AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Depth Electrode is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.

Device Description

The Depth Electrode consists of the Depth Electrode (electrode) and accessories. The accessories consists of Connection cable and Guidance screw. The Depth Electrodes are provided sterile after EO sterilization and are single use only. The electrodes are slender, cylindrical tubes inserted into the brain. The electrodes have conductive contacts at the distal end to monitor the brain's electrical signals or deliver electrical stimulation.

The electrode is a 0.8 mm diameter Polyurethane electrode with platinum-iridium alloy contacts and are available in varying numbers of contacts: 8 to 16 contacts, with an exploration length (recording depth) from 26.5 to 63.0 mm. The longitudinal width of each contact is 2.0 mm, and are spaced 1.5 mm apart (distance between contact groups is 10mm).

The Guidance screws are secured in the skull to provide an access point while stabilizing the electrode. It consists of a tube through which the electrode can pass. The proximal end is threaded, allowing the surgeon to screw it into the patient's skull via an appropriately sized burr hole. The distal end is threaded to accept the Sealing Cap of electrodes. Guidance screws are available in 20 mm, 25 mm, 30 mm, and 35 mm lengths, and can only be used by inserting through a small 2.1mm burr hole drilled in the skull. Guidance screws should be used only when depth electrodes are warranted.

The Connection cables are used for the transmission of electrical signals that connects to EEG monitoring equipment. When performing EEG monitoring, the connection cable of each model shall be used in conjunction with the electrode of the corresponding channel.

The Depth Electrode is temporarily placed (less than 30 days) at the subsurface level of the brain. The system is designed to be used in stereoelectroencephalography (sEEG) procedures.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - K250363

Page 1

November 5, 2025

Beijing Sinovation Medical Technology Co., Ltd.
Lina Wang
RA Specialists
Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road
Fengtai District
Beijing, 100070
China

Re: K250363

Trade/Device Name: Depth Electrode (RSDE-08); Depth Electrode (RSDE-10); Depth Electrode (RSDE-12); Depth Electrode (RSDE-14); Depth Electrode (RSDE-16); Depth Electrode (RSDE-16F); Depth Electrode (RSDE-08S); Depth Electrode (RSDE-10S); Depth Electrode (RSDE-12S); Depth Electrode (RSDE-14S); Depth Electrode (RSDE-16S); Depth Electrode (RSDE-16G); Depth Electrode (RSDE-18); Depth Electrode (RSDE-18F); Depth Electrode (RSDE-18T); Depth Electrode (RSDE-20); Depth Electrode (RSDE-20F); Depth Electrode (RSDE-20T); Guidance Screw (DX-20); Guidance Screw (DX-25); Guidance Screw (DX-30); Guidance Screw (DX-35); Connection Cable (LX-12); Connection Cable (LX-16); Connection Cable (LX-20); Connection Cable (LX-S); Connection Cable (LX-RF2)

Regulation Number: 21 CFR 882.1330
Regulation Name: Depth Electrode
Regulatory Class: Class II
Product Code: GZL
Dated: December 3, 2024
Received: February 10, 2025

Dear Lina Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database

Page 2

K250363 - Lina Wang Page 2

available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Page 3

K250363 - Lina Wang Page 3

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Patrick Antkowiak -S

for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K250363

Please provide the device trade name(s).

  • Depth Electrode (RSDE-08);
  • Depth Electrode (RSDE-10);
  • Depth Electrode (RSDE-12);
  • Depth Electrode (RSDE-14);
  • Depth Electrode (RSDE-16);
  • Depth Electrode (RSDE-16F);
  • Depth Electrode (RSDE-08S);
  • Depth Electrode (RSDE-10S);
  • Depth Electrode (RSDE-12S);
  • Depth Electrode (RSDE-14S);
  • Depth Electrode (RSDE-16S);
  • Depth Electrode (RSDE-16G);
  • Guidance Screw (DX-20);
  • Guidance Screw (DX-25);
  • Guidance Screw (DX-30);
  • Guidance Screw (DX-35);
  • Connection Cable (LX-12);
  • Connection Cable (LX-16);
  • Connection Cable (LX-20);
  • Connection Cable (LX-S);
  • Connection Cable (LX-RF2)

Please provide your Indications for Use below.

The Depth Electrode is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.

Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

510(k) Summary

I. Submitter

BEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.
Room 218, 2nd Floor, Building 1, No. 1, Hangfeng Road, Fengtai District, Beijing, P.R. China;100070
Tel:+86 10 80481789
Contact person:Lina Wang
Date Prepared:Oct 26, 2025

II. Device

Name of Device: Depth Electrode
Common or User name: Depth Electrode
Classification Name: Depth Electrode(21 CFR 882.1330)
Regulatory Class: II
Product Code: GZL

III. Predicate device

Predicate Device: Evo@sEEG System
Predicate 510(K) number: K222404
Manufacture: NeuroOne Medical Technologies Corp.
The 510(k) summary of the predicate device is provided as K222404.pdf
This predicate has not been subject to a design-related recall.

Predicate Device: DIXI Medical Microdeep Depth Electrode
Predicate 510(K) number: K170959
Manufacture: DIXI Medical.
The 510(k) summary of the predicate device is provided as K170959.pdf

Predicate Device: Depth Electrodes
Predicate 510(K) number: K163355
Manufacture: Ad-tech Medical Instrument Corporation
The 510(k) summary of the predicate device is provided as K163355.pdf

No reference devices were used in this submission.

IV. Description of Device

The Depth Electrode consists of the Depth Electrode (electrode) and accessories. The accessories consists of Connection cable and Guidance screw. The Depth Electrodes are provided sterile after EO sterilization and are single use only. The electrodes are slender, cylindrical tubes inserted into the brain. The electrodes have conductive contacts at the distal end to monitor the brain's electrical signals or deliver electrical stimulation.

Page 6

The electrode is a 0.8 mm diameter Polyurethane electrode with platinum-iridium alloy contacts and are available in varying numbers of contacts: 8 to 16 contacts, with an exploration length (recording depth) from 26.5 to 63.0 mm. The longitudinal width of each contact is 2.0 mm, and are spaced 1.5 mm apart (distance between contact groups is 10mm).

The Guidance screws are secured in the skull to provide an access point while stabilizing the electrode. It consists of a tube through which the electrode can pass. The proximal end is threaded, allowing the surgeon to screw it into the patient's skull via an appropriately sized burr hole. The distal end is threaded to accept the Sealing Cap of electrodes. Guidance screws are available in 20 mm, 25 mm, 30 mm, and 35 mm lengths, and can only be used by inserting through a small 2.1mm burr hole drilled in the skull. Guidance screws should be used only when depth electrodes are warranted.

The Connection cables are used for the transmission of electrical signals that connects to EEG monitoring equipment. When performing EEG monitoring, the connection cable of each model shall be used in conjunction with the electrode of the corresponding channel.

The Depth Electrode is temporarily placed (less than 30 days) at the subsurface level of the brain. The system is designed to be used in stereoelectroencephalography (sEEG) procedures.

V. Indication for use

The Depth Electrode is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.

The indications of Use statement for the Depth Electrode is identical to the predicate device.

VI. Comparison summary

The following table compares the Subject Device to the predicate device with respect to indications for use, and technological characteristics, forming the basis for the determination of substantial equivalence. The subject device does not raise any new question of safety or effectiveness as compared to the predicate device.

Device Classification Comparison

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Trade nameDepth ElectrodeEvo@sEEG SystemDIXI Medical Microdeep Depth ElectrodeDepth ElectrodesN/A
ManufacturerBEIJING SINOVATION MEDICAL TECHNOLOGY CO.,LTD.NeuroOne Medical Technologies Corp.DIXI MedicalAd-tech Medical Instrument CorporationN/A
Device882.1330882.1330882.1330882.1330Identical

Page 7

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Regulation Number: Device/Regulation Name:Electrode, DepthElectrode, DepthElectrode, DepthElectrode, DepthIdentical
Regulation Description/Common name:Depth electrodeDepth electrodeElectrode, DepthElectrode, DepthIdentical
Product Code:GZLGZLGZLGZLIdentical
Device Class/Regulation Classification:Class IIClass IIClass IIClass IIIdentical

Intended Use/Indication for Use Comparison

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Intended Use:The device is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.The device is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.The device are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brainIdentical
Indication for UseThe device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.The device is intended for temporary (less than 30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain.Identical

Electrode Technological and Performance Characteristics Comparison

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Electrode Point0.8 mm0.8 mm0.8 mm0.86-1.96 mmIdentical

Page 8

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Diameter
Electrode MaterialPolyurethanePolyamide//Different
Number of Contacts8-165-165-18Up to 16Minor different
Contact Height/Length2 mm2 mm2 mm/Same
Contact spacing1.5mm1.5-3.2 mm1.5-11 mm/Minor different
Contact Materialplatinum-iridiumPlatinumplatinum-iridiumPlatinumIdentical
Exploration Length (Recording Depth)26.5-63 mm16-80 mm16-80.5 mmMinor different
SterileEO sterilizationEO sterilizationEO sterilizationEO sterilizationSame
Single-UseyesyesyesYesSame
Environment of UseIntraoperative and Neurological monitoring locations at subsurface levels of the brainIntraoperative and Neurological monitoring locations at subsurface levels of the brainIntraoperative and Neurological monitoring locationsIntraoperative and NeurologicalIdentical
Duration of UseLess than 30 daysLess than 30 daysLess than 30 daysLess than 30 daysSame
Maximum Stimulation Charge Density≤ 30 μC/cm²/≤ 30 μC/cm²≤ 30 μC/cm²Identical
Where usedNeurosurgical operating roomNeurosurgical operating room//Identical
Principles of OperationDepth electrode are temporarily placed at subsurface levels of the brain to record, monitor, and stimulate electrical signalsDepth electrode are temporarily placed at subsurface levels of the brain to record, monitor, and stimulate electrical signalsDepth electrodes are temporarily placed at subsurface levels of the brain to record, monitor, and stimulate electrical signals/Identical

Guidance Screw Technological and Performance Characteristics Comparison

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Depth Electrode Diameter0.8 mm0.8 mm0.8 mm/Same

Page 9

Subject devicePredicate Device K222404Predicate Device K170959Predicate Device K163355Comparison
Compatibility Material(s)Titanium alloysTitaniumTitanium alloys/Same as K170959
Lengths Offered20-35 mm20-35 mm15-35 mm/Same as K222404
Guidance Screw ID0.9 mm0.9 mm0.8 mm/Same as K222404
Single-Useyesyesyes/same
SterileEO sterilizationEO sterilizationEO sterilization/same
Environment of UseIntraoperative and Neurological monitoring locationsIntraoperative and Neurological monitoring locationsIntraoperative and Neurological monitoring locations/same
Duration of UseLess than 30 daysLess than 30 daysLess than 30 days/same
Principles of OperationThreaded into a pre-drilled hole in the skullThreaded into a pre-drilled hole in the skullThreaded into a pre-drilled hole in the skull/Identical

VII. Performance Data

The Depth Electrode has the identical intended use and similar technological characteristics as the predicate, the minor different technological characteristics are Electrode Material, Number of Contacts, Contact Material, Recording Depth of Depth Electrode and Material(s) of Guidance Screw.

Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device.

The following performance data were provided in support of substantial equivalence.

Test/StandardTest Method SummaryResult
Biocompatibility Evaluation
ᵃDepth Electrode (Tissue/bone, brain and cerebrospinal fluid contact, >24h to 30 days)
ᵇSliding Seal Tube and Sealing Cap (Indirect Tissue/bone contact, >24h to 30 days)
ᶜGuidance Screw (Tissue/bone and cerebrospinal fluid contact, >24h to 30 days)
ᵃ'ᵇ'ᶜCytotoxicity ISO 10993-5MEM Elution using L-929 Cells (GLP)Pass - Non-cytotoxic
ᵃ'ᵇ'ᶜIrritation ISO 10993-23Intracutaneous Irritation test (GLP)Pass - Non-irritant
ᵃ'ᵇ'ᶜSkin sensitization ISO 10993-10Guinea pig maximization sensitization test (GLP)Pass - Non-sensitizing
ᵃ'ᵇ'ᶜPyrogenicity USP<151> ISORabbit material-mediatedPass - Non-Pyrogenic

Page 10

Test/StandardTest Method SummaryResult
10993-11pyrogenicity test (GLP)
ᵃ'ᵇ'ᶜAcute systemic toxicity ISO 10993-11Acute systemic toxicity effects in mice (GLP)Pass - Non-toxic
ᵃ'ᵇ'ᶜSubacute/Subchronic toxicity ISO 10993-11Rat IV and IP injection study (GLP) Chemical characterization study with toxicological risk assessment per ISO 10993-17 and ISO 10993-18Pass - Non-toxic
ᶜSubcutaneous Implantation ISO 10993-61 week and 30 day Subcutaneous test in Rabbits (GLP)Pass - Minimal or no reaction
ᶜBone Implantation ISO 10993-61 week and 30 day bone implantation in Rabbits (GLP)Pass - Minimal or no reaction
ᵃ'ᶜNeurotoxicityBrain Implantation Study - 1 week and 30 day Brain implantation in Rats (GLP)Pass
Paper-based risk assessment of materials and manufacturingAcceptable
In vitro Cytotoxicity Test using neuronal cell model - The activity of differentiated PC-12 Cells and morphology of undifferentiated PC-12 CellsPass
ᵃ'ᵇ'ᶜIndirect hemolysis ASTM F756-17Hemolytic test using rabbit blood (GLP)Pass - Non-hemolytic
ᵃ'ᵇ'ᶜGenotoxicity ISO 10993-3Bacterial reverse mutation – AMES Assay (GLP)Pass - Non-mutagenic
In Vitro mouse lymphoma assay (GLP)Pass - non-mutagenic, non-clastogenic

Other types of test /verification are as follows:

TestStandards/summaryResult
bench testVerify the mechanical, electrical and functional of the devicePass
Electrical safetyIEC 60601-1Pass
EMCIEC 60601-1-2Pass
shelf lifeASTM F1980Pass
transit testingASTM D4169Pass
EO Sterilization validationISO 11135-1Pass

Page 11

TestStandards/summaryResult
EO, ECH residueISO 10993-7Pass
Bacterial Endotoxin validationusp<161>Pass
Cleaning validationISO 19227Pass

VIII. Conclusion

The Depth Electrode has the identical intended use and similar technological characteristics as the predicate device, the minor different technological characteristics are Electrode Material, Number of Contacts, Contact Material, Exploration Length of Depth Electrode and the Material(s) of Guidance Screw.

Performance testing was conducted to evaluate and characterize the performance of the device to support a determination of substantial equivalence to the predicate device. The Depth Electrode has undergone bench test, electrical, biocompatibility, packaging and sterilization testing to demonstrate the differences in the sizes and materials do not raise questions of safety or effectiveness.

§ 882.1330 Depth electrode.

(a)
Identification. A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.(b)
Classification. Class II (performance standards).