AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0ml/s

Device Description

The HydroPICC 4F Single Lumen Catheter is comprised of a radiopaque hydrophilic catheter material with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the clamp. HydroPCC has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions f thrombus accumulation and thrombotic occusions were evaluated using in vitro and in vivo models. Pre-clinical in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.

AI/ML Overview

The provided text is a 510(k) summary for the HydroPICC 4Fr Single Lumen Marked catheter, where the purpose of the submission is to modify the power injection ratings in the indications for use. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than providing details of a study against specific acceptance criteria for a new device's performance.

Therefore, the document does not contain the information requested regarding acceptance criteria, a specific study proving device meets acceptance criteria, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth details.

The document discusses "comprehensive data demonstrating the safety and efficacy" and "rigorous testing protocols" including "in vitro" data to support the modified power injection rating, and states that "the form, fit, and function of these devices have not changed from their previous clearances." However, it does not provide specific details of these tests or their results in a measurable, quantifiable way that would allow for the completion of the requested table and study breakdown.

{0}------------------------------------------------

January 17, 2025

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, which includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Access Vascular, Inc. Brian Hanley Chief Operating Officer 749 Middlesex Turnpike Billerica, Massachusetts 01821

Re: K243941

Trade/Device Name: HydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - Basic Kit (70001201); HydroPICC 4Fr Single Lumen Marked catheter, 70 cm guidewire Basic Kit (70001202); HydroPICC 4Fr Single Lumen Marked catheter, Maximal Barrier kit (70001204); HydroPICC 4Fr Single Lumen Marked catheter, Mobile Maximal Barrier Kit (90001204) Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: December 20, 2024 Received: December 20, 2024

Dear Brian Hanley:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

David Walloschek

David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243941

Device Name

HydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - Basic Kit (70001201); HydroPICC 4Fr Single Lumen Marked catheter, 70 cm guidewire - Basic Kit (70001202); HydroPICC 4Fr Single Lumen Marked catheter, Maximal Barrier kit (70001204); HydroPICC 4Fr Single Lumen Marked catheter, Mobile Maximal Barrier Kit (90001204)

Indications for Use (Describe)

HydroPICC Single Lumen: Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0ml/s

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) #:510(k) SummaryPrepared on: 2025-01-14
Contact Details21 CFR 807.92(a)(1)
Applicant NameAccess Vascular, Inc.
Applicant Address749 Middlesex Turnpike Billerica MA 01821 United States
Applicant Contact Telephone781.538.6594
Applicant ContactMr. Brian Hanley
Applicant Contact Emailbhanley@accessvascularinc.com
Device Name21 CFR 807.92(a)(2)
Device Trade NameHydroPICC 4Fr Single Lumen Marked catheter, 130 cm guidewire - BasicKit (70001201);HydroPICC 4Fr Single Lumen Marked catheter, 70 cm guidewire - BasicKit (70001202);HydroPICC 4Fr Single Lumen Marked catheter - Maximal Barrier kit(70001204);HydroPICC 4Fr Single Lumen Marked catheter - Mobile Maximal BarrierKit (90001204)
Common NamePercutaneous, implanted, long-term intravascular catheter
Classification NameCatheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Regulation Number880.5970
Product Code(s)LJS
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K193015HydroPICCLJS

Device Description Summary

The HydroPICC 4F Single Lumen Catheter is comprised of a radiopaque hydrophilic catheter material with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in the manufacture of catheters are provided packaged in kit configurations with the appropriate accessories for placement in the respective clinical environments. The maximum power injection flow rate for the lumen is indicated on the clamp. HydroPCC has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions f thrombus accumulation and thrombotic occusions were evaluated using in vitro and in vivo models. Pre-clinical in vivo evaluations do not necessarily predict clinical performance with respect to thrombus formation.

21 CFR 807.92(a)(4)

21 CFR 807.92(a)(5)

The purpose of this 510(k) is to modify the power injection ratings in the indications for use.

Intended Use/Indications for Use

HydroPICC Single Lumen: Indicated for short-or long-tern peripheral access to the central venous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure

{5}------------------------------------------------

monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0ml/s

Indications for Use Comparison

Rationale Supporting 21 CFR 807.92(a)(5) for a Change in Indications for Use

  1. Technological Characteristics and Performance Data: The same technological characteristics as the predicate devices. In vitro data demonstrates that the devices will perform as intended with the modified power injection rating.

  2. Safety and Effectiveness The proposed device has been shown to be safe and effective for its intended use.

  3. Comparison with Predicate Devices The predicate device, HydroPCC (K193015) is indicated for similar uses, including short-term and long-term peripheral access to the venous therapy, administration of fluids, medications, and nutrients, and the sampling of blood.

  4. Intended Use and Indications The intended use of the proposed devices aligns with that of the proposed indications for use are consistent with the current clinical practices and address the same patient population and clinical needs as the predicate devices. The only modification is to the power injection rating.

  5. Regulatory Compliance The proposed changes comply with the requirements of 21 CFR 807.92(a)(5), which necessitates a comparison of the new device with a legally marketed predicate device to demonstrate that the new device is as safe and effective as the predicate device. The data provided support that the proposed devices meet these criteria.

Conclusion The proposed HydroPICC 4Fr Single Lumen Catheter provide for an increase to the power injection ratings, supported by data, justify the change in indications for use and demonstrate that the proposed devices are as safe and effective as the predicate devices.

Technological Comparison

21 CFR 807.92(a)(6)

The proposed device, HydroPCC 4F Single Lumen Catheter subject to this submission, has the same technological characteristics as the predicate device identified above. These characteristics include design, material, chemical composition, and principle of operation.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

This 510(k) submission provides comprehensive data demonstrating the safety and efficacy of the HydroPCC 4F Single Lumen Catheter. Rigorous testing protocols were employed to evaluate these devices performance under various conditions, ensuring they meet all applicable regulatory standards. The submission includes a summary of in vitro which substantiate the devices intended use and their ability to perform as claimed. Additionally, the form, fit, and function of these devices have not changed from their previous clearances. The data presented herein supports the conclusion that the HydroPlCC 4F Single Lumen Catheter is substantially equivalent to the predicate device(s) in terms of safety and effectiveness.

Upon reviewing the information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the HydroPICC 4F Single Lumen Catheter with the modified power injection rating is substantially equivalent to existing legally marketed devices.

21 CFR 807.92(a)(5)

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”