K Number
K250321

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-06-06

(122 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days.

GCI Sterilization Wrapper sizes and validated load limitations are as follows:

SKU / ModelSizeBorder ColorMaximum Validated Load Weight
P81818 KBRO18"x18"White1 lbs
P82424 KBRO24"x24"Pink1.82 lbs
P83030 KBRO30"x30"Yellow3.45 lbs
P83636 KBRO36"x36"Blue5.8 lbs
P84545 KBRO45"x45"Mint10.98 lbs
P85454 KBRO54"x54"Red18.5 lbs
P86060 KBRO60"x60"Green25 lbs

GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures.

Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated.

Device Description

GCI Sterilization Wrappers are intended to enclose a textile-based medical device for sterilization and maintain sterility of the enclosed device until used.

The proposed devices are medical devices that are provided non-sterile. They may be reused up to 38 times. They are available in multiple sizes to accommodate textile-based medical devices, ranging from 18"x18" to 60"x60". The size is indicated by color coding of the wrap borders for easy identification.

Load sizes should be large enough to prevent too much overlapping of the wrapper material (such that all but one side of the wrapper should be a single layer when properly folded) to allow appropriate sterilant penetration. Load sizes should be small enough to overlap by at least 2 inches on the remaining side to ensure there are no gaps.

AI/ML Overview

This document pertains to the 510(k) clearance for GCI Sterilization Wrappers, a Class II medical device. The information provided outlines the non-clinical performance data used to demonstrate substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

AttributeAcceptance CriteriaReported Device Performance (GCI Sterilization Wrappers)
Tearing Strength (ASTM D5587)>20 PSI Standard fabric direction> 21 PSI Standard fabric direction
Breaking Strength (ASTM D5034)>20 PSI> 163 PSI
Flammability (16 CFR 1610)Class 1Class 1
Water Resistance: Hydrostatic Pressure (AATCC 127)Water Resistant, >20 cm H2OWater Resistant, >27 cm H2O
Water Resistance: Impact Penetration (AATCC 42)< 4.5 g< 0.3 g
Linting< 5 log IPM< 5 log IPM
Sterility Maintenance (ANSI/AAMI/ISO 17665-1)Sterility maintained for labeled time28 Days
Cytotoxicity (ISO 10993-5)Non-CytotoxicNon-Cytotoxic
Irritation (ISO 10993-10)Non-IrritatingNon-Irritating
Sensitization (ISO 10993-10)Non-SensitizingNon-Sensitizing
Use Life and Reprocessing Effectiveness (AAMI TIR12 and TIR30)Reprocessing effectiveness confirmed for labeled use lifeUse life 38 cycles Reprocessing effectiveness confirmed
Sterilization Validation (ANSI/AAMI/ISO 17665-1)SAL of 10-6 shall be demonstrated.pass

Discussion of Study Proving Device Meets Acceptance Criteria:

The document states that performance testing was conducted in alignment with FDA guidance documents "Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities" and "Addendum to: Guidance on Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities."

Testing was also conducted to relevant standards including:

  • ASTM D5587: Tearing Strength
  • 16 CFR 1610: Flammability
  • AATCC 127: Water Resistance: Hydrostatic Pressure
  • AATCC 42: Water Resistance: Impact Penetration
  • ISO 10993-5: Biocompatibility - Cytotoxicity
  • ISO 10993-10: Biocompatibility - Irritation and Sensitization
  • ISO 9073: (Not directly referenced in table, but listed as a standard used for testing)
  • ASTM D5034: Breaking Strength
  • AAMI TIR 12: Use Life and Reprocessing Effectiveness
  • AAMI TIR 30: Use Life and Reprocessing Effectiveness
  • ANSI/AAMI/ISO 17665-1: Sterilization Validation and Sterility Maintenance

All testing results are reported to have "passed pre-determined acceptance criteria," supporting the performance of the subject device.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not explicitly state the specific number of samples used for each individual test. It mentions that "All testing results passed pre-determined acceptance criteria," suggesting that sufficient samples were tested to meet the statistical requirements of the respective standards.
  • Data Provenance: The document indicates that George Courey Inc. is located in Laval, Quebec, Canada. While the testing itself is described as "Performance testing was conducted," the specific location or country of origin where these tests were physically performed is not detailed. The data is retrospective in the context of the 510(k) submission, as these tests were completed prior to the submission for clearance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable to this type of device and study. The "ground truth" for the performance characteristics of sterilization wrappers is established by objective, standardized laboratory tests and international/national standards (e.g., ISO, ASTM, AAMI). Expert consensus by medical professionals is not typically used to establish the ground truth for properties like tearing strength, flammability, or water resistance of a material. For "Sterilization Validation" (SAL of 10-6), the "ground truth" is a measurable scientific endpoint achieved through validated sterilization cycles, not expert opinion.

4. Adjudication method for the test set:

This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image interpretation tasks where human readers make subjective assessments that might require a tie-breaking or consensus mechanism. For materials testing and performance validation of a sterilization wrapper, the results are objective measurements against defined acceptance criteria, not subjective interpretations requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable to this device. An MRMC study is relevant for diagnostic imaging AI or other applications where human interpretation is involved. GCI Sterilization Wrappers are a physical barrier device, and their performance is evaluated through laboratory-based physical, chemical, and microbiological testing, not human reader assessments.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not applicable. This is not an AI-powered diagnostic algorithm or software device. The product is a physical sterilization wrapper, and its performance is evaluated through material science and sterilization efficacy tests.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the performance of the GCI Sterilization Wrappers is based on objective measurements and adherence to established industry standards and regulatory guidance documents. This includes:

  • Physical Property Measurements: Tearing strength, breaking strength, water resistance, linting, flammability are all quantifiable physical properties measured using standardized test methods.
  • Biocompatibility Testing: Cytotoxicity, irritation, and sensitization are evaluated using established in vitro and in vivo models according to ISO 10993 standards, yielding objective results (e.g., "Non-Cytotoxic").
  • Sterilization Efficacy and Maintenance: Sterilization validation (achieving SAL of 10-6) and sterility maintenance are determined through microbiological testing and barrier integrity assessments under controlled conditions, adhering to ANSI/AAMI/ISO 17665-1.
  • Use Life and Reprocessing Effectiveness: Confirmed through repeated cycling and performance evaluation according to AAMI TIR12 and TIR30.

There is no subjective "ground truth" derived from expert consensus, pathology, or outcomes data in the traditional sense for this product.

8. The sample size for the training set:

This is not applicable. The GCI Sterilization Wrappers are a physical product, not a software algorithm that requires a "training set" in the context of machine learning or AI. The rigorous testing described is a validation process, not a training process.

9. How the ground truth for the training set was established:

This is not applicable for the same reasons as #8. There is no training set for this type of medical device.

FDA 510(k) Clearance Letter - GCI Sterilization Wrappers

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 6, 2025

George Courey Inc.
℅ Sarah Fitzgerald
Senior Consultant
Emergo by UL
2500 Bee Cave Road
Building 1, Suite 300
Austin, Texas 78746

Re: K250321
Trade/Device Name: GCI Sterilization Wrappers
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: FRG
Dated: May 6, 2025
Received: May 6, 2025

Dear Sarah Fitzgerald:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250321 - Sarah Fitzgerald Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250321 - Sarah Fitzgerald Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Christopher K. Dugard -S

Christopher K. Dugard, M.S.
Division Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250321

Device Name
GCI Sterilization Wrappers

Indications for Use (Describe)

GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days.

GCI Sterilization Wrapper sizes and validated load limitations are as follows:

SKU / ModelSizeBorder ColorMaximum Validated Load Weight
P81818 KBRO18"x18"White1 lbs
P82424 KBRO24"x24"Pink1.82 lbs
P83030 KBRO30"x30"Yellow3.45 lbs
P83636 KBRO36"x36"Blue5.8 lbs
P84545 KBRO45"x45"Mint10.98 lbs
P85454 KBRO54"x54"Red18.5 lbs
P86060 KBRO60"x60"Green25 lbs

GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures.

Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1

Page 5

K250321 510(k) Summary - GCI Sterilization Wrappers

1. Submission Sponsor

George Courey, Inc.
6620, Ernest-Cormier
Laval, Quebec, H7C 2T5
Canada
Contact: Cort Naab
Title: Director of Surgical Solutions
Telephone: 450-661-6620
E-mail: cnaab@georgecourey.com

2. Submission Correspondent

Emergo by UL
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327-9997
Email: lst.us.EmergoFDASubmissions@ul.com
Contact: Sarah Marie Fitzgerald
Title: Senior Consultant, Quality and Regulatory Affairs

3. Date Prepared

June 6, 2025

4. Device Identification

Trade/Proprietary Name: GCI Sterilization Wrappers
Common/Usual Name: Sterilization Wraps
Classification Name: Sterilization Wrap
Regulation Number: 880.6850
Product Code: FRG
Class: 2

5. Legally Marketed Predicate Device

Predicate Device name: Standard Supreme Sterilization Wrapper
510(k) number: K172207
Manufacturer: Standard Textile Co., Inc.
Common/Usual Name: Sterilization Wraps
Classification Name: Sterilization Wrap
Regulation Number: 880.6850
Product Code: FRG
Class: 2

6. Indication for Use Statement

K250321 Page 1 of 6

Page 6

K250321 Page 2 of 6

GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days.

GCI Sterilization Wrapper sizes and load limitation are as follows:

SKU / ModelSizeBorder ColorMaximum Validated Load Weight
P81818 KBRO18"x18"White1 lbs
P82424 KBRO24"x24"Pink1.82 lbs
P83030 KBRO30"x30"Yellow3.45 lbs
P83636 KBRO36"x36"Blue5.8 lbs
P84545 KBRO45"x45"Mint10.98 lbs
P85454 KBRO54"x54"Red18.5 lbs
P86060 KBRO60"x60"Green25 lbs

GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures.

Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated.

7. Device Description

GCI Sterilization Wrappers are intended to enclose a textile-based medical device for sterilization and maintain sterility of the enclosed device until used.

The proposed devices are medical devices that are provided non-sterile. They may be reused up to 38 times. They are available in multiple sizes to accommodate textile-based medical devices, ranging from 18"x18" to 60"x60". The size is indicated by color coding of the wrap borders for easy identification.

Load sizes should be large enough to prevent too much overlapping of the wrapper material (such that all but one side of the wrapper should be a single layer when properly folded) to allow appropriate sterilant penetration. Load sizes should be small enough to overlap by at least 2 inches on the remaining side to ensure there are no gaps.

8. Technological Characteristics Comparison

Table 1 compares the subject wraps to the predicate device with respect to intended use, principles of operation, technological characteristics, materials, and performance. Table 2 compares the specific indications for use.

The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

Table 1. Technological Characteristics Comparison Table

AttributeSUBJECT: K250321 GCI Surgical WrappersPREDICATE: K172207 Sterilization WrappersComparison
ClassIIIISame
Product CodeFRGFRGSame

Page 7

K250321 Page 3 of 6

AttributeSUBJECT: K250321 GCI Surgical WrappersPREDICATE: K172207 Sterilization WrappersComparison
Regulation (21 CFR)880.6850880.6850Same
Intended UseEnclose another medical device for sterilization and maintain sterility of the enclosed device until used.Enclose another medical device for sterilization and maintain sterility of the enclosed device until used.Same
Method of ActionPhysical barrierPhysical barrierSame
DurabilityReusableReusableSame
Use Life38 uses75 usesSimilar*
ColorsGreen with additional colors on edging (white, pink, yellow, blue, mint, red, green)Red, Yellow, Blue, or GreenSimilar* Colors provide easy reference.
Available Sizes- 18"x18" 24"x24" - 30"x30" 36"x36" 45"x45" 54"x54" - 60"x60"- 12"x12" 18"x18" 23"x23" 27"x27" 30"x30" 35"x35" 45"x45" 54"x54" 54"x72" - 70"x70"Within Same Range (Similar*)
Maximum Wrapped Package Content WeightsAppropriate wrapping described in IFU12.9 lbsSimilar* AAMI ST77 recommends maximum of 25 lbs, and both follow this.
Device Materials99% polyester, 1% carbonPolyester, cottonSimilar*
FabricWovenWovenSame
Cut-n-SewYesYesSame
Performance TestingConductedConductedSame See Table 3.
SterilizationSteam sterilization cycle – 4 minutes exposure at 132°C (~270°F), Dry time 30 minutesSteam sterilization cycle – 4 minutes exposure at 270°F (~132°C), Dry time unknownSimilar*
SterilityProvided non-sterileProvided non-sterileSame
BiocompatibilityAssessed per ISO 10993-1 and tested per the ISO 10993 SeriesAssessed per ISO 10993-1 and tested per the ISO 10993 SeriesSame

* Discussion on Similarity: In all cases of differences, one or more of the following are true:

  • The subject device was assessed in alignment with relevant standards and/or guidance documents and met expectations.

Page 8

K250321 Page 4 of 6

  • The subject device met the pre-determined acceptance criteria based on the expected use of the device.
  • The characteristic is not relevant to the performance, safety, or effectiveness of the device, but rather is provided for convenience (size, border colors).
  • The information for the predicate device is unknown based on publicly available information, but as the subject device meets expectations, no further comparison is needed.

Therefore, all pre-determined acceptance criteria have been met, the device performs as intended, and any differences are not relevant to the safety, effectiveness, or performance of the subject device as compared to the predicate device.

Table 2 provides a detailed comparison of the indications for use. The predicate has a significant amount of additional descriptive data.

Table 2. Indications for Use Comparison

Proposed Indications (K250321)Predicate K172207 Cleared Indications
GCI Sterilization Wrappers are intended to be used to enclose reusable textile-based medical devices, such as surgical gowns, towels, and drapes, for sterilization by a health care provider. They are designed to allow effective steam sterilization of the enclosed textile items and to maintain sterility until use, for up to 28 days. GCI Sterilization Wrapper sizes and load limitation are as follows:Standard Supreme Sterilization Wrappers are intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. Standard Supreme Sterilization Wrappers will function as a sterilization wrap when processed according to instructions. Standard Supreme Sterilization Wrappers are reusable through 75 wash, dry, and sterilization cycles. They are manufactured and distributed as non-sterile sterilization wraps that are intended to be sterilized and processed by health care facilities and/or contract sterilization / laundry companies.
SKU / ModelSizeBorder ColorMaximum Validated Load Weight
P81818 KBRO18"x18"White1 lbs
P82424 KBRO24"x24"Pink1.82 lbs
P83030 KBRO30"x30"Yellow3.45 lbs
P83636 KBRO36"x36"Blue5.8 lbs
P84545 KBRO45"x45"Mint10.98 lbs
P85454 KBRO54"x54"Red18.5 lbs
P86060 KBRO60"x60"Green25 lbs

GCI Sterilization Wrappers are reusable through 38 wash, dry, and sterilization cycles. Pack assembly should follow established folding, assembly, and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. Closure can include autoclave tape and elastomer closures.

Sterilization: A prevacuum steam sterilization cycle (4 minutes exposure at 132°C, Dry Time: 30 minutes) has been validated.

Pack Assembly:
Pack assembly should follow established folding, assembly and wrapping procedures as defined by the clinical staff for aseptic presentation of the pack and its contents. The oblong and envelope folds have been validated for use on textile packs. Closure can include autoclave tape and elastomer closures.

Sterilization:
A prevacuum sterilization cycle (4-minute exposure at 270°F) has been validated for use on textile packs. The maximum pack size validated for use was 0.58 cubic feet with a pack weight of 12.9 pounds – pack density of 22.4 lbs./cu.ft. The maximum chamber loading

Page 9

K250321 Page 5 of 6

Proposed Indications (K250321)Predicate K172207 Cleared Indications
configuration including 85.6 pounds of packs in a 9.1 cubic foot chamber – load density of 10.4 lbs./cu.ft. Table 1: Standard Supreme Sterilization Wrappers Color and Model #'s
SizesMistyJadeCeil
12"x12"21520502--
18"x18"215215022153150221541502
23"x23"215220022153200221542002
27"x27"-21532302-
30"x30"215228022153280221542802
35"x35"215230022153300221543002
45"x45"215250022153500221545002
54"x54"215260022153600221546002
54"x72"215282022153820221548202
70"x70"215288022153880221548802

9. Non-Clinical Performance Data

Performance testing was conducted in alignment with FDA guidance "Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities" and "Addendum to: Guidance on Premarket Notification [510(k)] Submissions for Sterilizers Intended for Use in Health Care Facilities."

Testing was conducted to relevant standards including ASTM D5587, 16 CFR 1610, AATCC 127, AATCC 42, ISO 10993-5, ISO 10993-10, ISO 9073, ASTM D5034, AAMI TIR 12, and AAMI TIR 30. All testing results passed pre-determined acceptance criteria. In alignment with requirements, supporting the performance of the subject device.

Table 3. Summary of Non-Clinical Testing (Averages)

AttributeAcceptance CriteriaSUBJECT: K250321 GCI Surgical WrappersPREDICATE: K172207 Sterilization WrappersComparison
Tearing Strength (ASTM D5587)>20 PSI Standard fabric direction> 21 PSI Standard fabric directionUnknownSimilar*
Breaking Strength (ASTM D5034)>20 PSI> 163 PSI>20 PSISimilar*
Flammability (16 CFR 1610)Class 1Class 1UnknownSimilar*
Water Resistance: Hydrostatic Pressure (AATCC 127)Water Resistant, >20 cm H2OWater Resistant, >27 cm H2OUnknownSimilar*
Water Resistance: Impact Penetration (AATCC 42)< 4.5 g< 0.3 gUnknownSimilar*
Linting< 5 log IPM< 5 log IPM< 5 log IPMSame

Page 10

K250321 Page 6 of 6

AttributeAcceptance CriteriaSUBJECT: K250321 GCI Surgical WrappersPREDICATE: K172207 Sterilization WrappersComparison
Sterility Maintenance (ANSI/AAMI/ISO 17665-1)Sterility maintained for labeled time28 DaysUnknownSimilar*
Cytotoxicity (ISO 10993-5)Non-CytotoxicNon-CytotoxicNon-CytotoxicSame
Irritation (ISO 10993-10)Non-IrritatingNon-IrritatingUnknownSimilar*
Sensitization (ISO 10993-10)Non-SensitizingNon-SensitizingUnknownSimilar*
Use Life and Reprocessing Effectiveness (AAMI TIR12 and TIR30)Reprocessing effectiveness confirmed for labeled use lifeUse life 38 cycles Reprocessing effectiveness confirmedUse life 75 cycles Reprocessing effectiveness unknownSimilar*
Sterilization Validation (ANSI/AAMI/ISO 17665-1)SAL of 10-6 shall be demonstrated.passpassSame

Note: PSI is equivalent to lbf/in²

10. Clinical Performance Data

N/A

11. Conclusion Statement

The conclusion drawn from the testing demonstrates that the subject device is at least as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K172207, Class II (21 CFR 880.6850, Product Code FRG).

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).