Intended Use

The HARBOR Occlusion Device is indicated for arterial embolization in the peripheral vasculature.

Device Description

The HARBOR Occlusion Device is a self-expanding braided nitinol arterial embolization implant (Figure 1) supplied with components used for implantation (Figure 2). The Device has a radiopaque marker band attached to the proximal end of the implant. The implant is packaged collapsed within an Introducer sheath and attached to a Delivery System provided within a hoop dispenser.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the HARBOR Occlusion Device outlines the device's technical specifications and the testing performed to demonstrate its substantial equivalence to predicate devices, rather than a clinical study establishing performance criteria in relation to human interpretation or outcomes. Therefore, several requested sections (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC comparative effectiveness) are not applicable to the information provided.

However, the document does detail the acceptance criteria (implicitly, "Pass" for each test) and the studies (bench and animal) that prove the device meets these criteria in the context of its substantial equivalence submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly "Pass" for each test listed, indicating that the device must perform acceptably in those areas to demonstrate substantial equivalence.

Acceptance Criteria CategorySpecific TestReported Device Performance
Implant Biocompatibility (ISO 10993-1)Medical Device Chemical Characterization and Extractables and Leachable for Compatibility of Materials (ISO 10993-18)In accordance with ISO 10993
Implant Hemolysis TestIn accordance with ISO 10993
Implant Complement Activation TestIn accordance with ISO 10993
Toxicological Risk Assessment (ISO 10993-17)In accordance with ISO 10993
Cytotoxicity (L929 MEM Elution Cytotoxicity Assay) (ISO 10993-5)In accordance with ISO 10993
Sensitization (Kligman Maximization Sensitization in Guinea pigs with two extracts) (ISO 10993-10)In accordance with ISO 10993
Irritation (Irritation/Intracutaneous Injection Reactivity) (ISO 10993-23)In accordance with ISO 10993
Pyrogenicity (Material Mediated Rabbit Pyrogen) (ISO 10993-11)In accordance with ISO 10993
Biocompatibility Evaluation Report (ISO10993-1)In accordance with ISO 10993
Delivery System & Introducer Sheath Biocompatibility (ISO 10993-1)Cytotoxicity (L929 MEM Elution Cytotoxicity Assay) (ISO 10993-5)In accordance with ISO 10993
Sensitization (Kligman Maximization Sensitization in Guinea pigs with two extracts) (ISO 10993-10)In accordance with ISO 10993
Irritation (Irritation/Intracutaneous Injection Reactivity) (ISO 10993-23)In accordance with ISO 10993
Systemic Toxicity (Acute Systemic Injection with two extracts) (ISO 10993-11)In accordance with ISO 10993
Pyrogenicity (Material Mediated Rabbit Pyrogen) (ISO 10993-11)In accordance with ISO 10993
Hemocompatibility (ASTM Hemolysis Complete, Indirect Contact) (ISO 10993-4)In accordance with ISO 10993
Preclinical Mechanical, Visual, and Bench TestingVisual/Dimension InspectionPass
Simulated Use (Preparation/Flush, Introduction, Tracking, Advancement, Kink Resistance, Flexibility, Microcatheter Compatibility, Deployment, Retraction, Detachment, Migration Resistance, Overall Performance)Pass
Radial ForcePass
Implant Tensile StrengthPass
Joint Tensile StrengthPass
Nickel Ion ReleasePass
Corrosion ResistancePass
Magnetic Resonance (MR) TestingPass
RadiopacityPass
Occlusion TimePass
Shelf-Life Study (Products/Packaging)Pass
PyrogenicityPass
SterilityPass
Packaging ValidationValidation per ASTM D4169, ASTM D4332, ASTM F88/F88M, ASTM F2096Pass
Sterilization ValidationAchieved SAL of 10^-6Pass
GLP Animal StudyEase of delivery (friction and tortuosity)Pass
Acute ComplicationsPass
Recanalization of the vessels/durability of occlusionPass
Local and systemic foreign body reactionsPass
Device migrationPass
EffectivenessPass

2. Sample size used for the test set and the data provenance

Test Set Sample Size:

  • Bench Tests: Not explicitly stated for each test, but implied to be sufficient for meeting ISO and ASTM standards.
  • Animal Study: Not explicitly stated, but the study was described as "GLP Animal Study."

Data Provenance: Not explicitly stated. The document refers to "pre-clinical testing" and "GLP animal study," which are typically conducted in a controlled laboratory environment but the country of origin is not specified. The studies are prospective in nature, as they were conducted specifically for this device's submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This document describes the technical and biological performance of a medical device (an occlusion device) rather than an AI/CADe system or a diagnostic imaging device that would typically involve expert readers for establishing ground truth on image interpretation. The "acceptance criteria" here relate to the device's physical and biological functioning.

4. Adjudication method for the test set

Not applicable. As above, this is not a study assessing observer performance or diagnostic accuracy.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document does not describe an AI-enabled device or an MRMC study. It pertains to a physical medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This describes a physical medical device, not an algorithm.

7. The type of ground truth used

The "ground truth" for the performance tests is established by:

  • Engineering specifications and standards: For bench testing (e.g., specific tensile strength values, radial force, dimensions, nickel ion release limits).
  • Biological responses: For biocompatibility tests (e.g., absence of cytotoxicity, irritation, sensitization, pyrogenicity, hemolysis, complement activation as per ISO 10993 standards).
  • Physiological and pathological observations in an animal model: For the GLP animal study (e.g., direct observation of delivery, acute complications, vessel recanalization, foreign body reactions, device migration, and effectiveness of occlusion).

8. The sample size for the training set

Not applicable. This device is not an AI/ML algorithm or system that requires a "training set."

9. How the ground truth for the training set was established

Not applicable. No training set exists for this type of device.

FDA 510(k) Clearance Letter - HARBOR Occlusion Device

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

July 9, 2025

Nuvascular Inc.
Meadow Wang
Director, RA/QA
141 Innovation Drive, Suite 100
Irvine, California 92617

Re: K250133
Trade/Device Name: HARBOR Occlusion Device
Regulation Number: 21 CFR 870.3300
Regulation Name: Vascular Embolization Device
Regulatory Class: Class II
Product Code: KRD
Dated: June 9, 2025
Received: June 9, 2025

Dear Meadow Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250133 - Meadow Wang
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K250133 - Meadow Wang
Page 3

Sincerely,

Misti L. Malone -S

Misti Malone, PhD
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K250133

Device Name: HARBOR Occlusion Device

Indications for Use (Describe):
The HARBOR Occlusion Device is indicated for arterial embolization in the peripheral vasculature.

Type of Use (Select one or both, as applicable):
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF

Page 5

K250133
Page 1 of 11

510(k) SUMMARY

SUBMITTER

Nuvascular Inc.
141 Innovation Dr. Suite 100, Irvine, CA 92617
Phone: 626-233-6758
Contact Person: Meadow S. Wang
Date Prepared: June 8, 2025

DEVICE

Trade Name of the Device: Harbor Occlusion Device
Device Common Name: Vascular Embolization Device
Classification Name: Device, Vascular, for promoting embolization
Regulatory Class: Class II, 21 CFR 870.3300
Product Code: KRD

PREDICATE DEVICE

AGA Medical Corporation (was acquired by Abbott): AMPLATZER® Vascular Plug 4, Arterial Embolization Device (KRD) (K113658) (primary predicate)

AGA Medical Corporation (was acquired by Abbott): AMPLATZER® Vascular Plug, Arterial Embolization Device (KRD) (K031810)

DEVICE DESCRIPTION

The HARBOR Occlusion Device is a self-expanding braided nitinol arterial embolization implant (Figure 1) supplied with components used for implantation (Figure 2). The Device has a radiopaque marker band attached to the proximal end of the implant. The implant is packaged collapsed within an Introducer sheath and attached to a Delivery System provided within a hoop dispenser.

Device implant sizing and dimensions are specified in Table 1. Based on its size, the device is designed to be used with commercially available microcatheters (Table 2) under fluoroscopy for delivery and implantation in the peripheral vasculature.

Figure 1 Harbor Occlusion Device – Implant

Figure 2. Harbor Occlusion Device – Delivery System

Page 6

K250133
Page 2 of 11

The implant comes in various diameters ranging from 4.0 to 13.5 mm Unconstrained OD and 2.0 to 5.4 mm Unconstrained Length. The Harbor Occlusion Device will be available in the following with delivery system configurations:

Table 1: Device Size

Product Model NumberHarbor Occlusion Device OD, Unconstrained, DIM A (mm)Harbor Occlusion Device Length, Unconstrained, DIM C (mm)Treatable Vessel Diameter Range (mm)Compatible Microcatheter (Inch)
04-OCCLUDE-0204.02.52.0 – 2.50.017
04-OCCLUDE-0254.02.52.0 – 2.50.017
04-OCCLUDE-0304.52.82.5 – 3.00.017
04-OCCLUDE-0355.03.03.0 – 3.50.017
04-OCCLUDE-0405.53.23.5 – 4.00.017
04-OCCLUDE-0456.03.44.0 – 4.50.017
04-OCCLUDE-0506.53.64.5 – 5.00.017
04-OCCLUDE-0557.03.85.0 – 5.50.017
04-OCCLUDE-0607.54.05.5 – 6.00.017
04-OCCLUDE-0658.04.26.0 – 6.50.017
04-OCCLUDE-0708.54.46.5 – 7.00.027
04-OCCLUDE-0809.54.67.0 – 8.00.027
04-OCCLUDE-09010.54.88.0 – 9.00.027
04-OCCLUDE-10011.55.09.0 – 10.00.027
04-OCCLUDE-11012.55.210.0 – 11.00.027
04-OCCLUDE-12013.55.411.0 – 12.00.027

The delivery system attached to the Harbor Occlusion Device implant is 185 cm in length and an outer diameter suitable for delivery through commercially available Microcatheter- Terumo Headway 17 with 0.017" ID Microcatheter or Terumo Headway 27 with 0.027" ID Microcatheter.

Table 2 Compatible with the available Commercial Microcatheter

DescriptionCatalog NumberInner Diameter (In)Usable Length (cm)
Terumo Headway 17MC172150SX0.017150
Terumo Headway 27MC172156S0.027156

INDICATIONS FOR USE

The Harbor Occlusion Device is indicated for arterial embolization in the peripheral vasculature.

Page 7

K250133
Page 3 of 11

DEVICE Attributes

Table 3 list the attributes of HARBOR Occlusion Device.

Table 3 HARBOR Occlusion Device Attributes

Device AttributesNuvascular Harbor Occlusion Device (Subject Device)
Device ClassificationClass II, KRD, 21 CFR 870.3300
Intended Use/Indications for Use StatementIndicated for arterial embolization in the peripheral vasculature
Device FunctionEmbolization in the arterial peripheral vasculature
Anatomical LocationPeripheral Arterial Vasculature
Harbor (Implant)Self-expanding, Nitinol mesh occlusion device with non-blood contact platinum core to improve radiopacity. The device has a radiopaque marker band at the proximal end and an uncoated Nitinol ball at one end for attaching to the delivery system.
Implant MaterialNitinol
Implant ShapeSingle Layer Braid Device Single Lobe with Cylindrical shape
Delivery System MaterialStainless Steel Hypotube with Nitinol Wire
Introducer SheathHDPE, High Density Polyethylene
Radiopaque MarkerPlatinum Marker Band at the proximal end of the Harbor Occlusion Device
Delivery MethodDelivered and compatible with Terumo Headway 17, 0.017" or Headway 27, 0.027" microcatheterTerumo Headway 17 MC172150SX:Inner Diameter (in): 0.017Usable Length (cm) 150Terumo Headway 27 MC272156S:Inner Diameter (in): 0.027Usable Length (cm): 156
Delivery system Length185 cm
Detachment SystemMechanical Detachment
Materials of Construction for ImplantNitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%)
Materials of Construction for Delivery systemStainless Steel Hypotube with inner Nitinol wire

Page 8

K250133
Page 4 of 11

Marker BandsPlatinum
PackagingHarbor Occlusion Device includes implant attached to 185 cm long delivery system, loaded in a hoop dispenser
Sterilization ProcessEbeam
Method of SupplySterile, single use

COMPARISION OF TECHNOLOGICAL CHARACTERISTICS OF THE HARBOR OCCLUSION DEVICE WITH THE PREDICATE DEVICE

The method of application of the Harbor Occlusion Device is the same as the predicate device, AMPLATZER Vascular Plug 4, Arterial Embolization Device (KRD) (K113658) and AMPLATZER Vascular Plug, Arterial Embolization Device (KRD) (K031810). They both utilize the same principle of operation to be delivered to a select target site for the vessel and embolic used. They are both available in a range of diameters to permit the selection of appropriate sizes for target vessel embolization. Accurate placement of the Harbor Occlusion Device is conducted with visualization using radiographic (fluoroscopy) imaging.

Table 4 compares the subject device, HARBOR Occlusion Device, and the predicate.

Table 4: Comparison of Technological Characteristics with the Predicate Device

Device AttributesNuvascular HARBOR Occlusion Device (Subject Device)Abbott AMPLATZER® Vascular Plug 4 (Primary Predicate Device) (K113658)Abbott AMPLATZER® Vascular Plug (Predicate Device) (K031810)Justification for Difference (If Present)
Device ClassificationClassII, KRD, 21 CFR 870.3300ClassII, KRD, 21 CFR 870.3300ClassII, KRD, 21 CFR 870.3300Same, Identical device classification
Intended Use/ Indications for Use StatementIndicated for arterial embolization in the peripheral vasculatureIndicated for arterial and venous embolization in the peripheral vasculatureIndicated for arterial and venous embolization in the peripheral vasculatureSame Intended Use/Indication for Use Statement, the intended use of the subject device is narrower than the predicate device

Principle of Operation/Fundamental Scientific Technology

Device FunctionArterial embolization in the peripheral vasculatureEmbolization in the peripheral vasculatureEmbolization in the peripheral vasculatureSame function, subject device function is narrower than the predicate device
Anatomical LocationPeripheral Arterial VasculaturePeripheral Vasculature including arterial and venousPeripheral Vasculature including arterial and venousSame, the anatomical location of subject device is narrower than the predicate device

Device Design

Page 9

K250133
Page 5 of 11

Harbor (Implant)Self-expanding, Nitinol mesh occlusion device with non-blood contact platinum core to improve radiopacity. The device has a radiopaque marker band at the proximal end and an uncoated Nitinol ball at one end for attaching to the delivery system.Self-expanding, Nitinol mesh occlusion device. The device has a radiopaque marker band at each end and a coated stainless steel micro screw attachment at one end for attaching to the delivery system.Self-expanding, Nitinol mesh occlusion device. The device has a radiopaque marker band at each end and a micro screw attachment at one end for attaching to the delivery system.Similar device design. The results of comparative bench testing, physician usability study, histopathology, biocompatibility testing, and animal study establish the substantial equivalency of the subject device with the predicate device.
Harbor OD, Unconstrained4.0 mm4.5 mm5.0 mm5.5 mm6.0 mm6.5 mm7.0 mm7.5 mm8.0 mm8.5 mm9.5 mm10.5 mm11.5 mm12.5 mm13.5 mm4.0 mm5.0 mm6.0 mm7.0 mm8.0 mm4.0 mm6.0 mm8.0 mm10.0 mm12.0 mm14.0 mm16.0 mmSame, the OD range of the subject device is narrower than the predicate device.
Harbor Unconstrained Length2.5 mm2.8 mm3.0 mm3.2 mm3.4 mm3.6 mm3.8 mm4.0 mm4.2 mm4.4 mm4.6 mm4.8 mm5.0 mm5.2 mm5.4 mm10.0 mm10.5 mm11.0 mm7.0 mm8.0 mmThe range of Unconstrained Length for the subject device is similar to the predicated deviceThe subject device has a smaller unconstrained length compared to the predicate device.Test data show the radial force met the radial force acceptance criteria, the simulated use test shows the subject device functions are the same as the predicate device, and the animal test met acceptance criteria.The subject device is

Page 10

K250133
Page 5 of 11

Harbor (Implant)Self-expanding, Nitinol mesh occlusion device with non-blood contact platinum core to improve radiopacity. The device has a radiopaque marker band at the proximal end and an uncoated Nitinol ball at one end for attaching to the delivery system.Self-expanding, Nitinol mesh occlusion device. The device has a radiopaque marker band at each end and a coated stainless steel micro screw attachment at one end for attaching to the delivery system.Self-expanding, Nitinol mesh occlusion device. The device has a radiopaque marker band at each end and a micro screw attachment at one end for attaching to the delivery system.Similar device design. The results of comparative bench testing, physician usability study, histopathology, biocompatibility testing, and animal study establish the substantial equivalency of the subject device with the predicate device.
Harbor OD, Unconstrained4.0 mm4.5 mm5.0 mm5.5 mm6.0 mm6.5 mm7.0 mm7.5 mm8.0 mm8.5 mm9.5 mm10.5 mm11.5 mm12.5 mm13.5 mm4.0 mm4.0 mmSame, the OD range of the subject device is narrower than the predicate device.
5.0 mm
6.0 mm6.0 mm
7.0 mm
8.0 mm8.0 mm
Harbor Unconstrained Length2.5 mm2.8 mm3.0 mm3.2 mm3.4 mm3.6 mm3.8 mm4.0 mm4.2 mm4.4 mm4.6 mm4.8 mm5.0 mm5.2 mm5.4 mm7.0 mm8.0 mmThe range of Unconstrained Length for the subject device is similar to the predicated deviceThe subject device has a smaller unconstrained length compared to the predicate device.Test data show the radial force met the radial force acceptance criteria, the simulated use test shows the subject device functions are the same as the predicate device, and the animal test met acceptance criteria.The subject device is
10.0 mm10.5 mm11.0 mm

Page 11

K250133
Page 6 of 11

12.5 mm13.5 mmsubstantially equivalent to the predicate device.
Target Vessel Diameter2.0 mm-2.5 mm2.5 mm-3.0 mm3.0 mm-3.5 mm3.5 mm-4.0 mm4.0 mm-4.5 mm4.5 mm-5.0 mm5.0 mm-5.5 mm5.5 mm-6.0 mm6.0 mm-6.5 mm6.5 mm-7.0 mm7.0 mm-8.0 mm8.0 mm-9.0 mm9.0 mm-10.0 mm10.0 mm-11.0 mm11.0 mm-12.0 mm2.5 mm-3.0 mm3.5 mm-4.0 mm4.0 mm-4.5 mm4.5 mm-5.5 mm5.5 mm-6.0 mm2.5 mm-3.0 mm4.0 mm-4.5 mm5.5 mm-6.0 mm6.5 mm-7.5 mm8.0 mm-9.0 mm9.5 mm-11.0 mm10.5 mm-12.5 mmThe range of the treated target vessel diameter for the subject device is similar to the predicate device.The subject device treats vessels 2.0 mm to 12.0 mm compared to 2.5 mm to 12.5 mm for the predicate device.90-day GLP animal data, histopathology, and bench testing show that the subject device can treat 2.0 mm vessels.Test data show the radial force met the radial force acceptance criteria, the simulated use test shows the subject device functions are the same as the predicate device, and the animal test met acceptance criteria.The subject device is substantially equivalent to the predicate device.
Implant ShapeSingle Layer Braid DeviceSingle Lobe with Cylindrical shapeSingle Layer Braid DeviceDouble-Lobed with low-profile embolizationSingle Layer Braid DeviceSingle Lobe with Cylindrical shapeSame single layer braid device with similar shape
Delivery System MaterialStainless Steel Hypotube with Nitinol WirePTFE-coated Stainless Steel Wire Coil with Nitinol CoreNitinolSimilar, the biocompatibility test results of the delivery system show the subject device met the acceptance criteria of ISO 10993.

Page 12

K250133
Page 6 of 11

Target Vessel Diameter2.0 mm-2.5 mm2.5 mm-3.0 mm3.0 mm-3.5 mm3.5 mm-4.0 mm4.0 mm-4.5 mm4.5 mm-5.0 mm5.0 mm-5.5 mm5.5 mm-6.0 mm6.0 mm-6.5 mm6.5 mm-7.0 mm7.0 mm-8.0 mm8.0 mm-9.0 mm9.0 mm-10.0 mm10.0 mm-11.0 mm11.0 mm-12.0 mm12.5 mm13.5 mm2.5 mm-3.0 mm2.5 mm-3.0 mmsubstantially equivalent to the predicate device.The range of the treated target vessel diameter for the subject device is similar to the predicate device.The subject device treats vessels 2.0 mm to 12.0 mm compared to 2.5 mm to 12.5 mm for the predicate device.
3.5 mm-4.0 mm4.0 mm-4.5 mm4.5 mm-5.5 mm4.0 mm-4.5 mm
5.5 mm-6.0 mm5.5 mm-6.0 mm6.5 mm-7.5 mm90-day GLP animal data, histopathology, and bench testing show that the subject device can treat 2.0 mm vessels.Test data show the radial force met the radial force acceptance criteria, the simulated use test shows the subject device functions are the same as the predicate device, and the animal test met acceptance criteria.The subject device is substantially equivalent to the predicate device.
8.0 mm-9.0 mm9.5 mm-11.0 mm10.5 mm-12.5 mm
Implant ShapeSingle Layer Braid DeviceSingle Lobe with Cylindrical shapeSingle Layer Braid DeviceDouble-Lobed with low-profile embolizationSingle Layer Braid DeviceSingle Lobe with Cylindrical shapeSame single layer braid device with similar shape
Delivery System MaterialStainless Steel Hypotube with Nitinol WirePTFE-coated Stainless Steel Wire Coil with Nitinol CoreNitinolSimilar, the biocompatibility test results of the delivery system show the subject device met the acceptance criteria of ISO 10993.

Page 13

K250133
Page 7 of 11

Loading Device MethodIntroducer Sheath, HDPELoader, PTFELoader, PTFESimilar, the biocompatibility test results of the delivery system show the subject device met the acceptance criteria of ISO 10993
Radiopaque MarkerPlatinum Marker Band at the proximal end of the Harbor Occlusion DevicePlatinum Marker Bands at each end of the plugPlatinum Marker Bands at each end of the plugSame material with the same principle
Delivery MethodDelivered and compatible with Terumo Headway 17, 0.017" or Headway 27, 0.027" MicrocatheterTerumo Headway 17 MC172150SX:Inner Diameter (in): 0.017Usable Length (cm): 150Terumo Headway 27 MC272156S:Inner Diameter (in): 0.027Usable Length (cm): 156Delivered utilizing a Diagnostic Catheter 0.038" meeting the minimum internal diameterCatheter Size:5F with length ≤125cm4F with length ≤100cmDelivered utilizing a Diagnostic Catheter 0.056", 0.066", and 0.087" meeting the minimum internal diameterCompatible with Guide Catheter size:Vessel size vs Minimum Guide Cather Size2.5mm-6.0mm, ≥ 0.066"6.5 mm-9.0mm, ≥ 0.087"9.5 mm-12.5mm, ≥ 0.11"Catheter Length: ≤100cmSimilar delivery method.The Harbor Occlusion Device is designed to be delivered through a smaller ID (0.017" or 0.027" inner diameter) microcatheter.The results of bench testing and physician usability study establish the substantial equivalence of the subject device has substantially equivalent function as the predicate device
Delivery system Length185 cm155 cm135 cmThe Harbor Occlusion Device is designed to be delivered through a longer-length delivery systemThe results of bench testing and physician usability study establish the substantial equivalence of subject device to the predicate device
Detachment SystemMechanical DetachmentMechanical DetachmentMechanical DetachmentSame principle

Page 14

K250133
Page 7 of 11

Loading Device MethodIntroducer Sheath, HDPELoader, PTFELoader, PTFESimilar, the biocompatibility test results of the delivery system show the subject device met the acceptance criteria of ISO 10993
Radiopaque MarkerPlatinum Marker Band at the proximal end of the Harbor Occlusion DevicePlatinum Marker Bands at each end of the plugPlatinum Marker Bands at each end of the plugSame material with the same principle
Delivery MethodDelivered and compatible with Terumo Headway 17, 0.017" or Headway 27, 0.027" MicrocatheterTerumo Headway 17 MC172150SX:Inner Diameter (in): 0.017Usable Length (cm): 150Terumo Headway 27 MC272156S:Inner Diameter (in): 0.027Usable Length (cm): 156Delivered utilizing a Diagnostic Catheter 0.038" meeting the minimum internal diameterCatheter Size:5F with length ≤125cm4F with length ≤100cmDelivered utilizing a Diagnostic Catheter 0.056", 0.066", and 0.087" meeting the minimum internal diameterCompatible with Guide Catheter size:Vessel size vs Minimum Guide Cather Size2.5mm-6.0mm, ≥ 0.066"6.5 mm-9.0mm, ≥ 0.087"9.5 mm-12.5mm, ≥ 0.11"Catheter Length: ≤100cmSimilar delivery method.The Harbor Occlusion Device is designed to be delivered through a smaller ID (0.017" or 0.027" inner diameter) microcatheter.The results of bench testing and physician usability study establish the substantial equivalence of the subject device has substantially equivalent function as the predicate device
Delivery system Length185 cm155 cmCatheter Length: ≤100cm135 cmThe Harbor Occlusion Device is designed to be delivered through a longer-length delivery systemThe results of bench testing and physician usability study establish the substantial equivalence of subject device to the predicate device
Detachment SystemMechanical DetachmentMechanical DetachmentMechanical DetachmentSame principle

Page 15

K250133
Page 8 of 11

Device Material

Materials of Construction for ImplantNitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%)Nitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%)Nitinol Wire (Main Substances: Nickel: 55-57%, Titanium: 43-45%)Same
Materials of Construction for Delivery systemStainless Steel Hypotube with inner Nitinol Delivery wirePTFE-coated Stainless Steel wire coil with inner Nitinol wireNitinol wireSimilar device material construction.The subject device has stainless steel Hypotube with inner Nitinol Delivery wire without PTFE- CoatingThe materials used for the Harbor Vessel Occlusion Device met the acceptance criteria of ISO 10993 for delivery system.
Marker BandsPlatinumPlatinumPlatinumSame
Micro ScrewN/AStainless SteelStainless SteelThe subject device has no micro screw

Other Attributes

PackagingHarbor Occlusion Device includes implant attached to 185 cm long delivery system, loaded in a hoop dispenserThe device is shipped attached to a 155 cm delivery system in a hoop dispenser and is preloaded in a loader. A plastic vise is also included and may be attached to the delivery system to facilitate device detachmentThe device is shipped attached to a 135 cm. delivery system in a hoop dispenser and is preloaded in a loader. A plastic vise is also included and may be attached to the delivery system to facilitate device detachmentSimilarThe packaging validation result for the subject device shows the packaging integrity can meet the requirements per ASTM D4169, ASTM D4332, ASTM F88/F88M, ASTM F2096.
Sterilization ProcessEbeamEtO (Ethylene Oxide)EtO (Ethylene Oxide)The sterilization validation result shows Harbor Occlusion Device met SAL of 10-6
Method of SupplySterile, single useSterile, single useSterile, single useSame

Page 16

K250133
Page 9 of 11

SUMMARY OF HARBOR OCCLUSION DEVICE PERFOMANCE TESTING:

1. Bench Test

Table 5: Verification and Test Summary

Pre-clinical TestingResult
Implant, Biocompatibility Testing (ISO 10993-1)In accordance with ISO 10993
- Medical Device Chemical Characterization and Extractables and Leachable for Compatibility of Materials (ISO 10993-18)
- Implant Hemolysis Test
- Implant Complement Activation Test
- Toxicological Risk Assessment (ISO 10993- 17)
- Cytotoxicity (ISO 10993-5)
➤ L929 MEM Elution Cytotoxicity Assay
- Sensitization (ISO 10993-10)
➤ Kligman Maximization Sensitization in Guinea pigs with two extracts
- Irritation (ISO 10993-23)
➤ Irritation/Intracutaneous Injection Reactivity
- Pyrogenicity (ISO 10993-11)
➤ Material Mediated Rabbit Pyrogen
- Biocompatibility Evaluation Report (ISO10993-1)
Delivery system with Introducer Sheath, Biocompatibility testing (ISO 10993-1)In accordance with ISO 10993
- Cytotoxicity (ISO 10993-5)
➤ L929 MEM Elution Cytotoxicity Assay
- Sensitization (ISO 10993-10)
➤ Kligman Maximization Sensitization in Guinea pigs with two extracts
- Irritation (ISO 10993-23)
➤ Irritation/Intracutaneous Injection Reactivity
- Systemic Toxicity (ISO 10993-11)
➤ Acute Systemic Injection with two extracts
- Pyrogenicity (ISO 10993-11)
➤ Material Mediated Rabbit Pyrogen
- Hemocompatibility (ISO 10993-4)
➤ ASTM Hemolysis Complete, Indirect Contact

Page 17

K250133
Page 10 of 11

Preclinical Testing including Mechanical, Visual, and Bench TestingPass
– Visual/Dimension Inspection
– Simulated Use
➤ Preparation/Flush
➤ Introduction
➤ Tracking
➤ Advancement
➤ Kink Resistance
➤ Flexibility
➤ Microcatheter Compatibility
➤ Deployment
➤ Retraction
➤ Detachment
➤ Migration Resistance
➤ Overall Performance
– Radial Force
– Implant Tensile Strength
– Joint Tensile Strength
– Nickel Ion Release
– Corrosion Resistance
– Magnetic Resonance (MR) Testing
– Radiopacity
– Occlusion Time
– Shelf-Life Study (Products/Packaging)
– Pyrogenicity
➤ Sterility
Packaging validation (Products/Packaging)Pass
Sterilization validation (Sterilization)Pass
Shelf-Life Study (Products/Packaging)Pass

2. GLP Animal Study

Table 6: GLP Animal Study Summary

Pre-clinical TestingResult
Animal testing – device compared to predicate device (Acute, 30 days and 90 days)Pass
– Ease of delivery (friction and tortuosity)
– Acute Complications
– Recanalization of the vessels/durability of occlusion
– Local and systemic foreign body reactions
– Device migration
– Effectiveness

Page 18

K250133
Page 11 of 11

SUMMARY OF SUBSTANTIAL EQUIVALENCE:

The data presented in this submission demonstrates the technological similarity and equivalency of the Harbor Occlusion Device when compared with the predicate device, AMPLATZER® Vascular Plug 4, Arterial Embolization Device (KRD) (K113658) and AMPLATZER® Vascular Plug, Arterial Embolization Device (KRD) (K031810).

The Subject device,

  • has the same intended use to the predicate devices,
  • has the intended use is narrower than the predicate devices,
  • uses same operating principle as the predicate devices,
  • uses similar materials, the types of material in subject device are less than the predicate devices,
  • incorporates similar basic design and construction to the predicate devices,
  • uses a similar packaging method to the predicate devices,
  • has similar unconstrained OD range to the predicate devices,
  • has similar unconstrained Length ranges to the predicate devices

In summary, the Harbor Occlusion Device described in this submission is substantially equivalent to the predicate devices.

§ 870.3300 Vascular embolization device.

(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).