(104 days)
This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, and angiography applications.
The digital radiography SKR 3000 performs X-ray imaging of the human body using an X-ray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data.
- This device is not intended for use in mammography
- This device is also used for carrying out exposures on children.
The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. The CS-7 is a software with basic documentation level. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation correction (Intelligent-Grid). The Intelligent-Grid is cleared in K151465.
The FPDs used in SKR 3000 can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection.
The SKR 3000 is distributed under a commercial name AeroDR 3.
The purpose of the current premarket submission is to add pediatric use indications for the SKR 3000 imaging system.
The provided FDA 510(k) clearance letter and summary for the SKR 3000 device focuses on adding a pediatric use indication. However, it does not contain the detailed performance data, acceptance criteria, or study specifics typically found in a clinical study report. The document states that "image quality evaluation was conducted in accordance with the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices'" and that "pediatric image evaluation using small-size phantoms was performed on the P-53." It also mentions that "The comparative image evaluation demonstrated that the SKR 3000 with P-53 provides substantially equivalent image performance to the comparative device, AeroDR System 2 with P-52, for pediatric use."
Based on the information provided, it's not possible to fully detail the acceptance criteria and the study that proves the device meets them according to your requested format. The document implies that the "acceptance criteria" likely revolved around demonstrating "substantially equivalent image performance" to a predicate device (AeroDR System 2 with P-52) for pediatric use, primarily through phantom studies, rather than a clinical study with human patients and detailed diagnostic performance metrics.
Therefore, many of the requested fields cannot be filled directly from the provided text. I will provide the information that can be inferred or directly stated from the document and explicitly state when information is not available.
Disclaimer: The information below is based solely on the provided 510(k) clearance letter and summary. For a comprehensive understanding, one would typically need access to the full 510(k) submission, which includes the detailed performance data and study reports.
Acceptance Criteria and Device Performance Study for SKR 3000 (Pediatric Use Indication)
The primary objective of the study mentioned in the 510(k) summary was to demonstrate substantial equivalence for the SKR 3000 (specifically with detector P-53) for pediatric use, compared to a predicate device (AeroDR System 2 with P-52).
1. Table of Acceptance Criteria and Reported Device Performance
Given the nature of the submission (adding a pediatric indication based on substantial equivalence), the acceptance criteria are not explicitly quantifiable metrics like sensitivity/specificity for a specific condition. Instead, the focus was on demonstrating "substantially equivalent image performance" through phantom studies.
| Acceptance Criteria (Inferred from Document) | Reported Device Performance (Inferred/Stated) |
|---|---|
| Image quality of SKR 3000 with P-53 for pediatric applications to be "substantially equivalent" to predicate device (AeroDR System 2 with P-52). | "The comparative image evaluation demonstrated that the SKR 3000 with P-53 provides substantially equivalent image performance to the comparative device, AeroDR System 2 with P-52, for pediatric use." |
| Compliance with "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices" for pediatric image evaluation using small-size phantoms. | "image quality evaluation was conducted in accordance with the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices'. Pediatric image evaluation using small-size phantoms was performed on the P-53." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Test Set): Not specified. The document indicates "small-size phantoms" were used, implying a phantom study, not a human clinical trial. The number of phantom images or specific phantom configurations is not detailed.
- Data Provenance: Not specified. Given it's a phantom study, geographical origin is less relevant than for patient data. It's an internal study conducted to support the 510(k) submission. Retrospective or prospective status is not applicable as it's a phantom study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not specified. Given this was a phantom study, ground truth would likely be based on physical measurements of the phantoms and expected image quality metrics, rather than expert interpretation of pathology or disease. If human evaluation was part of the "comparative image evaluation," the number and qualifications of evaluators are not provided.
- Qualifications: Not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Not specified. For a phantom study demonstrating "substantially equivalent image performance," adjudication methods like 2+1 or 3+1 (common in clinical reader studies) are generally not applicable. The comparison would likely involve quantitative metrics from the generated images.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
- MRMC Study: No. The document states "comparative image evaluation" and "pediatric image evaluation using small-size phantoms." This strongly implies a technical performance assessment using phantoms, rather than a clinical MRMC study with human readers interpreting patient cases. Therefore, no effect size of human readers improving with AI vs. without AI assistance can be reported, as AI assistance in image interpretation (e.g., CAD) is not the focus of this submission; it's about the imaging system's ability to produce quality images for diagnosis.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
- Standalone Performance: Not applicable in the traditional sense of an AI algorithm's diagnostic performance. The device is an X-ray imaging system. The "performance" being evaluated is its ability to generate images, not to provide an automated diagnosis. The "Intelligent-Grid" feature mentioned is an image processing algorithm (scattered radiation correction), but its standalone diagnostic performance is not the subject of this specific submission; its prior clearance (K151465) is referenced.
7. The Type of Ground Truth Used
- Ground Truth Type: For the pediatric image evaluation, the ground truth was based on phantom characteristics and expected image quality metrics. This is inferred from the statement "pediatric image evaluation using small-size phantoms was performed."
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable. The SKR 3000 is an X-ray imaging system, not an AI model that requires a "training set" in the machine learning sense for its primary function of image acquisition. While image processing algorithms (like Intelligent-Grid) integrated into the system might have been developed using training data, the submission focuses on the imaging system's performance for pediatric use.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as no training set (in the context of an AI model's image interpretation learning) is explicitly mentioned or relevant for the scope of this 510(k) submission for an X-ray system.
FDA 510(k) Clearance Letter - SKR 3000
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov
June 17, 2025
Konica Minolta, Inc.
℅ Jan Maniscalco
Executive Vice President QA/RA
Konica Minolta Inc.
411 Newark Pompton Turnpike
WAYNE, NJ 07470
Re: K250665
Trade/Device Name: SKR 3000
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-Ray System
Regulatory Class: Class II
Product Code: MQB, LLZ
Dated: May 13, 2025
Received: May 13, 2025
Dear Jan Maniscalco:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
June 17, 2025
Konica Minolta, Inc.
℅ Jan Maniscalco
Executive Vice President QA/RA
Konica Minolta Inc.
411 Newark Pompton Turnpike
WAYNE, NJ 07470
Re: K250665
Trade/Device Name: SKR 3000
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-Ray System
Regulatory Class: Class II
Product Code: MQB, LLZ
Dated: May 13, 2025
Received: May 13, 2025
Dear Jan Maniscalco:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Page 2
K250665 - Jan Maniscalco Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
Page 3
K250665 - Jan Maniscalco Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250665
Device Name
SKR 3000
Indications for Use (Describe)
This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, fluoroscopy, and angiography applications.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
Indications for Use
See PRA Statement below.
510(k) Number (if known)
K250665
Device Name
SKR 3000
Indications for Use (Describe)
This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, fluoroscopy, and angiography applications.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
Page 1 of 4
510(k) Summary
K250665
Complying with 21 CFR R §807.92
I. SUBMITTER
KONICA MINOLTA, INC.
1 Sakura-machi, Hino-shi,
Tokyo 191-8511, Japan
Contact Person: Jan Maniscalco, Executive Vice President of QA/RA
Konica Minolta Healthcare Americas, Inc.
Contact Telephone: 973.633.1500
Contact Email: jan.maniscalco@konicaminolta.com
Mailing Address: 411 Newark Pompton Turnpike, Wayne, New Jersey 07470
Date Prepared: June 13, 2025
II. DEVICE
Name of Device: SKR 3000
Common Name: Digital Radiography
Classification Name: Stationary x-ray system (21 CFR 892.1680)
Regulatory Class: Class II
Product Code: MQB, LLZ
III. PREDICATE DEVICES
Primary Predicate Device: K241319 - SKR 3000 (KONICA MINOLTA, INC.)
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary x-ray system
Regulatory Class: Class II
Product Codes: MQB, LLZ
Page 6
Page 2 of 4
Predicate Device: K151465 – AeroDR System2 (KONICA MINOLTA, INC.)
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary x-ray system
Regulatory Class: Class II
Product Codes: MQB, LLZ
IV. DEVICE DESCRIPTION
The digital radiography SKR 3000 performs X-ray imaging of the human body using an X-ray planar detector that outputs a digital signal, which is then input into an image processing device, and the acquired image is then transmitted to a filing system, printer, and image display device as diagnostic image data.
- This device is not intended for use in mammography
- This device is also used for carrying out exposures on children.
The Console CS-7, which controls the receiving, processing, and output of image data, is required for operation. The CS-7 is a software with basic documentation level. CS-7 implements the following image processing; gradation processing, frequency processing, dynamic range compression, smoothing, rotation, reversing, zooming, and grid removal process/scattered radiation correction (Intelligent-Grid). The Intelligent-Grid is cleared in K151465.
The FPDs used in SKR 3000 can communicate with the image processing device through the wired Ethernet and/or the Wireless LAN (IEEE802.11a/n and FCC compliant). The WPA2-PSK (AES) encryption is adopted for a security of wireless connection.
The SKR 3000 is distributed under a commercial name AeroDR 3.
The purpose of the current premarket submission is to add pediatric use indications for the SKR 3000 imaging system.
Page 7
Page 3 of 4
V. INDICATIONS FOR USE
This device is indicated for use in generating radiographic images of human anatomy. It is intended to a replace radiographic film/screen system in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, fluoroscopy, and angiography applications.
The subject device has the same Indications for Use statement as the primary predicate device except for the explicit inclusion of pediatric applications for the detector P-53 of the SKR 3000. This difference does not alter the intended use of the device. The subject device has the identical technological characteristics as the primary predicate device.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technological characteristics are identical in the proposed subject device and the primary predicate device. The sole purpose of this submission is to support the conditions of use for a pediatric population for the detector P-53 of the SKR 3000.
VII. PERFORMANCE DATA
Since this submission is to support the conditions of use for a pediatric population for the detector P-53 of the SKR 3000, image quality evaluation was conducted in accordance with the "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices". Pediatric image evaluation using small-size phantoms was performed on the P-53.
The comparative image evaluation demonstrated that the SKR 3000 with P-53 provides substantially equivalent image performance to the comparative device, AeroDR System 2 with P-52, for pediatric use.
Page 8
Page 4 of 4
VIII. CONCLUSIONS
It is concluded that the subject device is as safe and effective as the predicates and is substantially equivalent to the predicate devices to support the indication for pediatric imaging for the detector P-53 of the SKR 3000.
N/A