K Number
K250533

Validate with FDA (Live)

Date Cleared
2025-05-14

(79 days)

Product Code
Regulation Number
890.3800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Description

RS100 is a Mobility Scooter which provides mobility to a disabled or elderly person limited to a seated position. The Mobility Scooter is classified in the Class B and the maximum loading weight is 125kg. The scooter is a battery powered three-wheeled vehicle.

It consists Lithium-ion battery with an off-board battery charger, frame, controllers, motor, seat, back support, arm supports, control panel (including speed knob, battery gauge, power key switch, horn button, throttle control lever, charger port) two rear wheels, one front wheel, foot support.

For convenience of transportation and reduction of possible damage, the battery and arm supports can be dismantled and separately packaged. Users can also easily assemble these parts without use of the tools.

AI/ML Overview

The provided document is an FDA 510(k) clearance letter for a "Magnesium alloy scooter (RS100)". This type of device, a motorized three-wheeled vehicle, is a physical mobility aid, not an AI/Software as a Medical Device (SaMD). Therefore, the concepts of acceptance criteria in the context of AI performance metrics (like sensitivity, specificity, AUC), ground truth establishment by experts, adjudication methods, MRMC studies, or training/test sets for AI algorithms simply do not apply to this device.

The "acceptance criteria" for this device are based on its compliance with international performance and safety standards, primarily the ISO 7176 series and ISO 10993-1 for biocompatibility, and IEC 60601 for electrical safety and electromagnetic compatibility. The "study that proves the device meets the acceptance criteria" refers to the non-clinical bench testing conducted to demonstrate compliance with these standards and substantial equivalence to a predicate device.

Given this, I cannot extract the information required by your prompt regarding AI/SaMD performance. I will explain why each point in your prompt is inapplicable to this document:

Inapplicability of Prompt Points:

  1. A table of acceptance criteria and the reported device performance: While the document does compare the subject device's specifications to the predicate, and states that test results meet design specifications and ISO standards, it does not present these as a "table of acceptance criteria and reported device performance" in the way one would for AI metrics. The acceptance criteria are "compliance with ISO standards" and "meets design specifications," and the performance is implicitly satisfactory if it achieves this compliance.

  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): This would refer to a dataset used to evaluate an AI model. For a physical medical device like a scooter, the "test set" would be the physical prototypes tested. There's no "data provenance" in the sense of patient data. Testing is done on the device itself, likely within a lab setting.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Ground truth establishment by experts is relevant for diagnostic AI. For a scooter, the "ground truth" is whether it performs according to engineering specifications and safety standards, which is determined by objective physical measurements and adherence to specified test protocols (e.g., in the ISO standards), not expert interpretation of data.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Adjudication is for resolving disagreements among human experts labeling data for AI. Not applicable here.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: MRMC studies evaluate the impact of AI on human reader performance, typically in imaging diagnostics. This is entirely irrelevant for a physical mobility scooter.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Standalone performance refers to an AI algorithm operating without human intervention. Not applicable to a physical device.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): As explained in point 3, the "ground truth" for a physical device's performance is adherence to defined engineering and safety standards, measured objectively.

  8. The sample size for the training set: Training sets are for machine learning algorithms. This device is not an AI.

  9. How the ground truth for the training set was established: As explained in point 8, not applicable.

In conclusion, this FDA 510(k) clearance document pertains to a physical mechanical device, a scooter, and not to an AI/SaMD. Therefore, addressing the specific points of your prompt as if it were an AI device is not possible with the provided information.

FDA 510(k) Clearance Letter - Magnesium Alloy Scooter (RS100)

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

May 14, 2025

Taizhou Reach Technology Co., Ltd
℅ Luna Hu
Official Correspondent
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave.
Shanghai, China

Re: K250533
Trade/Device Name: Magnesium alloy scooter (RS100)
Regulation Number: 21 CFR 890.3800
Regulation Name: Motorized Three-Wheeled Vehicle
Regulatory Class: Class II
Product Code: INI
Dated: February 24, 2025
Received: February 24, 2025

Dear Luna Hu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250533 - Luna Hu Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250533 - Luna Hu Page 3

Sincerely,

Tushar Bansal -S

for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250533

Device Name
Magnesium alloy scooter (RS100)

Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Taizhou Reach Technology Co., Ltd 510(k) Summary

510(k) number: K250533
Document Prepared Date: 2025/05/14

1. Applicant

Company name: Taizhou Reach Technology Co., Ltd
Address: No.18 Xiangrong Road, Chengqu Industrial Zone, Taixing City, Jiangsu, China
Contact Person: Mr. Fan Rong
Tel: +8618852636388

Submission Correspondent:

Primary Contact: Luna Hu
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Central Ave., Shanghai 200122, China
Tel: 0086-21-58817802
Email: liu.hu@sungoglobal.com

Secondary Contact: Raymond Luo
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Central Ave., Shanghai 200122, China
Tel: 0086-21-68828050
Email: zxfda@sungoglobal.com

2. Device

Trade name: Magnesium alloy scooter (RS100)
Classification name: Motorized three-wheeled vehicle

Regulatory Information
Classification: Class II
Product code: INI
Regulation Number: 890.3800
Review Panel: Physical Medicine

3. Predicate Device

Manufacturer: Zhejiang Innuovo Rehabilitation Devices Co., Ltd.
Product Name: Scooter (W3331)
510(K) #: K240008

Page 6

4. Indication for Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

5. Device Description

RS100 is a Mobility Scooter which provides mobility to a disabled or elderly person limited to a seated position. The Mobility Scooter is classified in the Class B and the maximum loading weight is 125kg. The scooter is a battery powered three-wheeled vehicle.

It consists Lithium-ion battery with an off-board battery charger, frame, controllers, motor, seat, back support, arm supports, control panel (including speed knob, battery gauge, power key switch, horn button, throttle control lever, charger port) two rear wheels, one front wheel, foot support.

For convenience of transportation and reduction of possible damage, the battery and arm supports can be dismantled and separately packaged. Users can also easily assemble these parts without use of the tools.

6. Non-clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ISO 10993-1: 2018 Biological Evaluation of Medical Devices -- Part 1: Evaluation and testing within a risk management process
  • ISO 7176-1:2014, Wheelchairs - Part 1: Determination of static stability
  • ISO 7176-2:2017, Wheelchairs - Part 2: Determination of dynamic stability of electrically powered wheelchairs
  • ISO 7176-3:2012, Wheelchairs - Part 3: Determination of effectiveness of brakes
  • ISO 7176-4:2008, Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
  • ISO 7176-5:2008, Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
  • ISO 7176-6:2018, Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electrically powered wheelchairs
  • ISO 7176-7:1998, Wheelchairs - Part 7: Measurement of seating and wheel dimensions
  • ISO 7176-8:2014, Wheelchairs -Part 8: Requirements and test methods for static, impact and fatigue strengths
  • ISO 7176-9:2009, Wheelchairs - Part 9: Climatic tests for electric wheelchairs

Page 7

  • ISO 7176-10:2008, Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
  • ISO 7176-11:2012 Wheelchairs - Part 11: Test dummies
  • ISO 7176-13:1989, Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
  • ISO 7176-14:2008, Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
  • ISO 7176-15:1996, Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
  • ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs
  • ISO 16840-10: 2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method
  • ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
  • IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

7. Clinical Test Conclusion

No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the electric wheelchair to its predicate device.

8. Comparison technological characteristics with the predicate device

Table 1 General Comparison

Elements of ComparisonSubject DevicePredicate Device (K240008)Remark
ManufacturerTaizhou Reach Technology Co., LtdZhejiang Innuovo Rehabilitation Devices Co.,Ltd.--
Device nameMagnesium alloy scooterScooter--
Model(s)RS100W3331--
Indication for useIt is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated positionIt is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.Same
Overall dimension (mm)1130×550×8401110 x 560 x 840Similar
Frame MaterialMagnesium alloySteelDifferent
Front wheel diameter200mm × 60mm196mm × 66mmSimilar

Page 8

Elements of ComparisonSubject DevicePredicate Device (K240008)Remark
Front Wheels Quantity11Same
Rear wheel diameter8inch × 2inch196mm × 66mmSimilar
Rear Wheels Quantity22Same
Ground clearance80mm35mmDifferent
Max Loading(on level ground)125KG120KGSimilar
Turn Radius1212.5mm925mmDifferent
Motor output24 V 150WLINIX 24 V 180WDifferent
Drive SystemFront Wheel DriveRear Wheel DriveDifferent
BrakesElectromagnetic brakeElectromagnetic brakeSame
BatteryLi-ion battery 25.2V, 20AhLead-acid 12V12Ah*2Different
Charger24V, 3A24V, 2ADifferent
Max Speed7.56km/h6.83km/hDifferent
Max Slope10°Dynamic Stability: 6° Static Stability: 8°Different
Travel Distance25.3km15kmDifferent
Distance between armrests42cm37-50.5 cmDifferent
ControllerM3-YG8British PG 45ADifferent
Time to brake< 1 s< 1 sSame
Brake Distance- Normal operation (Horizontal-Forward- Max speed)≤1.5m≤1.3mDifferent
Total mass19KG40KGDifferent
Operating surface & environmentIndoor use and restricted outdoor use on pavements or paved footpaths only.Indoor use and restricted outdoor use on pavements or paved footpaths only.Same
Remote controlNoNoSame

Difference analysis

The design and technological characteristics of the proposed device scooter is similar to the predicate device. There are minor differences between the devices in frame material, size, load capacity, weight, ground clearance, turn radius, drive system, battery type, charger, controller, motor output, max speed, max slope, travel distance, Distance between Armrests, brake distance and time to brake. All of the parameters with differences have been tested according to ISO7176 series standards and the test records support its safety and effectiveness. There is no deleterious effect on safety and effectiveness due to the minor differences do not influence the intended use of the device. Therefore, the proposed Scooter is substantially equivalent (SE) to The Scooter (K240008).

Table 2 safety comparison

ItemSubject DevicePredicate Device (K240008)Results

Page 9

ItemSubject DevicePredicate Device (K240008)Results
BiocompatibilityAll user directly contacting materials are compliance with ISO10993-1All user directly contacting materials are compliance with ISO10993-1Same
EMCISO7176-21& IEC 60601-1-2:2014+A1:2020 &IEC TR 60601-4-2:2016ISO7176-21& IEC 60601-1-2:2014+A1:2020Same
PerformanceISO7176 seriesISO7176 seriesSame
Label and labelingConforms to FDA RegulatoryConforms to FDA RegulatorySame
ItemSubject DevicePredicate Device (K240008)Results
ISO7176-1The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification.The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification.Same
ISO7176-2The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification.The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification.Same
ISO7176-3The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification.The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification.Same
ISO7176-4The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification.The theoretical distance range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification.Same
ISO7176-5The dimensions, mass has been determined after the testing according to the ISO 7176-5.The dimensions, mass has been determined after the testing according to the ISO 7176-5.Same
ISO7176-6The maximum speed, acceleration and deceleration has been determined after the testing according to the ISO 7176-6.The maximum speed, acceleration and deceleration has been determined after the testing according to the ISO 7176-6.Same
ISO7176-7The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7.The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7.Same

Page 10

ItemSubject DevicePredicate Device (K240008)Results
ISO7176-8All test results meet the requirements in Clause 4 of ISO 7176-8.All test results meet the requirements in Clause 4 of ISO 7176-8.Same
ISO7176-9The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9.The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9.Same
ISO7176-10The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10.The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10.Same
ISO7176-11The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11.The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11.Same
ISO7176-13The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved.The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved.Same
ISO7176-14All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14.All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14.Same
ISO7176-15The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15.The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15.Same
ISO7176-16The performance of resistance to ignition meet the requirements of ISO 7176-16.The performance of resistance to ignition meet the requirements of ISO 7176-16.Same
ISO7176-21The EMC performance results meet the requirements of ISO 7176-21, IEC 60601-1-2:2014+A1:2020 and IEC TR 60601-4-2:2016.The EMC performance results meet the requirements of ISO 7176-21, IEC 60601-1-2:2014+A1:2020.Same

Page 11

ItemSubject DevicePredicate Device (K240008)Results
ISO7176-25The performance of batteries and charger of device meet the requirements in Clause 5 and 6 of ISO 7176-25.The performance of batteries and charger of device meet the requirements in Clause 5 and 6 of ISO 7176-25.Same

9. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Magnesium alloy scooter, model:RS100 , is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K240008.

§ 890.3800 Motorized three-wheeled vehicle.

(a)
Identification. A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.(b)
Classification. Class II (performance standards).