K Number
K250467

Validate with FDA (Live)

Date Cleared
2025-06-05

(107 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Red Light Hair Growth Cap is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV.

Device Description

The Red Light Hair Growth Cap is an over the counter (OTC) home-use, non-invasive device indicated to treat androgenetic alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV.

The Red Light Hair Growth Cap is a portable, therapeutic device whose purpose is to provide even, cool, narrow-band wavelengths of polychromatic light (660nm red light) produced by LEDs to treat androgenetic alopecia and promote hair grown.

The device comprises of a wearable soft textile cap, a controller, an AC adaptor and a USB cable. The device can be powered by the AC adaptor which connected to the mains directly, also it can be powered by a power bank. The device is operated by the controller. There are 2 buttons on controller, ON/Off button is used to turn on/off the device, time button is used for setting treatment time from 10 to 30 minutes, each press in 5-minute increments./decrements. The device will shut off automatically after reach treatment time.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the "Red Light Hair Growth Cap" do not contain the detailed clinical study information typically associated with acceptance criteria, reported device performance for clinical endpoints, sample sizes for test/training sets, expert qualifications, or adjudication methods for AI/software devices. This is because the device, an "Infrared Lamp" (a physical medical device), was cleared based on its substantial equivalence to predicate devices and extensive non-clinical (electrical safety, EMC, software verification/validation) testing, rather than a clinical effectiveness study.

The summary clearly states: "Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively."

Therefore, I cannot directly extract the requested information (points 1-9) in the manner typically expected for an AI/software medical device that undergoes rigorous clinical validation with specific performance metrics against a ground truth.

However, I can describe the acceptance criteria and study type that proves the device meets those criteria based on the provided document:


Acceptance Criteria and Study for the Red Light Hair Growth Cap

The Red Light Hair Growth Cap (LP-RJVGRW-BLK) received FDA 510(k) clearance based on its substantial equivalence to legally marketed predicate devices, combined with comprehensive non-clinical performance testing. The core "acceptance criteria" for this specific submission were centered around safety and effectiveness being comparable to existing devices and meeting established electrical and software safety standards.

1. A table of acceptance criteria and the reported device performance

Since this was a 510(k) based on substantial equivalence and non-clinical testing, the "acceptance criteria" are primarily related to safety standards and comparison to predicate device characteristics. There are no specific clinical performance metrics for "hair growth" reported in this document.

Acceptance Criteria CategorySpecific Criteria/Standard MetReported Device Performance/Compliance
Electrical SafetyIEC 60601-1: 2020Conforms to standard
IEC 60601-1-2: 2020 (EMC)Conforms to standard
IEC 60601-1-11: 2020 (Home Healthcare)Conforms to standard
IEC 60601-2-57: 2011 (Non-Laser Light Source)Conforms to standard
Photobiological SafetyIEC 62471: 2006Conforms to standard
Software Verification & ValidationFDA Guidance for Industry and Staff: "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Basic Documentation Level)Software verification and validation conducted; documentation provided.
Substantial Equivalence (Indications for Use)Treat Androgenetic Alopecia and promote hair growth in specific male (Norwood-Hamilton IIa-V) and female (Ludwig-Savin I-1 to I-4, II-1, II-2 or frontal) hair loss patterns, with Fitzpatrick Skin Types I-IV.Same as predicate devices (CurrentBody Skin™ Led Hair Regrowth and Biophotas Celluma RESTORE).
Substantial Equivalence (Technical Characteristics)Similar design principle, energy type (LED red light), peak wavelength, intensity, treatment dose, treatment time, control mechanisms, and electrical power.Similar to predicate devices. Differences in intensity, treatment dose (Note 1) and treatment time (Note 2) were deemed not to affect safety or effectiveness. Electrical power differences (Note 3) met safety standards.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set (Clinical): Not applicable. Clinical testing was explicitly stated as "not needed" for this 510(k). The evaluation relied on non-clinical engineering and software testing.
  • Data Provenance: The document does not specify country of origin for any safety testing data, only that the sponsor is Shenzhen Idea Light Limited, China. The testing itself is non-clinical (lab-based engineering/software V&V).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. There was no clinical "test set" requiring expert ground truth establishment for clinical endpoints.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. There was no clinical "test set" requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is not an AI/software diagnostic tool; it is a therapeutic device that does not involve human readers interpreting output or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical therapeutic device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • For the non-clinical testing, the "ground truth" refers to the established requirements and specifications of the referenced IEC standards (e.g., IEC 60601-1 for electrical safety) and FDA software guidance. Compliance with these standards served as the "ground truth" for safety and performance.
  • For substantial equivalence for indications for use, the "ground truth" was the cleared indications for use of the predicate devices.

8. The sample size for the training set

  • Not applicable. This device is not an AI/machine learning model and therefore does not have a "training set" in the context of data-driven algorithm development.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set.

FDA 510(k) Clearance Letter - Red Light Hair Growth Cap

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 5, 2025

Shenzhen Idea Light Limited
Nicole Liu
Manager
3rd Floor, Block B, Shengdelan Industry Park,
Longhua District
Shenzhen, Guangdong 518110
China

Re: K250467
Trade/Device Name: Red Light Hair Growth Cap (LP-RJVGRW-BLK)
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP,
Dated: May 13, 2025
Received: May 14, 2025

Dear Nicole Liu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250467 - Nicole Liu Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250467 - Nicole Liu Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by TANISHA L. HITHE -S
Date: 2025.06.05 14:55:53 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)
K250467

Device Name
Red Light Hair Growth Cap (LP-RJVGRW-BLK)

Indications for Use (Describe)
The Red Light Hair Growth Cap is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Sponsor: Shenzhen Idea Light Limited
Subject Device: Red Light Hair Growth Cap
Page 1 / 5

510(k) Summary of K250467

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92

1. Submitter's Information

Sponsor Name: Shenzhen Idea Light Limited
Establishment Registration Number: 3015287734
Address: 3rd Floor, Block B, Shengdelan Industry Park, Guanlan, Longhua District Shenzhen Guangdong, CN 518110
Contact Person (including title): Nicole Liu (Manager)
Tel: +86 0755 21055894
Fax: +86 0755 28020589
E-mail: Nicole@szidealight.com

Application Correspondent:
Application Correspondent:
Contact Person: Nicole Liu
Shenzhen Idea Light Limited
Address: 3rd Floor, Block B, Shengdelan Industry Park, Guanlan, Longhua District Shenzhen Guangdong, CN 518110
Tel: +86 0755 21055894
E-mail: Nicole@szidealight.com

2. Subject Device Information

Trade Name: Red Light Hair Growth Cap
Model: LP-RJVGRW-BLK
Regulation: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP

3. Primary Predicate Device Information

Sponsor: Light Tree Ventures Europe B.V.
Trade Name: CurrentBody Skin™ Led Hair Regrowth(Model: MZ-07)
Regulation: 21 CFR 890.5500

Page 6

Sponsor: Shenzhen Idea Light Limited
Subject Device: Red Light Hair Growth Cap
Page 2 / 5

510(k) Number: K230336
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP

Predicate Device 2 information:
Sponsor: Biophotas, Inc.
Trade Name: Biophotas Celluma RESTORE
Regulation: 21 CFR 890.5500
510(k) Number: K211038
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP

4. Device Description

The Red Light Hair Growth Cap is an over the counter (OTC) home-use, non-invasive device indicated to treat androgenetic alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV.

The Red Light Hair Growth Cap is a portable, therapeutic device whose purpose is to provide even, cool, narrow-band wavelengths of polychromatic light (660nm red light) produced by LEDs to treat androgenetic alopecia and promote hair grown.

The device comprises of a wearable soft textile cap, a controller, an AC adaptor and a USB cable. The device can be powered by the AC adaptor which connected to the mains directly, also it can be powered by a power bank. The device is operated by the controller. There are 2 buttons on controller, ON/Off button is used to turn on/off the device, time button is used for setting treatment time from 10 to 30 minutes, each press in 5-minute increments./decrements. The device will shut off automatically after reach treatment time.

5. Intended Use / Indications for Use

The Red Light Hair Growth Cap is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV.

6. Test Summary

6.1 Summary of Non-Clinical Performance Testing

Page 7

Sponsor: Shenzhen Idea Light Limited
Subject Device: Red Light Hair Growth Cap
Page 3 / 5

Performance Testing Summary

The Red Light Hair Growth Cap (Model: LP-RJVGRW-BLK) has been thoroughly evaluated for electrical safety and performance and has been found to conform to the following standards.

1) Electrical safety and electromagnetic compatibility (EMC)

  • IEC 60601-1: 2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
  • IEC 60601-1-2: 2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
  • IEC 60601-1-11:2020 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • IEC 60601-2-57:2011 Medical Electrical Equipment - Part 2-57: Particular Requirements for The Basic Safety And Essential Performance Of Non-Laser Light Source Equipment Intended For Therapeutic, Diagnostic, Monitoring And Cosmetic/Aesthetic Use
  • IEC 62471:2006 Photobiological safety of lamps and lamp systems.

2) Software verification and validation testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as Basic Documentation level, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

6.2 Clinical Performance

Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.

7. Comparison to predicate device and conclusion

Compared with predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate devices do not raise any new questions of safety or effectiveness.

Page 8

Sponsor: Shenzhen Idea Light Limited
Subject Device: Red Light Hair Growth Cap
Page 4 / 5

Elements of ComparisonSubject DevicePrimary Predicate DeviceReference deviceVerdict
CompanyShenzhen Idea Light LimitedLight Tree Ventures Europe B.V.Biophotas Inc--
Trade NameRed Light Hair Growth CapCurrentBody Skin™ Led Hair Regrowth(Model: MZ-07)Biophotas Celluma RESTORE--
510(k) NumberK250467K230336K211038--
Product CodeOAPOAPOAPSame
Device classificationClass IIClass IIClass IISame
Regulation number21 CFR 890.550021 CFR 890.550021 CFR 890.5500Same
Intended Use and Indications.The Red Light Hair Growth Cap is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I-IV.The CurrentBody Skin™ Led Hair Regrowth(Model: MZ-07) is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa – V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.The BioPhotas Celluma RESTORE is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa – V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV.Same
Intended location of useScalpScalpScalpSame
Energy typeLight emitting diodesVisible red light-emitting diodesLight emitting diodesSame
Peak Wavelength (FWHM)Red: 660 nm660nmRed: 640nm+/-25nm (615-665nm)Same
Intensity (mw/cm2)1.631.672.77Similar Note 1
Treatment Dose (J/cm2)0.98-2.9414.98Similar Note 1
Treatment timeEvery day 10-30 minutes.Every day 10 minsEach Treatment: 30 minutes; Total treatment: every other day, 16Same Note 2

Page 9

Sponsor: Shenzhen Idea Light Limited
Subject Device: Red Light Hair Growth Cap
Page 5 / 5

Elements of ComparisonSubject DevicePrimary Predicate DeviceReference deviceVerdict
weeks
ControlDevice uses a timer and software to control treatment durationNot public available.Device uses a timer and software to control treatment durationSame
Electrical power100-240VacNot public available.110-120VSimilar Note 3
Electrical safetyIEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57Complied with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 6060-2-5760601-1: 2012 60201-1-2: 2014Same

Comparison in Detail(s):

Note 1:
The "intensity" and "treatment dose" are very close to the predicate devices. Their wavelength is the same, the "intensity" and "treatment dose" are very closed to the primary predicate device. So, these differences will not affect the safety and effectiveness of the subject device compared to the predicate device.

Note 2:
The recommended treatment time for the subject device is 10-30 minutes each day, the 10 minutes limit is the same as K230336, and the 30 minutes limit is the same as the K211038. Moreover, the "intensity" and "treatment dose" are close to the two predicate devices. So, the slightly difference will not raise any safety or effectiveness issues.

Note 3:
The electrical power of subject device is different from predicate device, but it meets the requirement of IEC 60601-1. So, the difference will not raise any safety or effectiveness issues.

Final Conclusion:

The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated devices K211038 and K230336.

8. Date of the summary prepared: May 13, 2025

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.