K Number
K250327

Validate with FDA (Live)

Date Cleared
2025-04-15

(69 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the OptaBlate Radiofrequency (RF) Generator System is as follows:

  • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.
  • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standards therapy.
  • Ablation of benign bone tumors such as osteoid osteoma.
Device Description

The device description of the OptaBlate Radiofrequency (RF) Generator System remains consistent with that of the previously cleared K221074, with no modifications other than the incorporation of the OptaBlate Curve 10mm, 15mm, and 20mm Probe Single Kits.

The OptaBlate Curve Probe Kits provide flexible probes with microinfusers that are used with OmniCurve 10G Fracture Kits, the OptaBlate RF Generator and the Splitter Cable. Together they enable a unipedicular approach for bone tumor ablation. The access tools from the OmniCurve Kits create a curved, unipedicular pathway to the bone tumor. Based on tumor size, a flexible probe size is selected and inserted into the curved pathway. Then with the microinfuser (which reduces impedance events), the OptaBlate RF Generator and the Splitter Cable, the bone tumor is ablated.

AI/ML Overview

This document is a 510(k) clearance letter for the OptaBlate Radiofrequency (RF) Generator System. It's a medical device, not an AI/ML product. Therefore, the information requested to describe the acceptance criteria and study proving the device meets them, particularly regarding AI-specific criteria, is not present in this document.

The document discusses the substantial equivalence of the new device (including OptaBlate Curve 10mm, 15mm, and 20mm Probe Single Kits) to a previously cleared predicate device. The testing described focuses on non-clinical aspects typical for electrosurgical devices, rather than AI/ML performance.

Here's how the provided information relates to your request:

  • No AI-specific acceptance criteria or performance study is described. This is a clearance for an RF generator and associated probes, not an AI/ML-driven diagnostic or treatment planning system.
  • The "Performance Testing" section refers to design verification testing to confirm performance as intended and safety/effectiveness in accordance with FDA Guidance for Electrosurgical Devices, not a clinical study to demonstrate performance against a specific clinical endpoint using AI.
  • "Summary of Clinical Testing: Clinical testing was not required for this Special 510(k)." This explicitly states that no clinical trial data was submitted to support this particular 510(k) given its "Special 510(k)" nature, indicating minor modifications to an already cleared device.

Therefore, I cannot populate the table or answer the specific questions related to AI/ML device performance validation based on the provided text. The document focuses on regulatory compliance through substantial equivalence to a predicate device for an electrosurgical system, not on proving AI efficacy.

FDA 510(k) Clearance Letter - OptaBlate Radiofrequency (RF) Generator System

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 15, 2025

Stryker Instruments
Josh Measures
Staff Regulatory Affairs Specialist
1941 Stryker Way
Portage, Michigan 49002

Re: K250327
Trade/Device Name: OptaBlate Radiofrequency (RF) Generator System
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: February 5, 2025
Received: March 19, 2025

Dear Josh Measures:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

Re: K250327
Trade/Device Name: OptaBlate Radiofrequency (RF) Generator System
Regulation Number: 21 CFR 878.4400
Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories
Regulatory Class: Class II
Product Code: GEI
Dated: February 5, 2025
Received: March 19, 2025

Dear Josh Measures:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

Sincerely,

James H. Jang -S
Digitally signed by James H. Jang -S
Date: 2025.04.15 16:20:11 -04'00'

For
Long Chen, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250327

Device Name: OptaBlate Radiofrequency (RF) Generator System

Indications for Use (Describe)

The intended use of the OptaBlate Radiofrequency (RF) Generator System is as follows:

  • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.
  • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standards therapy.
  • Ablation of benign bone tumors such as osteoid osteoma.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.


This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(k) Summary K250327

Submitter Information

This Premarket Notification is submitted by:

510(k) OwnerStryker Instruments1941 Stryker WayPortage, MI 49002USA
Contact PersonJosh MeasuresStaff Regulatory Affairs SpecialistEmail: josh.measures@stryker.com
Registration Number3015967359
Date Summary PreparedApril 15, 2025

Device Name

Table 1 - Subject Device Information

Subject Device Information
Trade/Proprietary NameOptaBlate Radiofrequency (RF) Generator System
Classification NameElectrosurgical cutting and coagulation device and accessories (21 CFR 878.4400)
Classification PanelGeneral & Plastic Surgery
Product CodeGEI
Device ClassClass II

Predicate Device

The legally marketed predicate for the subject device is detailed in Table 2.

Table 2 - Predicate Device Information

Predicate Device Name510(k)Product CodeManufacturer
OptaBlate Radiofrequency (RF) Generator SystemK221074GEIStryker Instruments

Indications for Use

The intended use of the OptaBlate Radiofrequency (RF) Generator System is as follows:

  • Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.
  • Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standards therapy.
  • Ablation of benign bone tumors such as osteoid osteoma.

Contraindications

Use of the OptaBlate™ RF System is contraindicated in vertebral body levels of C1 through C7.

Page 6

Device Description

The device description of the OptaBlate Radiofrequency (RF) Generator System remains consistent with that of the previously cleared K221074, with no modifications other than the incorporation of the OptaBlate Curve 10mm, 15mm, and 20mm Probe Single Kits.

The OptaBlate Curve Probe Kits provide flexible probes with microinfusers that are used with OmniCurve 10G Fracture Kits, the OptaBlate RF Generator and the Splitter Cable. Together they enable a unipedicular approach for bone tumor ablation. The access tools from the OmniCurve Kits create a curved, unipedicular pathway to the bone tumor. Based on tumor size, a flexible probe size is selected and inserted into the curved pathway. Then with the microinfuser (which reduces impedance events), the OptaBlate RF Generator and the Splitter Cable, the bone tumor is ablated.

Substantial Equivalence Comparison

The predicate device has been identified as the OptaBlate Radiofrequency (RF) Generator System RF Generator (K221074). Both the subject and predicate device share the same intended use, and indications for use and technological characteristics. Results of verification and validation demonstrate that these differences do not introduce new questions of safety and effectiveness. The subject devices are at least as safe and effective as the predicate device.

Table 3 – Intended Use Comparison

DescriptionSubject DeviceOptaBlate Radiofrequency (RF) Generator SystemPredicate DeviceOptaBlate™ Radiofrequency (RF) Generator system
ClassificationClass IIClass II
Product CodeGEIGEI
Regulation21 CFR 878.440021 CFR 878.4400
Regulation NameElectrosurgical cutting and coagulation device and accessoriesElectrosurgical cutting and coagulation device and accessories
Indications for UseThe OptaBlate™ RF Ablation System is intended for:• Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.• Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.• Ablation of benign bone tumors such as osteoid osteoma.The OptaBlate™ RF Ablation System is intended for:• Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.• Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard therapy.• Ablation of benign bone tumors such as osteoid osteoma.
Anatomical SiteBoneBone
Method of AccessPercutaneousPercutaneous
ApproachUnipedicularBipedicular

Page 7

Table 4 – Technological Characteristics Comparison

DescriptionSubject DeviceOptaBlate Radiofrequency (RF) Generator SystemPredicate DeviceOptaBlate™ Radiofrequency (RF) Generator system
Energy TypeRadiofrequency EnergyRadiofrequency Energy
Principle of OperationOperator controlled; RF delivered from compatible generatorOperator controlled; RF delivered from compatible generator
Compatible RF GeneratorOptaBlate RF GeneratorOptaBlate RF Generator
Operating ModeBipolar RF energyBipolar RF energy
Generator Maximum Output EnergySystem: 30 WPer Channel: 7.5 WSystem: 30 WPer Channel: 7.5 W
Generator Output Frequency500 kHz500 kHz
Feedback MechanismTemperature ControlledTemperature Controlled

Summary of Non-Clinical Testing

The following tests were conducted to demonstrate substantial equivalence with the predicate device:

  • Sterilization
  • Biocompatibility
  • Packaging
  • Electrical Safety and EMC Testing

Testing requirements were determined by relevant standards and FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued March 9, 2020.

Performance Testing

A range of design verification testing was conducted to confirm the subject devices perform as intended and are safe and effective for their intended use in accordance with FDA Guidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, issued March 9, 2020.

Summary of Clinical Testing

Clinical testing was not required for this Special 510(k).

Conclusion

The subject devices share the same intended use and fundamental operating principles as the legally marketed predicate devices. The modifications introduced do not raise new or different questions of safety or effectiveness. Testing results support a determination of substantial equivalence to the predicate device for the requested indications for use.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.