K Number
K244000

Validate with FDA (Live)

Date Cleared
2025-04-30

(125 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

Device Description

The Arm Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.

The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low.

AI/ML Overview

The FDA 510(k) clearance letter for the Arm Blood Pressure Monitor does not explicitly state acceptance criteria in a quantitative table or the specific performance metrics achieved in a typical "device performance" section as one might expect for a software-based or diagnostic device. Instead, it refers to compliance with established medical device standards, particularly ISO 81060-2:2018 (and its amendment from 2020) for clinical accuracy.

However, based on the context of the clinical data and the requirements of ISO 81060-2, we can infer the acceptance criteria for accuracy.

Here's a breakdown of the requested information based on the provided document:


Implied Acceptance Criteria and Reported Device Performance

The device's performance is demonstrated by its adherence to the clinical accuracy requirements of ISO 81060-2 Third edition 2018-11 (including AMD1:2020).

Implied Acceptance Criteria of ISO 81060-2:2018 (and AMD1:2020):

The ISO 81060-2 standard consists of two parts for accuracy assessment:

  • Criterion 1 (Mean Difference and Standard Deviation):
    • The mean difference between the device and the reference measurement for both systolic and diastolic blood pressure shall be $\leq \pm 5$ mmHg.
    • The standard deviation of these differences shall be $\leq 8$ mmHg.
  • Criterion 2 (Cumulative Percentage of Differences):
    • For each subject, the absolute difference between the device and the reference measurement should be calculated.
    • The cumulative percentage of devices should have absolute differences within certain thresholds:
      • $\leq 5$ mmHg for at least 65% of measurements.
      • $\leq 10$ mmHg for at least 85% of measurements.
      • $\leq 15$ mmHg for at least 95% of measurements.

Table of Acceptance Criteria and Reported Device Performance (Inferred from ISO 81060-2)

Acceptance Criterion (from ISO 81060-2)Systolic Blood Pressure (SYS)Diastolic Blood Pressure (DIA)Reported Device Performance (Overall)
Criterion 1 (Mean $\pm$ SD)
Mean difference $\leq \pm 5$ mmHg$\leq \pm 5$ mmHg$\leq \pm 5$ mmHg"not over the limits of ISO 81060-2"
Standard deviation $\leq 8$ mmHg$\leq 8$ mmHg$\leq 8$ mmHg"not over the limits of ISO 81060-2"
Criterion 2 (Cumulative Differences)
Differences $\leq 5$ mmHg for $\geq 65$%$\geq 65$%$\geq 65$%"not over the limits of ISO 81060-2"
Differences $\leq 10$ mmHg for $\geq 85$%$\geq 85$%$\geq 85$%"not over the limits of ISO 81060-2"
Differences $\leq 15$ mmHg for $\geq 95$%$\geq 95$%$\geq 95$%"not over the limits of ISO 81060-2"

Note: The document explicitly states, "All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2 Third edition 2018-11 [Including AMD1:2020]." This confirms the device met the numerical precision requirements of the standard for mean difference and standard deviation. While it doesn't provide the exact numbers for each group, it confirms compliance with the key statistical thresholds.


Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Total Sample Size: The 28 models were grouped into 7 categories for clinical accuracy testing. Each of these 7 groups had a test set of 100 subjects, except for Group 7, which had 92 subjects.
      • Group 1: 100 subjects
      • Group 2: 100 subjects
      • Group 3: 100 subjects
      • Group 4: 100 subjects
      • Group 5: 100 subjects
      • Group 6: 100 subjects
      • Group 7: 92 subjects
    • Male/Female Distribution: Each group had a participant distribution of at least 30% male and at least 30% female.
    • Age Range: Subjects were "adult person" with an age range of "> 12 years old," specifically noted ranges from 15 to 80 years old across the groups.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical data. It also does not explicitly state if the study was retrospective or prospective, but clinical accuracy studies adhering to ISO 81060-2 are typically prospective and involve simultaneous, sequential measurements.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • The document does not specify the number or qualifications of experts (e.g., radiologists) for establishing ground truth in this context. For blood pressure monitors, "ground truth" is typically established by trained technicians or clinicians using a validated reference method (e.g., mercury sphygmomanometer) following a rigorous protocol (like the auscultatory method per ISO 81060-2). The standard requires independent observers for reference measurements.
  3. Adjudication Method for the Test Set:

    • The document states, "The clinical accuracy test report and data analysis followed the requirements of the ISO 81060-2 Third edition 2018-11 [Including AMD1:2020]. The Same Arm Sequential Method was chosen for all studies."
    • ISO 81060-2 outlines a specific method for ground truth establishment for blood pressure. It typically involves multiple independent observers (at least two trained observers) who simultaneously or sequentially measure blood pressure using a validated reference device. Discrepancies between these observers are resolved in accordance with the standard, which usually involves specific rules for averaging or discarding measurements based on predefined variability criteria, rather than a formal "adjudication panel" in the sense of image interpretation.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant for diagnostic imaging AI devices where human readers interpret medical images with and without AI assistance. For a non-invasive blood pressure monitor, the study focus is on the device's measurement accuracy against a reference standard, not on how it assists human interpretation.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, in essence, a "standalone" or "algorithm only" performance was evaluated. The device (Arm Blood Pressure Monitor) is an automated system that measures blood pressure. Its clinical accuracy study measures its direct output (systolic and diastolic pressure) against a reference standard. There isn't a "human-in-the-loop" component in the operational measurement process that would affect its core accuracy, as it's an automated device.
  6. The Type of Ground Truth Used:

    • The ground truth was established by comparison with a validated reference measurement method, as prescribed by ISO 81060-2. This typically involves the auscultatory method using a mercury sphygmomanometer or an equivalent validated reference device, with measurements performed by trained observers.
  7. The Sample Size for the Training Set:

    • Not applicable / Not provided. This device is a hardware-based blood pressure monitor that uses an oscillometric technique and a "blood pressure core algorithm." It is not described as a machine learning or AI algorithm that requires a separate "training set" in the conventional sense of deep learning. Its performance is based on its engineering design and the embedded validated algorithm.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable / Not provided. As mentioned above, there isn't a "training set" in the common understanding for a typical medical AI device. The core algorithm is likely developed and validated internally by the manufacturer through engineering principles and extensive testing, not necessarily through a separate clinical "training set" with ground truth in the AI context. The clinical data presented is for validation (test set) of the final device, demonstrating its accuracy.

FDA 510(k) Clearance Letter - Arm Blood Pressure Monitor

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 30, 2025

Shenzhen AOJ Medical Technology Co., LTD
Jack Wang
Deputy Chief
Room 301&4F, Block A, Bldg A,ingfa Intelligent Manufacturing
Park Xiaweiyuan,Gushu Community, Xixiang Str Bao'an District
Shenzhen, Guangdong 518126
China

Re: K244000

Trade/Device Name: Arm Blood Pressure Monitor (Models ARM-30A+, ARM-30G2, ARM-30Q, ARM-30E+, ARM-3010, ARM-3011, ARM-3012, ARM-3013, ARM-3110, ARM-30G+, ARM-30M, ARM-30S, ARM-30T, ARM-HA101, ARM-3040, ARM-30V, AMR-30W, SBM 15, BU 518, ARM-30U, ARM-3030, ARM-30Y, BU 520, ARM-3120, ARM-3020, ARM-3021, ARM-3022 and ARM-3023)

Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive Blood Pressure Measurement System
Regulatory Class: Class II
Product Code: DXN
Dated: March 28, 2025
Received: March 28, 2025

Dear Jack Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

Page 2

K244000 - Jack Wang Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-

Page 3

K244000 - Jack Wang Page 3

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S

LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.

510(k) Number (if known): K244000

Device Name: Arm Blood Pressure Monitor (Models ARM-30A+, ARM-30G2, ARM-30Q, ARM-30E+, ARM-3010, ARM-3011, ARM-3012, ARM-3013, ARM-3110, ARM-30G+, ARM-30M, ARM-30S, ARM-30T, ARM-HA101, ARM-3040, ARM-30V, AMR-30W, SBM 15, BU 518, ARM-30U, ARM-3030, ARM-30Y, BU 520, ARM-3120, ARM-3020, ARM-3021, ARM-3022 and ARM-3023)

Indications for Use (Describe):
The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

Type of Use (Select one or both, as applicable):

  • ☐ Prescription Use (Part 21 CFR 801 Subpart D)
  • ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20) Page 1 of 1

Page 5

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2025/02/26

1. Submission sponsor

Name: Shenzhen AOJ Medical Technology Co., Ltd.
Address: Room 301&4F, Block A, Building A, Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, 518126 Shenzhen, PEOPLE'S REPUBLIC OF CHINA
Contact person: Jack Wang
Title: Deputy Chief
TEL: 86 755-27786026

2. Subject Device Information

FieldValue
Trade/Device NameArm Blood Pressure Monitor
Modelmodels ARM-30A+,ARM-30G2,ARM-30Q,ARM-30E+, ARM-3010,ARM-3011,ARM-3012,ARM-3013,ARM-3110, ARM-30G+,ARM-30M,ARM-30S,ARM-30T,ARM-HA101, ARM-3040,ARM-30V,AMR-30W,SBM 15,BU 518,ARM-30U, ARM-3030,ARM-30Y,BU 520,ARM-3120,ARM-3020,ARM-3021, ARM-3022 and ARM-3023
Common NameAutomatic Blood Pressure Monitor
Regulatory ClassClass II
Product CodeDXN
Submission typeTraditional 510(K)

3. Predicate Device

Manufacturer: Shenzhen AOJ Medical Technology Co., Ltd.
Device name: Arm Blood Pressure Monitor, models AOJ-30A, AOJ-30B, AOJ-30C, AOJ-30D, AOJ-30E, AOJ-30F and AOJ-30G.
510(K) Number: K222125

4. Device Description

The Arm Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.

The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low.

K244000

Page 6

The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients.

The product is provided non-sterile, and not to be sterilized by the user prior to use.

All the models included in this submission follow the the same intended use, same measurement principle, same blood pressure core algorithm and similar product design. The main differences are appearance, Dimensions and some function which will not affect the safety and effectiveness of the device.

5. Intended use & Indication for use

The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

6. Comparison to the Predicate Device

ITEMProposed DevicePredicate DeviceComparison Result
Arm Blood Pressure Monitor ARM-30A+,ARM-30G2,ARM-30Q, ARM-30E+,ARM-3010,ARM-3011, ARM-3012,ARM-3013,ARM-3110, ARM-30G+,ARM-30M,ARM-30S, ARM-30T,ARM-HA101,ARM-3040, ARM-30V,AMR-30W,SBM 15, BU 518,ARM-30U,ARM-3030, ARM-30Y,BU 520,ARM-3120, ARM-3020,ARM-3021,ARM-3022, ARM-3023 / K244000Arm Blood Pressure Monitor AOJ-30A, AOJ-30B, AOJ-30C, AOJ-30D, AOJ-30E, AOJ-30F and AOJ-30G / K222125
ManufacturerShenzhen AOJ Medical Technology Co., Ltd.Shenzhen AOJ Medical Technology Co., Ltd.Same
Intended Use/Indications for UseThe Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.The Arm Blood Pressure Monitor is intended to measure the systolic and diastolic blood pressure as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.Same
ContraindicationsNo known contraindication existingNot KnownSame
Application scenarioMedical Facilities and Home UseMedical Facilities and Home UseSame

Operational Specifications

ParameterProposed DevicePredicate DeviceComparison Result
PrincipleOscillometricOscillometricSame
Measurement ItemSYS, DYS, Pulse RateSYS, DYS, Pulse RateSame
Patient populationAdultAdultSame
Measurement siteUpper armUpper armSame
Accuracy± 3 mmHg± 3 mmHgSame

K244000

Page 7

ParameterProposed DevicePredicate DeviceComparison Result
Heart rate measurement range40-199 bpm40-199 bpmSame
Accuracy± 5%± 5%Same
Cuff size22 - 42 cm22 - 42 cmSame
DisplayBlood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record NumberBlood Pressure (Systolic and Diastolic), Pulse rate, Date, Time, WHO BP Classification Indicating Bar, Low Battery Icon, Heart Icon, Memory Record NumberSame
Auto shutdownYESYESSame
Operating environmentTemperature: 5°C~40°C Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPaTemperature: 5°C~40°C Humidity: 15%-90% RH, Atmospheric pressure: 70 kPa -106 kPaSame
Storage environmentAmbient Temperature: -20°C to 55°C Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPaAmbient Temperature: -20°C to 55°C Relative Humidity: 10-93% RH, Atmospheric pressure: 70 kPa -106 kPaSame
Battery type1) 6Vdc (41.5V AAA batteries): ARM-30A+,ARM-30G2, ARM-30Q,ARM-30E+,ARM-3010, ARM-3011,ARM-3012,ARM-3013, ARM-3110,ARM-30G+,ARM-30M, ARM-30S,ARM-30T,ARM-HA101, ARM-3040,ARM-30V,AMR-30W, SBM 15,ARM-30U,ARM-3030, ARM-30Y,ARM-3120,ARM-3020, ARM-3021,ARM-3022,ARM-3023 2) 6Vdc (41.5V AA batteries): BU 518,BU 5206Vdc (4*1.5VAAA batteries)Different¹
WeightSBM 15: About 175g ARM-30V / ARM-30W: About 209g ARM-30Q: About 193g ARM-30A+: About 316g ARM-30G2: About 263g ARM-30G+: About 234g ARM-30M: About 280g ARM-HA101: About 290g ARM-3040: About 235g ARM-30Y:About 265g ARM-3030:About 266g BU 518: About 270g BU 520: About 286g ARM-30S / ARM-30T / ARM-30U: About 221gAOJ-30A /AOJ-30B /AOJ-30G: (262 ±5)g without battery AOJ-30C: (316 ±5)g without battery AOJ-30D: (349 ±5)g without battery AOJ-30E: 220 g without battery AOJ-30F: 220 g without batteryDifferent²

K244000

Page 8

ParameterProposed DevicePredicate DeviceComparison Result
ARM-30E+ / ARM-3010 / ARM-3011 / ARM-3012 / ARM-3013 / ARM-3110: About 220g ARM-3120/ARM-3020/ ARM-3021/ ARM-3022/ ARM-3023:About 230g
DimensionsARM-30A+ / ARM-3030: 108mm×139mm× 62mm ARM-30G2 / ARM-30G+: 130mm×100mm×60mm ARM-30Q: 100mm×100mm×59mm ARM-30M : 129mm×112mm×58.5mm ARM-HA101 : 130mm×112mm×58.3mm ARM-3040: 136mm×113mm×28mm ARM-30V / ARM-30W: 130mm×100mm×51mm SBM 15: 115.4mm×92mm×44.5mm ARM-30Y:140.8mm×110mm×61mm BU 518 / BU 520: 130mm×111.5mm×56mm ARM-30S / ARM-30T / ARM-30U: 130mm×100mm×49mm ARM-30E+ / ARM-3010 / ARM-3011 / ARM-3012 / ARM-3013 / ARM-3110: 118mm×98mm×62.5mm ARM-3120 /ARM-3020/ ARM-3021/ ARM-3022 / ARM-3023: 118mm×98mm×61mmAOJ-30A /AOJ-30B /AOJ-30G: 127mm×93mm×73mm AOJ-30C: 108mm×139mm× 62mm AOJ-30D:136mm×113mm×68mm AOJ-30E /AOJ-30F: 118mm×98mm×61mm
Patient ContactingSurface-contacting, Less than 24 hSurface-contacting, Less than 24 hSame
Biocompatibility evaluationCytotoxicity, skin sensitization and irritationCytotoxicity, skin sensitization and irritationSame
Electrical safetyIEC 60601-1 IEC 60601-1-11 ISO 80601-2-30IEC 60601-1 IEC 60601-1-11 ISO 80601-2-30Same
EMCIEC 60601-1-2IEC 60601-1-2Same
BiocompatibilityISO 10993-1 ISO 10993-5 ISO 10993-10 ISO 10993-23ISO 10993-1 ISO 10993-5 ISO 10993-10Different³
Data transmissionNot availableHas wireless function with Bluetooth and available for AOJ-30A and AOJ-30BDifferent⁴

Justification for the differences:

1) Different kind of power supply
The subject device powered by internal battery (AA or AAA) ,which both belong to alkaline battery. It will not affect the safety and effective of subject device. The test results have been verified by verification testing.

K244000

Page 9

2) Different Physical Specifications
The subject and the predicate device are of' different size but proximity. Moreover, such engineering design has been verified during the international standards, so such minor different ill not raise any safety and effectiveness questions.

3) Different Biocompatibility standard
The subject device has been tested 10993-5,10993-10,10993-23 according to the update of the standard.

4) Different data transmission function
The subject device has no data transmission function with Bluetooth.

7. Non-clinical Data

Biocompatibility testing

The biocompatibility evaluation for the Arm Blood Pressure Monitor were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices − Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:

  • Cytotoxicity
  • Skin Sensitization
  • Skin Irritation

Bench testing

The device has been tested according to the following standards:

  • IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
  • IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • IEC 80601-2-30: Medical electrical equipment – Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.
  • IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • FDA Guidance for Non-Automated Sphygmomanometer.

K244000

Page 10

8. Clinical data

According to similar clinical characteristics, 28 submitted models were divided into 7 groups for clinical accuracy research.

The clinical accuracy test report and data analysis followed the requirements of the ISO 81060-2 Third edition 2018-11 [Including AMD1:2020] Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type [Including: Amendment 1 (2020)]. The Same Arm Sequential Method was chosen for all studies.

Group No.Test modelNumber of Subjects (a minimum of 85 subjects)Male (at least 30 %)Female (at least 30 %)Age range (> 12 years old)
1ARM-3030, ARM-30Y100495117 to 78
2ARM-3120, ARM-3020, ARM-3021, ARM-3022, ARM-3023100544615 to 80
3ARM-30A+, ARM-30G2, ARM-30E+, ARM-30Q, ARM-3010, ARM-3011, ARM-3012, ARM-3013, ARM-3110100415919 to 78
4ARM-30M, ARM-30G+, ARM-HA101, ARM-30S, ARM-30T, ARM-30U100544617 to 71
5ARM-3040100534715 to 78
6BU 520, BU 518100495115 to 78
7SBM 15, ARM-30V, ARM-30W92474515 to 70

All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2 Third edition 2018-11 [Including AMD1:2020]. No adverse effect and/or complication is found in this study.

9. Conclusion

It is concluded form the non-clinical and clinical tests that demonstrate that the subject devices are as safe, as effective, and performs as well as the legally marketed predicate device identified above.

K244000

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).