K Number
K250069

Validate with FDA (Live)

Date Cleared
2025-04-29

(109 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient.

Device Description

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim of 50 Micrograms per dm2 or Less Protein are single use. Non sterile, natural rubber latex powder-free examination gloves. They are available in one color (blue) and come in five sizes: Small, Medium, Large, Extra Large, and Extra Extra Large. The subject gloves meet all the requirements of ASTM D3578-19 Standard Specification for Rubber Examination Gloves.

AI/ML Overview

This document is a 510(k) clearance letter for medical examination gloves. It is not a document describing a study for a medical device that uses AI or requires expert ground truth. Therefore, I cannot provide the requested information.

The prompt specifically asks for details related to:

  • A table of acceptance criteria and device performance for an AI/ML device.
  • Sample sizes for test and training sets (relevant to AI/ML).
  • Number and qualifications of experts for ground truth (relevant to AI/ML).
  • Adjudication methods (relevant to AI/ML).
  • MRMC studies (relevant to AI/ML).
  • Standalone algorithm performance (relevant to AI/ML).
  • Type of ground truth used (relevant to AI/ML).
  • How ground truth was established for training (relevant to AI/ML).

The provided text discusses the non-clinical performance testing of rubber examination gloves. The "acceptance criteria" and "results" tables in the document pertain to physical and chemical properties of gloves (e.g., length, width, thickness, tensile strength, protein content, biocompatibility) tested against established ASTM and ISO standards for medical gloves.

There is no mention of an algorithm, AI, machine learning, human-in-the-loop performance, or expert review for diagnostic purposes. The "ground truth" in this context refers to the measured physical and chemical properties of the gloves conforming to established industry standards, not expert interpretations of medical images or data.

U.S. Food & Drug Administration 510(k) Clearance Letter

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 29, 2025

Comfort Rubber Gloves Industries Sdn. Bhd.
Sumathi Saravana Sami
RA Manager
Lot 821
Jalan Matang
Matang, Perak 34750
Malaysia

Re: K250069
Trade/Device Name: Powder Free Natural Rubber Latex Examination Gloves - Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-Powdered Patient Examination Glove
Regulatory Class: Class I, reserved
Product Code: LYY
Dated: March 26, 2025
Received: March 26, 2025

Dear Sumathi Saravana Sami:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K250069 - Sumathi Saravana Sami Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250069 - Sumathi Saravana Sami Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026

Indications for Use

See PRA Statement below.

510(k) Number (if known)
K250069

Device Name
Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein

Indications for Use (Describe)
Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1

Page 5

510(k) SUMMARY

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 1 of 9

K250069

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein

1.0 Submitter

Name: Comfort Rubber Gloves Industries Sdn. Bhd.
Address: Lot 821, Jalan Matang, 34750 Matang, Perak Malaysia
Phone No: 605-847 2777
Fax No: 605-847 9108
Contact Person: Sumathi d/o Saravana Sami (Miss.)
Designation: Regulatory Affair Manager
Email: s.sumathi@comfortrubber.com
Date of Preparation: 26th April 2025

2.0 Name of Device

Device Name: Powder Free Natural Rubber Latex Examination Gloves – Blue Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Common Name: Patient Examination Gloves
Classification Name: Patient Examination Gloves (21 CFR 880.6250 product code LYY)
Device Classification: Class I
510(K) Number: K250069

3.0 Predicate Device

Device Name: Blue Latex Examination Powder Free Gloves
Company: JR Engineering & medical Technologies (M) sdn. Bhd.
510(K) Number: K192329
Device Classification: Class I
Product Code: LYY

4.0 Description of the Device

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim of 50 Micrograms per dm2 or Less Protein are single use. Non sterile, natural rubber latex powder-free examination gloves. They are available in one color (blue) and come in five sizes: Small, Medium, Large, Extra Large, and Extra Extra Large. The subject gloves meet all the requirements of ASTM D3578-19 Standard Specification for Rubber Examination Gloves.

5.0 Indication for Use of the Device

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient.

Page 6

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 2 of 9

6.0 Summary of the Technological Characteristics of the Device

The technological characteristics of Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein are compared to the predicate device as shown in Table 1.

Table 1: Device General Comparison Table

CHARACTERISTICSSTANDARDSPREDICATE DEVICE (K192329)SUBJECT DEVICECOMPARISONEXPLANATION
Manufacturer(s)-JR Engineering & medical Technologies (M) sdn. Bhd.Comfort Rubber Gloves Industries Sdn. Bhd--
Product Code-LYYLYYSame-
Classification-Class IClass ISame-
Regulation Number-21 CFR 880.625021 CFR 880.6250Same-
Name of Device-Blue Latex Examination Powder Free GlovesPowder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein.--
Indication for Use-Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient.Same-
MaterialASTM D3578 – 19Natural LatexNatural Rubber LatexSame-
Color-BlueBlueSame-
Surface Appearance-Finger TexturedFinger TexturedSame-

Page 7

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein

CHARACTERISTICSSTANDARDSPREDICATE DEVICE (K192329)SUBJECT DEVICECOMPARISONEXPLANATION
Surface Treatment Powder/Powder Free-Powder FreePowder FreeSame-
SizeASTM D3578 – 19Small, Medium, Large, Extra Large, Extra Extra LargeSmall, Medium, Large, Extra Large, Extra Extra LargeSimilar-
Single useMedical Glove Guidance Manual – LabellingSingle UseSingle UseSame-
Sterile/Non-Sterile-Non-SterileNon-SterileSame-
Watertight Test (1000ml)21 CFR 800.20, ASTM D5151 – 19Pass AQL-2.5Pass AQL-2.5Same-
Dimension - LengthASTM D3578 – 19Length >230mmLength>230mmSimilarBoth the subject and predicate devices meet the same acceptance criteria.

Size Average Size Length:

  • Small: 304 vs Small: 285-290
  • Medium: 304 vs Medium: 285-294
  • Large: 305 vs Large: 286-294
  • X-Large: 305 vs X-Large: 286-290
  • XX-Large: 305 vs XX-Large: 296-299

| Dimension - Width | ASTM D3578 – 19 | Width 95±10 mm (for Medium Size) | Width 95±5 mm (for Medium Size) | Similar | Both the subject and predicate devices meet the same acceptance criteria. |

Size Average Size Width:

  • Small: 84 vs Small: 80-81
  • Medium: 94 vs Medium: 92-93
  • Large: 105 vs Large: 103-105
  • X-Large: 114 vs X-Large: 112-114
  • XX-Large: 123 vs XX-Large: 123-125

| Thickness | ASTM D3578 – 19 | Palm - >0.08mm, Finger- >0.08mm | Palm - >0.08mm, Finger- >0.08mm | Similar | Both the subject and predicate devices meet the same acceptance criteria. |

Size Palm (Actual Value) Finger (Actual Value):

  • Small: 0.31, 0.38 vs Small: 0.27-0.29, 0.33-0.39
  • Medium: 0.31, 0.38 vs Medium: 0.28-0.31, 0.34-0.40
  • Large: 0.31, 0.38 vs Large: 0.27-0.29, 0.35-0.38
  • X-Large: 0.31, 0.38 vs X-Large: 0.26-0.28, 0.36-0.39
  • XX-Large: 0.31, 0.38 vs XX-Large: 0.26-0.29, 0.33-0.35

Page 8

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 3 of 9

CHARACTERISTICSSTANDARDSPREDICATE DEVICE (K192329)SUBJECT DEVICECOMPARISONEXPLANATION
Physical Properties – Tensile StrengthASTM D3578 – 19Before Aging Tensile Strength>18MPaBefore Aging Tensile Strength>18MPaSimilarBoth the subject and predicate devices meet the same acceptance criteria.

Size Actual Value vs Size Tensile Strength (MPa):

  • Small: 33.0 vs Small: 26.20
  • Medium: 32.9 vs Medium: 30.04
  • Large: 32.2 vs Large: 29.91
  • X-Large: 31.9 vs X-Large: 26.15
  • XX-Large: 31.1 vs XX-Large: 28.93

After Aging Tensile Strength>14MPa:

  • Small: 30.0 vs Small: 26.77
  • Medium: 30.6 vs Medium: 24.89
  • Large: 29.9 vs Large: 25.71
  • X-Large: 29.7 vs X-Large: 23.43
  • XX-Large: 28.2 vs XX-Large: 25.86

| Physical Properties – Ultimate Elongation | ASTM D3578 – 19 | Before Aging Ultimate Elongation>650% | Before Aging Ultimate Elongation 650% Min | Similar | Both the subject and predicate devices meet the same acceptance criteria. |

Size Actual Value vs Size Elongation (%):

  • Small: 1322 vs Small: 850
  • Medium: 1250 vs Medium: 853
  • Large: 1392 vs Large: 888
  • X-Large: 1130 vs X-Large: 893
  • XX-Large: 1149 vs XX-Large: 833

After Aging Ultimate Elongation>500%:

  • Small: 1046 vs Small: 927
  • Medium: 1122 vs Medium: 751
  • Large: 1257 vs Large: 790
  • X-Large: 1011 vs X-Large: 845
  • XX-Large: 1110 vs XX-Large: 841

| Powder Residual Content | ASTM D6124-06 (2022) | ≤2.0mg/glove | ≤2.0mg/glove | Similar | Both the subject and predicate devices meet the same acceptance criteria. |

Size Residual Powder Content (mg/glove):

  • Small: 0.20 vs Small: 0.24
  • Medium: 0.21 vs Medium: 0.38
  • Large: 0.22 vs Large: 0.40
  • X-Large: 0.22 vs X-Large: 0.54
  • XX-Large: 0.23 vs XX-Large: 0.60

Page 9

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 4 of 9

CHARACTERISTICSSTANDARDSPREDICATE DEVICE (K192329)SUBJECT DEVICECOMPARISONEXPLANATION
Protein contentASTM D5712 – 15 (2020)Max 200µg/dm2<50µg/dm2SimilarThe subject device has a lower acceptance criterion to support a low protein claim.

Size Protein Content (µg/dm2):

  • Small: 45 vs Small: 43.65
  • Medium: 44 vs Medium: 40.05
  • Large: 44 vs Large: 44.71
  • X-Large: 43 vs X-Large: 42.51
  • XX-Large: 45 vs XX-Large: 48.33

| Biocompatibility | Primary Skin Irritation (Predicate Device) ISO 10993-10:2010(E) / Animal Irritation Test (Subject Device) ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation | Under the conditions of study not an irritant | Under the conditions of the study, not an irritant | Same | - |
| Dermal Sensitization Assay | ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization | Under the conditions of the study, not a sensitizer | Under the condition of this study, not a sensitizer. | Same | - |

Page 10

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 5 of 9

CHARACTERISTICSSTANDARDSPREDICATE DEVICE (K192329)SUBJECT DEVICECOMPARISONEXPLANATION
BiocompatibilityIn Vitro Cytotoxicity ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicityUnder the conditions of the study, cytotoxic which is to be expected as latex is the positive control for this test.Under the conditions of the study, exhibited cytotoxicity reactivity from 100.0% to 6.25% and no cytotoxicity reactivity at 3.125% extract concentrations.Same-
Acute Systemic ToxicityISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicityNo data providedUnder the condition of this study, the single dose acute systemic toxicity of extracts from test material using both normal saline and cottonseed oil, did not demonstrate any adverse toxic reaction.DifferentAcute systemic toxicity testing was conducted to address the in vitro cytotoxicity
Shelf LifeMedical Glove Guidance ManualNo data provided3 YearsDifferentDoes not impact substantial equivalence because the shelf life is determined by specific stability testing conducted under relevant conditions
Type of Use21 CFR Part 801 Subpart C – Labeling Requirements for Over-the-Counter DevicesOver the Counter (OTC)Over the Counter (OTC)Same-

Page 11

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 6 of 9

7.0 Summary of Non-Clinical Performance Testing

Test MethodPurposeAcceptance CriteriaResult
ASTM D3578-19 Standard Specification for Rubber Examination GlovesTo determine the length of the glovesLength: S: >220mm M-XXL:>230mmSize Length: Small 285-290, Medium 285-294, Large 286-294, X-Large 286-290, XX-Large 296-299
ASTM D3578-19 Standard Specification for Rubber Examination GlovesTo determine the width of the glovesWidth: S: 80±10mm M: 95±10mm L: 111±10mm XL: 120±10mm XXL: 130±10mmSize Width: Small 80-81, Medium 92-93, Large 103-105, X-Large 112-114, XX-Large 123-125
ASTM D3578-19 Standard Specification for Rubber Examination GlovesTo determine the thickness of the glovesPalm - >0.08mm Finger- >0.08mmSize Palm(mm) Finger(mm): Small 0.27-0.29 0.33-0.39, Medium 0.28-0.31 0.34-0.40, Large 0.27-0.29 0.35-0.38, X-Large 0.26-0.28 0.36-0.39, XX-Large 0.26-0.29 0.33-0.35
ASTM D3578-19 Standard Specification for Rubber Examination GlovesTo determine the physical properties – Tensile strengthBefore Aging Tensile Strength>18MPa After Aging Tensile Strength>14MPaSize Before Aging After Aging: Small 26.20 26.77, Medium 30.04 24.89, Large 29.91 25.71, X-Large 26.15 23.43, XX-Large 28.93 25.86
To determine the physical properties – Ultimate elongationBefore Aging Ultimate Elongation>650% After Aging Ultimate Elongation >500%Size Before Aging After Aging: Small 850 927, Medium 853 751, Large 888 790, X-Large 893 845, XX-Large 833 841
ASTM D5151-19 Standard Test Method for Detection of Holes in Medical GlovesTo determine the holes in the glovesAQL 2.5S – Pass AQL 2.5, M – Pass AQL 2.5, L – Pass AQL 2.5, XL – Pass AQL 2.5, XXL – Pass AQL 2.5

Page 12

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 7 of 9

Test MethodPurposeAcceptance CriteriaResult
ASTM D6124-06(2022) Standard Test Method for Residual Powder on Medical GlovesTo determine the residual powder in the gloves≤2.0mg/gloveSize Residual Powder Content (mg/glove): Small 0.24, Medium 0.38, Large 0.40, X-Large 0.54, XX-Large 0.60
ASTM D5712-15(2020) Standard Test Method for Analysis of Aqueous Extractable Protein in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry MethodTo determine the extractable protein in the gloves<50µg/dm2Size Protein Content (µg/dm2): Small 43.65, Medium 40.05, Large 44.71, X-Large 42.51, XX-Large 48.33
ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritationAssessment of medical devices and their constituent materials with regard to their potential to produce irritationNon-irritatingUnder the conditions of the study, not an irritant.
ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitizationAssessment of medical devices and their constituent materials with regard to their potential to induce skin sensitizationNon-sensitizingUnder the condition of this study, not a sensitizer.
ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicityTo determine the biological response of mammalian cells in vitro using appropriate biological parametersNon-cytotoxicUnder the conditions of the study, exhibited cytotoxicity reactivity from 100.0% to 6.25% and no cytotoxicity reactivity at 3.125% extract concentrations
ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicityEvaluation of the potential for medical device materials to cause adverse systemic reactionsNo acute systemic toxicityUnder the condition of this study, the single dose acute systemic toxicity of extracts from test material using both normal saline and cottonseed oil, did not demonstrate any adverse toxic reaction.

Non-clinical tests were conducted to demonstrate that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

  • ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves

Page 13

Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein
Page 8 of 9

  • ASTM D6124-06(2022) Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D5712-15(2020) Standard Test Method for Analysis of Aqueous Extractable Protein in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
  • ASTM D3578-19 Standard Specification for Rubber Examination Gloves
  • ISO 2859-1 Sampling Procedures and Tables for Inspection by Attributes
  • ISO 10993-5 Biological evaluation of medical devices-Part 5 Tests for in vitro cytotoxicity
  • ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization
  • ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity
  • ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation

8.0 Clinical Performance Data

No clinical study is included in this submission.

9.0 Conclusion

The conclusion drawn from the nonclinical test demonstrate that the proposed device Powder Free Natural Rubber Latex Examination Gloves – Blue, Non-Sterile and Protein Claim Of 50 Micrograms per dm2 Or Less Protein is as safe, as effective, and performs as well as or better than the legally marketed predicate device, (K192329: Blue Latex Examination Powder Free Gloves).

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.