K Number
K250581

Validate with FDA (Live)

Date Cleared
2025-05-30

(92 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the red and near-infrared spectrum and is intended for the use in the treatment of full-face wrinkles.

The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the blue region of the light spectrum and is intended for the use in the treatment of mild to moderate acne vulgaris of the face.

Device Description

KALA Red Light Face Mask is a home wearable light-emitting diode phototherapy device with three proven wavelengths of light 630nm Red light, 830nm Near infrared red light and 465nm blue light, all of these lights are known to treat wrinkles and mild to moderate acne vulgaris of the face. Among them, the device emits energy in red and near-infrared spectrum to treat full-face wrinkles, emits energy in the blue region of the light spectrum to treat mild to moderate acne vulgaris of the face.

The system consists of a flexible silicone mask that contains LEDs and a controller. The mask is worn on the face and is held in place by Head Straps. The mask compromises of 2 surfaces. An inner surface that contacts the skin and an outer surface. Both surfaces are constructed of silicone.

The controller contains a rechargeable Lithium battery, the power supply (adaptor) is used to charge the Lithium battery and be connected to a suitable mains outlet via a 2 or 3 pin input socket and wall plug. The KALA Red Light Face Mask cannot be operated while charging. The controller switches the LEDs ON/OFF and controls power to the mask. Switch the controller ON and allow the mask to run for Red + NIR or Blue modes and 10- or 20-minutes treatment time program. The cable for connecting with the controller is detachable.

The device is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The device does not require checks on the light output as the LEDs do not dim with age to any practical extent.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the KALA Red Light Face Mask (KALA-01) primarily focus on non-clinical performance testing and substantial equivalence to predicate devices, rather than a detailed study demonstrating the device meets clinical acceptance criteria.

The document states, "Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively." This indicates that the regulatory clearance was based on demonstrating the device's technical performance, safety, and similarity to already approved devices, rather than a clinical trial with specific acceptance criteria related to efficacy (e.g., wrinkle reduction or acne treatment efficacy percentages).

Therefore, the following information will be based on the provided non-clinical performance testing and will highlight the absence of clinical study data as explicitly stated in the document.


Acceptance Criteria and Device Performance (Non-Clinical)

Since no clinical study focusing on efficacy endpoints (like wrinkle reduction percentage or acne lesion count reduction) was presented as part of the 510(k) summary, the acceptance criteria and performance reported here are for the non-clinical performance tests conducted to ensure the device's safety and fundamental operational characteristics.

1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

Title of the TestApplicable StandardAcceptance CriteriaReported Device Performance
General requirements for basic safety and essential performanceIEC 60601-1:2005/AMD1:2012/AMD2:2020The test result is within the test acceptance range of the standard.Pass
Electromagnetic disturbancesIEC 60601-1-2:2014+A1:2020No degradation of performance found during test or Lower than limits of measurement.Pass
Requirements for medical electrical equipment and systems used in home healthcare environmentIEC 60601-1-11:2015/AMD1:2020The device operates normally and can provide basic safety and essential performance.Pass
Particular Requirements for The Basic Safety and Essential Performance of Non-Laser Light Source Equipment (Therapeutic, Diagnostic, Monitoring, Cosmetic/Aesthetic Use)IEC 60601-2-57:2011The test result is within the test acceptance range of the standard.Pass
Photobiological safety of lamps and lamp systemsIEC 62471:2006The test result is within the test acceptance range of the standard.Pass
Performance Test (Power Density, Leakage current)Internal Performance Test ReportThe device can meet the requirement of the performance test, Power Density test and Leakage current test.Pass
BiocompatibilityISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-23 (Irritation)Compliance with biocompatibility requirements based on identical materials to previously cleared devices. Based on K202390.Complies (via justification of identical materials)
Usability TestingIEC 62366-1 and IEC 60601-1-6The device complies with the standards for usability.Complies
Software Verification & ValidationFDA'S Guidance for Industry and FDA StaffDocumentation provided as recommended by FDA's guidance for device software functions. (Implicit acceptance: software functions as intended and safely).Documentation provided

Study Details (Based on Non-Clinical Testing)

As explicitly stated in the 510(k) summary, no clinical testing was performed or required for this clearance. The "study" here refers to the collection of non-clinical tests performed to demonstrate safety and technical performance.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not specify a "sample size" in the context of a patient cohort. For the non-clinical tests, it would typically refer to a statistically relevant number of devices or components tested. This information is not provided in detail, but it's understood that typical engineering and safety tests involve a representative sample of manufactured units or materials.
  • Data Provenance: The tests were conducted in a lab setting, likely by the manufacturer or accredited testing facilities. The country of origin for the data (tests) is not explicitly stated beyond the manufacturer being in Shenzhen, China. The nature of these tests is prospective in the sense that they are planned tests on the device or its components to demonstrate compliance with standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • This concept is not applicable to the non-clinical performance and safety tests described. These tests involve objective measurements against established engineering and safety standards (e.g., electrical safety, optical power, material biocompatibility). "Ground truth" in this context is the standard itself (e.g., IEC 60601-1 requirements).
  • For biocompatibility, the equivalence relies on the materials being identical to those of a previously cleared device (K202390), implicitly accepting that the ground truth for those materials' biocompatibility was established in the prior clearance.

4. Adjudication Method for the Test Set:

  • None in the sense of clinical adjudication by multiple human experts. The compliance for non-clinical tests is determined by adherence to predefined, objective criteria outlined in the relevant international standards (e.g., IEC, ISO). Test reports are typically reviewed by regulatory bodies like the FDA.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No. An MRMC study is relevant for evaluating the performance of diagnostic devices or AI systems where human readers interpret medical images or data. This device is a therapeutic light-emitting mask, and no such study was conducted or presented for its 510(k) clearance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done:

  • Not applicable. This device does not involve a diagnostic algorithm or AI for interpretation that would necessitate a "standalone" or "human-in-the-loop" performance evaluation. Its performance is based on its physical emission characteristics and safety features.

7. The Type of Ground Truth Used:

  • For the non-clinical tests, the "ground truth" is defined by international safety and performance standards (e.g., IEC 60601 series, IEC 62471, ISO 10993 series) and the specified technical requirements for the device (e.g., power density, leakage current limits).
  • For biocompatibility, the ground truth is established by the prior clearance (K202390), based on the assumption that the identical materials have already demonstrated acceptable biocompatibility.

8. The Sample Size for the Training Set:

  • Not applicable. The provided documentation does not describe any training set, as it does not involve machine learning or AI models that require specific training data.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as there was no training set.

U.S. Food & Drug Administration - 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.0 7 . 0 5
Silver Spring, MD 20993
www.fda.gov

May 30, 2025

Shenzhen Kaiyan Medical Equipment Co. Ltd.
Alain Dijkstra
CEO
Building#3 and Building#5, 40th of Fuxin Street
Huaide Community, Fuyong Town, Baoan District
Shenzhen, Guangdong 518103
China

Re: K250581
Trade/Device Name: KALA Red Light Face Mask (KALA-01)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: April 29, 2025
Received: April 29, 2025

Dear Alain Dijkstra:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn

Page 3

(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by TANISHA L. HITHE -S
Date: 2025.05.30 15:14:04 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K250581

Please provide the device trade name(s).
KALA Red Light Face Mask (Model: KALA-01)

Please provide your Indications for Use below.
The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the red and near-infrared spectrum and is intended for the use in the treatment of full-face wrinkles.

The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the blue region of the light spectrum and is intended for the use in the treatment of mild to moderate acne vulgaris of the face.

Please select the types of uses (select one or both, as applicable).

  • Prescription Use (Part 21 CFR 801 Subpart D)
  • Over-The-counter Use (21 CFR 801 Subpart C)

KALA Red Light Face Mask
Page 9 of 39

Page 5

510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

Sponsor

Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd.
Establishment Registration Number: 3011644607
Address: Building #3, and Building#5, 40th of Fuxin Street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China
Contact Person (including title): Alain Dijkstra (CEO)
Tel: 0755-82129361
Fax: 0755-25024651
E-mail: alaindijkstra@kaiyanmedical.com

Distributor

Distributor Name: Kala Therapy Inc
Address: 1100 Courtneypark Dr. E. Mississauga, ON L5T 1S7, Canada
Contact Person: Cam Stajer
E-mail: cam@kalaredlight.com

Manufacturer:

Manufacturer Name: Shenzhen Kaiyan Medical Equipment Co., Ltd.
Establishment Registration Number: 3011644607
Address: Building #3, and Building#5, 40th of Fuxin Street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China
Contact Person (including title): Alain Dijkstra (CEO)
Tel: 0755-82129361
Fax: 0755-25024651
E-mail: alaindijkstra@kaiyanmedical.com

Application Correspondent:

Contact Person: Alain Dijkstra
Company: Shenzhen Kaiyan Medical Equipment Co., Ltd.
Address: Building #3, and Building#5, 40th of Fuxin Street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China
Tel: +86 755 82129361
Fax: +86 755 25024651
Email: registrar@kaiyanmedical.com

Page 6

2. Date of the summary prepared: May 30, 2025

3. Subject Device Information

Classification Name: Light Based Over-the-Counter Wrinkle Reduction (OHS)
Trade Name: KALA Red Light Face Mask
Model Name: KALA-01
Review Panel: General & Plastic Surgery
Product Code: OHS
Subsequent Product Codes: OLP
Regulation Number: 878.4810
Regulatory Class: II

4. Predicate Device Information

Predicate Device 1 Information

Sponsor: Light Tree Ventures Europe B.V
Trade Name: LED Light Therapy Mask
Classification Name: Light based over the counter wrinkle reduction
510(K) Number: K221775
Review Panel: General & Plastic Surgery
Product Code: OHS, OLP
Regulation Number: 878.4810
Regulation Class: II

Predicate Device 2 Information

Sponsor: Dongguan Boyuan Intelligent Technology Co., Ltd.
Trade Name: LED Light Therapy Device (KFB290, KFB291, KFB265, KFB293)
Classification Name: Light Based Over the Counter Wrinkle Reduction
510(K) Number: K241857
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 878.4810
Regulation Class: II

5. Device Description

KALA Red Light Face Mask is a home wearable light-emitting diode phototherapy device with three proven wavelengths of light 630nm Red light, 830nm Near infrared red light and 465nm blue light, all of these lights are known to treat wrinkles and mild to moderate acne vulgaris of the face. Among them, the device emits energy in red and near-infrared spectrum to treat full-face wrinkles, emits energy in the blue region of the light spectrum to treat mild to moderate acne vulgaris of the face.

Page 7

The system consists of a flexible silicone mask that contains LEDs and a controller. The mask is worn on the face and is held in place by Head Straps. The mask compromises of 2 surfaces. An inner surface that contacts the skin and an outer surface. Both surfaces are constructed of silicone.

The controller contains a rechargeable Lithium battery, the power supply (adaptor) is used to charge the Lithium battery and be connected to a suitable mains outlet via a 2 or 3 pin input socket and wall plug. The KALA Red Light Face Mask cannot be operated while charging. The controller switches the LEDs ON/OFF and controls power to the mask. Switch the controller ON and allow the mask to run for Red + NIR or Blue modes and 10- or 20-minutes treatment time program. The cable for connecting with the controller is detachable.

The device is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The device does not require checks on the light output as the LEDs do not dim with age to any practical extent.

6. Intended Use / Indications for Use

The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the red and near-infrared spectrum and is intended for the use in the treatment of full-face wrinkles.

The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the blue region of the light spectrum and is intended for the use in the treatment of mild to moderate acne vulgaris of the face.

Page 8

7. Comparison to predicate device and conclusion

Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.

Elements of ComparisonSubject DevicePredicate Device 1 K221775Predicate Device 2 K241857Remark
CompanyShenzhen Kaiyan Medical Equipment Co.,LtdLight Tree Ventures Europe B.V.Dongguan Boyuan Intelligent Technology Co., Ltd.--
Trade NameKALA Red Light Face MaskLED Light Therapy MaskLED Light Therapy Device (KFB290, KFB291, KFB265, KFB293)--
Classification NameLight Based Over the Counter Wrinkle ReductionLight Based Over the Counter Wrinkle ReductionLight Based Over the Counter Wrinkle Reduction--
510(k) NumberTBDK221775K241857--
Product CodeOHS, OLPOHS, OLPOHS, OLP, ILYSame
FDA Device ClassificationClass IIClass IIClass IISame
UseOver the CounterOver the CounterOver the counterSame
Intended Use / Indications for UseThe KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the red and near-infrared spectrum and is intended for the use in the treatment of full-face wrinkles.The KALA Red Light Face Mask (Model: KALA-01) is an over-the-counter device that is intended to emit energy in the blue region of the light spectrum and is intended for the use in theThe LED Light Therapy Mask (Models: MK-78, MK-04) is an over-the-counter device that is intended for the use in the treatment of full-face wrinkles.The LED Light Therapy Mask (Models: MK66-H, EL00003) is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face.The LED Light Therapy Mask (Models: MK66-H, EL00003) is an over-the-counter deviceKFB290, KFB291:Red light: Treatment of full-face wrinkles.Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.Red+Infrared Light: Treatment of full-face wrinkles.Amber light: Treatment of full-face wrinkles.Blue light: Treatment of mild to moderate inflammatory acne.Same

Page 9

Elements of ComparisonSubject DevicePredicate Device 1 K221775Predicate Device 2 K241857Remark
treatment of mild to moderate acne vulgaris of the face.intended to emit energy in the red and Near Infra-red spectrum and is intended for the use in the treatment of full-face wrinkles.Mixed light (Red+Blue +Infrared): Treatment of mild to moderate inflammatory acne.KFB265, KFB293:Red+Infrared Light: Treatment of wrinkles.Blue light: Treatment of mild to moderate inflammatory acne.Amber light: Treatment of wrinkles.
Intended location of useFaceFaceFacesame
Energy TypeLight emitting diodesLight emitting diodesLight Emitting DiodesSame
WavelengthsRed: 630nm±10nmNIR: 830nm±10nmBlue: 465nm±10nm1.MK-78, MK-04:Red: 630±5 nmNIR: 830nm2.MK66-H, EL00003:Blue: 415nm,Red: 630nm +/- 5nm,NIR: 830nm635nm ± 5nm visible red light;850nm±5nm Invisible red light;465±5nm blue light;605±5nm amber lightSame
Total Intensity (mW/cm²)Mode1:Red: 20mw/cm²Infrared Red: 10mw/cm²,Total: 30mw/cm²1.MK-78: 20-30 mw/cm²2.MK-04: 30mw/cm²3.MK66-H, EL00003:(1) Blue/Red 44 mw/cm²(2) Red/NIR 30 mw/cm²KFB290, KFB291Red: 25mW/cm²;IR: 3mW/cm²;Red+IR: 30mW/cm²;Blue: 18mW/cm²;Amber: 20mW/cm²;Mixed light: 9mW/cm²KFB265, KFB293Red+IR: 25.5mW/cm²Blue: 1.36mW/cm² Amber: 20mW/cm²SimilarNote 1
Mode2:Blue: 10mw/cm²
Treatment Time10, 20 minutes treatment time program10 minutesFor red, blue and red+infrared: 10, 20, 30 minutes for infrared, amber light and mixed light: 10, 20 minutes.Same
Treatment protocol3-5 times /weekAcne: 4 x weekly, 6 weeks;Wrinkles: 5 x weekly, 6 weeksAs aboveSame

Page 10

Elements of ComparisonSubject DevicePredicate Device 1 K221775Predicate Device 2 K241857Remark
Software controllerDevice uses a timer and software to control treatment durationDevice uses a timer and software to control treatment duration-Same
Power supplyRechargeable Lithium batteryRechargeable Lithium batteryRechargeable Lithium-ion batterySame

Note 1:
The intensity of mode 1 in the subject device is the same as predicate device 1 (K221775), which only emits two kinds of light: red light and infrared light. The intensity of mode 2 is similar to predicate device 2 (K241857). Additionally, the subject device is the same as predicate device 2 (K241857), both have single blue light mode to treat acne. The intensity of the subject device is similar to that of the predicate device 2 (18mW/cm2, K241857), with a small difference 8mW/cm2. In addition to this, model KFB265, KFB293 has an intensity 1.36mW/cm2, the intensity is much lower than that of the subject device. This indicate that the subject device is as safe and effective as products already on the market.

Furthermore, the subject device has already passed the testing regarding standard IEC 60601-1 and IEC 60601-2-57. In conclusion, the device is safe and effective. The slight differences between the subject device and the predicate devices will not raise any safety or effectiveness issues.

Final Conclusion:
The subject device is the same or similar to the legally marketed predicate device K221775 and K241857.

Page 11

8. Test Summary

8.1 Summary of Non-Clinical Performance Testing

1) Performance Testing Summary

KALA Red Light Face Mask (Model: KALA-01) has been evaluated the safety and performance by lab bench testing as following:

Title of the testTest Method/Applicable StandardsAcceptance criteriaUnexpected Results/Significant DeviationsTest results
General requirements for basic safety and essential performanceIEC 60601-1:2005/AMD1:2012/AMD2:2020The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard.NAPass
Electromagnetic disturbancesIEC 60601-1-2:2014+A1:2020No degradation of performance was found during test or Lower than limits of measurementNAPass
Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.IEC 60601-1-11:2015/AMD1:2020The device operates normally, and can provide basic safety and essential performance.NAPass
Particular Requirements for The Basic Safety and Essential Performance of Non-Laser Light Source Equipment Intended for Therapeutic, Diagnostic, Monitoring and Cosmetic/Aesthetic UseIEC 60601-2-57:2011The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard.NAPass
Photobiological safety of lamps and lamp systems.IEC 62471:2006The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard.NAPass
Performance TestThe Performance Test Report performs the following tests on the finished product: Power Density Test; Leakage current test.The device can meet the requirement of the performance test, Power Density test and Leakage current test.NAPass

Page 12

2) Biocompatibility testing

The component materials of the subject device are identical to the corresponding component materials of the previously cleared devices (K202390) in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents).

Here is no change in biocompatibility since the previously cleared devices. Therefore, based on this information, the subject device can comply with the biocompatibility requirements of ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), and ISO 10993-23 (Irritation).

3) Usability Testing

Usability testing was conducted on KALA Red Light Face Mask, the device complies with IEC 62366-1 and IEC 60601-1-6.

4) Software verification and validation testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions: Guidance for Industry and Food and Drug Administration Staff"

8.2 Clinical Performance

Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.

9. Final Conclusion:

The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K221775 and K241857.

N/A