K Number
K250532

Validate with FDA (Live)

Date Cleared
2025-05-22

(87 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage.

Device Description

The Solawave 2-in-1 Skincare Mini (Model: 61043) is a hand-held, battery-powered device that reduce wrinkles by emitting LED red light (630nm) and infrared light (830nm). The device is powered by a Lithium-Ion rechargeable battery, and it comes with a charging cable, storage bag, and instruction manual.

The Solawave 2-in-1 Skincare Mini is a revolutionary device that offers two simultaneous approaches for your skin routine in one handheld device:

  1. Red and Near-Infrared Light Therapy
  2. Warming function (This function is not for medical purpose)
AI/ML Overview

The provided FDA 510(k) clearance letter for the Solawave 2-in-1 Skincare Mini (K250532) indicates that no clinical performance studies were conducted or deemed necessary for its clearance.

Therefore, it is not possible to describe acceptance criteria and associated study results for device performance based on the provided document, as no such studies were submitted or reviewed by the FDA for this particular clearance.

The document states:

  • "Clinical testing was not needed for this 510(k)." (Page 10, Section 7.2)
  • The clearance is based on the device being "substantially equivalent" to legally marketed predicate devices. This means the FDA determined the device has the same intended use and technological characteristics as a predicate device, or if there are differences, those differences do not raise different questions of safety and effectiveness.

Breakdown of the requested information based on the provided document:

  1. A table of acceptance criteria and the reported device performance:

    • Not applicable. No clinical performance studies were conducted or provided in this 510(k) submission to establish acceptance criteria for device performance or report device performance against such criteria. The "performance" assessment was primarily through comparison to predicate devices and non-clinical testing.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not applicable. No clinical test set was used for performance evaluation in this 510(k).
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. No clinical test set requiring expert ground truth was used for performance evaluation in this 510(k).
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical test set requiring adjudication was used for performance evaluation in this 510(k).
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a light-based therapy device, not an AI-powered diagnostic or assistive tool, and no MRMC study was conducted.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical light-based therapy device, not an algorithm, and no standalone performance study was conducted.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. No clinical ground truth was established as no clinical studies were performed. The "ground truth" for substantial equivalence was based on existing predicate device characteristics and regulatory standards.
  8. The sample size for the training set:

    • Not applicable. This device is a physical light-based therapy device, not a machine learning model, and therefore does not have a "training set" in the context of AI/ML.
  9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set for an AI/ML model was used.

Summary of what was reported instead:

The 510(k) submission relied on:

  • Non-Clinical Tests:
    • Electrical safety (e.g., IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-1-2)
    • Photobiological safety (IEC 62471)
    • Battery safety (IEC 62133-2)
    • Biocompatibility (ISO 10993-5, ISO 10993-10, ISO 10993-23)
    • Software verification and validation (classified as "moderate" concern)
  • Comparison to Predicate Devices: The document extensively compares the Solawave 2-in-1 Skincare Mini with several predicate and reference devices regarding features like intended use, regulation class, product code, power source, irradiance source, wavelengths, irradiances, and treatment time to establish substantial equivalence. The argument for equivalent "treatment effect (wrinkle treatment)" is made based on similar wavelengths, and irradiances/treatment times being "very close to those of reference devices 1 and 2."

In conclusion, for this specific 510(k), the acceptance criteria and proof of device performance were primarily established through substantial equivalence to existing legally marketed devices and successful completion of non-clinical safety testing, rather than through new clinical performance studies.

FDA 510(k) Clearance Letter - Solawave 2-in-1 Skincare Mini

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

May 22, 2025

Shenzhen Kaiyan Medical Equipment Co., Ltd
Alain Dijkstra
Official Correspondent
Building#3 and Building#5, 40th of Fuxin Street,
Huaide Community Fuyong Town, Baoan District
Shenzhen, Guangdong 518103
China

Re: K250532
Trade/Device Name: Solawave 2-in-1 Skincare Mini (61043)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS
Dated: February 22, 2025
Received: February 24, 2025

Dear Alain Dijkstra:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by TANISHA L. HITHE -S
Date: 2025.05.22 15:29:33 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23)

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250532

Device Name: Solawave 2-in-1 Skincare Mini (Model: 61043)

Indications for Use (Describe):
The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)
  • Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

K250532 510(k) Summary

This summary of 510(k) is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd
Establishment Registration Number: 3011644607
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Contact Person (including title): Alain Dijkstra (Manager)
Tel: +86-135-10378748
Fax: +86-755-25024651
E-mail: registrar01@kaiyanmedical.com

Distributor:
Company: Solawave Inc.
Address: 3633 Lenawee Avenue Suite 100 Los Angeles, CA 90016 USA

Application Correspondent:
Contact Person: Alain Dijkstra
Company: Shenzhen Kaiyan Medical Equipment Co., Ltd
Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
Tel: +86 755 82129361
Fax: +86 755 25024651
Email: registrar01@kaiyanmedical.com

2. Subject Device Information:

Common/Trade Name: Solawave 2-in-1 Skincare Mini(Model: 61043)
Classification Name: Light Based Over the Counter Wrinkle Reduction
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 21 CFR 878.4810
Regulation Class: II

3. Predicate and Reference Device Information

Predicate Device (K213841)
Sponsor: Premier North America Inc
Trade Name: ENEO TOTALE
Classification Name: Light Based Over the Counter Wrinkle Reduction
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 21 CFR 878.4810
Regulation Class: II

Page 6

Reference Device 1 (K241102)
Sponsor: FOREO, Inc.
Trade Name: Luna™ 4 plus
Classification Name: Light Based Over the Counter Wrinkle Reduction;Transcutaneous electrical nerve stimulator for pain relief
Review Panel: General & Plastic Surgery
Product Code: OHS,NFO
Regulation Number: 21 CFR 878.4810, 21 CFR 882.5890
Regulation Class: II

Reference Device 2 (K202055)
Sponsor: Heat In A Click LLC
Trade Name: Looper
Classification Name: Light Based Over the Counter Wrinkle Reduction;over-the-counter powered light based laser for acne
Review Panel: General & Plastic Surgery
Product Code: OHS,OLP
Regulation Number: 21 CFR 878.4810
Regulation Class: II

Reference Device 3 (K242382)
Sponsor: ISMART Developments Ltd
Trade Name: décoLITE
Classification Name: Light Based Over the Counter Wrinkle Reduction;
Review Panel: General & Plastic Surgery
Product Code: OHS
Regulation Number: 21 CFR 878.4810
Regulation Class: II

4. Device Description

The Solawave 2-in-1 Skincare Mini (Model: 61043) is a hand-held, battery-powered device that reduce wrinkles by emitting LED red light (630nm) and infrared light (830nm). The device is powered by a Lithium-Ion rechargeable battery, and it comes with a charging cable, storage bag, and instruction manual.

The Solawave 2-in-1 Skincare Mini is a revolutionary device that offers two simultaneous approaches for your skin routine in one handheld device:

  1. Red and Near-Infrared Light Therapy
  2. Warming function (This function is not for medical purpose)

5. Intended Use / Indications for Use

The Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletage.

Page 7

6. Comparison to predicate and reference devices

Compare with the predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, and the applicable standards.

Elements of ComparisonSubject devicePredicate device K213841Reference device 1 (K241102)Reference device 2 K202055Reference device 3 K242382Remark
ManufacturerShenzhen Kaiyan Medical Equipment Co., LtdPremier North America IncFOREO, Inc.Heat In A Click LLCISMART Developments Ltd--
510(k) NumberK250532K213841K241102K202055K242382--
Device NameSolawave 2-in-1 Skincare MiniENEO TOTALELuna™ 4 plusLooperdécoLITE--
Model61043unknownUnknownZX-579S/--
OTC/RxOTCOTCOTCOTCOTCSame
Regulation ClassClass IIClass IIClass IIClass IIClass IISame
Product CodeOHSOHSOHS,NFOOHS, OLPOHSSame
Regulation Number21 CFR 878.481021 CFR 878.481021 CFR 878.4810, 21 CFR 882.589021 CFR 878.481021 CFR 878.4810Same
Indications for Use / Intended useThe Solawave 2-in-1 Skincare Mini (Model: 61043) is an over-the-counter device that emits energy in the red and infrared spectrum for treating wrinkles on the face and decolletageENEO TOTALE is an over the counter device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles1. The Red + IR light is intended to treat periorbital wrinkles 2. The microcurrent is intended to treat facial stimulationLooper (Model ZX-579S) is a handheld device for over-the-Counter aesthetic purposes. The Photon mode (red light) is indicated for the use in Treating wrinkles on the faceThe décoLITE LED device is an over-the-counter device that is intended for the use in the treatment of wrinkles in the decolletage areaSimilar (Note 1)
Intended Location UseFace, decolletageFaceFaceFaceDecolletage (upper chest)Similar (Note 2)

Page 8

Elements of ComparisonSubject devicePredicate device K213841Reference device 1 (K241102)Reference device 2 K202055Reference device 3 K242382Remark
Power SourceLithium battery: DC 3.7V, 450mAh, 1.665WhAdaptor:100~240V AC 50/60Hz 2.4A Lithium battery: 3.75Vdc, 3000 mAhInternal rechargeable Lithium batteryDC 3.7V 1000mA Li batteryLi batterySimilar (Note 3)
Irradiance sourceLEDLEDLED, micro currentLEDLEDSame
HandheldYesYesYesYesNo(Mask type)Similar
Dose/per time(J/cm²)7.2~9.937.5~52.53.78~15.129.918Similar (Note 4)
LED wavelengthRed: 630nm ± 10nm Infrared: 830nm ± 10nmRed: 633nm± 5nm Infrared:830nm± 5nmRED light:633nm± 10nm Red+IR: 633+10nm/850nm±10nmRed: 630±00nmRed: 630nm ± 10nm NIR: 830nm ± 10nmSimilar (Note 4)
IrradiancesRED/IR: 40-55 mW/cm²69mW/cm² for 633nm 56mW/cm² for 830nmRED+IR: 63 mW/cm²Red light: 55mW/cm²±10%30mW/cm² totalSimilar (Note 4)
Treatment Time3 minutes per treatment area5-7 minutes on each treatment zone For the first month (4 weeks)treatment should be performed 3 times a week for 15-20 minutes each time1-4 min (adjustable) per treatment3 minutes per target area600 seconds (10 minutes) 5× weekly, 6 weeksSimilar (Note 4)

Page 9

Elements of ComparisonSubject devicePredicate device K213841Reference device 1 (K241102)Reference device 2 K202055Reference device 3 K242382Remark
Electrical SafetyCompliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57 IEC 62471Compliant with IEC 60601- 1, IEC 60601-1-2, IEC 60601-1-11, IEC 62471Compliant with IEC 60601-1, IEC 60601- 1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 60601-2-10Compliant with IEC 60601- 1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57 IEC 60601-2-83 IEC 62471Same
BiocompatibilityISO 10993-1 ISO10993-5 ISO10993-10 ISO 10993-23ISO10993-5 ISO10993-10ISO10993-5 ISO10993-10 ISO 10993-23ISO10993-5 ISO10993-10unknownSame

Comparison in Details:

Note 1:
The Intended use of subject device is to treat wrinkle on the face, which is slightly different from the predicate device and reference device 1 which treat periorbital wrinkles. However, it is the same as reference device 2 which is indicated for the entire face wrinkle treatment, and both of them are hand held LED device.

Note 2:
The intended location use of the subject device is for face and decolletage, which is same as the predicate device, reference device 1, reference device 2 for face and reference device 3 for decolletage.

Note 3:
The power supply for the subject device is a little different from that of the predicate device, however the lithium battery of the subject device has been tested under standard IEC 62133-2, so this difference should not raise any safety/effectiveness issues.

Note 4:
The subject device is a little different from the predicate device in terms of "Treatment Time","Dose/per time" and "Irradiances", however they have the same LED wavelengths. And the "Treatment time", "Dose/per time" and "Irradiances" of the subject device are very close to those of reference devices 1 and 2, so it can be concluded that the subject device can achieve the same treatment effect (wrinkle treatment) as the predicate and reference devices. In addition, all of them have passed the tests of IEC 60601-1,IEC 60601-1-11 and IEC 60601-2-57, these differences will not raise any new safety or effectiveness issues.

7. Test Summary

7.1 Non-Clinical Tests Performed

1) Electrical safety, and electromagnetic compatibility Test

Page 10

Non-clinical tests were performed on the subject device to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:

  • IEC 60601-1 2020-08 Edition 3.2 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-11 Edition 2.1 2020-07 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
  • IEC 60601-2-57 Edition 2.0 2023 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
  • IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
  • IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems
  • IEC 62133-2:2017+AMD1:2021 Edition1.1 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications – Part 2: Lithium systems

2) Biocompatibility Test

The component of the Solawave 2-in-1 Skincare Mini (Model: 61043) has been conformed to ISO 10993-5, ISO 10993-10, and ISO 10993-23.

3) Software verification and validation

Software verification and validation testing was conducted and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered as a "moderate" level concern since a malfunction of or a latent design flaw in the Software Device leads to an erroneous diagnosis or a delay in the delivery of appropriate medical care that would likely lead to Minor Injury.

7.2 Summary of Clinical Performance

Clinical testing was not needed for this 510(k).

8. Date Prepared: May 22, 2025

9. Final Conclusion

The subject device is as safe as effective and performs as well as the legally marketed predicate device cleared under K213841, Class II (21 CFR 878.4810), product code OHS.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.