K Number
K250002

Validate with FDA (Live)

Date Cleared
2025-04-24

(112 days)

Product Code
Regulation Number
868.1720
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Smart Check O2 is a tool used to measure oxygen purity, flow and pressure at the outlet of an oxygen concentrator. The Smart Check is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.

Device Description

The Smart Check O2 is an ultrasonic oxygen analyzer, used to verify the performance of oxygen concentrators. The device is typically used by durable medical equipment technicians while servicing concentrators in their workshops or while visiting patient homes, but it is not left with the patient and is not patient contacting.

The device measures the performance of the oxygen concentrator by making ultrasonic time-of-flight measurements, both upstream and downstream. Gas flow rate and oxygen concentration are determined using the resulting data. Temperature and pressure sensors inside the sample cell allow for accurate flow and oxygen readings over the range of specified operating environment conditions.

Oxygen and flow readings are shown to the user on the display. The Smart Check O2 can be toggled into a mode for testing pulsing (conserving) oxygen concentrators wherein it displays oxygen concentration and pulse volume instead of flow rate. The user may initiate a pressure check mode by stopping the sample exhaust port with their finger during which the Smart Check O2 measures and displays the maximum pressure generated by the concentrator. The user may also enter a calibration check mode and deliver pure oxygen to the Smart Check O2 to verify its performance. However, it does not require regular calibration after manufacturing.

The Smart Check O2 is for prescription use only.

Key Components:

  • Handheld analyzer unit
  • Removable battery door
  • Replaceable sample tube
  • Two Alkaline AA cells
AI/ML Overview

The provided FDA 510(k) Clearance Letter for the Smart Check O2 device does not contain the specific details of a study proving the device meets acceptance criteria, nor does it present acceptance criteria in a structured table format with reported performance. The document focuses on demonstrating substantial equivalence to a predicate device, and the "Non-Clinical Performance Data" section primarily lists compliance with consensus standards, rather than presenting test results against performance specifications.

Therefore, for the purpose of answering your request, I will extract relevant performance specifications from the "Substantial Equivalence Comparison Table" and present them as "Acceptance Criteria" and "Reported Device Performance." Please note that these are derived from the comparison table and are not explicitly stated as "acceptance criteria" within the provided text.

Here's an attempt to reconstruct the requested information based on the available text:


1. Table of Acceptance Criteria and Reported Device Performance

As the document does not explicitly present acceptance criteria or detailed study results in a table, the following table is constructed from the performance specifications listed in the "Substantial Equivalence Comparison Table" where the Subject Device (K250002) performance characteristics are taken as the "Reported Device Performance" and are implicitly assumed to have met an internal "Acceptance Criteria" derived from either design requirements or predicate device performance.

Performance MetricAcceptance Criteria (Derived from Subject Device Spec)Reported Device Performance (Subject Device K250002)
Oxygen Measurement Range (concentrator)20.9 - 96%20.9 - 96%
Oxygen Measurement Accuracy+/- 1.5%+/- 1.5%
Oxygen Measurement Resolution0.1%0.1%
Flow Measurement Range0-10 LPM0-10 LPM
Flow Measurement Accuracy+/- 0.2 LPM+/- 0.2 LPM
Flow Measurement Resolution0.1 LPM0.1 LPM
Pressure Measurement Range0 - 40 PSI0 - 40 PSI
Pressure Measurement Accuracy+/- 0.5% of reading+/- 0.5% of reading
Pressure Measurement Resolution0.1 PSI0.1 PSI
Pulse Volume Measurement Range3 - 200 ml3 - 200 ml
Pulse Volume Measurement Accuracy+/- 3 ml+/- 3 ml
Pulse Volume Measurement Resolution0.1 ml up to 100 ml, 1 ml above 100 ml0.1 ml up to 100 ml, 1 ml above 100 ml
Response Time≤ 10 seconds (implicitly improved from predicate)10 seconds
Start-up (Warm-up) Time≤ 2.5 seconds (implicitly acceptable despite predicate being faster)2.5 seconds
Operating Temperature5˚C - 40˚C5˚C - 40˚C
Storage Temperature-25˚C - 70˚C-25˚C - 70˚C
Atmospheric Pressure700 – 1060 hPa (mbar)700 – 1060 hPa (mbar)
Humidity0 - 90% (non-condensing)0 - 90% (non-condensing)
Battery Life> 16,000 read cycles> 16,000 read cycles
Electromagnetic Emissions ComplianceGroup 1, Class BGroup 1, Class B
Ingress RatingIP22IP22

The provided document describes a 510(k) submission, which primarily focuses on demonstrating "substantial equivalence" to a predicate device. It does not describe a clinical study involving human subjects or interpretation of medical images. Instead, it details the technical performance of a medical device (an oxygen gas analyzer) used for servicing other medical equipment (oxygen concentrators).

Therefore, many of the requested items related to clinical studies, expert consensus, and human reader performance are not applicable to this type of device submission as described in the provided text.

Based on the information given, here's what can be inferred for the other requested points:

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document refers to "non-clinical tests" and demonstrates compliance with standards. Testing would involve a number of units of the Smart Check O2 device under various conditions.
  • Data Provenance: Not explicitly stated. Given it's a device performance test, the data would likely be generated in a lab setting by the manufacturer, Life Spark Medical. It would be prospective data collected during device testing and verification.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • Not applicable. This device measures physical parameters (oxygen purity, flow, pressure, pulse volume). The 'ground truth' is established by calibrated reference standards and test equipment, not by human experts interpreting clinical data.

4. Adjudication Method for the Test Set

  • Not applicable. Since the ground truth for device performance is based on measurements against calibrated physical standards, there is no human interpretation or adjudication process as would be seen in, for example, a radiology study.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size of AI vs. Human Assistance

  • Not applicable. This is not an AI-powered diagnostic device that assists human readers. It's a measurement tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The performance metrics (accuracy, range, resolution, response time, etc.) are characteristics of the device itself (its internal sensors and algorithms) without human intervention in the measurement process, beyond initiating the test. The "Substantial Equivalence Comparison Table" lists these standalone performance characteristics.

7. The Type of Ground Truth Used

  • The ground truth for the performance metrics (oxygen purity, flow, pressure, pulse volume) would be established by calibrated reference standards and test equipment (e.g., precise gas mixtures for oxygen purity, calibrated flow meters, and pressure gauges).

8. The Sample Size for the Training Set

  • Not applicable/Not stated. This document describes a physical measurement device, not a machine learning or AI model. Therefore, there is no "training set" in the context of data for model training. The device's underlying technology is ultrasonic time-of-flight measurements, which is a physics-based method, not a data-driven learning algorithm.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As a non-AI/ML device, the concept of a training set and its ground truth is not relevant here. The device's accuracy relies on its design, component quality, and calibration procedures.

FDA 510(k) Clearance Letter - Smart Check O2

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

April 24, 2025

Life Spark Medical, LLC
Darryl Zitting
Biomedical Engineer
2430 W 350 N
Washington, Utah 84737

Re: K250002
Trade/Device Name: Smart Check O2 (MA0236)
Regulation Number: 21 CFR 868.1720
Regulation Name: Oxygen Gas Analyzer
Regulatory Class: Class II
Product Code: CCL
Dated: December 30, 2024
Received: January 2, 2025

Dear Darryl Zitting:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K250002 - Darryl Zitting Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

Page 3

K250002 - Darryl Zitting Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K250002

Device Name: Smart Check O2 (MA0236)

Indications for Use (Describe):

The Smart Check O2 is a tool used to measure oxygen purity, flow and pressure at the outlet of an oxygen concentrator. The Smart Check is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary – K250002

Life Spark Medical
2430 W 350 N, Hurricane, UT 84737
Tel (435) 254-3010
E-mail – info@lifesparkmed.com

Contact Person: Darryl Zitting

Date Summary Prepared: Mar-27-2025

Subject Device Trade Name: Smart Check O2
Common Name: Oxygen Gas Analyzer

Classification: Analyzer, Gas, Oxygen, Gaseous-Phase (21 CFR 868.1720, Product Code CCL, Class 2)

Predicate Device: Maxtec Ultramax Oxygen Analyzer, 510(k) # K112402

Subject Device Description

The Smart Check O2 is an ultrasonic oxygen analyzer, used to verify the performance of oxygen concentrators. The device is typically used by durable medical equipment technicians while servicing concentrators in their workshops or while visiting patient homes, but it is not left with the patient and is not patient contacting.

The device measures the performance of the oxygen concentrator by making ultrasonic time-offlight measurements, both upstream and downstream. Gas flow rate and oxygen concentration are determined using the resulting data. Temperature and pressure sensors inside the sample cell allow for accurate flow and oxygen readings over the range of specified operating environment conditions.

Oxygen and flow readings are shown to the user on the display. The Smart Check O2 can be toggled into a mode for testing pulsing (conserving) oxygen concentrators wherein it displays oxygen concentration and pulse volume instead of flow rate. The user may initiate a pressure check mode by stopping the sample exhaust port with their finger during which the Smart Check O2 measures and displays the maximum pressure generated by the concentrator. The user may also enter a calibration check mode and deliver pure oxygen to the Smart Check O2 to verify its performance. However, it does not require regular calibration after manufacturing.

The Smart Check O2 is for prescription use only.

Key Components:

  • Handheld analyzer unit
  • Removable battery door
  • Replaceable sample tube
  • Two Alkaline AA cells

Page 6

Key Performance Characteristics

All key performance characteristics are listed in the subsequent Substantial Equivalence Comparison Table.

Indication for Use Statement

The Smart Check O2 is a tool used to measure oxygen purity, flow and pressure at the outlet of an oxygen concentrator. The Smart Check is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.

Technological Characteristics

The technological characteristics of the Smart Check O2 are substantially equivalent to the predicate device including:

  • Physical Form: Handheld
  • Energy Source: Battery operated
  • Method of operation: Ultrasonic, upstream/downstream time-of-flight measurement
  • Control Mechanism: Firmware controlled microprocessor
  • Material composition: Piezo-electric ultrasonic transducers, injection molded structural components, and common PCBA materials / integrated circuit components.

Non-Clinical Performance Data

The Smart Check O2 complies with the following consensus standards:

ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021]
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]

ANSI AAMI HA60601-1-11:2015 [Including AMD1:2021]
Medical Electrical Equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment (IEC 60601-1-11:2015 MOD) [Including Amendment1 (2021)]

ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021]
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]

ISO 80601-2-55 Second edition 2018-02 [Including AMD1:2023]
Medical electrical equipment - Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitor [Including Amendment 1 (2023)]

Page 7

Substantial Equivalence Comparison Table

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Model:Smart Check O2UltraMaxO2SameNo
Manufacturer:Life Spark MedicalMaxtecSameNo
Product Code:CCL (Class 2) - Analyzer, Gas, Oxygen, Gaseous-PhaseCCL (Class 2) - Analyzer, Gas, Oxygen, Gaseous-PhaseSameNo
Patient Population:Not for use on patients, only oxygen concentrators.Not for use on patients, only oxygen concentrators.SameNo
Indications for Use:The Smart Check O2 is a tool used to measure oxygen purity, flow, and pressure at the outlet of an oxygen concentrator. The Smart Check is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.The UltramaxO2 Oxygen Analyzer is a tool used to measure oxygen purity, flow, and pressure of an oxygen concentrator. The UltraMaxO2 Oxygen Analyzer is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.SameNo
User Interface:Input: Tactile dome switches (buttons) Output: Segmented LCD with backlightInput: Tactile dome switches (buttons) Output: Segmented LCD without backlightSimilarNo: Improved visibility in low light without adding risks.
Use Environment:Hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.Hospitals, nursing homes, extended care facilities, patient homes, and respiratory device service and repair centers.SameNo

Page 8

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Concentrator Compatibility:Continuous flow concentrators and pulsing (conserving) concentrators.Continuous flow concentrators only.New feature using same technology.No: See following Substantial Equivalence Discussion.
Oxygen Measurement Range: (concentrator)20.9 - 96%20.9 - 96%SameNo
Oxygen Measurement Accuracy:+/- 1.5%+/-1.5%SameNo
Oxygen Measurement Resolution:0.1%0.1%SameNo
Flow Measurement Range:0-10LPM0-10LPMSameNo
Flow Measurement Accuracy:+/-0.2 LPM+/-0.2 LPMSameNo
Flow Measurement Resolution:0.1 LPM0.1 LPMSameNo

Page 9

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Pressure Measurement Range:0 - 40 PSI0.5 - 50 PSISimilarNo: The highest pressure generated by concentrators, currently on the market, is 35 psi. Future concentrators offering higher supply pressures are not anticipated due to patient safety considerations.
Pressure Measurement Accuracy:+/-0.5% of reading+/-0.5% of readingSameNo
Pressure Measurement Resolution:0.1 PSI0.1 PSISameNo
Pulse Volume Measurement Range:3 - 200 mlDoes not measure pulse volumeNew feature using same technology.No: See Substantial Equivalence Discussion.
Pulse Volume Measurement Accuracy:+/- 3 mlDoes not measure pulse volumeNew feature using same technology.No: See Substantial Equivalence Discussion.

Page 10

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Pulse Volume Measurement Resolution:0.1 ml up to 100 ml, 1 ml above 100 mlDoes not measure pulse volumeNew feature using same technology.No: See Substantial Equivalence Discussion.
Response Time:10 seconds≤17 secondsSimilarNo: Subject device response time is improved compared to predicate.
Start-up (Warm-up) Time:2.5 seconds< 1 secondSimilarNo: No new risks associated with adding 1.5 seconds to the analyzer boot-up time.
Operating Temperature:5˚C - 40˚C15˚C - 40˚CIncremental ImprovementNo: Subject device operating temp. range is wider, and compliant with the requirements of ANSI/AAMI HA60601-1-11 subclause 4.2.3.1.

Page 11

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Storage Temperature:-25˚C - 70˚C-15˚C - 60˚CIncremental ImprovementNo: Subject device storage temp. range is wider, and compliant with the requirements of ANSI/AAMI HA60601-1-11 subclause 4.2.2
Atmospheric Pressure:700 – 1060 hPa (mbar)800 - 1000 mbar (hPa)SimilarNo: Subject device operating pressure range is wider, and compliant with the requirements of ANSI/AAMI HA60601-1-11 subclause 4.2.3.1.
Humidity:0 - 90% (non-condensing)0 - 95% (non-condensing)SimilarNo: Subject device operating humidity range is compliant with the requirements of ANSI/AAMI HA60601-1-11 subclause 4.2.3.1.
Power Source:2x AA Alkaline Batteries2x AA Alkaline BatteriesSameNo
Battery Life:> 16,000 read cycles16,500 read cyclesSameNo

Page 12

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Dimensions:75 mm x 147 mm x 27 mm80.3 mm x 129.5 mm x 26.4 mmSimilarNo: Sizes of predicate and subject devices are very similar and both suitable for the use environment. No new risks associated with the minor differences in size.
Weight:230 g181 gSimilarNo: Weight of predicate and subject devices are similar and both suitable for the user population. No new risks associated with the minor differences in weight.

Page 13

Substantial Equivalence Comparison TableSubject Device (K250002)Predicate device (K112402)ComparisonDoes the difference raise new questions of safety and effectiveness?
Electromagnetic Emissions compliance:Group 1, Class BGroup 1, Class ASimilarNo: Subject device was tested to the more challenging CISPR 11 Class B requirements for residential environments per ANSI/AAMI HA60601-1-11 clause 12.
Ingress rating:IP22IPX1Incremental ImprovementNo: Subject device complies with the more stringent ingress rating requirements for residential environments per ANSI/AAMI HA60601-1-11 subclause 8.3.1.

Page 14

Substantial Equivalence Discussion and Conclusion

The Smart Check O2 is substantially equivalent in environment, user population, application and performance when compared to the predicate device. Minor differences, as noted by "similar" or "incremental improvement" in the comparison column of the table, exist primarily as a result of direct compliance with the recommendations of the homecare standard ANSI/AAMI HA60601-1-11. All noted differences are minor and do not result in new or increased risks.

Pulse Mode:

In addition to displaying % oxygen and continuous flow rate, the subject device can calculate and display pulse volume, for pulsing ("conserving") concentrators, using multiple flow reading measurements. This feature is not significantly different than measuring continuous flow rate because it uses the exact same measurement methods and technology to make repeated flow measurements, then calculates the additional data (pulse volume) by integrating flow rate over the time period of the pulse. This difference does not introduce new or different risks compared to the predicate.

The conclusions drawn from the non-clinical tests (discussed above) demonstrate that the device is substantially equivalent to the predicate device.

§ 868.1720 Oxygen gas analyzer.

(a)
Identification. An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.(b)
Classification. Class II (performance standards).