K Number
K250548

Validate with FDA (Live)

Date Cleared
2025-06-26

(121 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.

Device Description

The Arm-type Blood Pressure Monitor (BPM) series is an automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the Oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats per minute.

The DBP-6286B embed bluetooth module to transfer data to APP. With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS.

AI/ML Overview

Based on the provided FDA 510(k) clearance letter for the JOYTECH Healthcare Co., Ltd. Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP-6186), here's a description of the acceptance criteria and the study proving the device meets them:

Acceptance Criteria and Device Performance for Blood Pressure Monitors

The acceptance criteria for blood pressure monitors are typically defined by recognized international standards, primarily ISO 81060-2:2018+AMD2020, which pertains to the clinical investigation of intermittent automated measurement type for non-invasive sphygmomanometers. This standard specifies the accuracy requirements for blood pressure measurements.

The key acceptance criterion is that the accuracy of the blood pressure monitor must be within the acceptable scope specified in ISO 81060-2:2018+AMD2020. While the exact numerical thresholds for group means and standard deviations aren't explicitly stated in this clearance letter, the general requirement is clear.

The reported device performance confirms that the device meets these criteria.

Acceptance CriterionReported Device Performance
Blood Pressure Accuracy (Systolic and Diastolic) within acceptable scope of ISO 81060-2:2018+AMD2020For cuff with arm circumference 22-42cm: The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020. This was covered by a representative study on DBP-6279B, which shares the same key factors (principle, measurement type, core algorithm, pressure sensor, MCU, cuff size, and material) affecting clinical accuracy.
Blood Pressure Accuracy (Systolic and Diastolic) within acceptable scope of ISO 81060-2:2018+AMD2020For cuff with arm circumference 40-56cm: The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020. This was demonstrated by a study on DBP-6286B.

Study Details Proving Device Meets Acceptance Criteria:

  1. A table of acceptance criteria and the reported device performance: (See table above)

  2. Sample sizes used for the test set and the data provenance:

    • Cuff (22-42cm): 91 valid subjects' data were used for analysis.
    • Cuff (40-56cm): 85 valid subjects' data were used for analysis.
    • Data Provenance: The document does not explicitly state the country of origin of the data, but it implies a prospective clinical study ("All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect").
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document implies the use of human observers for the reference measurements, as it states: "The manual Mercury Sphygmomanometer was used as a reference device." This suggests that the ground truth was established by human operators taking readings from a mercury sphygmomanometer against which the automated device's readings were compared. However, the exact number and qualifications of these "experts" (human observers) are not specified in the provided document.
  4. Adjudication method for the test set:

    • The document states, "Same arm sequential method was adopted during clinical testing." This method involves comparing simultaneous or near-simultaneous measurements from the test device and a reference device on the same arm. While this describes the measurement procedure, it does not describe an adjudication method for discrepancies between readings, as would be typical for expert consensus. Given that the ground truth is a direct measurement from a mercury sphygmomanometer, an adjudication process in the traditional sense (e.g., for image interpretation) is not directly applicable.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device clearance for a blood pressure monitor, not an AI-assisted diagnostic imaging device. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are irrelevant and were not performed.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The study design (clinical validation against a reference standard) inherently evaluates the "standalone" performance of the automated blood pressure monitor's algorithm. The device itself is designed to operate without continuous human intervention during the measurement process, making its performance essentially "algorithm only" in terms of blood pressure determination from the oscillometric data.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth was established by direct measurement using a manual Mercury Sphygmomanometer. This is considered the clinical gold standard for non-invasive blood pressure measurement in this context.
  8. The sample size for the training set:

    • The document does not specify a separate training set or its sample size. For traditional medical devices like blood pressure monitors, the "training" (development and calibration) often occurs during the device engineering and manufacturing phases, and clinical validation is then performed on a separate, independent test set as described.
  9. How the ground truth for the training set was established:

    • As no specific "training set" is mentioned in the context of a separate clinical dataset, the method for establishing ground truth for such a set is not provided. The device's underlying algorithms would have been developed and refined through engineering and calibration processes, likely using extensive internal testing and validation against standardized pressure sources and reference measurements, rather than a distinct "clinical training set" in the AI/ML sense.

FDA 510(k) Clearance Letter - K250548

Page 1

June 26, 2025

JOYTECH Healthcare Co., Ltd.
Cong Jing
RA Manager
No.365, Wuzhou Road
No.502, Shunda Road
Hangzhou, Zhejiang 311100
China

Re: K250548
Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP6186)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: February 25, 2025
Received: May 27, 2025

Dear Cong Jing:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K250548 - Cong Jing Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

K250548 - Cong Jing Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Robert T. Kazmierski -S

for

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Submission Number (if known): K250548

Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitors (DBP-6286B, DBP-6186)

Indications for Use (Describe)

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

Page 5

Substantial Equivalence Comparison

1. Sponsor's Identification

Name: JOYTECH Healthcare Co., Ltd.
Address 1: No. 365. Wuzhou Road, Hangzhou, Zhejiang 311100, China.
Address 2: No. 502. Shunda Road, Hangzhou, Zhejiang 311100, China.
Contact Person: Cong Jing
Email address: Jingc@sejoy.com

2. Name of the device

Trade Name: Arm-type Fully Automatic Digital Blood Pressure Monitor

Arm-typeDBP-6186, DBP-6286B

Common Name: Blood Pressure Monitor
Classification name: Non-invasive blood pressure measurement System

3. Classification information

Regulation Number: 21 CFR 870.1130
Product Code: DXN
Device Class: II
Panel: 74 Cardiovascular

4. Predicate device information

The Arm-Type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following predicate devices:

510(k) numberPredicate device modelProduct codeManufacturer
K212115DBP-6286BDXNJOYTECH Healthcare Co., Ltd.

K250548 Summary Page 1 of 10

Page 6

Note: As for the subject device in this submission are derived from the DBP-6286B, DBP-6186 in cleared K212115. Compared to the previous device, the subject device has added two cuffs and also the device unit has changed the power supply voltage, air pump, bluetooth module, and appearance. Therefore we resubmitted a new traditional 510 (k).

5. Device description

The Arm-type Blood Pressure Monitor (BPM) series is an automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use in home and clinical environment. The systolic and diastolic pressures are determined using the Oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg, and the pulse rate range from 30 to 180 beats per minute.

The DBP-6286B embed bluetooth module to transfer data to APP. With the use of software (including APP) and Bluetooth communication module, the wireless software function and hardware function are solely intended to transfer, store, convert formats, or display medical device data and results (blood pressure and pulse rate readings), without controlling or altering the functions or parameters of any connected medical devices, which is not be intended for active patient monitoring, therefore, based on the FDA guidance titled "Medical Device Data Systems, Medical Image Storage Devices, and Medical Image Communications Devices" (issued on September 28, 2022.), this software function is belong to Non-device-MDDS.

6. Indications for use

K250548 Summary Page 2 of 10

Page 7

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cm.

7. Comparison to the predicate device

Comparison itemsSubject devices (DBP-6186, DBP-6286B)Predicate device DBP-6286B (K212115)Judgement
ManufacturerJOYTECH Healthcare Co., Ltd.JOYTECH Healthcare Co., Ltd.Same
Trade nameArm-type Fully Automatic Digital Blood Pressure MonitorArm-type Fully Automatic Digital Blood Pressure MonitorSame
Recommended classification regulation21CFR 870.113021CFR 870.1130Same
Regulatory classClass IIClass IISame
Product codeDXNDXNSame
Indications for use/ Intended useThe Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cmThe Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age.Same Note 1

K250548 Summary Page 3 of 10

Page 8

Comparison itemsSubject devices (DBP-6186, DBP-6286B)Predicate device DBP-6286B (K212115)Judgement
ManufacturerJOYTECH Healthcare Co., Ltd.JOYTECH Healthcare Co., Ltd.Same
Trade nameArm-type Fully Automatic Digital Blood Pressure MonitorArm-type Fully Automatic Digital Blood Pressure MonitorSame
Recommended classification regulation21CFR 870.113021CFR 870.1130Same
Regulatory classClass IIClass IISame
Product codeDXNDXNSame
Indications for use/ Intended useThe Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cmThe Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm and 40cm-56cmSame Note 1

K250548 Summary Page 4 of 10

Page 9

Intended patientAdults and adolescents over 12 years of ageAdults and adolescents over 12 years of ageSame
Intended use environmentHome and Professional healthcare facility environment.Home and Professional healthcare facility environment.Same
Working principleOscillometric methodOscillometric methodSame
Measurement typeDetermined during inflationDetermined during inflationSame
Cuff locationUpper armUpper armSame
Cuff circumferenceFits arm circumference 22-42cm and 40-56cmFits arm circumference 22-36 cmSimilar Note 1
AppearanceSimilar Note 2
Solenoid valveSolenoid valveSolenoid valveSame
SensorSilicon pressure transducerSilicon pressure transducerSame
PumpAir pumpAir pumpSimilar Note 3
Display screenLCDLCDSame

K250548 Summary Page 4 of 10

Page 10

Power supply source4 x1.5V AAA batteries or Medical AC adaptor (DC 5V, 1A, recommend but not provide)3x1.5V AAA battery or Medical AC adaptor (DC 5V, 1A, recommend but not provide)Similar Note 4
Bluetooth moduleDBP-6286B applicable Frequency range: 2.4 GHz (2402-2480 MHz) Modulation type: GFSK Version 5.2 (BLE only)Frequency range: 2.4 GHz (2402-2480 MHz) Modulation type: GFSK Version 5.0 (BLE only)Similar Note 5
Measuring rangeSystolic Pressure: 60mmHg ~ 260 mmHg Diastolic Pressure:40mmHg ~ 200 mmHg Pressure: 0mmHg~299 mmHg Pressure: ±3mmHg Pulse:30~180 Beats/MinuteSystolic Pressure: 60mmHg ~ 260 mmHg Diastolic Pressure:40mmHg ~ 200 mmHg Pressure: 0mmHg~299 mmHg Pressure: ±3mmHg Pulse:30~180 Beats/MinuteSame
AccuracyPressure: ±3mmHg Pulse: 40-180Beats/Minute, ±5%; 30-39Beats/Minute, ±5bpmPressure: ±3mmHg Pulse: 40-180Beats/Minute, ±5%; 30-39Beats/Minute, ±5bpmSame
Main functionMeasure blood pressure value Measure pulse rate Irregular Heartbeat Detection Memory WHO Classification Indicator Last 3 Test Average Low Battery Detection Automatic Power-Off Voice abroad Backlight Arm shake indicatorMeasure blood pressure value Measure pulse rate Irregular Heartbeat Detection Memory WHO Classification Indicator Last 3 Test Average Low Battery Detection Automatic Power-Off Voice abroad Backlight Arm shake indicatorSame

K250548 Summary Page 5 of 10

Page 11

Cuff loose indicatorCuff loose indicator
Bluetooth function (DBP-6286B applicable)Bluetooth function (DBP-6286B applicable)
Operating condition10℃-40℃(50℉~104℉) 15%-93% RH 800hPa-1060hPa10℃-40℃(50℉~104℉) 15%-93% RH 800hPa-1060hPaSame
Storage condition-25℃-55℃ (-13℉~158℉) ≤93% RH-25℃-55℃ (-13℉~158℉) ≤93% RHSame
BiocompatibilityISO 10993-1 Bio-compatibleISO 10993-1 Bio-compatibleSame
Applicable standardsIEC 60601-1 IEC 60601-1-11 IEC 60601-1-2 ISO 80601-2-30 ISO 81060-2IEC 60601-1 IEC 60601-1-11 IEC 60601-1-2 ISO 80601-2-30 ISO 81060-2Same

Note 1: The subject device has the same intended purpose and principle with cleared devices (K212115 ). Compare with predicate device, the modification of cuff is occur in only add two cuffs, 22-42cm and 40-56cm. Adding new cuffs generally does not result in changes in electrical safety and EMC, and its corresponding clinical accuracy has been confirmed through clinical trials according to ISO 81060-2 . Thus, the difference between the subject device and the cleared predicate device do not raise different issues of safety and effectiveness.

Note 2: The subject device was added number of buttons based on predicate device. At the same time, The USB adapter jack type has no changed, the interface location of the power adapter has been changed from the right side to the top. These changes have passed electrical safety testing and will not cause new safety and effectiveness issues.

Note 3: Compared with cleared predicate device, the external dimensions and rated working voltage of the both air pump are not consistent, but the structural composition of the both air pump is the same.This subject device has passed IEC 60601-1, IEC 60601-1-2,

K250548 Summary Page 6 of 10

Page 12

and ISO 80601-2-30 standard testing,which will not result in new safety and effectiveness issues.

Note 4: The battery type of subject devices and predicates are the same,which all of them are AAA batteries. The main difference lies in the voltage. The voltage of the subject device is 6V, and the risks caused by different voltages have been controlled through electrical safety and electromagnetic compatibility testing, without bringing new safety and effectiveness issues.

Note 5: The difference of model DBP-6286B and DBP-6186 is only bluetooth. The BLE module version of DBP-6286B was updated from V5.0 to V5.2 by adding a shield cover. The subject device DBP-6286B has been verified the safety by IEC 60601-1, IEC 60601-11, IEC 60601-1-2, FCC and Wireless coexistence test report, which those of difference between subject device and predicate device do not raise new issues of safety and effectiveness comparing with the predicated device.

8. Performance Data

Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted. The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.

The following National and International Standards were utilized for testing the subject device.

Electrical Safety and performance requirements:

IEC 60601-1:2005+AMD1:2012+AMD2:2020 , Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.

AAMI ES60601-1:2005+AMD1:2012+AMD2:2021 , Medical Electrical Equipment Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance

K250548 Summary Page 7 of 10

Page 13

IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

Home-used medical equipment requirements and environmental test:

IEC 60601-1-11:2015+AMD1:2020 , General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

Electromagnetic compatibility requirements:

IEC 60601-1-2:2014+AMD1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests

Bio-compatibility Evaluation for patient contacting components:

ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization

ISO 10993-23:2021, Biological evaluation of medical devices Part 23: Tests for irritation

FCC Test

FCC Part15 Subpart C

RF Exposure valuation

Guidance Document:

The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"

K250548 Summary Page 8 of 10

Page 14

The test result all meet or exceed the requirement of these standards.

9. Discussion of Clinical Tests Performed

Clinical Validation:

  • ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers Part 2: Clinical investigation of intermittent automated measurement type.

For cuff with arm circumstance 22-42cm

Compared with Arm-type Fully Automatic Digital Blood Pressure Monitor DBP-6279B which manufactured by our company, although the subject device is different in appearance from DBP-6279B, the key factors affecting the clinical accuracy are the same. For example, both of them have the same principle and measurement type (determined by inflation), and they have the same core algorithm, same pressure sensor, same MCU, and same cuff size and material. Therefore, we believe that the clinical report of DBP-6279B can cover the subject devices.

Model DBP-6279B was selected as representative for testing. It was conducted according to ISO 81060-2:2018+AMD2020 standard. Same arm sequential method was adopted during clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The subject demographics include 91 valid subjects data were used to analysis. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

For cuff with arm circumstance 40-56cm

Model DBP-6286B was selected as representative for testing. It was conducted according to ISO 81060-2:2018+AMD2020 standard. Same arm sequential method was adopted during clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The subject demographics include 85 valid subjects data were used to analysis. The results showed

K250548 Summary Page 9 of 10

Page 15

the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

10. Conclusion

The subject devices and predicate devices have the same intended use, measuring principle, measuring method, performance and most technical characteristic.

The difference between the subject and the predicate devices does not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device. The minor difference between the subject devices and the predicate devices have been evaluated and determined to not raise any new issues of safety or effectiveness.

Therefore, the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.

K250548 Summary Page 10 of 10

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).