K Number
K250411

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-03-12

(27 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval Devices are indicated for use as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgery.

Device Description

The DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System are sterile single patient use devices, which comprise of a flexible plastic bag with a deployment mechanism. The bag is supplied pre-loaded in the introducer assembly and consists of a large, easily accessible opening and a closure suture that facilitates closure of the specimen bag after the specimen(s) have been collected. The deployment mechanism consists of a push-pull rod and an introducer assembly. The deployment mechanism allows for easy insertion through the cannula and full deployment of the bag with the use of the biasing arms.

AI/ML Overview

The provided FDA 510(k) clearance letter for the DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System (K250411) does not contain information related to software, artificial intelligence (AI), diagnostic performance, or studies involving human readers or ground truth for diagnostic accuracy.

The document describes a medical device, a specimen retrieval system, and its non-clinical testing to demonstrate substantial equivalence to a predicate device. The information requested regarding acceptance criteria, device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone AI performance, and ground truth establishment is not applicable to a submission for a physical surgical device like this.

Therefore, I cannot populate the requested table or answer the specific questions related to AI-driven diagnostic studies based on the provided text.

Here's a summary of what is available regarding acceptance criteria and studies, framed according to the nature of this physical device:

1. Table of Acceptance Criteria and Reported Device Performance

For this device, "acceptance criteria" and "device performance" are primarily related to its physical characteristics, functionality, and safety as compared to a predicate device, rather than diagnostic accuracy.

Characteristic/CriterionAcceptance MethodologyReported Device Performance (as demonstrated by testing)
Intended UseComparison to predicateMatches predicate: "receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgery."
Material BiocompatibilityConformance to ISO 10993 series standards (1, 5, 7, 10, 11, 23)Conforms to specified ISO 10993 standards.
Sterilization EfficacyConformance to ISO 11135:2014 for Ethylene Oxide (EtO) sterilization to an SAL of 10^-6Sterilized using EtO to an SAL of 10^-6.
Aging StabilityAging StudySuccessfully passed aging study (details of specific criteria not provided in this document).
General Device OperationPerformance Studies (e.g., deployment, function)Met design specifications (details of specific metrics not provided).
Bag OpeningPerformance StudiesSuccessfully demonstrates bag opening (details of specific metrics not provided).
Bag VolumePerformance StudiesAvailable volumes from 120 to 1500 mL, matching predicate.
Bag Integrity & Seam StrengthPerformance StudiesSuccessfully demonstrates integrity and seam strength (details of specific metrics not provided).
Introducer DiameterDesign Specification5, 10, and 12mm, matching predicate.
Prescription UseRegulatory ClassificationPrescription Use (Part 21 CFR 801 Subpart D), matching predicate.
Intended EnvironmentDesign SpecificationProfessional Healthcare Facility (Surgical Room or Operating theatre), matching predicate.

Given that this is a 510(k) for a physical medical device (a specimen retrieval system) and not an AI/software-as-medical-device, the following points are not applicable as they pertain to AI/diagnostic studies:

  1. Sample sizes for test set and data provenance: Not applicable. No test set of diagnostic data was used.
  2. Number of experts used to establish ground truth & qualifications: Not applicable. No ground truth for diagnostic accuracy was established.
  3. Adjudication method: Not applicable.
  4. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable. No human readers were evaluated for improvement with AI assistance.
  5. Standalone (algorithm only) performance: Not applicable. There is no algorithm.
  6. Type of ground truth used: Not applicable.
  7. Sample size for the training set: Not applicable. There is no training set for an AI/algorithm.
  8. How the ground truth for the training set was established: Not applicable.

The "studies" conducted for this device were non-clinical performance and biocompatibility tests to demonstrate the device's physical and material safety and functionality, and its substantial equivalence to a legally marketed predicate device.

FDA 510(k) Clearance Letter - DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

March 12, 2025

Dannik, LLC
Olga Haberland
President
941 W Morse Blvd. Suite 100
Winter Park, FL 32789

Re: K250411
Trade/Device Name: DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: GCJ
Dated: February 13, 2025
Received: February 14, 2025

Dear Olga Haberland:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250411 - Olga Haberland Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250411 - Olga Haberland Page 3

Sincerely,

James H. Jang -S Digitally signed by James H. Jang -S Date: 2025.03.12 22:31:22 -04'00'

Long Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

K250411

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K250411

Device Name
DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System

Indications for Use (Describe)
The DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval Devices are indicated for use as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgery.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

www.dannik.us
941 West Morse Blvd. Suite 100, Winter Park, FL 32789 U.S.A. 007-1

510(k) TBD
510(k) SUMMARY

1. SUBMITTER'S CONTACT INFORMATION

Company: DANNIK, LLC
Address: 941 West Morse Blvd. Suite #100 Winter Park, FL 32789
Contact Person: Olga Haberland, President
Phone: (407) 745-1698

2. DEVICE NAME

Trade Name – DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System
Common Name – Retrieval Bag, Specimen Retrieval Bag
Regulation Number – 21 CFR 876.1500
Classification Name – Laparoscope, General & Plastic Surgery
Product Code – GCJ
Device Classification – Class II
Classification Panel – General and Plastic Surgery

3. PREDICATE DEVICE

The DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval Devices claim Substantial Equivalence to the Unimax Specimen Retrieval System cleared under 510(k) K103510.

4. DEVICE DESCRIPTION

The DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System are sterile single patient use devices, which comprise of a flexible plastic bag with a deployment mechanism. The bag is supplied pre-loaded in the introducer assembly and consists of a large, easily accessible opening and a closure suture that facilitates closure of the specimen bag after the specimen(s) have been collected. The deployment mechanism consists of a push-pull rod and an introducer assembly. The deployment mechanism allows for easy insertion through the cannula and full deployment of the bag with the use of the biasing arms.

5. INDICATIONS FOR USE

The DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval Devices are indicated for use as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgery.

K250411

Page 6

www.dannik.us
941 West Morse Blvd. Suite 100, Winter Park, FL 32789 U.S.A. 007-2

6. SUBSTANTIAL EQUIVALENCE TABLE

CategorySubject DevicePrimary Predicate
DeviceDANNIK® Laparoscopic Single-Use Poly Specimen Retrieval System (510(k) TBD)Unimax Specimen Retrieval System (510(k) K103510)
Intended UseThe DANNIK® Laparoscopic Single-Use Poly Specimen Retrieval Systems are intended to be used as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgery.The Unimax Specimen Retrieval Systems are intended to be used as a receptacle for the collection and extraction of tissue, organs, and calculi during general and laparoscopic surgery.
Product Picture[Images showing Introducer Shaft, Handle, Push Rod, and Specimen Bag components][Images showing Handle, Push Rod, Introducer Shaft, and Specimen Bag components]
DesignThese devices include a handle attached to an introducer shaft with the specimen bag pre-loaded inside designed to fit down standard 5 to 15mm trocar sheaths. Advancing the push rod deploys the bag from the stored position into the operating field. Metal arms secured to the distal end of the push rod automatically open the mouth of the bag upon deployment providing a large easily accessible opening to capture the tissue. Retracting the push rod removes the arms and simultaneously cinches the closure suture to fully close the mouth of the bag.Same
Bag Volume120 to 1500 mLSame
Introducer Diameter5, 10 and 12mmSame
BiocompatibilityConforms to ISO 10993Unknown
SterilizationSterilized using Ethylene Oxide for single patient use in accordance with ISO 11135 to an SAL of 10^-6.Sterilized using Ethylene Oxide
Prescription UseYesYes
Intended EnvironmentProfessional Healthcare Facility (Surgical Room or Operating theatre)Same

7. NONCLINICAL TESTS

Nonclinical testing has been conducted to verify that these devices met all design specifications and are substantially equivalent to the predicate device. Testing included the following:

  • Biocompatibility Testing performed in accordance with the following:
    • ISO 10993-1: 2018 – Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process (Recognition No. 2-258)
    • ISO 10993-5:2009 – Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (Recognition No. 2-245)
    • ISO 10993-7:2008 – Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (Recognition No. 14-408)
    • ISO 10993-10:2021 – Biological evaluation of medical devices - Part 10: Tests for skin sensitization (Recognition No. 2-296)
    • ISO 10993-11:2017 – Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (Recognition No. 2-255)

Page 7

www.dannik.us
941 West Morse Blvd. Suite 100, Winter Park, FL 32789 U.S.A. 007-3

  • ISO 10993-23:2021 – Biological evaluation of medical devices - Part 10: Tests for irritation (Recognition No. 2-291)
  • Aging Study
  • Ethylene Oxide Sterilization Validation per ISO 11135:2014 – Sterilization of health-care products - Ethylene Oxide (Recognition No. 14-529)

In addition, these devices have been compared to the predicate device through various performance studies designed to test the general device operation and bag opening, volume, integrity and seam strength.

8. CLINICAL TESTS

There were no clinical trials performed on these devices.

9. CONCLUSIONS

The subject device has equivalent indications for use as the predicate device. There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.