluoride (NaF), 2.1% (wt/wt)Sodium fluoride (NaF), 2.5% (wt/wt)IDENTICAL Subject device and Varnish America are both offered in 5% versions. SIMILAR Subject device and ClinPro and SmartCoat / FluoriMax have are range between 2.1% and 2.5%.
FlavoringAvailable in MintAvailable in MintAvailable in MintAvailable in MintIDENTICAL
Amount of fluoride Ion22,600 ppm for 5% version 9500 ppm for 2.1% version22,600 ppm for 5% version9,500 ppm for 2.1% version11,300 ppm for 2.5% versionThe fluoride ions are identical between 5% versions and 2.1% versions.
Releases FluorideYesYesYesYesIDENTICAL
SterilityNon-sterileNon-SterileNon-sterileNon-sterileIDENTICAL
Shelf Life24 months24 months24 months24 monthsIDENTICAL
MaterialsSodium Fluoride Xylitol Rosin, Denatured Alcohol, Sodium Hydroxide, Beeswax, Sucralose, Flavor, ColoringSodium Fluoride Xylitol Rosin, Denatured Alcohol, Beeswax, Sucralose, FlavorSodium Fluoride Xylitol Polyacrylic acid, hydroxyethyl cellulose, calcium salt, phosphate salt, pH buffer, flavor, water, potassium sorbateSodium Fluoride Xylitol Shellac, propylene glycol, ethyl alcohol, dibasic sodium phosphate, ammonium phosphate, sucralose, calcium hydroxyapatite, flavorIDENTICAL All of the products contain Sodium Fluoride, Xylitol and Flavoring. The subject device and Varnish America device have almost identical ingredients. DIFFERENCE The reference predicates have some different ingredients. The essential design of these products are the same as they are formulated with sodium fluoride with some different additives.

Page 10

Substantial Equivalency – Performance Bench Studies

Physical PropertiesTest MethodSpecificationCentrix FluorodoseK040098 VarnishAmerica Primary Predicate3M Clinpro 2.1% (K231338) Reference predicateSmartCoat 2.5% NaF Varnish (K131376) Secondary Predicate
Dentin tubule occlusion (In-vitro)SEMComparisonPassPassPassPass
Dentin PermeabilityInternal TestComparison No statistically significant differences between the varnishes tested2.1% = Average 99.19% 5% = Average 99.99%N/AN/A2.5% = Average 99.89%
3 Day Fluoride ReleaseCentrix Internal Test MethodComparison for release of fluoride during first 2 hours2.1% and 5% Release burst of fluoride in first 2 hoursNot tested Data not knownNot tested Data not knownRelease burst of fluoride in first 2 hours.
Fluoride ReleaseISO 17730-2020Pass/Fail: Fluoride release per ISO 17330 *Confirmed through internal testing ++Confirmed through 510(k) Summary / Available InformationPass*Pass++Pass++Pass*

Page 11

7. PERFORMANCE TESTING

The following testing was performed and where applicable consensus standards were utilized.

Summary of Testing

StandardDescription of TestResults
ISO 10993-5 (2009) CytotoxicityBiological Evaluation of Medical Devices – Part 5: Tests for in vitro cytotoxicityThe testing found the subject device was classified as cytotoxic but performed similarly to the predicate devices. The Biological Evaluation Report examined biocompatibility in its entirety and concluded that the materials did not exhibit concerns.
ISO 10993-10 (2021)Biological Evaluation of Medical Devices – Part 10: Tests for Skin SensitizationThe testing concluded that the subject device is NOT considered to be a sensitizer.
ISO 10993-23 (2021)Biological Evaluation of Medical Devices – Part 23: Tests for IrritationThe testing concluded that the subject device is considered to be a Non-Irritant to oral mucosa.
ISO 10993-11 (2017)Biological Evaluation of Medical Devices – Part 11: Tests for Systemic ToxicitySystemic Injection testing concluded that the device did not induce significantly greater biological reaction than the control extracts following a single dose to albino Swiss mice. The testing met requirements of ISO 10993-11.
N/ASEM Analysis of Subject Device (both 2.1% and 5%) and Predicate DeviceThe SEM report summarizes comparison between subject device and predicate device on human tooth samples. The testing concluded that all of the varnishes tested provided occlusion of open tubules. The Centrix product deposited observable varnish within tubules.
ISO 17730Fluoride Release Testing. Evaluation of fluoride release including the subject device (2.1% and 5%) and predicate device.The study concluded that all varnishes tested passed requirements of ISO 17730 for fluoride release.
ISO 17730Total Fluoride Content testing of subject device (2.1% and 5%)The study concluded that the fluoride content matched the labeled amount (for 2.1% and 5%) as required by ISO 17730.
N/ADentin Permeability. Human molar teeth samples were evaluated to calculate the % reduction of dentin permeability. The testing included the subject device (2.1% and 5%) and predicate device.The results found that there was no statistically significant difference for reduction of dentin permeability between subject device (2.1% and 5%) and subject device.
N/A3 Day Fluoride Release test for subject device (2.1% and 5%) and predicate device.The testing concluded that all varnishes tested released burst of fluoride within the first 2 hours.

Page 12

8. SAFETY AND EFFICACY

The subject device is found to be equivalent to the primary predicate device, secondary predicate and reference predicate based on comparative evaluation and testing. The overall safety and efficacy of the subject device has been demonstrated through testing conducted on the subject device as well as comparative testing between subject and predicate device.

9. CONCLUSION

Information presented supports substantial equivalence of the subject device to the predicate devices based on similarities in intended use, design, principles of operation, performance specifications, and testing.

Centrix Inc believes that based on the indications for use, technological characteristics, and comparison to predicate devices the subject device has been shown to be substantially equivalent to the predicate and the minor differences do not impact safety and effectiveness.

§ 872.3260 Cavity varnish.

(a)
Identification. Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.(b)
Classification. Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.