K Number
K243998

Validate with FDA (Live)

Date Cleared
2025-04-25

(120 days)

Product Code
Regulation Number
876.5880
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DCX Disposable Cassette, as part of the Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.

Device Description

The DCX Disposable Cassette, provides a sterile fluid pathway, houses and protects the kidneys during perfusion and it has a gravity flow system which allows circulation of perfusate through the kidneys.

The kidneys are placed in an organ chamber filled with perfusate. The perfusate flows from the organ chamber to the pumphead. Then, it is pumped through the heat exchanger to the bubble trap where it is delivered to the cannulated kidney(s). The perfusate then returns to the organ chamber via the kidney vein to repeat the perfusion cycle.

The cassette can provide circulation of perfusate to one or two kidneys, attached individually or in bloc.

AI/ML Overview

The provided FDA 510(k) clearance letter and summary for the DCX Disposable Cassette (K243998) primarily focuses on demonstrating substantial equivalence to its predicate device through non-clinical testing. It does not detail a clinical study with acceptance criteria often associated with device performance in diagnostic or therapeutic efficacy. Instead, the acceptance criteria are met through various non-clinical tests verifying safety, sterility, biocompatibility, and functional performance.

Here's a breakdown of the requested information based on the provided document:


Acceptance Criteria and Reported Device Performance

The acceptance criteria for the DCX Disposable Cassette are primarily focused on safety, sterility, biocompatibility, and functional performance, aligning with regulatory standards and ensuring it operates as intended without posing new risks compared to its predicate.

Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Sterility
Sterility Assurance Level (SAL)10⁻⁶ (meets standard)
Sterilization MethodEthylene Oxide sterilization (validated according to ISO 11135 and ISO 10993-7)
Biocompatibility
CytotoxicityFailed to mention specific results but stated "DCX is safe for the intended biocontact" based on ISO 10993 compliant testing.
Skin SensitizationFailed to mention specific results but stated "DCX is safe for the intended biocontact" based on ISO 10993 compliant testing.
Primary Skin IrritationFailed to mention specific results but stated "DCX is safe for the intended biocontact" based on ISO 10993 compliant testing.
HemolysisFailed to mention specific results but stated "DCX is safe for the intended biocontact" based on ISO 10993 compliant testing.
Acute Systemic ToxicityFailed to mention specific results but stated "DCX is safe for the intended biocontact" based on ISO 10993 compliant testing.
Non-pyrogenicSupplied as non-pyrogenic
Shelf Life/Stability
Shelf Life24 months
Storage Conditions2°C to 25°C (with temporary 40°C outing tolerated)
Stability Testing ResultsAging of test articles at recommended storage conditions does not affect product specifications for 24 months.
Functional Performance
Compatibility with RM4 Control UnitPerformance tests with perfusion accessories and RM4 Control Unit showed the device "performs as expected and meets the requirements set for the device."
Design FeaturesConnection luer for perfusion accessories; oxygenation directly in organ chamber; removable partition for 1L perfusate for single organ perfusion; two lids. (These are design changes/features compared to the predicate, and their functional performance was implied to be satisfactory through comprehensive testing.)

Study Details

The provided document describes a non-clinical testing study rather than a clinical trial focused on diagnostic accuracy or therapeutic outcomes in patients.

  1. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify a "test set" in the context of patient data or samples. The testing described is primarily for device components and materials (e.g., biomaterial samples for ISO 10993 testing, samples for sterility testing, units for stability testing).
    • The provenance of the data is from Institut Georges Lopez (IGL), based in France, the manufacturer of the device.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not applicable as the described study is a non-clinical device performance and safety evaluation (e.g., sterilization validation, biocompatibility testing, stability testing), not a study requiring expert consensus on clinical findings.
  3. Adjudication Method for the Test Set:

    • This information is not applicable as the described study is a non-clinical device performance and safety evaluation. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies for interpretation of imaging or clinical outcomes where multiple experts assess the same case.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of an AI-assisted diagnostic tool on human reader performance, which is beyond the scope of this device (a disposable cassette for organ perfusion).
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • No, this refers to an algorithm's performance, which is not applicable to a disposable cassette.
  6. The Type of Ground Truth Used:

    • The "ground truth" for this device's evaluation is based on established regulatory standards and engineering specifications. For example:
      • Sterility: Achieved by verifying results against ISO 11135 and ISO 10993-7.
      • Biocompatibility: Achieved by verifying results against ISO 10993 (cytotoxicity, sensitization, irritation, hemolysis, acute systemic toxicity).
      • Stability: Achieved by verifying that product specifications remain within acceptable limits after aging tests for the specified shelf life.
      • Functional Performance: Achieved by demonstrating that the device "performs as expected and meets the requirements set for the device" when used with compatible components (RM4 Control Unit and perfusion accessories).
  7. The Sample Size for the Training Set:

    • This information is not applicable. The DCX Disposable Cassette is a physical medical device, not a machine learning algorithm that requires a training set.
  8. How the Ground Truth for the Training Set Was Established:

    • This information is not applicable for the same reason as point 7.

FDA 510(k) Clearance Letter - DCX Disposable Cassette

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 25, 2025

Institut Georges Lopez (IGL)
Matthieu Prouteau
Director of Quality & Regulatory Affairs
Parc Tertiaire du Bois Dieu
1 Allée des Chevreuils
Lissieu, 69380
France

Re: K243998
Trade/Device Name: DCX Disposable Cassette (DCX)
Regulation Number: 21 CFR§ 876.5880
Regulation Name: Isolated kidney perfusion and transport system and accessories
Regulatory Class: II
Product Code: KDN
Dated: March 27, 2025
Received: March 27, 2025

Dear Matthieu Prouteau:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243998 - Matthieu Prouteau Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

Page 3

K243998 - Matthieu Prouteau Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maura Rooney -S

Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K243998

Device Name
DCX Disposable Cassette (DCX)

Indications for Use (Describe)
The DCX Disposable Cassette, as part of the Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Traditional 510(k) Premarket Notification – DCX Disposable Cassette
Confidential Page 1 of 4

510(k) Summary

[As required by 21 CFR 807.92]

Written on December, 17 2024

I. Submitter

Institut Georges Lopez (IGL)
Parc Tertiaire du Bois Dieu
1 allée des Chevreuils
69380 Lissieu
France
Telephone: +33 437 646 332
Official Contact: Matthieu Prouteau – Director of Quality and Regulatory Affairs

II. Device Information

Trade Name: DCX Disposable Cassette (DCX)
Common Name: System, Perfusion, Kidney
Regulation Description: Isolated kidney perfusion and transport system and accessories
Regulation Number: 21 CFR 876.5880
Class: Class II
Product Code: KDN

III. Predicate Device Identification

K NumberPredicate Device NameManufacturer
Predicate K211224RM4 control unitInstitut Georges Lopez - IGL (France)

IV. Device Description

The DCX Disposable Cassette, provides a sterile fluid pathway, houses and protects the kidneys during perfusion and it has a gravity flow system which allows circulation of perfusate through the kidneys.

The kidneys are placed in an organ chamber filled with perfusate. The perfusate flows from the organ chamber to the pumphead. Then, it is pumped through the heat exchanger to the bubble trap where it is delivered to the cannulated kidney(s). The perfusate then returns to the organ chamber via the kidney vein to repeat the perfusion cycle.

The cassette can provide circulation of perfusate to one or two kidneys, attached individually or in bloc.

V. Indication for Use

The DCX Disposable Cassette, as part of the RM4 Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.

K243998

Page 6

Traditional 510(k) Premarket Notification – DCX Disposable Cassette
Confidential Page 2 of 4

VI. Comparison of Technological Characteristics

The DCX Disposable Cassette and the DCM-100 Disposable Cassette used with the Kidney Perfusion system have identical technical specifications.

DCX Disposable Cassette is the new version of the DCM-100 Disposable Cassette, compatible with the RM4 Perfusion system (control unit). In order to improve the system and the condition of kidney storage, IGL develops for the new cassette, following technical characteristics:

  • connection luer for perfusion accessories
  • in the DCX, the oxygenation is provided directly in the organ chamber, whereas on the DCM100 the perfusate circulates through a membrane where O2 exchange occurs (two different means to bring oxygen to the perfusate).
  • the placement of a removable partition to allow the immersion of the organ using 1L of perfusate instead of 2L of perfusate for single organ perfusion in the DCX. In DCM-100, even if it for a single perfusion, 2L of solution have to be poured inside the system.
  • the number of lids. Indeed, two lids are on the DCX whereas only one lid is on the DCM-100.

Both cassettes are packaged into a primary packaging which is considered as the sterile barrier system, then sterilized with the Ethylene Oxide sterilization method and intended for single-use only. Finally, both cassettes are labeled as sterile, single use and with a 24 months shelf life.

The difference on the labelling is on the storage conditions. Indeed, the DCX is labeled with storage conditions between 2°C and 25°C with temporary 40°C outing tolerated instead of 5°C to 50°C for DCM-100. Thus, even if the storage condition of both cassettes are different, the range of temperature is similar.

Comparison with Predicate Device

Subject DevicePrimary Predicate Device
RM4 Kidney Perfusion System with DCXRM4 Kidney Perfusion System with DCM-100 (K211224).
Regulatory Specifications
Intended UseKidney storage and preservation for transplantationIdentical
Indication for UseThe DCX Disposable Cassette, as part of the Kidney Perfusion System, is intended to be used for the pulsatile hypothermic machine perfusion (RM4 Kidney Perfusion System) of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.The RM4 control unit, as part of the RM4 Kidney Perfusion System is intended to be used for the pulsatile hypothermic machine perfusion of kidneys for preservation and eventual transplantation into a recipient only in healthcare professional environment.
Principle of OperationSingle use, disposable cassette used with a control unit which provides continuous pulsatile hypothermic physiologic perfusionIdentical
ManufacturerInstitut Geoges Lopez, FranceIdentical
Regulation number21 CFR 876.5880Identical

K243998
Page 2 of 4

Page 7

Traditional 510(k) Premarket Notification – DCX Disposable Cassette
Confidential Page 3 of 4

Comparison with Predicate Device (continued)

Subject DevicePrimary Predicate Device
RM4 Kidney Perfusion System with DCXRM4 Kidney Perfusion System with DCM-100 (K211224).
Regulation descriptionIsolated kidney perfusion and transport system and accessoriesIdentical
Product CodeKDNIdentical
ClassClass IIIdentical
Technical Specifications
Organ TypeKidneyIdentical
Organ CapacityOne or two kidneysIdentical
Solution use to preserve the organPERF-GEN or equivalent solutionIdentical
TransportabilityShort transportation within the healthcare facility by trained personnel.Identical
Compatible ElementsRM4 control UnitIdentical
Disposable compatible elementsCannulas and clampsIdentical
Cassette mountingTop of control unitIdentical
Sterilization, Shelf Life, and Storage
Sterilization MethodEthylene Oxide sterilizationIdentical
Single useYesIdentical
Sterility Assurance level (SAL)10⁻⁶Identical
Shelf life24 monthsIdentical

VII. Summary of Non-Clinical Testing

No performance standard has been established by FDA for the DCX Disposable Cassette.

In order to evaluate the performance and safety of the subject device DCX:

  • The DCX is supplied sterile and non-pyrogenic in order to assure safety for transplant recipients. The validation of the sterilizing was carried out according to ISO 11135 and ISO 10993-7.
  • The biomaterial safety of the DCX has been evaluated through ISO 10993 compliant testing, which included cytotoxicity test, skin sensitization test in guinea pigs, primary skin irritation, hemolysis test and acute systemic toxicity test in mice. Results of this testing, showed the DCX is safe for the intended biocontact.
  • The stability testing has showed that aging of test articles at the recommended storage conditions of 2-25°C (35.6° - 77°F) does not affect the product specifications for the DCX labeled with 24 months shelf life.
  • Performances tests have been performed also with perfusion accessories, and a RM4 Control Unit, in accordance with the recommendations outlined in the IFU of the subject device and in the instructions for use of the RM4 Control Units. For each test performed, the subject performs as expected and meets the requirements set for the device.

The results of these tests indicate that DCX Disposable Cassette is substantially equivalent to the

K243998
Page 3 of 4

Page 8

Traditional 510(k) Premarket Notification – DCX Disposable Cassette
Confidential Page 4 of 4

predicate devices.

VIII. Conclusion

The technological differences between the subject and predicate devices were evaluated through non-clinical testing. The results of these tests demonstrated that the subject device does not raise new issues of safety and effectiveness compared to the predicate device. The indications for use, technological characteristics, and performance characteristics of DCX Disposable Cassette stored for up to 24 months assessed to be substantially equivalent to the predicate device.

K243998
Page 4 of 4

§ 876.5880 Isolated kidney perfusion and transport system and accessories.

(a)
Identification. An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.(b)
Classification. Class II (performance standards).