K Number
K243978

Validate with FDA (Live)

Device Name
LED Facial Mask
Date Cleared
2025-06-20

(179 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LED Facial Mask is an over-the-counter (OTC) device intended for the following uses:

For Red Mode(s): The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.

For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.

For Blue Mode: The device emits light in the blue region of the spectrum and is specifically indicated for the treatment of mild to moderate acne on the face.

For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

Device Description

The LED Facial Mask is a device used for treatment of full-face wrinkles, mild to moderate acne on the face, and providing topical heating. LEDs housed inside the device emit light onto the face. These LEDs generate blue, red, and infrared wavelengths.

This product is composed of the main device, controller, fastening straps, storage bag, blindfold, and charging cable.

AI/ML Overview

This FDA 510(k) clearance letter pertains to an LED Facial Mask, which is categorized as a Class II medical device. The information provided outlines the manufacturing, intended use, and comparison to predicate devices, but it does not contain the specific performance data one would expect for a study proving the device meets acceptance criteria.

The document primarily focuses on demonstrating substantial equivalence to already cleared predicate devices based on technological characteristics and intended use. The "Performance Data" section lists various recognized consensus standards to which the device was tested (e.g., electrical safety, electromagnetic compatibility, usability, biological evaluation), rather than clinical study results establishing efficacy.

Therefore, I cannot create the detailed table of acceptance criteria and reported device performance, nor can I answer questions about sample size, ground truth establishment, expert adjudication, or MRMC studies, as this information is not present in the provided FDA clearance letter.

It explicitly states: "Clinical/animal testing was not performed for the proposed device as part of the submission." This confirms the absence of efficacy studies that would typically define acceptance criteria for performance outcomes.

The acceptance criteria for this device, as implied by the FDA 510(k) process for a Class II device, are largely related to safety, performance against engineering standards, and substantial equivalence to legally marketed predicate devices. The clearance is based on the premise that if the new device is sufficiently similar to existing cleared devices in terms of its technology and intended use, and meets relevant safety standards, it does not require new clinical efficacy data.

Here's a breakdown of what can be inferred from the document regarding acceptance criteria and the "study" (non-clinical testing) that proves the device meets them:


Inferred Acceptance Criteria and Device "Performance" (Non-Clinical)

Since no clinical efficacy studies were performed, the "acceptance criteria" are not related to a specific clinical outcome (e.g., X% reduction in wrinkles, Y% clearance of acne). Instead, they relate to safety, electrical performance, and biocompatibility, as demonstrated through testing against consensus standards.

Acceptance Criterion (Inferred from Standards)Reported Device Performance (Implied by Clearance)
Electrical Safety (IEC 60601-1, -2-57, -2-83)Device meets general requirements for basic safety and essential performance, particular requirements for non-laser light source equipment, and home light therapy equipment. Implies safe power delivery, insulation, and fault protection.
Electromagnetic Compatibility (IEC 60601-1-2)Device meets requirements for electromagnetic disturbances. Implies suitable electromagnetic emissions and immunity to interference.
Usability (IEC 60601-1-6)Device meets usability requirements. Implies the device is designed for safe and effective user interaction.
Home Healthcare Environment Suitability (IEC 60601-1-11)Device meets requirements for use in the home healthcare environment. Implies robustness and safety for lay users.
Photobiological Safety (IEC 62471)Device meets photobiological safety standards. Implies emitted light is safe for the eyes and skin under specified use conditions.
Software Life Cycle Processes (IEC 62304)Device software development followed medical device software life cycle processes. Implies software is engineered to be safe and reliable.
Biocompatibility (ISO 10993-1, -5, -10, -23)Materials in contact with the user are biologically evaluated against standards for cytotoxicity, skin sensitization, and irritation. Implies materials are safe for human contact.
Wavelength AccuracyBlue: 410 nm, 460 nm; Red: 610 nm, 630 nm, 660 nm; NIR: 830 nm, 850 nm, 880 nm. (Measured against specified ranges)
Intensity/Irradiance (mW/cm²)Red1 (Y) Mode: 36.0 mW/cm²; Red2 (R) Mode: 19.0 mW/cm²; Blue (B) Mode: 6.1 mW/cm²; NIR Mode: 45.0 mW/cm²; Red1 (Y)+NIR Mode: 81.0 mW/cm²; Red2 (R)+NIR Mode: 64.0 mW/cm². (Measured against specified output)
Substantial Equivalence to Predicate DevicesDevice has similar intended use, technological characteristics (light source, anatomical location, wavelength, intensity/irradiance), and safety profile to previously cleared devices.

Study Details (Based on the Provided Document)

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable in the context of clinical data for efficacy. The "test set" here refers to the device itself being tested against engineering and safety standards. There is no patient sample size.
    • Data Provenance: The tests conducted are in vitro (bench testing) and in silico (software analysis) or conducted on the physical device itself, not on human subjects. Therefore, there's no country of origin for patient data, nor is it retrospective or prospective clinical data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The "ground truth" for the engineering and safety tests is defined by the consensus standards themselves (e.g., a certain voltage tolerance, a specific emission limit). The "experts" would be the engineers and technicians performing the tests and verifying compliance with the standards, not clinical experts establishing ground truth for medical conditions.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This concept pertains to clinical studies where human interpretation of medical data needs to be harmonized. For engineering tests, results are typically objective measurements against a standard.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This type of study is for AI/CADe devices and involves human readers reviewing cases. This device is an LED facial mask, not an AI diagnostic tool.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This applies to AI/CADe devices.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this clearance is adherence to recognized consensus standards for safety, electrical performance, photobiological safety, and biocompatibility, as well as demonstrated technical similarity and intended use alignment with predicate devices. There is no clinical ground truth (e.g., biopsy results, clinician's diagnosis) involved in this substantial equivalence determination.
  7. The sample size for the training set:

    • Not applicable. There is no AI model or "training set" for this device.
  8. How the ground truth for the training set was established:

    • Not applicable.

In summary, the provided FDA 510(k) clearance document for the LED Facial Mask demonstrates that the device was cleared based on its substantial equivalence to existing devices and its adherence to a series of non-clinical safety and performance standards. No clinical efficacy studies were conducted or required for this particular clearance, as explicitly stated in the document.

FDA 510(k) Clearance Letter - LED Facial Mask

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 20, 2025

Shenzhen Sunsred Technology Co.,Ltd
℅ Jun Da Huang
Project Manager
Registrar Corp
Room 502, Bldg A, Phase I, Qianhai Economic and Trade Center
No. 151 Free Trade West Street, Nanshan Sub-district,Qianhai
Shenzhen, Guangdong Province 518066
China

Re: K243978
Trade/Device Name: LED Facial Mask
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, ILY, OLP
Dated: December 10, 2024
Received: December 23, 2024

Dear Jun Da Huang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243978 - Jun Da Huang Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243978 - Jun Da Huang Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

YAN FU -S
Digitally signed by YAN FU -S
Date: 2025.06.20 15:28:03 -04'00'

for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known): K243978

Device Name: LED Facial Mask

Indications for Use (Describe)

The LED Facial Mask is an over-the-counter (OTC) device intended for the following uses:

For Red Mode(s): The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.

For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.

For Blue Mode: The device emits light in the blue region of the spectrum and is specifically indicated for the treatment of mild to moderate acne on the face.

For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

LED Facial Mask

Date of 510(k) Summary Preparation: 2024-12-20
Date of Latest 510(k) Summary Modification: 2025-06-17

1. Applicant Information

Shenzhen Sunsred Technology CO.,Ltd
201, No. 22, Shijing Road, Fumin Community
Fucheng Street, Longhua District
Shenzhen, Guangdong, China

Contact Person: Xiufeng Cai
Telephone Number: 0086 18682165796
Email: nancyhuang@sunsred.com

2. Submission Correspondent

Registrar Corp
Room 502, Bldg A, Phase I, Qianhai Economic and Trade Center, No. 151 Free Trade West Street, Nanshan Sub-district, Qianhai Shenzhen-Hong Kong Cooperation Zone, Shenzhen, Guangdong Province, 518066, People's Republic of China

Contact Person: HUANG Jun Da Ken
Telephone Number: +86 137 6074 9097
Email: adnil1330@vip.qq.com

3. Proposed Device

Trade Name: LED Facial Mask
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Infrared lamp.
Model(s): SR-M4
Classification Regulation Number: 21 CFR 878.4810, 21 CFR 890.5500
Product Code: OHS (Light Based Over The Counter Wrinkle Reduction), OLP (Over-The-Counter Powered Light Based Laser For Acne), ILY (Lamp, Infrared, Therapeutic Heating)
Device Class: II
FDA CDRH Review Panel: General & Plastic Surgery, Physical Medicine

Page 6

4. Predicate/Reference Device(s)

Predicate Device:
510(k) Number: K223544
Applicant: Guangdong Newdermo Biotech Co.,Ltd
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Infrared lamp
Trade Name: LED Light Therapy Mask
This predicate device has not been subject to a design-related recall.

Reference Device #1:
510(k) Number: K162652
Applicant: Guangzhou LETA Testing Technology Co., Ltd
Classification Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief. Laser surgical instrument for use in general and plastic surgery and in dermatology.
Trade Name: Smart Photon Micro-current Device
This reference device has not been subject to a design-related recall.

Reference Device #2:
510(k) Number: K163329
Applicant: Pulsaderm LLC
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Trade Name: Pulsaderm Wrinkle Mask 28 and Wrinkle Mask 72
This reference device has not been subject to a design-related recall.

Reference Device #3:
510(k) Number: K162489
Applicant: Zhongshan Bisen Plastic Electronic Products Co.,Ltd.
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Trade Name: RED Light Device
This reference device has not been subject to a design-related recall.

5. Device Description

The LED Facial Mask is a device used for treatment of full-face wrinkles, mild to moderate acne on the face, and providing topical heating. LEDs housed inside the device emit light onto the face. These LEDs generate blue, red, and infrared wavelengths.

This product is composed of the main device, controller, fastening straps, storage bag, blindfold, and charging cable.

Page 7

6. Indications for Use

The LED Facial Mask is an over-the-counter (OTC) device intended for the following uses:

For Red Mode(s): The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.

For Red + NIR Mode(s): The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.

For Blue Mode: The device emits light in the blue region of the spectrum and is specifically indicated for the treatment of mild to moderate acne on the face.

For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

7. Comparison to the Predicate Device(s)

Indications for Use/Intended Use:

The subject LED Facial Mask is an Over-the-Counter (OTC) device mainly intended for treatment of full-face wrinkles, mild to moderate acne on the face, and providing topical heating. The predicate/reference devices have the same indications for use. All devices share the similar anatomical location and intended use for treatment. The proposed device can be used at home and healthcare settings.

Technological Characteristics:

The subject LED Facial Mask has the similar principle of operation and characteristics as the predicate/reference devices. Any minor differences between the subject device and the listed predicate/reference devices do no raise any issues of safety or efficacy. Performance data supports that the device is as safe and effective as the predicate/reference devices for its intended use.

The subject LED Facial Mask is substantially equivalent to the predicate/reference devices in that it has the same/similar intended use and technological characteristics, as demonstrated in the following Table 1 Substantial Equivalence Comparison.

Page 8

Table 1 Substantial Equivalence Comparison

FeatureProposed Device (K243978)Predicate Device (K223544)Reference Device #1 (K162652)Reference Device #2 (K163329)Reference Device #3 (K162489)Remarks
ManufacturerShenzhen Sunsred Technology CO.,LtdGuangdong Newdermo Biotech Co.,LtdGuangzhou LETA Testing Technology Co., LtdPulsaderm LLCZhongshan Bisen Plastic Electronic Products Co.,Ltd./
Device NameLED Facial MaskLED Light Therapy MaskSmart Photon Micro-current DevicePulsaderm Wrinkle Mask 28 and Wrinkle Mask 72RED Light Device/
Model(s)SR-M4FM-01, FM-02, FM-03EP-300Pulsaderm Wrinkle Mask 28, Pulsaderm Wrinkle Mask 72BZ-0606/
Intended UseThe LED Facial Mask is an over-the-counter (OTC) device intended for the following uses: For Red Mode(s): The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles. For Red + NIR Mode(s): The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles. For Blue Mode: The device emits light in the blue region of the spectrum and is specifically indicated for the treatment of mild to moderate acne on the face. For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.Red light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne.For micro current stimulation mode: The Smart Photon Micro-current Device is intended for facial stimulation and is indicated for over-the-counter aesthetic use. For red light irradiation mode: The red light is intended for the treatment of periorbital wrinkles, For blue light irradiation mode: The blue light is for the treatment of mild to moderate acne.The Pulsaderm Wrinkle Masks 28 and 72 are intended for the use in the treatment of facial wrinkles and for people with Fitzpatrick Skin Types I, II and III.The RED Light Device is an OTC device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.Same
Rx or OTCOTCOTCOTCOTCOTCSame
Light SourceLEDLEDLEDLEDLEDSame
Product CodeOHS, ILY, OLPOHS, ILY, OLPNFO, OHS, OLPOHSOHSSame
Anatomical LocationFaceFace and bodyFaceFaceFaceSame
WavelengthBlue: 410 nm, 460 nm; Red: 610 nm, 630 nm, 660 nm; NIR: 830 nm, 850 nm, 880 nmBlue: 460 nm; Red: 620 nm; Infrared: 850 nmBlue: 415±10 nm; Red: 630±10 nmRed: 620~630 nm; Infrared: 850 nmRed: 633±5 nm; Infrared: 830±5 nmSame Within the range of the predicate devices
Intensity / Irradiance (mW/cm²)Red1 (Y) Mode: 36.0 mW/cm²; Red2 (R) Mode: 19.0 mW/cm²; Blue (B) Mode: 6.1 mW/cm²; NIR Mode: 45.0 mW/cm²; Red1 (Y)+NIR Mode: 81.0 mW/cm²; Red2 (R)+NIR Mode: 64.0 mW/cm²Red: 2.0~3.0 mW/cm²; Blue: 2.0~4.0 mW/cm²; Infrared: 2.0~4.0 mW/cm²; Mixed: 9.0~12.0 mW/cm²Red Light Irradiation Mode: 80 mW/cm²; Blued Light Irradiation Mode: 50 mW/cm²; Micro Current Stimulation Mode: N/APulsaderm Wrinkle Mask 28: Red: 15.94 mW/cm²; Infrared: 5.24 mW/cm²; Total: 21.18 mW/cm²; Pulsaderm Wrinkle Mask 72: Red: 18.71 mW/cm²; Infrared: 6.61 mW/cm²; Total: 25.35 mW/cm²Red: 70 mW/cm²; Infrared: 55 mW/cm²; Total: 125 mW/cm²Same Total accumulation of treatment energy is within the range of the predicate devices

Page 9

FeatureProposed Device (K243978)Predicate Device (K223544)Reference Device #1 (K162652)Reference Device #2 (K163329)Reference Device #3 (K162489)Remarks
ModeRed1 (Y) Mode: 630 nm(55 LEDs), 660 nm(5 LEDs); Red2 (R) Mode: 610 nm(5 LEDs), 630 nm(55 LEDs); Blue (B) Mode: 410 nm(5 LEDs), 460 nm(55 LEDs); NIR Mode: 830 nm(5 LEDs), 850 nm(50 LEDs), 880 nm(5 LEDs); Red1 (Y)+NIR Mode: 630 nm(55 LEDs), 660 nm(5 LEDs), 830 nm(5 LEDs), 850 nm(50 LEDs), 880 nm(5 LEDs); Red2 (R)+NIR Mode: 610 nm(5 LEDs), 630 nm(55 LEDs), 830 nm(5 LEDs), 850 nm(50 LEDs), 880 nm(5 LEDs)Red: 620 nm; Blue: 460 nm; Infrared: 850 nm; Mixed: 460 nm, 620 nm, 850 nm; The number of LEDs is unclear.Red Light Irradiation Mode: 630±10 nm; Blued Light Irradiation Mode: 415±10 nm; The number of LEDs is unclear. Micro Current Stimulation Mode: 5 levels of output intensity; Mode configuration is unclear.Red: 620~630 nm; Infrared: 850 nm; Pulsaderm Wrinkle Mask 28: Total 28 LEDs; Pulsaderm Wrinkle Mask 72: Total 72 LEDs; Mode configuration is unclear.Red: 633±5 nm; Infrared: 830±5 nm; 25 LEDs over 17 cm²The difference in number of LEDs does not affect the safety or efficacy of the device as there are a wide range of number of LEDs devices cleared under device code OHS, ILY and OLP.
Treatment Time5 minutes per Treatment. DO NOT exceed 5 minutes in one session. 3 days per week for 8 weeksManual Mode: 15 minutes each time; Automatic Mode: 10 minutes each time. 3-4 treatment a week, reduce to 1-2 treatment a week once the results shown.For light irradiation of red light, the recommend treatment session is 3 minutes / 2-3 times per week. And for blue light, the recommend treatment session is 4 minutes / 2 times per week on each treatment area.15 minutes everydayFor the first month (4 weeks), treatment should be performed 3 times a week for 15-20 minutes each time. (5-7 minutes on each treatment zone).Same Total accumulation of treatment energy is within the range of the predicate devices

Page 10

8. Performance Data

The following performance data are provided in support of the substantial equivalence determination:

The proposed device performs testing in accordance with the following recognized consensus standards:

  • IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
  • IEC 60601-1-2:2014+AMD1:2020 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
  • IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
  • IEC 60601-1-11:2015+AMD1:2020 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
  • IEC 60601-2-57:2023 Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring, cosmetic and aesthetic use
  • IEC 60601-2-83: 2019+AMD1:2022 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
  • IEC 62471:2006 Photobiological safety of lamps and lamp systems
  • IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
  • ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for in vitro Cytotoxicity
  • ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests for Skin Sensitization

Page 11

  • ISO 10993-23:2021 Biological Evaluation of Medical Devices - Part 23: Tests for Irritation

9. Clinical/Animal Testing

Clinical/animal testing was not performed for the proposed device as part of the submission.

10. Conclusions

Based on the above performance as documented in this application, the LED Facial Mask is found to have a safety and effectiveness profile that is similar to the predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.